Average CT in PET studies of colorectal cancer patients with metastasis in the liver and esophageal cancer patients

Average CT in PET studies of colorectal cancer patients with metastasis in the liver and esophageal cancer patients

Elena Tonkopi 1, Pai-Chun Melinda Chi 2, Osama Mawlawi 1, Adam C. Riegel 1, Eric M. Rohren 3, Homer A. Macapinlac 3, Tinsu Pan 1a
Department of Imaging Physics, 1 Department of Radiation physics, 2 Department of Nuclear Medicine, The University of Texas M. D. Anderson Cancer Center, Houston TX, USA
tpan@mdanderson.org .
3

Received 30 March 2009 : Accepted 15 September 2009

Average CT (ACT) and PET have a similar temporal resolution and it has been shown to improve registration of the CT and PET data for PET/CT imaging of the thorax. The purpose of this study was to quantify the effect of ACT attenuation correction on PET for gross tumor volume (GTV) delineation with standardized uptake value (SUV) for liver and esophageal lesions. Our study included 48 colorectal cancer patients with metastasis in the liver and 52 esophageal cancer patients. These patients underwent a routine PET/CT scan followed by a cine CT scan of the thoracic region for ACT. Differences between the two PET data sets (PET HCT and PET ACT ) corrected with the helical CT (HCT) and ACT were quantified by analyzing image alignment, maximum SUV (SUV max ), and GTV. The 67% of the colorectal and 73% of the esophageal studies demonstrated misregistration between the PET HCT and HCT data. ACT was effective in removing misregistration artifacts in 65% of the misregisted colorectal and in 76% of the misregisted esophageal cancer patients. Misregistration between the CT and PET data affected GTVs due to the change in SUV max with ACT. A change of SUV max greater than 20% between PET HCT and PET ACT was found in 15% of the colorectal and 17% of the esophageal cases. Our results demonstrated a more pronounced effect of misregistration for the smaller lesions (< 5 cm 3 ) near the diaphragm (< 5 cm). ACT was effective in improving registration between the CT and PET data in PET/CT for the colorectal and esophageal cancer patients.

PACS number: 87.58.Fg

Key words: respiration-averaged CT , PET/CT , respiration-induced artifact

I. INTRODUCTION

The emergence of positron emission tomography/computed tomography (PET/CT) made it possible to provide anatomic and functional information in a single examination and to improve the sensitivity and specificity of tumor detection over PET or CT alone.(1–3) PET/CT imaging is now part of routine clinical practice in oncology for diagnosis, staging, and monitoring of tumor response to therapy.(4–6) PET/CT has also gained acceptance in cardiology for coronary artery imaging and myocardial functional assessment.(7–13) Standardized uptake value (SUV) is normally used as an indicator of malignancy.(14,15) Over 50% of all patients with cancer receive radiation therapy and FDG-PET has been shown to influence the decision of target volumes for non-small–cell lung cancers (NSCLC), esophageal tumors, and head-and-neck squamous cell carcinomas.(16–24) Incorporation of PET data into treatment volume delineation for NSCLC can increase or decrease the treated volume between 15% to 60%.(19) Since CT data is used for attenuation correction (AC) of PET data, spatial misregistration between the CT and PET data may compromise the quantification of PET. Respiratory motion during imaging – in combination with the difference in temporal resolution between the PET and CT data – could cause misregistration between the PET and CT data, compromising the advantages of this hybrid imaging modality.(25) This misregistration problem did not exist in the stand-alone PET scanners, which used transmission rod sources to acquire transmission data for AC. However, stand-alone PET scanners suffered from a long transmission scan time. As a result, most patients were scanned in the position of arms down, different from the arms up position typically used in the radiotherapy simulation or treatment, making it more complex in registration of the PET and CT data. In addition, the PET and CT data were acquired as two different sessions and at two different times. Issues pertaining to software registration of the PET and CT data acquired in two different scanners have been reduced significantly with hardware registration of the PET/CT scanner.(26)

