A quality assurance tool for helical tomotherapy using a step-wedge phantom and the on-board MVCT detector

Vincent Althof, Paul van Haaren, Rik Westendorp, Tonnis Nuver, Dinant Kramer, Marijke Ikink, Arjen Bel, Andre Minken

Abstract


The purpose of this study was to develop and evaluate filmless quality assurance (QA) tools for helical tomotherapy by using the signals from the on-board megavoltage computed tomography (MVCT) detector and applying a dedicated step-wedge phantom. The step-wedge phantom is a 15 cm long step-like aluminum block positioned on the couch. The phantom was moved through the slit beam and MVCT detector signals were analyzed. Two QA procedures were developed, with gantry fixed at 0°: 1) step-wedge procedure: to check beam energy consistency, field width, laser alignment with respect to the virtual isocenter, couch movement, and couch velocity; and 2) completion procedure: to check the accuracy of a field abutment made by the tomotherapy system after a treatment interruption. The procedures were designed as constancy tool and were validated by measurement of deliberately induced variations and comparison with a reference method. Two Hi-Art II machines were monitored over a period of three years using the step-wedge procedures. The data acquisition takes 5 minutes. The analysis is fully automated and results are available directly after acquisition. Couch speed deviations up to 2% were induced. The mean absolute difference between expected and measured couch speed was 0.2% ± 0.2% (1 standard deviation SD). Field width was varied around the 10 mm nominal size, between 9.7 and 11.1 mm, in steps of 0.2 mm. Mean difference between the step-wedge analysis and the reference method was < 0.01 mm ± 0.03 mm (1 SD). Laser (mis)alignment relative to a reference situation was detected with 0.3 mm precision (1SD). The step-wedge profile was fitted to a PDD in water. The PDD ratio D20/D10, measured at depths of 20 cm and 10 cm, was used to check beam energy consistency. Beam energy variations were induced. Mean difference between step-wedge and water PDD ratios was 0.2% ± 0.3% (1SD). The completion procedure was able to reveal abutment mismatches with a mean error of –0.6 mm ± 0.2 mm (1SD). The trending data over a period of three years showed a mean deviation of 0.4% ± 0.1% (1 SD) in couch speed. The spread in field width was 0.15 mm (1 SD). The sagittal and transverse lasers showed a variation of 0.5 mm (1 SD). Beam energy varied 1.0% (1 SD). A mean abutment mismatch was found of –0.4 mm ± 0.2 mm (1 SD) between interrupted treatments. The on-board MVCT detector, in combination with the step-wedge phantom, is a suitable tool for a QA program for helical tomotherapy. The method allowed frequent monitoring of machine behavior for the past three years.

Keywords


quality assurance; helical tomotherapy; step-wedge phantom; megavoltage CT detector

Full Text: PDF HTML