Analysis of couch position tolerance limits to detect mistakes in patient setup
Department of Radiation Oncology Physics, The University of Michigan Medical School,
Ann Arbor, MI, USA swhadley@umich.eduReceived 21 April, 2008; accepted 18 May, 2009
This work investigates the use of the tolerance limits on the treatment couch position to detect mistakes in patient positioning and warn users of possible treatment errors. Computer controlled radiotherapy systems use the position of the treatment couch as a surrogate for patient position, and a tolerance limit is applied against a planned position. When the couch is out of tolerance, a warning is sent to a user to indicate a possible mistake in setup. A tight tolerance may catch all positioning mistakes while at the same time sending too many warnings; a loose tolerance will not catch all mistakes. We developed a statistical model of the absolute position for the three translational axes of the couch. The couch position for any fraction is considered a random variable xi. The ideal planned couch position xp is unknown before a patient starts treatment and must be estimated from the daily positions of xi. As such, xp is also a random variable. The tolerance, tol, is applied to the difference between the daily and planned position, di = xi - xp. The di is a linear combination of random variables and therefore the density of di is the convolution of distributions of xi and xp. Tolerance limits are based on the standard deviation of di such that couch positions that are more than two standard deviations away are considered out of tolerance. Using this framework, we investigated two methods of setting xp and tolerance limits. The first, called first day acquire (FDA), is to take couch position on the first day as the planned position. The second is to use the cumulative average (CumA) over previous fractions as the planned position. The standard deviation of di shrinks as more samples are used to determine xp and, as a result, the tolerance limit shrinks as a function of fraction number when a CumA technique is used. The metrics of sensitivity and specificity were used to characterize the performance of the two methods to correctly identify a couch position as in- or out-of-tolerance. These two methods were tested using simulated and real patient data. Five clinical sites with different indexed immobilization were tested. These were whole brain, head and neck, breast, thorax, and prostate. Analysis of the head and neck data shows that it is reasonable to model the daily couch position as a random variable in this treatment site. Using an average couch position for xp increased the sensitivity of the couch interlock and reduced the chances of acquiring a couch position that was a statistical outlier. Analysis of variation in couch position for different sites allowed the tolerance limit to be set specifically for a site and immobilization device. The CumA technique was able to increase the sensitivity of detecting out-of-tolerance positions while shrinking tolerance limits for a treatment course. Making better use of the software interlock on the couch positions could have a positive impact on patient safety and reduce mistakes in treatment delivery.
PACS number: 87.55.Ne, 87.55.Qr, 87.55.tg, 87.55.tm
Key words: patient safety, treatment errors, record and verify, immobilization
I. INTRODUCTION
Computer controlled radiotherapy and record and verify systems (R&V) were introduced to
allow for complex treatments and increase the safety of radiation delivery.(1–7) It was found
that these systems improve patient safety and reduce treatment errors when used properly.(8-14)
The work presented here focuses on the use of one aspect of the R&V system related to the
software interlock applied to the position of the treatment couch. This software interlock is used
to provide a level of automatic oversight to the patient setup by applying a tolerance limit to a
baseline position of the treatment couch. When used properly this interlock may aid in warning
users of potential mistakes in the patient setup.
We use the term “mistake” to mean that the process to setup the patient has a failure that
puts the patient in the wrong position for treatment. We use this term to distinguish this type of
error from the more common “setup error” or “positioning error”, which is often used to mean
the systematic and random geometric displacement of a target from its planned position. Some
examples of setup mistakes are: setting the incorrect source to surface distance, setting up to
the wrong mark on a patient with multiple isocenters, not applying a necessary shift from a
setup mark or even using the wrong treatment plan as a result of a patient identification error.
Correct setups would be those that used the correct information and in which all setup instructions
were followed correctly.
This aspect of the R&V systems uses the treatment couch location as a surrogate for patient
position and the software interlock acts as a binary classifier to determine correct and incorrect
setups. The software will allow treatment to continue if the couch position is within a preset
tolerance limit, or it can trigger a software interlock if the position is out of tolerance. In this
work, we investigated statistical methods to determine baseline values and tolerance limits for
the position of the treatment couch for a patient's course. The goal is to make the best use of
this software system to improve patient safety.