To address the registration problem from respiration in PET/CT, there are multiple studies investigating different approaches to correct respiratory motion artifacts. Some investigated different CT scanning protocols(25,27–41) including deep inspiration breath hold,(30–32) mid-expiration breath-hold,(42) shallow breathing,(33,43) respiratory gating,(35–39,41) or using a slow CT scan.(40) Each of the recommendations has its advantages and disadvantages. For instance, complying with the breathing protocol can be difficult for some patients.(44) Gated PET images usually exhibit reduced statistics; therefore, they need a longer acquisition time.(45) Slow CT protocol is unreliable in depicting the average motion of respiration.(46) AC of the PET images with CT data on phase-to-phase basis requires 4D acquisition of both PET and CT, which may not be practical.

We have proposed the use of respiration-averaged CT (ACT) to match the temporal resolution of CT and PET, and have shown that ACT reduces respiratory artifacts in PET/CT.(46,47) ACT acquisition has been optimized for clinical implementation and is available on the PET/CT scanners in our institution. In order to quantify the effect of ACT, data was acquired under the institution review board protocol DR07-0560 for three months and results have been analyzed. An extensive analysis on the 229 lung cancer patients(48) has shown an improvement of the PET and CT data registration with the ACT technique and its potential effect on GTV delineation. The other anatomical areas prone to respiratory induced artifacts are the esophagus and the upper abdomen,(49–51) and this study focused on these particular regions. The aim of our investigation was to demonstrate a potential improvement in the PET/CT data registration and the effect on GTV delineation by the usage of ACT in patients with esophageal cancer or colorectal cancer with metastasis in the liver.

II. MATERIALS AND METHODS

A. Data acquisition

Our study included 48 colorectal cancer patients with metastasis in the liver and 52 esophageal cancer patients. These patients had undergone a routine PET/CT examination for diagnosis or post-therapy evaluation. All the data were acquired on three different PET/CT scanners (DST/RX/DSTE; GE Healthcare, Waukesha, WI). The patients were injected with 555–740 MBq of 18F-FDG and scanned 1 h after injection. A helical CT (HCT) scan was acquired in 16 and 32 sec over 90 cm coverage by a 16-slice and 8-slice CT, respectively, followed by a 2D PET scan of 3 min per bed and a cine CT scan of the thorax (approximately 20 cm) for ACT. Patients were free breathing during all scans without any coaching of respiration. Both HCT and ACT were used for AC of the PET data, resulting in two data sets: PET corrected with HCT (PET HCT ) and PET corrected with ACT (PET ACT ).

HCT data were acquired at 120 kVp, 300 mA, 1.35 pitch for 8-slice and 1.375 pitch for 16-slice, and 0.5 sec gantry rotation. X-ray collimation was 8 × 1.25 mm on the 8-slice DST and 16 × 1.25 mm on the 16-slice RX and DSTE. Cine duration was 5.9 sec per 2 cm coverage for capturing at least one respiratory cycle.(46,47,52) The other parameters for cine scan were 120kVp and collimation of 8 × 2.5 mm for both 8- and 16-slice CTs. Tube current was adjusted according to the patients’ weight: 10 mA for < 70 kg, 15 mA for 70–100 kg, and 20 mA for >100 kg. Processing time of ACT for PET attenuation correction was less than 5 min; additional radiation dose from the cine CT scan for ACT was 5–10 mGy(46) or 1.7–3.4 mSv.

B. Data analysis

The method described in our previous study of the lung cancer patients(48) was adopted for data analysis. The PET HCT and PET ACT images were compared in quantitation using viewing software (Advantage Workstation (AW) 4.2, GE Healthcare, Waukesha, WI). Misalignment between the CT and PET data due to respiratory motion was measured as the thickness of a photopenic region or white band at the diaphragm level. An example is shown in Fig. 1. Another criterion to quantify the difference was SUV max , measured on both PET HCT and PET ACT with the AW workstation. The percent changes in SUV max were calculated in reference to PET HCT .

