JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS

Application of a Monte Carlo-based method for total scatter factors of small beams to new solid state micro-detectors

Paolo Francescon, Stefania Cora, Carlo Cavedon,a Paolo Scalchi
Department of Medical Physics, ULSS6 "Vicenza", Vicenza, Italy

carlo.cavedon@ulssvicenza.it

Received 4 June 2008; accepted 17 October 2008

The scope of this work was to apply a method for estimation of total scatter factors of the smallest beams of the Cyberknife radiosurgery system to newly available solid-state detectors: the PTW 60008 diode, the SunNuclear EdgeDetectorTM diode, and the Thomson and Nielsen TN502RDM micromosfet. The method is based on a consistency check between Monte Carlo simulation of the detectors and experimental results, and was described in a recent publication. Corrected total scatter factors were in excellent agreement with the findings of the former study. The results showed that the diodes tend to overestimate the total scatter factor of small beams, probably due to excessive scatter from the material surrounding the active layer. The correction factor for diodes and for the micromosfet, however, was found to be independent of the electron beam width. This is a desirable characteristic because it allows standard correction factors to be used for treatment units of the same type, without the need of case-by-case Monte Carlo simulation.

PACS numbers: 87.55.kh; 87.55.ne; 87.56.Fc.

Key words: small beams, total scatter factor, Monte Carlo, radiosurgery, Cyberknife

I. Introduction

The scope of this work was to apply a recently published method for determination of total scatter factors (sc,p) of the smallest beams of the Cyberknife radiosurgery system (Accuray, Inc., Sunnyvale, CA) by means of multiple detectors.(1) A survey among users of the system revealed differences in sc,p values that are not consistent with actual differences between treatment units, thus calling for a method capable of accounting for effects that might alter the detector's response in very small fields. The method published in Francescon et al.(1) provided details for two microchambers, a diode and a diamond detector, but did not include the objects of this study. These are the PTW 60008 diode, that had been considered less appropriate with respect to the 60012 model because of the presence of a metal filter,(2) but nevertheless is widely used in clinical practice, and the SunNuclear EdgeDetectorTM model 1118 diode, that became available only recently. The Thomson and Nielsen TN502RDM micromosfet was also included because of its small active volume compared to other solid-state detectors and to the field size.

We believe that the method reported in Francescon et al.(1) and briefly described below would be useful to institutions that do not have access to Monte Carlo resources or that prefer to avoid the whole simulation process. In fact, the simulated range of beam parameters should cover any existing treatment unit of the same type. However, the method requires that the details of the detectors used for measurement be simulated within the framework of the study, to make simulation results available to other users. This study was aimed at providing such simulation for the three new detectors.

II. Materials and Methods

The characteristics of the three detectors are summarized in Table 1.

Table 1. Characteristics of the detectors.

Sc,p was defined as Dcoll/D60, where Dcoll was the dose measured with each collimator and D60 was the reference dose measured with the 60 mm collimator. Measurements were performed in water with the effective point of measurement placed at 80 cm from the source and 1.5 cm depth (78.5 source to surface distance), corresponding to the point of maximum dose along the depth dose curve of the 60 mm collimator. The PTW 60008 diode was used with the stem parallel to the beam axis. The SunNuclear diode was used with the stem perpendicular to the beam axis. However due to the design of the detector, this orientation would expose the active layer equivalently to the PTW diode. The micromosfet was exposed both with the flat side and the round side (bulb) towards the radiation source, and results were averaged. The average deviation between the two orientations was 1.1% (max 2.4%).

The detectors were centered on the horizontal plane by means of a laser pointer, then their position was finely tuned (within ±0.2 mm) to achieve the maximum signal intensity using the 5 mm collimator. All measurements were made with 50 monitor units and averaged over a series of at least 5 repeated runs. The standard deviation between repeated measurements was on average 0.9% (max 1.6%) for the micromosfet (within each orientation); the standard deviation was less than 0.1% for the diodes.

The method applied in this work and extended to the three detectors was described in Francescon et al.(1) and is summarized below. It consists of using both simulation and experiment to assess total scatter factors of circular beams obtained by means of the 5 mm, 7.5 mm and 10 mm collimators. Simulations were performed in the hypothesis of three different values of beam energy (6.5, 7.0 and 7.5 MeV) and four values of the full width at half maximum (FWHM) of the electron beam incident on the target (1.4, 1.8, 2.2 and 2.6 mm). The specification provided by the manufacturer was FWHM in the range 1.5 mm - 2.0 mm. Energy was chosen first, based on tissue-maximum ratios (TMR) data; the optimal value resulted to be 7.0 MeV.