X-ray imaging is the main method to correct patient positioning errors. Much work has been
done on online and offline strategies for correcting systematic and random errors in patient position.
(15-18) Imaging may or may not be used every day depending on what strategy is used. There is
also a possibility that the interpretation of the image is incorrect and a mistake in patient position
is undetected. Some radiation therapy treatments don't use X-ray imaging at all, and thus don't
benefit from these methods. Errors in the patient setup may be present when imaging is not used.
In these situations the R&V system software interlock on the couch position may be the only
electronic and automatic check of patient setup. Given that R&V systems are installed and in use
in the vast majority of radiation therapy clinics, careful analysis of their use seems warranted.
Much work has been done to improve immobilization of patients for many different treatment
sites.(19-26) When an immobilization device is indexed to the table, it attaches rigidly in
the same place for each fraction and, theoretically, improves the coupling between the digital
position readout of the couch and the patient's placement with respect to the isocenter. This
would improve the ability of the R&V system to detect mistakes in setup.
Previous work in radiation therapy investigated the use of the R&V system to understand
variations in machine parameters and their possible impact on patient safety. Podmaniczky et
al.(4) recognized the utility of the R&V system to collect data on patient setup, and analyzed
the variations that exist in the axes of the treatment machine. They used statistical analysis of
recorded histories to set tolerance limits for patient setups. Patton et al.(11) compiled a review
of errors and determined, among other things, that indexed immobilization along with couch
position tolerance limits was an important part of achieving correct patient setup. They also
recognized the interplay between couch tolerance limits and usability by therapists to detect
incorrect patient positioning. Klein et al.(27) used different tolerance limits based on indexed
immobilization and the treatment type.
In this work, we critically analyze the use of the couch digital position readout as a surrogate
for the patient position. We investigate the question of how the couch digital can be used as
part of the quality assurance process to eliminate gross mistakes in treatments. Couch positions
from patient treatment records are used to determine the variation in couch locations for
different treatment sites and immobilization devices. The ability of indexed immobilization to
reduce variation in digital readout of the treatment couch is investigated by calculating standard
deviations from mean couch positions for a course of treatment. We develop a statistical analysis
of patient data in order to set planned couch positions that improve the ability of the tolerance
limit to detect out-of-tolerance patient setups. To test our methods, we employ the metrics of
sensitivity and specificity to quantify the ability of different techniques to correctly classify
couch positions as being in or out of tolerance. This type of quantitative analysis may allow
users to adjust the tolerance limits to control the expected number of warnings sent to users.
II. MAtERIALS AND METHODS
We consider the position of the couch on any given fraction to be an independent random variable
drawn from the probability density function p(x) where x is one of the translational axes
of the treatment couch. The density function p(x) will have both correct and incorrect patient
setups. The actual form of p(x) will depend on the isocenter location, treatment site, positioning
errors, and the indexed immobilization that is used for the treatment. The presence of systematic
positioning errors is not explicitly dealt with in the following method. Instead, we rely on the
patient positioning and imaging protocols to deal with possible systematic errors in position
that may be present in a patient setup.
We take a statistical approach to the problem of setting planned couch values and tolerance
limits for patient treatments. Let the random variable x represent one axis of the couch, which
is sampled from the density function p(μ,σ) with a mean μ and standard deviation σ. We assume
that the couch position x for each fraction is independent and identically distributed. The
ideal planned couch position, xp, must be μ. We assume that couch positions that are far from
μ indicate a possible mistaken setup. The variance σ indicates the ability of the immobilization
device to reposition the patient on the table, as well as random error that may exist in the patient
positioning. Current treatment delivery software allows for a planned value xp to be set for each
field, while the tolerance limit is set from a smaller set of tables that cannot be patient-specific.
The tolerance limit is applied symmetrically around the planned position.