 


 

Fig. 1 . The PET HCT (left) and PET ACT images (right) of a patient study. The misalignment at the diaphragm level was measured as the thickness of a photopenic region or a white band. The artifact was reduced in the PET ACT .

Both PET data sets were analyzed on a commercial treatment planning system (Pinnacle3 7.6, Philips Medical Systems, Andover, MA). The images were displayed in SUV using a software tool developed in-house. The gross tumor volumes (GTVs) were generated based on the SUV thresholds. The initial threshold for the delineation was 2.5, which usually indicates potential malignancy of a lesion.(53) In some cases when the SUV was too close to the background uptake, the threshold had to be increased until the GTV delineation could be performed by SUV threshold. Typical values were 3.0–3.5 for the colorectal and 3.0–3.3 for the esophageal lesions. The same SUV threshold was chosen for both PET HCT and PET ACT images of the same patient. In this study, the following parameters were analyzed: the centroid shift between the GTVs corresponding to the PET HCT and PET ACT , the percent change in the GTV volume, and the concordance index. The centroid location and the GTV volume were measured. The concordance index was defined as the intersection of the two GTVs corresponding to the PET HCT and PET ACT divided by their union,(54) and it was calculated using a MATLAB program written to read the regions of interest defined in the treatment planning system. Colorectal and esophageal cancer studies were analyzed and presented separately.

III. RESULTS

A. Respiratory artifacts evaluation and SUV max comparison

On PET HCT images, 32 colorectal (67%) and 38 esophageal (73%) cancer patients demonstrated the photopenic region or white band artifacts as a result of mismatch between the PET and HCT data. PET ACT was effective in removing this artifact for 21 colorectal and for 29 esophageal cancer patients. In those cases when the misalignment was not completely removed, PET ACT was still effective in reducing it. The results of the comparison between PET HCT and PET ACT are shown in Fig. 2. The mean values of the misalignment on PET HCT for the colorectal and esophageal patients were respectively 16 ± 6 mm and 14 ± 5 mm, with the maximum magnitude up to 33 mm for one colorectal patient. The mean values of the misalignment on PET ACT for the cases when it was not completely removed were 11 ± 4 mm for the colorectal and 10±2mm for the esophageal patients.

 


 

Fig. 2 . Two histograms showing the misalignment between PET HCT and PET ACT : (a) 67% of the 48 colorectal and (b)73% of the 52 esophageal cancer patients demonstrated misalignment of over 5 mm between the CT and PET data with maximum magnitude of 33 mm for the colorectal and 31 mm for the esophageal cancers. There was less misregistration in PET ACT than in PET HCT .

In the analysis of SUV max change for the patients with multiple lesions only the lesion with the highest percentage change in the SUV max was included. Therefore, there were 48 colorectal and 52 esophageal lesions considered. Figure 3 shows the results of the SUV max change. There were 15% of the colorectal and 17% of the esophageal cancer patient studies demonstrating a percent change of over 20% in SUV max , with the most extreme cases of an esophageal lesion showing 50% increase and a liver lesion showing 29% increase in SUV max on PET ACT data. In the majority of cases, a higher SUV max value was measured on PET ACT ; however, in five esophageal patient studies, SUV max was higher on PET HCT . Absolute values of the percent change were included in the analysis.

 


 

Fig. 3 . Two histograms showing the difference in SUV max between PET HCT and PET ACT : (a) 15% of the colorectal cancer patients and (b) 17% of the esophageal cancer patients showed a difference in SUV max of over 20%.