The FWHM of the electron beam was determined indirectly by solving a linear system whose variables are a correction factor to be applied to raw sc,p measurements for each detector (Fcorr) and the estimated total scatter factor (sc,p):

(1)

where sc,pm is the measured total scatter factor and a and b are the parameters of a linear fit between the four possible values of the pair (Fcorr, sc,p) corresponding to the four values of the FWHM. A graphical interpretation of the linear system above is given in Figure 1. For the calculation of the linear fit between the pairs, sc,p is the Monte Carlo-calculated value in a small volume of water (i.e., without simulating the detector(1)) and Fcorr is the appropriate value taken from Table 2. Obviously, the solution to the linear system in terms of Fcorr and sc,p (denoted as F*corr and s*c,p) already responds to the goal of the study. In the graphical representation of Figure 1, the abscissa of the point of intersection is the estimated s*c,p, while the ordinate is the estimated F*corr.

   
Fig. 1. Graphical interpretation of the method based on a linear system. The figures describ the simulated values of sc,p and Fcorr in the hypothesis of the 4 different values of FWHM of the electron beam (dotted lines) and the result of the experimental measurement (continuous line). The abscissa of the point of intersection is the estimated s*c,p , while the ordinate is the estimated Fcorr. Fig. 1(a): PTW 60008 diode; the estimated s*c,p value is 0.678. Fig. 1(b): TN micromosfet; the estimated s*c,p value is 0.680. Fig. 1(c): SunNuclear "edge" diode; the estimated s*c,p value is 0.678.

Table 2. Fcorr of the 3 detectors for the 5, 7.5, and 10 mm collimators (as a function of the FWHM).

In this method, estimation of the FWHM is indirect, in other words it is a consequence rather than a prerequisite for simulation.

The absorbed dose in the active volume of the detectors embedded in a water phantom was simulated by means of CAVITY, an EGSnrc-based code(3,4) and the C++ class library, egspp, for use with the EGSnrc package, developed by Kawrakow.(5) All simulations were performed until the achieved uncertainty was less than 0.15% (one standard deviation). The transport parameters were ECUT = 512 keV (total energy for electrons), PCUT = 1 keV (total energy for photons), ESTEPE = 0.25, and XIMAX = 0.5. The electron multiple scattering and boundary crossing algorithm was PRESTA-II with the EXACT option. The composition of the detectors (whose details were provided by the manufacturers) has been simulated using PEGS4,(3) with a cutoff energy of AE = 512 keV for electrons and a photon energy cutoff, AP, of 1 keV. Density effect correction factors reported in ICRU Report 37(6) have been used in the Monte Carlo model of the detectors. The active volume of the PTW diode was described as a 60μm-thick layer, while for the SunNuclear diode a 22μm layer was used.(7,8). In fact, though the depletion region specified by the manufacturer is smaller (see Table 1), the active region is the diffusion length of the minority carriers (according to the studies reported in Rikner et al.(7) and Shi et al.(8)). For the micromosfet, the active volume was identified with the 0.5 μm thick SiO2 layer. Backscatter from the collimating system to the monitor chamber was not taken into account for all simulations. However, such effect was initially tested and found to be negligible (less than 0.5% on calculated sc,p). The effect is lower than in conventional linear accelerators, probably due to the peculiar geometry of the collimating system, the smaller volume of the monitor chambers, and the smaller reference field (28 cm2 circular versus 100 cm2 square).(9)

III. Results

Table 2 shows the values of Fcorr to be applied to raw measurements of sc,p as a function of the FWHM of the electron beam. This Table should be used to estimate F*corr and s*c,p for a treatment unit different from the one investigated in this study, according to the flowchart in Francescon et al.(1)

In other words, and to better explain the method, Fcorr in Table 2 is the ratio between the "supposed true" sc,p (simulated in water) and the expected sc,p (simulated in the active volume of the actual detector). In this sense, Fcorr is thus already determined by simulation without need of measurement; however, Fcorr is a function of the electron beam width (FWHM). The role of the measured sc,p is to compare with the simulated detector response (through linear system(1)) in order to pick up the correct value of beam width, thus univocally determining the Fcorr value and the "supposed true" sc,p.