A. Average Couch Position
For any given patient treatment, we assume that μ and σ are not known prior to treatment. The planned couch position, xp, must be estimated from positions obtained for each fraction. Because we assume that the couch position from any fraction is a random variable, then the estimated planned couch position is also a random variable. The ideal unknown planned couch position μ can be estimated using an average over previous fractions during a patient's course of treatment,
(1) |
where xi is the couch position on the ith fraction and n is the number of previous fractions. The estimation of the average position, xp, improves with each fraction such that the standard deviation of xp is σ/√n. xp is a random variable with density function p(μ, σ/√n).
B. Couch deviations from a planned position
The tolerance limit must be compared to the difference from the planned position and the couch position for that fraction, di = xi - xp. Because di is the difference of two random variables, the density function of di is the convolution of p(μ, σ) with p(μ, σ/√n) with zero mean and the standard deviation, σd, taken in quadrature.
(2) |
The density function of differences di always has a standard deviation equal to or larger than the
couch positions xi and is a function of the fraction number n. Equation 2 provides a method for
shrinking the tolerance limit as more couch positions are averaged together, and the estimate of
xp improves if the assumption that xi is independent and identically distributed holds for a given
patient treatment. This assumption can be violated by clinically relevant situations — such as
adjustments to patient setup (due to weight loss that requires the couch to move systematically
to compensate), or the use of an imaging protocol that adjusts position on days when imaging
is used but not on other days.
C. Strategies for setting planned couch values and tolerance limits
We attempt to balance two goals when setting tolerance limits. One is to set tolerance limits
with tight geometric constraints, and the other is to set the limit at a level that will not send too
many false warnings to an operator that may be a clinical burden and lead him/her to mistrust
the warning. We chose to set the tolerance limit to trigger a warning on 5% of the fractions
treated on average.
Two strategies for setting planned couch values and tolerance limits were investigated. The
first is based on what current treatment control software allows for in a clinical setting. For
example, our software system, Varis (Varian Medical Systems, Palo Alto CA), allows for each
field to have its own planned couch position with a tolerance table attached to it. A limited
number of tolerance tables can be defined and used in the R&V system. Based on these limitations,
the couch position on the first fraction is used as the planned position for the remaining
fractions. The tolerance limit, tol, is set to twice the population standard deviation of the differences
di based on Eq. 2, where n = 1, tol = 2σpop√2. With this tolerance, couch positions
should be identified as out of tolerance 4.5% of the fractions on average. We call this method
the First Day Acquire (FDA) method.
The second strategy uses averaging of couch positions to improve the planned couch value
and bring it closer to the unknown average position for a patient. We call this the cumulative
average (CumA) method. Once again the tolerance limit is set equal to twice σd but we use Eq.
2 to shrink the limit as the number of samples in the average increase.
(3) |
Using this method, the tolerance limit becomes a function of the fraction number when n is greater than 1 and will shrink as more patient specific information is derived. For the first fraction, the difference di is always zero. Therefore, the tolerance limit has no meaning. There is no ability to determine if there is a mistake in the setup based on the couch position on the first fraction.
D. Performance evaluation
The software interlock is acting like a binary classifier to make a decision about setups that may or may not have errors. Specificity and sensitivity are often used to characterize the performance of binary classifiers. To evaluate the performance of these two methods, we compare specificity and sensitivity estimates. For the clinical data, couch positions that were more than 2σpop√2 away from the patient's mean overall fractions were chosen to represent positions that were out of tolerance. This defined the ground truth data to determine which couch positions were in or out of tolerance. When the FDA and CumA techniques were applied to the datasets, their ability to detect out-of-tolerance couch positions was tested against the ground truth data. Specificity and sensitivity were calculated from the number of true positives (NTP), false positives (NFP), true negatives (NTN), and false negatives (NFN). The percentage of couch positions that were determined as out of tolerance was also calculated.
(4) |
E. Testing using clinical and artificial data
These two methods were tested using clinical and artificial data. The simulation data was used
to test the two methods under ideal conditions and served as a basis to compare to the performance
using clinical data. For any given dataset, the couch positions that were more than two
standard deviations away from the mean were considered out of tolerance.
