Journal of Applied Clinical Medical Physics

Comparison of IPSM 1990 photon dosimetry code of practice with IAEA TRS-398 and AAPM TG-51.

Silvia Vargas Castrillóna and Francisco Cutanda Henríquez

NW Medical Physics, Christie Hospital NHS Foundation Trust, Manchester, UK.

stvcastrillon@hotmail.com

Received 2 January 2008; accepted 2 September 2008

Several codes of practice for photon dosimetry are currently used around the world, supported by different organizations. A comparison of IPSM 1990 with both IAEA TRS-398 and AAPM TG-51 has been performed. All three protocols are based on the calibration of ionization chambers in terms of standards of absorbed dose to water, as it is the case with other modern codes of practice. This comparison has been carried out for photon beams of nominal energies: 4 MV, 6 MV, 8 MV, 10 MV and 18 MV.

An NE 2571 graphite ionization chamber was used in this study, cross-calibrated against an NE 2611A Secondary Standard, calibrated in the National Physical Laboratory (NPL). Absolute dose in reference conditions was obtained using each of these three protocols including: beam quality indices, beam quality conversion factors both theoretical and NPL experimental ones, correction factors for influence quantities and absolute dose measurements. Each protocol recommendations have been strictly followed. Uncertainties have been obtained according to the ISO Guide to the Expression of Uncertainty in Measurement. Absorbed dose obtained according to all three protocols agree within experimental uncertainty. The largest difference between absolute dose results for two protocols is obtained for the highest energy: 0.7% between IPSM 1990 and IAEA TRS-398 using theoretical beam quality conversion factors.

PACS number: 87.55.tm

Key words: photon beam dosimetry, code of practice, absorbed dose to water, IPSM, IAEA, AAPM

I. Introduction

IPSM (Institute of Physical Sciences in Medicine) 1990,(1) AAPM (American Association of Physicist in Medicine) Task Group 51,(2) and the IAEA (International Atomic Energy Agency) Technical Reports Series 398(3) codes of practice are based on the calibration of ionization chambers in terms of absorbed dose to water. This approach is not limited to these codes of practice but is a worldwide trend which includes: German DIN 1997,(4) Japanese Association of Radiological Physics,(5) and the Swiss Society of Radiobiology and Medical Physics.(6) DIN in Germany(4,7) and IPSM in the United Kingdom(1,8) were pioneers in the use of standards of absorbed dose to water in their dosimetry protocols. The new standards of calibration of ionization chambers in terms of absorbed dose to water offer the possibility of reducing the uncertainty in the dosimetry of radiotherapy beams, provide a more robust system of primary standards than previous air-kerma based standards,(3) and allow the use of a more straightforward formalism.


There are similarities and differences among these protocols concerning the choice of quality index, calibration setup, standards of ionization chamber calibration and nomenclature. AAPM TG-51 and IAEA TRS-398 have already been discussed in great detail in the literature,(9-13) particularly concerning the advantages and disadvantages of the recommended photon beam quality indices.(14-17)

1A. Calibration

There are different methods of establishing primary standards of absorbed dose to water: ionization chamber (Bureau International des Poids et Mesures in France), water calorimetry (National Research Council in Canada, National Institute of Standards and Technology in the USA), graphite calorimetry (National Physical Laboratory in the UK, Ente per le Nuove Tecnologie, l'Energia e l'Ambiente in Italy, Australian Radiation Protection and Nuclear Safety Agency in Australia, Bundesamt für Eich- und Vermessungswesen in Austria, Laboratoire Primaire de Métrologie des Rayonnements Ionisants in France).

The three codes of practice under study here are based on standards of absorbed dose to water; nevertheless, there are some peculiarities regarding the way calibrations are dealt with. The IPSM 1990 protocol is based on the UK National Physical Laboratory ion chamber calibration service, which provides calibration coefficients for a range of beam quality indices.(8,18-21) Their primary standard is a graphite calorimeter, and three ion chambers (NE 2561/2611, NE Technology Ltd, Reading, UK) are used as reference standards in graphite. TG-51 is based on ion chambers with absorbed dose to water calibration coefficients for 60Co quality Q0 and sets of beam quality conversion factors. IAEA TRS-398 provides the most general and flexible framework for calibration, allowing four very detailed possibilities that include the use of experimental or theoretical beam quality conversion factors (see Section 4.1, IAEA TRS-398).

