Monte Carlo dosimetric
characterization of the
IsoAid ADVANTAGE 103Pd
brachytherapy source
Keith T. Sowards
Department of Radiation Oncology, University of Louisville, James Brown Cancer Center, Louisville, Kentucky
ktsowa01@louisville.edu
Received 13 October 2006; accepted 5 January 2007
For roughly 25 years, 125I and 103Pd sources have been used in the treatment of various malignant diseases such as prostate cancer. Various new sources have been marketed and produced to meet the demand for new sources to use in treatment. Recently, IsoAID LLC created the ADVANTAGE 103Pd source. Various dosimetric parameters must be determined to facilitate treatment planning using this source. Theoretical determination of dosimetric characteristics, dose rate constant, radial dose function, and anisotropy function for this new source followed the American Association of Physicists in Medicine (AAPM) Task Group 43U1 recommendations. Theoretical calculations were performed in liquid water using the PTRAN Monte Carlo code version 7.44. The radial dose function of the new source was calculated in liquid water at distances up to 10.0 cm, and the anisotropy function, at distances ranging from 0.5 cm to 7.0 cm. The anisotropy factors and anisotropy constant were derived from the anisotropy function. The results in water indicate that the dose rate constant is 0.709 ± 0.014 cGy·h1·U1 and that the anisotropy constant is 0.880 ± 0.040. The dosimetric characteristics of this new source compare favorably with those of other commercially available 103Pd sources.
PACS: 87.53.Jw
Key words: 103Pd, dosimetry, Monte Carlo, brachytherapy
Keith T. Sowards, University of Louisville, James Brown Cancer Center, Department of Radiation Oncology, 529 South Jackson Street, Louisville, KY 40202 U.S.A.; phone: 502-5612700; fax: 502-5612709; email: ktsowa01@louisville.edu
I. INTRODUCTION
For approximately 25 years, 103Pd brachytherapy sources have been produced for use in interstitial implants in various tumor sites. Sources using 103Pd are favored because their low-energy photon emissions provide a rapid decrease in dose with increasing distance, minimizing the dose to normal tissues.
Use of ultrasound-guided brachytherapy seed implantation for prostate cancer has increased greatly since the technique was developed in the early 1980s. The reasons for the increase are many, including patient convenience (one treatment versus many for external-beam radiotherapy), reduced side effects as compared with radical prostatectomy, and greater cost effectiveness.(14) With a shortage of available seeds and an increase in the number of procedures being performed nationally, several manufacturers, including IsoAid (IsoAid LLC, Port Richey, FL), have developed new 103Pd sources to meet the increasing demand.
II. MATERIALS AND METHODS
A. Source of 103Pd
The ADVANTAGE 103Pd source has a physical length of 4.5 mm and an outer diameter of 0.8 mm. To create the four active 0.5-mm polystyrene spheres, 103Pd isotope is absorbed throughout the spheres, which are then encapsulated in a 0.05-mmthick titanium capsule (Fig. 1). A silver rod, 0.5 mm in diameter and 1.25 mm in length, serves as an X-ray marker and is placed between the pair of spheres at each end of the capsule. An active length of 3.4 mm was assumed for this source during calculation of the source parameters.
B. Monte Carlo simulation
Dose distributions for the ADVANTAGE source were calculated in liquid water using the PTRAN Monte Carlo code.(57) Simulations were performed for up to 10 million histories divided into one hundred batches. By combining this number of histories with use of a distance-and-attenuation-averaged, bounded, next-flight point-kerma estimator(5) standard errors about the mean (67% confidence intervals) ranging from 1.5% (near the source: r < 3 cm) to 5%6% (far from the source: r > 5 cm) were achieved.
One assumption made in the Monte Carlo simulation of the dose distributions around the source was that a uniform distribution of the 103Pd isotope is present within each polystyrene sphere. A density of 1.046 g/cm3 was assumed for each polystyrene sphere. The X-ray marker at the center of the source is 1.25 mm in length and 0.5 mm in diameter. The overall physical length of the source is 4.5 mm, with an outer diameter of 0.8 mm. The internal cavity of the source is filled with dry air. The PTRAN code used the 103Pd photon spectrum extracted from the Medical Internal Radiation Dosimetry (MIRD) pamphlet 10.(6)
The Monte Carlosimulated dose rate constant was obtained by calculating the kerma rate to water at the reference point (1 cm, π/2) in a medium and then dividing that result by the simulated air kerma (SK) strength of the source. The SK was determined by calculating the air kerma rate at 10 cm distance and correcting for the inverse square of the distance to obtain the value at 1 cm while suppressing characteristic X-ray production. It is understood, but impractical, to simulate at the distance that the National Institute of Standards and Technology (NIST) uses (100 cm). Such a simulation would have required extremely long run times to yield even marginal uncertainty levels. Using 10 cm as a reference yields good uncertainty at a great enough distance to fairly approximate true NIST calibration standards. In the calculations, simulations in air were performed with the titanium characteristic X-ray production suppressed. The air kerma rate strength per unit contained activity is given in cGy·cm2·h1· mCi1.