B. GTVs analysis in Pinnacle

Proper GTV delineation could not be performed for 13 colorectal and 6 esophageal cancer patients due to low tumor-to-background ratio. Those patients were excluded from the GTV analysis. Some patients had multiple lesions. Overall, there were 52 esophageal and 59 liver lesions included in the study. Figure 4 shows the transverse, sagittal, and coronal views of the PET ACT images for two patients: one from the colorectal cancer group and the other one from the esophageal cancer group. The green contour was generated from PET HCT and the blue contour was generated from PET ACT . For the first example with a liver lesion, the percent SUV max change was 29%, volume change was 246% with an increase in absolute value of 10.6cm3, centroid shift was 2 mm and concordance index was 0.336. For the second example with an esophageal lesion, the percent SUV max change was 22%, volume change was 90% with an increase in absolute value of 4.1 cm3, centroid shift was 14 mm and concordance index was 0.569. The scatter plots of the parameter changes versus lesion size are shown in Fig. 5 for the colorectal cancer and in Fig. 6 for the esophageal cancer patients. Both patient data sets demonstrate a more pronounced effect of misalignment for the smaller lesions. For instance, for the liver lesions, the highest percent volume change of 640% corresponds to the 1.6 cm3 absolute change of the lesion size in PET HCT of 0.25 cm3. The same lesion shows the smallest concordance index of 0.09 in the colorectal patient population, and one of the largest changes in the centroid shift of 7.1 mm. The same trend was also noticed for the esophageal lesions. The highest percent volume change of 294% corresponding to the 2.3 cm3 change of the lesion size in PET HCT of 0.78 cm3 was found for the lesion with the lowest concordance index of 0.23, and with one of the highest centroid shift of 3.6 mm.

 


 

Fig. 4 . Examples of segmented GTVs on transverse, sagittal, and coronal views of the PET ACT images: (a) the liver lesion of a colorectal cancer patient and (b) an esophageal tumor. The green contour is generated from PET HCT and the blue contour is generated from PET ACT .

 


 

Fig. 5 . Scatter plots for (a) percent SUV max change, (b) GTV volume change, (c) centroid shift, and (d) concordance index for the colorectal cancer patients between PET HCT and PET ACT .

 


 

Fig. 6 . Scatter plots for (a) percent SUV max change, (b) GTV volume change, (c) centroid shift, and (d) concordance index for the esophageal cancer patients between PET HCT and PET ACT . All lesions were less than 50 cm3.

C. Correlation with the lesion size and location

The liver and esophageal lesions were smaller in size than the lung lesions in our previous study.(48) In current study, the average liver lesion was 10.8 cm3 and the average esophageal lesion was 8.5 cm3. The criterion to define “small” and “large” lesions was chosen to be 5 cm3, which is approximately half of the average size for both patient data sets. The mean values of the analyzed parameters for those distributions are shown in Table 1. A two-tailed, nonpaired t-test was used to evaluate the difference between the small and the large lesions. The results for both sets of patients showed that at p-value 0.05, the difference is not significant for the means of the following parameters: percent SUV max change ( p = 0.392 for the liver and p =0.781 for the esophageal lesions), percent volume change ( p = 0.068 for the liver, p = 0.057 for the esophageal lesions), and centroid shift ( p = 0.456 for the liver, p = 0.864 for the esophageal lesions). The only significant difference was found for the concordance index for both patient data sets: p = 0.005 for the liver lesions and p = 0.018 for the esophageal lesions.

Table 1 .  Distribution of the parameters according to the lesion size.

 

To analyze correlation with lesion location, all lesions were divided in three groups according to the absolute distance from the diaphragm: 0 < d <5 cm, 5 cm < d < 10 cm, and d >10 cm, as shown in Table 2. The maximum distance 16.6 cm was measured for the esophageal lesions, located superior to the diaphragm. A one-way ANOVA test at 0.05 p-value level was used to determine any difference between those groups. The only significant difference between variations in different groups ( p =0.04) was found for ΔSUV max for the liver lesions. Figures 7 and 8 demonstrate the mean values of percent SUV max change, percent volume change, centroid shift, and concordance index for each group of the liver and esophageal lesions.

Table 2 .  Distribution of the parameters according to the lesion location.