IAEA TRS-398 recommends the use of generic experimental beam quality conversion factors only if they have been determined in a standards laboratory, such as NPL. Values must be obtained from a large sample of ionization chambers and the standard deviation of chamber to chamber variations should be small.(3)

Beam quality conversion factors are not included in the IPSM 1990 protocol, however NPL provides experimental values together with ND,w in their most recent calibration certificates (2007). ND,w factors are based on determinations at three or more beam qualities. Based on NPL×s vast experience with ion chamber types NE 2561/NE 2611, a generic fit of kQ versus TPR20,10 (tissue phantom ratio at 20 cm deep, normalized to 10 cm) is generally used and the ion chamber to be calibrated is checked at several beam qualities. No generic fit is used for other chamber types, and measurements for a full range of beam qualities are carried out instead.

Despite the use of different protocols, all reference dose measurements in this work are traceable to NPL standards. Thus, this comparison is free from any influence related to differences among primary standards and methodologies used in standards laboratories. In a more general situation, when comparing results obtained following different protocols, each institution could be using ionization chambers traceable to different standards and, therefore, dose results for each institution would also include primary standard differences. These differences among standards could lead to differences in calibration factors up to 0.66% according to a BIPM report.(3,21-22)

1B. Quality Indices and Reference Conditions

The IAEA TRS-398 and IPSM 1990 recommend TPR20,10 as the quality index in order to choose appropriate calibration coefficients (ND,wQ) or beam quality conversion factors, while AAPM TG-51 uses %dd(10)x (photon component of the percentage depth dose at 10 cm depth).

According to the IPSM 1990 protocol, dose should be measured at 5 cm deep for TPR20,10: 0.58-0.75 and 60Co; at 7 cm deep for TPR20,10: 0.75-0.81. The AAPM TG-51 and IAEA TRS-398 recommend a depth of 10 cm, although IAEA TRS-398 allows 5 cm if TPR20,10 is less than 0.7.

In all the protocols, the energy determination is done with a 10 × 10 cm2 field at 100 cm SSD. The calibration can be done at the nominal SSD/isocenter location of the accelerator (80 or 100 cm) according to IAEA TRS-398 and AAPM TG-51.

1C. Beam Quality Conversion Factors

Beam quality conversion factors play an important role in IAEA TRS-398 (kQ,Qo, when Q0 is 60Co: kQ = kQ,Qo) and AAPM TG-51 (kQ) protocols. Both AAPM TG-51 and IAEA TRS-398 provide sets of theoretically derived kQ factors for a number of ionization chambers, although IAEA TRS-398 tables provide factors for a larger number of ionization chambers. IAEA has made an effort to include a wide range of ionization chambers used worldwide in TRS-398; details on the calculation procedures that lead to these kQ values are given in Appendix B of the IAEA TRS-398 code of practice, along with uncertainty estimates for each component. The kQ values are based on Bragg-Gray theory with suitable corrections. The combined standard uncertainty in the values for kQ is 1.0%.(3,23,24) TG-51 concentrates on those chambers used commonly in North America at the time of its publication and includes recommendations on how to use chambers not listed in the protocol.

For TRS-398 beam quality conversion factors can also be determined experimentally from experimental calibration coefficients for the beam qualities Q and Q0 using the definition given by Hohlfeld,(7)

(1)

With this method, variations among individual ionization chambers of the same model can be taken into account.

Standards laboratories can take advantage of the cancellation of factors in both calibration coefficients in this equation to reduce the uncertainty in kQ,Qo (IAEA TRS-398 notation). Comparison between experimental and theoretical data from IAEA TRS-398 along with their uncertainties can be found in the comprehensive review by Andreo.(24)

A user provided with a set of calibration coefficients NDQ,w for different beam qualities Q can compute kQ,Qo from them, dividing every ND,Qw by NDQ0,w, in fact IAEA TRS-398 recommends this practice. If kQ,Qo factors are computed following this procedure, their uncertainty has to be derived from the one given for ND,Qw and NDQ0,w in the calibration certificate. This uncertainty value could be larger than the one in standards laboratory measurements, because the user has to assume that ND,Qw and NDQ0,w have independent uncertainty.(25-27)

1D. The IPSM 1990 code of practice

There are additional similarities and differences between IPSM 1990 and AAPM TG-51/IAEA TRS-398 that are interesting to highlight:

  1. IPSM 1990 does not include detailed recommendations about several issues: desired characteristics of field ion chambers, design characteristics of the waterproof sleeve to be used with non-waterproof field ion chambers, water phantoms, wall thickness when horizontal beams are used, and practical considerations such as limits on leakage current and influence quantities (only generic recommendations are provided).
  2. IPSM 1990 and IAEA TRS-398 use TPR20,10 as quality index, while AAPM TG-51 uses %dd(10)x. It should be noted that TRS-398 includes an empirical relationship between TPR20,10 and PDD20,10.
  3. IPSM 1990 is based on the calibration of ionization chambers at NPL. Since this code of practice does not include beam quality conversion factors, it would not be possible to use IPSM 1990 for ionization chambers calibrated in standards laboratories which do not provide a calibration at several beam qualities, except for the calibration of a 60Co unit.
  4. Neither IPSM 1990 nor AAPM TG-51 offers any explicit guidelines to estimate measurement uncertainty in a user's beam. IAEA TRS-398 provides detailed guidelines.
  5. IPSM 1990 does not give recommendations about the chamber type to be used when measuring central axis depth dose distributions; specifically, no information is provided about the effective point of measurement of an ionization chamber. TRS-398 and TG-51 recommend shifting cylindrical chambers by 0.6 times the inner radius of a cylindrical ion chamber for photon beams.
  6. Recombination factors are treated in two different ways in IPSM 1990: the experimental formula by Burns and Rosser(28) (valid only for the NE2561/2611 Secondary Standard [-200 V polarizing voltage]), and Boag's theory.(29) There is no mention of fits like that by Weinhous and Meli(30), or any approximate relationship such as the ones reported in the other two protocols.

The aims of this paper are to obtain absolute dose results for a set of five photon beam qualities according to the three different codes of practices, and discuss the differences between IPSM 1990 and IAEA TRS-398/AAPM TG-51. Attention has been paid to the different methods mentioned in IAEA TRS-398 and the ones by Rogers(31) and by Kalach and Rogers(32) to obtain TPR20,10 and their effect on the value of the beam quality conversion factor and, therefore, in absorbed dose to water. Uncertainty has been evaluated according to ISO Guide to the Expression of Uncertainty in Measurement (GUM) and is discussed below.

II. Materials and Methods

Measurements were performed on two Elekta Synergy (Fleming Way, Crawley, UK) and two Elekta Precise linear accelerators. Photon energies used were 4 MV, 6 MV, 8 MV, 10 MV and 18 MV.

Measurements were carried out with a computer controlled scanner in a large water tank (MP3 beam analyzer, PTW Freiburg, Germany). For central axis depth dose, measurements were performed with a PTW 34001 (Roos type) plane parallel chamber and a 0.125 cm3 PTW 31002 cylindrical chamber; both chambers were connected to a PTW Tandem electrometer. The reference setup corresponds to a 10 × 10 cm2 field size and SSD = 100 cm.

IAEA TRS-398 recommends the use of plane parallel chambers for depth-ionization measurements. Both chamber types were used in order to compare their measurements since no chamber type is explicitly recommended by IPSM 1990 or by AAPM TG-51. Cylindrical chambers were shifted 0.6 times their cavity radius in order to place their effective point of measurement at the measuring point, and plane parallel chambers were placed with the centre of the front surface of the chamber air cavity at the measuring point.(2,3)

An NE 2571 graphite chamber connected to a Keithley 35040 Therapy Dosimeter electrometer (Keithley Instruments Inc., Cleveland, OH) was used for absolute dosimetry and TPR20,10 measurements. This ionization chamber is not waterproof; therefore it was used with a waterproofing sleeve, meeting the requirements of both IAEA TRS-398 and AAPM TG-51 (a PMMA sleeve with a wall not thicker than 1 mm and an air gap of less than 0.2 mm between the chamber and the sleeve).(2,3)

A 0.125 cm3 PTW 31002 waterproof thimble chamber was placed inside the phantom at the same depth of the NE 2571 but 4 cm away, and the ratio of both readings was recorded to account for beam output instabilities, following IAEA TRS-398 recommendations.