The estimated F*corr and s*c,p for the three solid state detectors are reported in Table 3.

Figure 1 shows a graphical interpretation of the linear system whose solutions are the values given in Table 3. Only the data relative to the 5 mm collimator are reported. The estimated s*c,p values resulting from solution of the linear system are 0.678, 0.680 and 0.678 for the PTW 60008 diode, the TN micromosfet and the SunNuclear "edge" diode. Raw measurements gave 0.718, 0.712 and 0.714, respectively (see also Table 4).

Table 3. Estimated F*corr and s*c,p for the 5 mm, 7.5 mm, and 10 mm collimators for the 3 detectors simulated in this study.

Table 4. Measured and MC-simulated sc,p, for the 3 detectors and for the 5, 7.5, and 10 mm collimators for the various FWHM of the Gaussian spatial distribution of the electron source.

Table 4 reports measured and MC-simulated sc,p (that is, the expected measured value as a result of the Monte Carlo simulation of the detector) for the 3 detectors and for the 5, 7.5, and 10 mm collimators, in the hypothesis of the four different values of the FWHM of the Gaussian distribution of the electron source.

IV. Discussion & Conclusions

The results of this study confirmed some findings obtained in previous investigations.

When compared with Francescon et al.(1), the estimated values of sc,p in this study were on average 0.679 ± 0.002 versus 0.677 ± 0.004 for the 5 mm collimator, 0.822 ± 0.008 versus 0.820 ± 0.008 for the 7.5 mm collimator, and 0.875 ± 0.006 versus 0.871 ± 0.008 for the 10 mm collimator.

The correction factors for small solid state detectors were found to be lower than 1,which is consistent with previous studies and with the hypothesis of excessive scatter reaching the active layer due to the relatively high atomic number of the silicon substrate (and metal filters, if present) compared to water.

The study reported in Francescon et al. included one diode detector, for which the variation of Fcorr was found to be practically independent of the FWHM of the electron beam. The study reported here confirmed this behaviour for two different diodes, a desirable characteristic that allows users to apply a standard correction factor to the measured sc,p independently of fine tuning details of the treatment unit. This feature is shared by the TN micromosfet which, however, is affected by higher statistical uncertainty. On the other hand, correction factors for the micromosfet are closer to 1 compared to diodes _ apart for the smallest field (5 mm) for which even a small detector like the micromosfet produces a non-negligible perturbation. The microchambers and diamond detector previously investigated did not show independence of the correction factor from the size of the electron source, thus requiring a more complicated approach to the correction of experimental results.

References

1. Francescon P, Cora S, Cavedon C. Total scatter factors of small beams: a multidetector and Monte Carlo study. Med Phys. 2008; 35(2):504-13.

2. Griessbach I, Lapp M, Bohsung J, Gademann G, Harder D. Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams. Med Phys. 2005;32(12):3750-54.

3. Kawrakow I, Rogers DWO. The EGSnrc Code System: Monte Carlo simulation of electron and photon transport. Ottawa (ON): National Research Council of Canada; 2003. Technical Report PIRS-701 (4th printing).

4. Kawrakow I. Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc the new EGS4 version. Med Phys. 2000; 27(3):485-98.

5. Kawrakow I. EGSnrc C++ Class Library. Ottawa (ON): National Research Council of Canada; 2005. Technical Report PIRS-898.

6. International Commission on Radiation Units & Measurements. Stopping powers for electrons and positrons. Washington (DC); 1984. ICRU Report 37.

7. Rikner G, Grusell E. Effects of radiation damage on p-type silicon detectors. Phys Med Biol. 1983;28(11):1261-67.

8. Shi S, Simon WE. Modeling the instantaneous dose rate dependence of radiation diode detectors. Med Phys. 2003;30(9):2509-19.

9. Ding GX. Using Monte Carlo simulations to commission photon beam output factors—a feasibility study. Phys Med Biol. 2003;48(23):3865-74.


a Corresponding author: Carlo Cavedon, Department of Medical Physics, ULSS6 "Vicenza", Viale Rodolfi 37, 36100 Vicenza, Italy; phone: +39 0444 75 7308; fax: +39 0444 75 2557; email: carlo.cavedon@ulssvicenza.it

(Return to Top)

Journal of Applied Clinical Medical Physics, Vol. 10, No. 1, Winter 2009