Clinical patient data was obtained under an institutional review board approved retrospective
study. Couch coordinates from patients treated in our department were used to test the two
methods of setting planned couch positions and tolerance limits. The position of the couch was
taken from the treatment field history, and represents the actual couch position for treatment
after patient setup and any image guidance. Patients were stratified by treatment site and type
of indexed immobilization device. The treatment sites were whole brain, head and neck (H&N)
intensity-modulated radiation therapy (IMRT), breast, thorax and prostate IMRT. The immobilization
devices used were the Sinmed Posifix for whole brain and H&N setups, Posiboard
for breast setups, and the Posirest (all by Civco Medical Solutions, Kalona, Iowa) for thorax
setups. All immobilization devices were indexed to the table top. Prostate patients did not use
an immobilization device and were not indexed to the treatment table. The IMRT treatments
used daily imaging to correct for systematic and random errors. Prostate patients were aligned
to implanted gold markers and H&N cases were aligned to bony anatomy. Non-IMRT cases
used an imaging protocol to detect systematic patient positioning errors within the first five
fractions of treatment and then used weekly imaging thereafter. Table 1 summarizes the number
of patients and fractions for the five groups.
The head and neck IMRT data provided the best opportunity to analyze the statistical nature
of the digital couch position due to the use of good indexed immobilization and daily image
guidance. The mean of each patient for each couch axis was subtracted for the raw couch
coordinates. Statistical analysis was done on this dataset to characterize the variation between
patients by calculating each patient's standard deviation for each axis of the treatment couch.
A population standard deviation was calculated for each treatment fraction to study trends in
variation over the course of treatment.
For each of the five sites, the population standard deviation, spop, was calculated and used
to determine the couch positions that were greater than 2spop from the patient's mean overall
fractions. This was used as ground truth data to which the FDA and CumA techniques were
compared. For each of five sites and two techniques, the specificity and sensitivity were calculated
across all fractions.
For the simulated data, a single translation of the treatment couch was modeled as a zero mean
Gaussian distribution with a standard deviation of 1 cm. A 35-fraction treatment was simulated
for 1,000,000 patients. The couch position for each fraction was determined using a random
number generator, and the FDA and CumA techniques were applied to the same dataset. The
sensitivity, specificity, and percentage of out-of-tolerance warnings were calculated for the two
methods as a function of fraction number as well as overall fractions.
III. RESULTS
A. Clinical data
The standard deviation for each H&N patient was calculated and plotted as a histogram in Fig. 1. The population standard deviation within a fraction is graphed in Fig. 2. For this well-indexed and immobilized treatment site with daily image guidance there is no indication that the table position on any given fraction is more correct than any other.
Population standard deviations and tolerance limits for the two methods for the three
translational axes of the couch for the five treatment sites are shown in Table 2. The tolerance
limit for the FDA technique is calculated from Eq. 3, using n=1. For the CumA technique, the
tolerance limit is a function of the fraction number. The value in Table 2 is the limit near the
end of a typical number of fractions for that body site. Histograms of couch positions after the
patient average has been subtracted can be seen in Fig. 3. The two sites shown are H&N and
prostate IMRT. These two sites used daily image guidance to reduce systematic and random
setup errors for each fraction. Despite image guidance, variation still exists in the position of
the treatment couch during a course of treatment.
An example of the FDA and CumA technique applied to the vertical couch axis of a selected
breast patient's daily couch position is shown in Fig. 4. This case was chosen because the first
fraction had the largest deviation from the average for the vertical position of the couch. For
the FDA method, each day except for one is out of tolerance. The CumA technique incorporates
the planned value each day to the average of the previous fractions. The two methods disagree
with respect to the couch being in or out of tolerance for most fractions.