2A. Quality indices

Several procedures were used for quality index calculation. For 4 MV, 6 MV and 8 MV, %dd(10)x was obtained as a measured %dd(10). For higher energy beams the %dd(10) needs to be corrected for signals resulting from electron contamination at Dmax. This can be done either by using a generic formula for a typical linac, or by using a 1 mm lead filter to remove the contaminating electrons and replace them with the better known contamination from the lead itself. For 10 MV and 18 MV %dd(10)Pb was measured, placing a lead filter at 52.5 cm distance from water surface and using the equations in the protocols to remove the signal from electrons generated in the lead from the PDD measurements. For the highest energy, the alternative equation for the electron contamination from a typical linac (eq. 15 in TG-51) was also used to compute %dd(10)x from open beam %dd(10) (18 MV is the only beam for which it was applicable).

TPR20,10 was directly measured, keeping the chamber in the same position during the measurements and changing the water level instead. TPR20,10 was also computed with the relationship given in TRS-398 (Section 6.3.1) from PDD(10) values, the fit in the publication by Followill(33) which uses the ratio PDD20,10 and the ones proposed by Rogers(31) and Kalach and Rogers(32) using %dd(10)x.

2B. Absolute dosimetry

A graphite chamber NE 2571 (cavity length 24.0 mm, cavity radius 3.2 mm, wall thickness 0.065 gcm-2 and aluminum central electrode) was chosen for these measurements in order to fulfill all three protocol recommendations. The reference instrument in our institution is a NE 2611A Secondary Standard cylindrical chamber (cavity length 9.2 mm, cavity radius 3.7 mm, graphite wall, wall thickness 0.090 gcm-2 and aluminum central electrode), the one recommended by the National Physical Laboratory (Teddington, Middlesex, UK), along with a PTW UNIDOS electrometer. Cross-calibrations of field chambers take place on a regular basis to ensure traceability to the secondary standard. A regular schedule of constancy checks with Sr-90 check sources is carried out for both NE 2571 and NE 2611A. All ionization chambers were X-rayed before and after this study, and were found in proper condition.

The side-by-side method was used to cross-calibrate the NE 2571 ion chamber against the Secondary Standard.(1) Both ion chambers were placed in a perspex phantom meeting the requirements described in IPSM 1990, with two built-in holes and appropriate perspex inserts to allow chambers to fit in. The inserts can be exchanged along with the chambers. Three sets of measurements were performed; after each, the inserts were exchanged. Cross-calibrations were carried out at three different depths, corresponding to the reference conditions in the three protocols: 5 cm for 4 MV, 6 MV, 8 MV and 10 MV; 7 cm for 18 MV; 10 cm for all energies. The independence of ND,w with depth has been evaluated in several publications.(34,35) In our measurements, variations in ND,w between different depths were approximately 0.1% for all energies, well within experimental uncertainty.

Reference conditions for absolute dose measurements were 100 cm SSD, 10 × 10 cm2 field size, 10 cm depth in water for IAEA TRS-398 and AAPM TG-51. For IPSM 1990, the measurement depths were 5 cm for 4 MV, 6 MV, 8 MV and 10 MV and 7 cm for 18 MV. Corrections for environmental conditions (pressure, temperature) were applied as recommended. Humidity was always within 20% to 70% and temperature within 15° C and 25° C and no humidity correction was necessary. Correction factors (polarity, ion recombination) were measured at the corresponding depth for each beam quality and protocol. No shift was applied, so that in every case the center of the ion chamber was placed at the reference depth according to the different protocols.

Ion recombination correction factors were measured with the two voltage technique according to each protocol. Polarity correction factors were computed as

(2)

(default polarity is negative). After every change of polarity or absolute voltage, the electrometer and ion chamber were allowed to stabilize for 20 minutes. Four non-trending readings were taken for each polarity and voltage. The recombination factors were computed according to the quadratic fit by Weinhous and Meli using the approximation given in IAEA TRS-398 and AAPM TG-51. A factor of 1+X% for a X% decrease in the reading when halving the voltage is the recombination factor recommended by IPSM 1990 and it has been used for the NE 2571 ion chamber.(1)

For the secondary standard chamber NE 2611A with -200 V polarizing voltage, NPL and IPSM 1990 protocol provide a formula to obtain the ion recombination factor from the value of the dose per pulse: k = 1.0014 + 0.23p (where p is the dose per pulse in the chamber in cGy.(28)) The dose per pulse can be obtained using the PRF (Pulse Repetition Frequency) setting for each beam and the dose rate. This relationship was used for the Secondary Standard chamber as it is the one used at the time of calibration at NPL.

