A seed specific dose kernel method for low-energy brachytherapy dosimetry

John J. DeMarco, Timothy D. Solberg, Nzhde Agazaryan

Abstract


We describe a method for independently verifying the dose distributions from pre- and post-implant brachytherapy source distributions. Monte Carlo calculations have been performed to characterize the three-dimensional dose distribution in water phantom from a low-energy brachytherapy source. The calculations are performed in a voxelized, Cartesian coordinate geometry and normalized based upon a separate Monte Carlo calculation for the seed specific air-kerma strength to produce an absolute dose grid with units of cGyhr1U1. The seed-specific, three-dimensional dose grid is stored as a text file for processing using a separate visual basic program. This program requires the coordinate positions of each seed in the pre- or post-plan and sums the kernel file for a three-dimensional composite dose distribution. A kernel matrix size of 818181 with a voxel size of 1.01.01.0mm3 was chosen as a compromise between calculation time, kernel size, and truncation of the stored dose distribution as a function of radial distance from the midpoint of the seed. Good agreement is achieved for a representative pre- and post-plan comparison versus a commercial implementation of the TG-43 brachytherapy dosimetry protocol. 2003 American College of Medical Physics.