 

 


 

Fig. 7 . Mean values of (a) percent SUV max change, (b) GTV volume change, (c) centroid shift, and (d) concordance index for each group of the liver lesions between PET HCT and PET ACT , defined according to their distance to the diaphragm.

 


 

Fig. 8 . Mean values of (a) percent SUV max change, (b) GTV volume change, (c) centroid shift, and (d) concordance index for each group of the esophageal lesions between PET HCT and PET ACT , defined according to their distance to the diaphragm.

IV. DISCUSSION

The influence of respiratory motion in PET/CT on liver and esophageal lesions was investigated. The PET data of 48 colorectal cancer patients with metastasis in the liver and 52 esophageal cancer patients were attenuation corrected with both HCT and ACT. Respiratory artifacts were present in PET HCT data in 67% of the colorectal and 73% of the esophageal patients. ACT was effective in removing misregistration artifacts in 65% of the misregisted colorectal and 76% of the misregisted esophageal cancer patients. Another parameter used for quantification of the differences between the two methods was SUV max value, commonly used in PET imaging for the treatment response monitoring and the baseline evaluation of the patients. A reduction of at least 15 – 25% in tumor uptake after one cycle of chemotherapy is classified as a partial metabolic response.(55,56) Variations in SUV max of over 20% have been found in 15% of the colorectal and in 17% of the esophageal patient populations. Therefore, under-estimation as well as over-estimation of the SUV max due to misregistration of the PET and CT data could potentially change the patient management. These results are consistent with the results reported in our previous study of the lung cancer patients.(48)

Treatment planning based on PET/CT could also be affected by misalignment between the PET and CT data. It was shown in this study that localization of the tumor and delineation of the GTVs could be impacted by the use of ACT. Since ACT has a better registration with PET data than HCT, as it reduced misregistration, the GTV generated on PET ACT image is expected to be more accurate. The changes influenced by misregistration are dependent on the lesion size and location: the smaller lesions located closer to the diaphragm were typically affected more by misregistration. This result is similar to the conclusion of our previous investigation of lung patients.(48) However, there were some differences in the two studies due to the different anatomic regions. The average liver lesion was 10.8 cm3 and the average esophageal lesion was 8.5 cm3, therefore our criterion for the definition of a small lesion was chosen to be less than 5 cm3. It was 50 cm3 for the lung lesions.(48) Both the liver and esophageal lesions showed a significant difference between the “small” and “large” groups only for concordance index, the lung lesions also demonstrated a significant difference in percent SUV max change. For the lung cancer patients, the correlation between the variations and the lesion locations showed a significant difference for the tumors located below the dome of the diaphragm (d≤0). In this study the absolute distance from the diaphragm was chosen as a criterion because all the liver lesions are located inferior to the diaphragm level. Three different groups (0 < d < 5, 5 < d < 10, d > 10 cm) were considered and the only significant difference was found to be the percent change in SUV max for the liver lesions. There was the same general trend in both studies that the parameter variation decreases as tumor-diaphragm distance increases. Overall the liver and the esophageal lesions demonstrated less dependence on the respiratory artifacts, because the majority of those lesions are located further away from the diaphragm level than the lung lesions.

V. CONCLUSIONS

Our study demonstrated that using ACT for attenuation correction of the PET data could successfully reduce misregistration between the PET and CT data due to respiratory motion. The respiration artifacts can affect SUV max values and segmented GTVs in delineation of the esophagus and the liver lesions, having a potential to change the patient’s management and treatment. Smaller lesions located near the diaphragm are affected the most by respiratory motion. Based on our previous and current studies, ACT has become part of a standard clinical protocol in our institution for PET/CT acquisition for treatment planning of the lung and esophageal cancer patients.

REFERENCES

1. Beyer T, Townsend DW, Brun T, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med. 2000;41(8):1369–79.

2. Kluetz PG, Meltzer CC, Villemagne VL, et al. Combined PET/CT imaging in oncology: impact on patient management. Clin Positron Imaging. 2000;3(6):223–30.