Cross-calibrations were carried out for the NE 2571 field chamber against the Secondary Standard chamber leading to the corresponding ND,w field for each beam quality. Corrections for influence quantities were taken into account: pressure, temperature, ion recombination, polarity effect, and electrometer calibration. Recombination factors were obtained for the field instrument using the two voltage technique for the field chamber, and using the NPL equation (based on dose per pulse) for the Secondary Standard. Since the NE 2571 ionization chamber was cross-calibrated with the Secondary Standard for each beam, only kQ factors for the Secondary Standard NE 2611A are needed.

2C. Uncertainty

Uncertainties were computed according to the guidelines given in IAEA TRS-398 Appendix B and D, in accordance with ISO GUM.(25) Uncertainties are reported as standard uncertainty (k = 1) within this study. In general, a Type A evaluation was performed for each series of raw measurements. Whenever required, uncertainties were composed using the law of propagation of uncertainty where independent uncertainties are added in quadrature.(25)

III. Results & Discussion

Standard uncertainty propagation was used for direct measurements of TPR20,10. The fit given in IAEA TRS-398 (section 6.3.1) for TPR20,10 derived from PDD20,10 measurement has a maximum difference with regard to measured data of 0.6% for beam qualities below 50 MV. That means that the Type B standard uncertainty due to the fit in this case is 0.35%, assuming a rectangular probability distribution, and this uncertainty was composed with the Type B uncertainty from the PDD measurements. The same approach was used for the fit by Kalach and Rogers,(32) who report a maximum uncertainty for their fit of measured data on clinical beams of 0.011. On the other hand, we could not find any figure for uncertainty in the quality index relationship given by Followill,(33) although the uncertainty interval (k = 2) in the measured values used to design the fit (from nearly 700 machines) is shown in a graph. Thus, only Type B uncertainty from PDD measurements is reported in our results. Finally, no uncertainty estimate was found for the fit by Rogers.(31) Therefore, only Type B uncertainty from PDD measurements is reported in our results for these two fits.

The uncertainty for the AAPM TG-51 quality index %dd(10)x is the Type A one derived from the law of propagation of uncertainty for energies where the use of a lead filter is not recommended. When the quality index is obtained from measurements with a lead filter, an additional Type B uncertainty from the equation is used. Rogers(36) found that by using the fit, the worst case scenario would be that the measured value of %dd(10)x is wrong by 0.50%, leading to a standard uncertainty of 0.29%.

Quality indices are used to determine beam quality correction factors, kQ. The variation of kQ over the whole range of quality indices in Table I in AAPM TG-51 is 0.047 (%dd(10)x from 50% to 93%),(2) and in Table 6.III in IAEA TRS-398 it is 0.057 (TPR20,10 from 0.50 to 0.84).(3) Therefore, quality index uncertainty yields a smaller uncertainty in kQ.

Concerning beam quality conversion factors kQ in a previous publication, Andreo(37) reports uncertainty estimates of 1.2% to 1.5% for theoretical kQ. These values were reduced in TRS-398 to 1% and reported by Andreo.(24) For kQ factors provided by NPL, we have considered an uncertainty value of 0.7%, based on the report by Sharpe.(38)

AAPM quality indices %dd(10)x are shown in Table 1. Agreement between depth dose percentage values measured with a plane parallel and a cylindrical ion chamber was found to be within ± 0.1% for all beam qualities; PDDs measured with plane parallel chamber were used. As mentioned previously, an additional estimate for %dd(10)x for 18 MV has been obtained using equation 15 in TG-51. The result, 79.74%, is close to the one obtained from %dd(10)Pb, 80.04%, and agreement is within experimental uncertainty. The kQ factors computed with each of these values are 0.975 for %dd(10)Pb = 79.74%, and 0.976 for %dd(10)Pb = 80.04%.

Table 1. Values for the quality index %dd(10)x.

Values for the quality index TPR20,10 can be found in Table 2. The largest discrepancy (1.7%) between quality indices obtained with different procedures corresponds to 18 MV for the TPR20,10 values obtained using Rogers' and the IAEA relationships from PDD data – which are 0.8% higher and 1.0% lower than measured TPR20,10 respectively. Uncertainty for TPR20,10 associated with the fits by Followill(33) and Rogers(31) is not available. Their Type B uncertainty due to the PDD measurement is shown in the table, along with values of uncertainty for the IAEA, Kalach and Rogers(32) fits, and the experimental values.

