Comparison of IPSM 1990 photon dosimetry code of practice with IAEA TRS-398 and AAPM TG-51.
Abstract
A NE 2571 graphite ionization chamber was used in this study, cross-calibrated against a NE 2611 Secondary Standard, calibrated in the National Physical Laboratory (NPL). An absolute dose study has been performed according to each of these three protocols: beam quality indices, beam quality conversion factors both theoretical and NPL experimental ones, correction factors for influence quantities and absolute dose measurements. Each protocol recommendations have been strictly followed. Uncertainties have been obtained according to the ISO Guide to the expression of Uncertainty in Measurement. Absorbed dose obtained according to all three protocols agree within experimental uncertainty. The largest difference between absolute dose results for two protocols is obtained for the highest energy: 0.7% between IPSM 1990 and IAEA TRS-398 using its own theoretical beam quality conversion factors.