3. Townsend DW, Beyer T. A combined PET/CT scanner: the path to true image fusion. Br J Radiol. 2002;75 Spec No:S24–30.

4. Bar-Shalom R, Yefremov N, Guralnik L, et al. Clinical performance of PET/CT in evaluation of cancer: additional value for diagnostic imaging and patient management. J Nucl Med. 2003;44(8):1200–09.

5. Hany TF, Steinert HC, Goerres GW, Buck A, von Schulthess GK. PET diagnostic accuracy: improvement with in-line PET-CT system: initial results. Radiology. 2002;225(2):575–81.

6. Weber WA, Petersen V, Schmidt B, et al. Positron emission tomography in non-small-cell lung cancer: prediction of response to chemotherapy by quantitative assessment of glucose use. J Clin Oncol. 2003;21(14):2651–57.

7. Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. J Nucl Med. 2007;48(7):1112–21.

8. Gould KL, Pan T, Loghin C, Johnson NP, Sdringola S. Reducing radiation dose in rest-stress cardiac PET/CT by single poststress cine CT for attenuation correction: quantitative validation. J Nucl Med. 2008;49(5):738–45.

9. Koepfli P, Hany TF, Wyss CA, et al. CT attenuation correction for myocardial perfusion quantification using a PET/CT hybrid scanner. J Nucl Med. 2004;45(4):537–42.

10. Di Carli MF, Dorbala S. Integrated PET/CT for cardiac imaging. Q J Nucl Med Mol Imaging. 2006;50(1):44–52.

11. Di Carli MF, Dorbala S, Meserve J, El Fakhri G, Sitek A, Moore SC. Clinical myocardial perfusion PET/CT. J Nucl Med. 2007;48(5):783–93.

12. Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. J Nucl Med. 2007;48(2):188–93.

13. Namdar M, Hany TF, Koepfli P, et al. Integrated PET/CT for the assessment of coronary artery disease: a feasibility study. J Nucl Med. 2005;46(6):930–35.

14. Keyes JW, Jr. SUV: standard uptake or silly useless value? J Nucl Med. 1995;36(10):1836–39.

15. Zasadny KR, Wahl RL. Standardized uptake values of normal tissues at PET with 2-[fluorine-18]-fluoro-2-deoxy-D-glucose: variations with body weight and a method for correction. Radiology. 1993;189(3):847–50.

16. Grégoire V, Haustermans K, Geets X, Roels S, Lonneux M. PET-based treatment planning in radiotherapy: a new standard? J Nucl Med. 2007;48 Suppl 1:68S–77S.

17. Mac Manus MP, Hicks RJ. Impact of PET on radiation therapy planning in lung cancer. Radiol Clin North Am. 2007;45(4):627–38, v.

18. Mac Manus MP, Hicks RJ, Ball DL, et al. F-18 fluorodeoxyglucose positron emission tomography staging in radical radiotherapy candidates with nonsmall cell lung carcinoma: powerful correlation with survival and high impact on treatment. Cancer. 2001;92(4):886–95.

19. Macapinlac HA. Clinical applications of positron emission tomography/computed tomography treatment planning. Semin Nucl Med. 2008;38(2):137–40.

20. Nestle U, Walter K, Schmidt S, et al. 18F-deoxyglucose positron emission tomography (FDG-PET) for the planning of radiotherapy in lung cancer: high impact in patients with atelectasis. Int J Radiat Oncol Biol Phys. 1999;44(3):593–97.

21. Paulino AC, Johnstone PA. FDG-PET in radiotherapy treatment planning: Pandora’s box? Int J Radiat Oncol Biol Phys. 2004;59(1):4–5.

22. Lavrenkov K, Partridge M, Cook G, Brada M. Positron emission tomography for target volume definition in the treatment of non-small cell lung cancer. Radiother Oncol. 2005;77(1):1–4.

23. Leong T, Everitt C, Yuen K, et al. A prospective study to evaluate the impact of FDG-PET on CT-based radiotherapy treatment planning for oesophageal cancer. Radiother Oncol. 2006;78(3):254–61.