Table 2. Values for the quality index TPR20,10: directly measured (obtained with Followill's equation from PDD20,10, with the equation in TRS-398 from PDD10, and with the fits given by Rogers and Kalach).

Results for beam quality conversion factors (NE 2611) for the five beam qualities, according to theoretical tables in IAEA TRS-398 and AAPM TG-51 and experimental results from NPL, are plotted in Figure 1. Differences in the quality index TPR20,10 obtained by different procedures would lead to a kQ difference of 0.4% for the case of TRS-398 theoretical kQ factors and of 0.6% for experimental kQ (Tables 3 and 4). Differences between experimental and theoretical kQ are found to be larger for the highest energies as reported by Andreo(24) but the differences are well within the error bars for all energies.

   

Fig. 1. Beam quality conversion factors kQ as a function of nominal energy. Graphs are shown for theoretical kQ from IAEA TRS-398 (using measured TPR20,10), AAPM TG-51 and experimental kQ from NPL.

Table 3. Theoretical kQ factors from IAEA TRS-398 and AAPM TG-51 tables, computed with each of the TPR20,10 obtained by different methods.

Table 4. Experimental kQ factors from NPL calibration certificate for the ion chamber NE2611, computed with each of the TPR20,10 obtained by different methods.

Polarity factors and ion recombination factors were obtained at the corresponding depths (10 cm for IAEA TRS-398 and AAPM TG-51, 7 cm and 5 cm for IPSM 1990). The maximum difference for ion recombination factors, 0.2%, has been found between IPSM 1990 and AAPM TG-51 for the 18 MV beam (1.008 for TG-51 and 1.010 for IPSM 1990). Despite the difference in depth, recombination factors for IPSM are within 0.1% of the values for the other two protocols and the remaining beam qualities. The differences in recombination factors between IAEA TRS-398 and AAPM TG-51, measured at the same depth, are 0.1% or less for all beam qualities. Uncertainty in recombination factor determination is below 0.2% when care is taken to ensure the measuring system is given adequate time to settle after changes in polarizing voltage. The polarity factor is identical for TRS-398 and TG-51 and is less than 0.1% different for IPSM 1990, and its uncertainty is below 0.3%.

Absolute absorbed dose to water in reference conditions has been finally computed in four different ways, using the common formalism given by:

(3)

where M* is the reading corrected for all influence quantities. It has been divided by the appropriate PDD value to obtain the absorbed dose at maximum. The four methods are:

  1. IAEA TRS-398, with NDCo,w and experimental beam quality conversion factors for the secondary standard provided by NPL.
  2. IAEA TRS-398, with NDQ,w factors computed applying theoretical kQ factors (from IAEA TRS-398 tables) to the NDCo,w provided by NPL.
  3. AAPM TG-51, with NDQ,w factors computed applying theoretical kQ factors (from AAPM TG-51 tables) to the NDCo,w provided by NPL.
  4. IPSM 1990, measuring at 5 or 7 cm deep instead of 10 cm deep, and using NDQ,w calibration coefficients from the NPL calibration certificate.

The ratios of absorbed dose obtained with IAEA TRS-398 and AAPM TG-51 to IPSM 1990 were found to be close to 1.005 (Table 5). Huq(9) reported overall differences between IAEA TRS-398 and AAPM TG-51 of 0.2% with a single case of 0.3% for 18 MV. Our results are in agreement with these values for these two codes of practice (AAPM TG-51 results compared with IAEA TRS-398 computed from theoretical beam quality correction factors). The work by Huq does not address the case of experimental beam quality correction factors. The reason for the differences between the first and second column in Table 5 (IAEA absolute dose results using theoretical and experimental kQ) is related to the differences in the beam quality correction factors (see Tables 3 and 4).

Table 5. Absolute dose quotients of IAEA TRS-398/IPSM 1990 and AAPM TG-51/IPSM 1990. For IAEA TRS-398, results for experimental and theoretical kQ factors are included.

IV. Conclusions

IPSM 1990 photon dosimetry code of practice has been compared with AAPM TG-51 and IAEA TRS-398, and the similarities and differences between IPSM 1990 and AAPM TG-51/IAEA TRS-398 have been discussed.