Sensitivity and specificity for the FDA and CumA techniques are summarized in Tables 3
and 4. The results of the simulation data are included for comparison. Results for the five sites
generally show that sensitivity increased when the CumA technique was used. The exception
is the whole brain dataset where sensitivity decreased. This dataset was investigated and it
was found that the first fraction represented the mean better than the cumulative average up to
about the 6th fraction treated. This may be a result of the specifics of the patient setup where
the marks on the immobilization mask are set to lasers, and setup imaging is only used on the
first day and then every 5th fraction thereafter.
The percentage of treatments marked as out of tolerance are presented in Table 5. The Table
shows the percentage of fractions that were determined to be out of tolerance by the two techniques.
Additionally, the percentage of fractions that have any of the axes as out of tolerance
is presented. The cumulative average technique reduced the number of out-of-tolerance warnings
for each individual axis, as well as overall for any axis. For four of the five sites studied,
the out-of-tolerance warning occurred on about 10% of all fractions treated. This would mean
that, on average, fractions treated would have an axis out of tolerance for one out of every 10
fractions treated.
B. Artificial data
A graph of the sensitivity and specificity of the simulation data as a function of fraction number
is shown in Fig. 5. It can be seen that on the first fraction, when no information is known about
where the couch should be, the tolerance limit is insensitive to mistakes in patient setup. By
the second fraction, the first day acquire (FDA) method has set the planned couch position, and
the sensitivity has moved to 0.33 and is unchanged for the remaining fractions. The cumulative
average technique, CumA, is identical to the FDA method for the first two fractions. With
the third fraction, the cumulative average increased the sensitivity of the couch tolerance limit
to detect out-of-tolerance couch positions. By the last fraction, 35, the CumA technique had
increased the sensitivity to 0.84, while the FDA method was at 0.33.
The sensitivity over all 35 fractions for the FDA and CumA techniques was 0.32 and 0.72,
respectively. The specificity for FDA and CumA was 0.97 and 0.99, respectively. The specificity
for both methods started off at 1.0 and then dropped to 0.97 on the second fraction where
the FDA remains for all 35 fractions. The CumA method improves the specificity over the 35
fractions to 0.99. Both techniques maintained a rate of out-of-tolerance warnings of 4.6% for
each fraction after the first. The CumA technique had the advantage of a shrinking tolerance
level while maintaining a constant rate of out-of-tolerance warnings.
IV. DISCUSSION
Record and verify systems were created to increase the accuracy and safety of radiation therapy
treatments. Much of the work on quality assurance and R&V systems has been to verify that
the correct data is entered into the system. While many users report on the quality assurance
of the R&V systems, few have focused on how the R&V system participates in the quality assurance
of treatment delivery. In this work, we focused on a software interlock applied to the
daily couch position that is intended to exert control over possible mistakes made in the patient
setup process. In order to understand how well this system can reduce possible mistakes in
patient setup, we analyzed the variation that exists in the system by using recorded treatment
histories from patients treated in our department.
The results in Table 2 clearly indicate that different sites require different tolerance limits if
an operator wants to maintain the same rate of warning to the users. If one tolerance limit were
set for all sites, then some sites would trigger more interlocks and receive additional attention
while another site may not benefit at all. The policy in our department is that, when a couch
parameter is out of tolerance, a procedure is triggered such that the radiation therapists review
the essential parts of the patient setup to check for any mistakes. Additional X-ray imaging may
be performed to verify the positioning. If a tolerance were set too tight, therapists would be
investigating a large number of patient setups that have no mistakes. This would be a burden on
a busy clinic for only limited improvement in the quality of patient setups. Knowing the actual
variation that exists in couch positions for a given immobilization device would allow one to
set tolerance limits to balance the needs of clinical efficiency with good oversight of patient
setup. How to set those limits is a policy issue that should involve discussion with physicians
and administrators, and should consider the ability of the couch digital readout — in conjunction
with the immobilization device — to provide good information about possible mistakes in
patient setup.