24. Geets X, Daisne JF, Tomsej M, Duprez T, Lonneux M, Grégoire V. Impact of the type of imaging modality on target volumes delineation and dose distribution in pharyngo-laryngeal squamous cell carcinoma: comparison between pre- and per-treatment studies. Radiother Oncol. 2006;78(3):291–97.

25. Beyer T, Antoch G, Blodgett T, Freudenberg LF, Akhurst T, Mueller S. Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology. Eur J Nucl Med Mol Imaging. 2003;30(4):588–96.

26. Townsend DW, Beyer T, Blodgett TM. PET/CT scanners: a hardware approach to image fusion. Semin Nucl Med. 2003;33(3):193–204.

27. Osman MM, Cohade C, Nakamoto Y, Marshall LT, Leal JP, Wahl RL. Clinically significant inaccurate localization of lesions with PET/CT: frequency in 300 patients. J Nucl Med. 2003;44(2):240–43.

28. Beyer T, Antoch G, Müller S, et al. Acquisition protocol considerations for combined PET/CT imaging. J Nucl Med. 2004;45 Suppl 1:25S–35S.

29. Beyer T, Rosenbaum S, Veit P, et al. Respiration artifacts in whole-body (18)F-FDG PET/CT studies with combined PET/CT tomographs employing spiral CT technology with 1 to 16 detector rows. Eur J Nucl Med Mol Imaging. 2005;32(12):1429–39.

30. Kawano T, Ohtake E, Inoue T. Deep-inspiration breath-hold PET/CT of lung cancer: maximum standardized uptake value analysis of 108 patients. J Nucl Med. 2008;49(8):1223–31.

31. Meirelles GS, Erdi YE, Nehmeh SA, et al. Deep-inspiration breath-hold PET/CT: clinical findings with a new technique for detection and characterization of thoracic lesions. J Nucl Med. 2007;48(5):712–19.

32. Nehmeh SA, Erdi YE, Meirelles GS, et al. Deep-inspiration breath-hold PET/CT of the thorax. J Nucl Med. 2007;48(1):22–26.

33. Allen-Auerbach M, Yeom K, Park J, Phelps M, Czernin J. Standard PET/CT of the chest during shallow breathing is inadequate for comprehensive staging of lung cancer. J Nucl Med. 2006;47(2):298–301.

34. Aston JA, Cunningham VJ, Asselin MC, Hammers A, Evans AC, Gunn RN. Positron emission tomography partial volume correction: estimation and algorithms. J Cereb Blood Flow Metab. 2002;22(8):1019–34.

35. Nehmeh SA, Erdi YE, Ling CC, et al. Effect of respiratory gating on quantifying PET images of lung cancer. J Nucl Med. 2002;43(7):876–81.

36. Nehmeh SA, Erdi YE, Ling CC, et al. Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. Med Phys. 2002;29(3):366–71.

37. Nehmeh SA, Erdi YE, Pan T, et al. Four-dimensional (4D) PET/CT imaging of the thorax. Med Phys. 2004;31(12):3179–86.

38. Nehmeh SA, Erdi YE, Pan T, et al. Quantitation of respiratory motion during 4D-PET/CT acquisition. Med Phys. 2004;31(6):1333–38.

39. Boucher L, Rodrigue S, Lecomte R, Bénard F. Respiratory gating for 3-dimensional PET of the thorax: feasibility and initial results. J Nucl Med. 2004;45(2):214–19.

40. Nye JA, Esteves F, Votaw JR. Minimizing artifacts resulting from respiratory and cardiac motion by optimization of the transmission scan in cardiac PET/CT. Med Phys. 2007;34(6):1901–06.

41. Nagel CC, Bosmans G, Dekker AL, et al. Phased attenuation correction in respiration correlated computed tomography/positron emitted tomography. Med Phys. 2006;33(6):1840–47.