Uncertainties have been computed for all experimental results according to the ISO Guide for the Expression of Uncertainty in Measurement recommendations(25). Absorbed dose to water determinations according to the three protocols agree within experimental uncertainty, and this uncertainty is similar to the one reported in IAEA TRS-398. The maximum difference in absorbed dose to water determination is obtained for 18 MV: IPSM 1990 result is 0.7% lower than the IAEA TRS-398 one using its theoretical beam quality conversion factors. This maximum difference is mainly related to the use of experimental beam quality conversion factors for IPSM 1990 and theoretical ones for IAEA TRS-398 (Fig. 1). It must be stressed that this comparison has been performed with a single calibration from NPL and that the use of different protocols and calibrations traceable to different standards laboratories will add further differences increasing or lowering results (see Section 2.2 and Table 2.2 in IAEA TRS-398).

Acknowledgements

The authors are grateful to Huey Yang for his careful revision of the manuscript and to Lisa Rodan for grammar revision.

References

1. IPSM (Institute of Physical Sciences in Medicine). Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service. Phys. Med. Biol. 1990;35:1355–60.

2. Almond PR, Biggs PJ, Coursey BM et al. AAPM Task Group 51: Protocol for clinical reference dosimetry of high-energy photon and electron beams. Med. Phys.1999(9);26: 1847–70.

3. Andreo P, Burns DT, Hohlfeld K et al. Absorbed dose determination in external beam radiotherapy: an international Code of Practice for dosimetry based on standards of absorbed dose to water. Vienna (Austria): IAEA Technical Report Series No. 398; 2000.

4. DIN (Deutsches Institut fur Normung). Dosismessverfahren Nach der Sondenmethode fur Photonen- und Elektronenstrahlung Teil 2: Ionisationsdosimetrie Deutsche Norm. Berlin (Germany): DIN 6800-2;1997.

5. JARP Japanese Association of Radiological Physics. A practical code for the dosimetry of high-energy photon and electron beams in radiotheraphy. Tokyo (Japan): Tsusho-sangyo-kenkyusya;1986.

6. Swiss Society of Radiobiology and Medical Physics. High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers. Bern (Switzerland): SSRMP Recommendations No. 8;2000.

7. Hohlfeld K. The standard DIN 6800: Procedures for absorbed dose determination in radiology by the ionization method (IAEA-SM-298/31). Vienna (Austria): Dosimetry in Radiotherapy Vol 1;1988:13–22.

8. Burns JE, Dale JWG, Dusautoy AR et al. New calibration service for high-energy X-irradiation at NPL. Vienna (Austria): Dosimetry in Radiotherapy Vol 2;1988:125–32.

9. Huq MS, Andreo P, Song H. Comparison of the IAEA TRS-398 and AAPM TG-51 absorbed dose to water protocols in the dosimetry of high-energy photon and electron beams. Phys. Med. Biol. 2001;46(11):2985-3006.

10. Araki F, Kubo HD. Comparison of high-energy photon and electron dosimetry for various dosimetry protocols. Med. Phys. 2002;29(5):857-68.

11. Landoni V, Malatesta T, Capparella R, Fragomeni R, Prioelli A, and Begnozzi L. Comparison of the IAEA TRS-398 Code of Practice and the AAPM TG51 Protocol for dosimetry calibration of high energy photon and electron beam. Physica Medica 2003;XIX(1):31–35.

12. Huq MS, Andreo P. Advances in the determination of absorbed dose to water in clinical high-energy photon and electron beams using ionization chambers. Phys. Med. Biol. 2004;49:R49-R104.

13. López-Medina A, Teijeiro A, Salvador F, et al. Comparison between TG-51 and TRS-398: electron contamination effect on photon beam-quality specification. Phys. Med. Biol. 2004;49(1):17-32.

14. Kosunen A, Rogers DWO. Beam quality specification for photon beam dosimetry. Med. Phys. 1993;20(4):1181-88.

15. Andreo P. On the beam quality specification of high-energy photons for radiotherapy dosimetry. Med. Phys. 2000;27(3):434-40.

16. Rogers DWO. Comment on `On the beam quality specification of high-energy photons for radiotherapy dosimetry'. Med. Phys. 2000;27(3):441-44.

17. Andreo P. Reply to Comment on "On the beam quality specification of high-energy photons for radiotherapy dosimetry". Med. Phys. 2000;27(7):1693-95.