The results of the H&N data validate the idea that the position of the treatment couch can
be considered a random variable with an unknown average and standard deviation. There is no
evidence in the data to suggest that the position of the couch on any given fraction is better or
more correct than for another fraction. Given that, it would seem reasonable to use averaging
to estimate a better baseline value for the couch position for the software interlock. The result
of estimating a better baseline value for the software interlock makes the system more sensitive
to positions that are out of tolerance. Using an average position also guards against acquiring
a table position that is an outlier in the distribution of table positions even though the target
position may be correct. The simulation experiments showed large increase in sensitivity due
to averaging and shrinking tolerance limits. This could be an advantage to users of the system
for it increases their trust that the system is providing accurate information.
The sensitivities reported in Table 3 show that the interlock system is much less sensitive
than the ideal (1.0). How sensitive the system is depends on its tolerance limit. In this work, we
used Eq. 3 to set tolerance limits that were statistically consistent with the underlying data. For
example, the tolerance limit on the lateral couch position for prostate patients is over 4.5 times
larger than for a well-indexed head and neck patient. If a more sensitive system is desired, one
either needs to improve the quality of the immobilized indexing or shrink the tolerance limit.
Shrinking the tolerance limit without improving the indexed immobilization would decrease the
specificity. Despite the relativity loose tolerances in Table 2 and the low sensitivities in Table
3, Table 5 shows that anywhere from 4.5% to 20% of fractions will have an out-of-tolerance
warning, with most sites triggering an interlock on over 10% of fractions treated.
Much work has been done on the management of systematic and random errors in target
position for conformal radiation therapy. The results of serial imaging have been used to determine
the average positioning error and to apply a shift to future treatments to reduce systematic
errors. Random errors in target position are often dealt with by using daily imaging to correct
the target position. It is not the purpose of the software interlock on the couch readout to
provide oversight of the target positioning, but rather provide some level of quality assurance
against mistakes in positioning. In this work, we ignored the presence of systematic errors in the
target position. If a systematic error were found and corrected by use of an imaging protocol,
that same shift could be applied to the average couch position to produce a baseline value that
incorporates the systematic error.
V. CONCLUSIONS
In this work, we investigated a new method to set planned couch positions and adapt the tolerance limits based on updated information about the patient setup. It was shown that by adapting the planned position and tolerance limit based on data obtained during treatment, the R&V system could make better use of the software interlock. Increases in patient safety may be possible by modifying R&V systems to take into account possible errors.
ACKNOWLEDGEMENTS
This work was supported by NIH grant P01CA59827.
REFERENCES
- Perry H, Mantel J, Lefkofsky M. A programmable calculator to acquire, verify and record radiation treatment parameters from a linear acceleration. Int J Radiat Oncol Biol Phys. 1976;1(9-10):1023–26.
- Dickof P, Morris P, Getz D. Vrx: a verify-record system for radiotherapy. Med Phys. 1984;11(4):525–27.
- Mohan R, Podmaniczky K, Caley R, Lapidus A, Laughlin J. A computerized record and verify system for radiation treatments. Int J Radiat Oncol Biol Phys. 1984;10(10):1975–85.
- Podmaniczky K, Mohan R, Kutcher G, Kestler C, Vikram B. Clinical experience with a computerized record and verify system. Int J Radiat Oncol Biol Phys. 1985;11(8):1529–37.
- Glisch C, Limberg C, Barber-Derus S, Gillin M. Clinical application of a record and verify system in radiation oncology. Radiol Technol. 1988;59(3):251–53.
- Miller R, van de Geijn J. The use of a bar code scanner to improve the utility and flexibility of record and verify systems used in radiation therapy. Med Phys. 1988;15(4):611–13.
- Fraass B, McShan D, Kessler M, Matrone G, Lewis J, Weaver T. A computer-controlled conformal radiotherapy system. I: Overview. Int J Radiat Oncol Biol Phys. 1995;33(5):1139–57.
- Leunens G, Verstraete J, Van den Bogaert W, Van Dam J, Dutreix A, van der Schueren E. Human errors in data transfer during the preparation and delivery of radiation treatment affecting the final result: “garbage in, garbage out”. Radiother Oncol. 992;23(4):217-22.
- Fraass B, Lash K, Matrone G, et al. The impact of treatment complexity and computer-control delivery technology on treatment delivery errors. Int J Radiat Oncol Biol Phys. 1998;42(3):651–59.