42. Grgic A, Nestle U, Schaefer-Schuler A, Kremp S, Kirsch CM, Hellwig D. FDG-PET-based radiotherapy planning in lung cancer: optimum breathing protocol and patient positioning – an intraindividual comparison. Int J Radiat Oncol Biol Phys. 2009;73(1):103–11.

43. Gilman MD, Fischman AJ, Krishnasetty V, Halpern EF, Aquino SL. Optimal CT breathing protocol for combined thoracic PET/CT. AJR Am J Roentgenol. 2006;187(5):1357–60.

44. de Juan R, Seifert B, Berthold T, von Schulthess GK, Goerres GW. Clinical evaluation of a breathing protocol for PET/CT. Eur Radiol. 2004;14(6):1118–23.

45. Nehmeh SA, Erdi YE. Respiratory motion in positron emission tomography/computed tomography: a review. Semin Nucl Med. 2008;38(3):167–76.

46. Pan T, Mawlawi O, Luo D, et al. Attenuation correction of PET cardiac data with low-dose average CT in PET/CT. Med Phys. 2006;33(10):3931–38.

47. Pan T, Mawlawi O, Nehmeh SA, et al. Attenuation correction of PET images with respiration-averaged CT images in PET/CT. J Nucl Med. 2005;46(9):1481–87.

48. Chi PC, Mawlawi O, Luo D, Liao Z, Macapinlac HA, Pan T. Effects of respiration-averaged computed tomography on positron emission tomography/computed tomography quantification and its potential impact on gross tumor volume delineation. Int J Radiat Oncol Biol Phys. 2008;71(3):890–99.

49. Vogel WV, van Dalen JA, Wiering B, et al. Evaluation of image registration in PET/CT of the liver and recommendations for optimized imaging. J Nucl Med. 2007;48(6):910–19.

50. Papathanassiou D, Becker S, Amir R, Menéroux B, Liehn JC. Respiratory motion artefact in the liver dome on FDG PET/CT: comparison of attenuation correction with CT and a caesium external source. Eur J Nucl Med Mol Imaging. 2005;32(12):1422–28.

51. Miyata H, Doki Y, Yasuda T, et al. Evaluation of clinical significance of 18F-fluorodeoxyglucose positron emission tomography in superficial squamous cell carcinomas of the thoracic esophagus. Dis Esophagus. 2008;21(2):144–50.

52. Pan T, Lee TY, Rietzel E, Chen GT. 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT. Med Phys. 2004;31(2):333–40.

53. Bradley J, Thorstad WL, Mutic S, et al. Impact of FDG-PET on radiation therapy volume delineation in non-small-cell lung cancer. Int J Radiat Oncol Biol Phys. 2004;59(1):78–86.

54. Fox JL, Rengan R, O’Meara W, et al. Does registration of PET and planning CT images decrease interobserver and intraobserver variation in delineating tumor volumes for non-small-cell lung cancer? Int J Radiat Oncol Biol Phys. 2005;62(1):70–75.

55. Weber WA, Ziegler SI, Thödtmann R, Hanauske AR, Schwaiger M. Reproducibility of metabolic measurements in malignant tumors using FDG PET. J Nucl Med. 1999;40(11):1771–77.

56. Young H, Baum R, Cremerius U, et al. Measurement of clinical and subclinical tumour response using [18F]-fluorodeoxyglucose and positron emission tomography: review and 1999 EORTC recommendations. European Organization for Research and Treatment of Cancer (EORTC) PET Study Group. Eur J Cancer. 1999;35(13):1773–82.


ACKNOWLEDGEMENTS

The authors would like to thank Rebecca Marsh, Ph.D. for the very helpful comments on thispaper.


a Corresponding author: Tinsu Pan, Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, 1515 Holcomb Blvd, Houston TX 77030, USA; phone: (713) 563-2714; fax: (713) 563-2720; email: tpan@mdanderson.org

(Return to Top)



Journal of Applied Clinical Medical Physics , Vol. 11 , No. 1 , Winter 2010