18. Burns DT, McEwen MR, Williams AJ. An NPL absorbed dose calibration service for electron beam radiotherapy. In Measurement Assurance in Dosimetry. Vienna: International Atomic Energy Agency, 1994:61–71.

19. Morgan AM, Aird EGA, Aukett RJ et al. IPEM guidelines on dosimeter systems for use as transfer instruments between the UK primary dosimetry standards laboratory (NPL) and radiotherapy centres. Phys. Med. Biol. 2000;45(9):2445-57.

20. Burns JE. Absorbed-dose calibrations in high-energy photon beams at the National Physical Laboratory: conversion procedure. Phys. Med. Biol. 1994;39(10):1555-75.

21. Allisy-Roberts PJ, Burns DT. Comparisons and calibrations at the BIPM, Report of the Consultative Committee for Ionizing Radiation CCRI(I)/99-1. Sevres (France): BIPM, Pavillon de Breteuil, 1999.

22. Kessler C, Allisy-Roberts PJ, Burns DT, Krauss A, and Kapsch R-P. Comparison of the standards for absorbed dose to water of the PTB, Germany and the BIPM for 60Co gamma rays, 2006. Metrologia, 2006;43:Tech Suppl. 06005.

23. Nuttbrown RF, Duane S, Shipley DR, Thomas RAS. Evaluation of factors to convert absorbed dose calibrations for graphite to water for the NPL high-energy photon calibration service. Phys. Med. Biol. 2002;47(3):441-54.

24. Andreo P. A comparison between calculated and experimental kQ photon beam quality correction factors. Phys. Med. Biol. 2000;45(9):L25–L38.

25. International Standards Organization. Guide to the Expression of Uncertainty in Measurement. Geneva (Switzerland); 1995.

26. Taylor BN, Kuyatt CE. Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results. NIST Technical Note 1297. Gaithersburg (MD): National Institute for Standards and Technology; 1994.

27. United Kingdom Accreditation Service. The Expression of Uncertainty and Confidence in Measurement. Feltham, Middlesex (UK): UKAS; 2007.

28. Burns JE, Rosser KE. Saturation correction for the NE 2560/1 dosimeter in photon dosimetry. Phys. Med. Biol. 1990;35(5):687-93.

29. Boag JW. Ionization measurements at very high intensities. Pulsed radiation beams. Br J Radiol 1950;23(274):601-11.

30. Weinhous MS, Meli JA. Determining Pion, the correction factor for recombination losses in an ionization chamber. Med. Phys. 1984;11(6):846-849.

31. Rogers DWO. Fundamentals of dosimetry based on absorbed dose standards Teletherapy Physics, Present and Future. In: Palta JR, Mackie TR, editors. AAPM Summer School 1996:319–56.

32. Kalach N I, Rogers D W O. Which accelerator photon beams are ``clinic-like'' for reference dosimetry purposes? Med. Phys. 2003;30(7):1546-55.

33. Followill DS, Tailor RC, Tello VM, and Hanson WF. An empirical relationship for determining photon beam quality in TG-21 from a ratio of percent depth doses. Med. Phys. 1998;25(7 Pt 1):1202–05.

34. Boutillon M, Perroche AM. Determination of calibration factors in terms of air kerma and absorbed dose to water in the 60Co gamma rays. IAEA/WHO Network of Secondary Standards Dosimetry Laboratories (SSDL) Newsletter. IAEA 1993;32:3-13.

35. Booth A, Rogers DWO. Monte Carlo study of effects of phantom size, radial position, and depth on photon beam calibration. NRC Report PIRS–507. Ottawa (Canada): NRC Canada; August 1995.

36. Rogers DWO. Correcting for electron contamination at dose maximum in photon beams. Med. Phys. 1999;26(4):533–37.

37. Andreo P. Absorbed dose beam quality factors for the dosimetry of high-energy photon beams. Phys. Med. Biol. 1992;37(12):2189-2211.

38. Sharpe P. Progress report on radiation dosimetry at NPL. Report CCRI(I)/99-20.


a Corresponding author: Silvia Vargas Castrillón, NW Medical Physics, Christie Hospital NHS Foundation Trust, Wilmslow Road, Withington, M20 4BX, Manchester, UK; phone: 01614463531; fax: 01614463545; email: stvcastrillon@hotmail.com

(Return to Top)


Journal of Applied Clinical Medical Physics, Vol. 10, No. 1, Winter 2009