- Barthelemy-Brichant N, Sabatier J, Dewé W, Albert A, Deneufbourg J. Evaluation of frequency and type of errors detected by a computerized record and verify system during radiation treatment. Radiother Oncol. 1999;53(2):149–54.
- Patton G, Gaffney D, Moeller J. Facilitation of radiotherapeutic error by computerized record and verify systems. Int J Radiat Oncol Biol Phys. 2003;56(1):50–57.
- Huang G, Medlam G, Lee J, et al. Error in the delivery of radiation therapy: results of a quality assurance review. Int J Radiat Oncol Biol Phys. 2005;61(5):1590–95.
- Klein E, Drzymala R, Purdy J, Michalski J. Errors in radiation oncology: a study in pathways and dosimetric impact. J Appl Clin Med Phys. 2005;6(3):81–94.
- Williams M. Radiotherapy near misses, incidents and errors: radiotherapy incident at Glasgow. Clin Oncol. 2007;19(1):1–3.
- Bortfeld T, van Herk M, Jiang SB. When should systematic patient positioning errors in radiotherapy be corrected? Phys Med Biol. 2002;47(23):N297–302.
- de Boer HC, Heijmen BJ. A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload. Int J Radiat Oncol Biol Phys. 2001;50(5):1350–65.
- Bel A, van Herk MB, Bartelink H, Lebesque JV. A verification procedure to improve patient set-up accuracy using portal images. Radiother Oncol. 1993;29(2):253–60.
- Litzenberg DW, Balter JMB, Lam KL, Sandler HM, Ten Haken RK. Retrospective analysis of prostate cancer patients with implanted gold markers using off-line and adaptive therapy protocols. Int J Radiat Oncol Biol Phys. 2005;63(1):123–33.
- Thornton A, Ten Haken R, Gerhardsson A, Correll M. Three-dimensional motion analysis of an improved head immobilization system for simulation, CT, MRI, and PET imaging. Radiother Oncol. 1991;20(4):224–28.
- Bentel G, Marks L, Sherouse G, Spencer D. A customized head and neck support system. Int J Radiat Oncol Biol Phys. 1995;32(1):245–48.
- Bentel G, Marks L, Hendren K, Brizel D. Comparison of two head and neck immobilization systems. Int J Radiat Oncol Biol Phys. 1997;38(4):867–73.
- Carter D, Marks L, Bentel G. Impact of setup variability on incidental lung irradiation during tangential breast treatment. Int J Radiat Oncol Biol Phys. 1997;38(1):109–15.
- Fuss M, Salter B, Rassiah P, Cheek D, Cavanaugh S, Herman T. Repositioning accuracy of a commercially available double-vacuum whole body immobilization system for stereotactic body radiation therapy. Technol Cancer Res Treat. 2004;3(1):59–67.
- Kim S, Akpati H, Kielbasa J, et al. Evaluation of intrafraction patient movement for CNS and head & neck IMRT. Med Phys. 2004;31(3):500–06.
- Engelsman M, Rosenthal S, Michaud S, et al. Intra- and interfractional patient motion for a variety of immobilization devices. Med Phys. 2005;32(11):3468–74.
- Lovelock D, Hua C, Wang P, et al. Accurate setup of paraspinal patients using a noninvasive patient immobilization cradle and portal imaging. Med Phys. 2005;32(8):2606–14.
- Klein EE, Drzymala RE, Williams R, Westfall LA, Purdy JA. A change in treatment process with a modern record and verify system. Int J Radiat Oncol Biol Phys. 1998;42(5):1163–68.
a Corresponding author: Scott W. Hadley, Department of Radiation Oncology Physics, The University of Michigan Medical School, 1500 E. Medical Center Drive, Ann Arbor, MI, USA; phone: 1.734.936.4309; fax: 1.734.936.7859 email: swhadley@umich.edu
Journal of Applied Clinical Medical Physics, Vol. 10, No. 4, Fall 2009