Fig. 1. Schematic diagram of the ADVANTAGE 103Pd brachytherapy source (courtesy of IsoAid LLC). The area between the capsule and the X-ray marker is filled with air. |
||
C. Dosimetry technique
Characteristics of the source were determined theoretically according to Task Group Report No. 43 (TG-43U1)(810) from the American Association of Physicists in Medicine (AAPM). Under that protocol, the dose distribution around a sealed brachytherapy source can be determined using the formalism
(1) |
where SK is the air kerma strength of the source, Λ is the dose rate constant at a reference point (r0, θ0) or (1 cm, π/2), G(r,θ) is the geometry function, g(r) is the radial dose function, and F(r,θ) is the anisotropy function. The above quantities are thoroughly defined and discussed in great detail in TG-43(U1) and that discussion will therefore not be repeated here.
The goal of the present project was to compare the calculated dosimetric parameters of the IsoAid ADVANTAGE 103Pd brachytherapy source with those of other commercially available 103Pd sources. The determinations were performed according to the methodology outlined in TG-43U1(810) and in accordance with the AAPM recommendations for source calibration.(11)
III. RESULTS
The dose rate constant, Λ, of the ADVANTAGE 103Pd source was determined using the equation
(2) |
As shown in Table 1, the Monte Carlo calculations yielded a value of 0.709 ± 0.014 cGy·h1·U 1 in liquid water. The uncertainty of the Monte Carlo simulation was determined by combining the uncertainties of the dose rate calculated in medium and the calculated air kerma rate. The air kerma rate was obtained from the Monte Carlo data by simulating the dose rate at 1 cm in phantom and then dividing that result by the simulated dose rate at 10 cm in air with characteristic X-ray production suppressed. The resulting value was then corrected for the effects of the inverse square law.
Table 1. Comparison of the dose rate constant, L, of the ADVANTAGE 103Pd brachytherapy source with the dose rate constants of the Theragenics Model 200, Best Industries 103Pd, and NAS MED3633 sources |
a 1U = 1
cGy·cm2·h1 .
|
The radial dose function of the source was calculated in water from 0.1 cm to 10 cm. The uncertainty of the calculated data is ±3%. Fig. 2 shows a comparison between the calculated g(r) of the ADVANTAGE source and a selection of other commercially available sources. Table 2 presents the values of g(r) in water. Those values were obtained by simulating the dose rate at each g(r) distance and then normalizing to the simulated dose rate at 1 cm. The result was then corrected for inverse square relation as per TG-43.
Fig. 2. Comparison in water of the Monte Carlocalculated radial dose function of the ADVANTAGE 103Pd source with radial dose functions of other commercially available sources. |
||
Table 2. The calculated radial dose function of the ADVANTAGE 103Pd brachytherapy source in liquid water |
Many treatment planning systems require a polynomial fit to
g(r). For use of this data in those various treatment planning systems, the calculated
g(r) in water in the range of 0.1 cm
to 10 cm was fitted to a fifth-order polynomial function defined as follows:
(3) |
where a0 = 1.1983, a1 = 7.3502E2, a2 = 3.1789E1, a3 = 9.1913E2, a4 = 9.9569E3, and a5 = 3.7557E4. This fifth-order polynomial fit has been found to fail to accurately reproduce the g(r) at radial distances greater that 10 cm; however, it is accurate at distances less than 10 cm.
The anisotropy function of the source was calculated in liquid water for distances ranging from 0.5 cm to 7 cm. Those values were obtained by simulating the dose rate at each angle and normalizing to the dose rate at 90 degrees and at the radial distance in question. The result was then corrected using the geometry function relationship as defined in TG-43. The uncertainties of the calculated values range from ±5% to ±6%.
Fig. 3 shows the variation of
F(r,θ) in water as a function of distance from the
source. Figs. 4 and 5 compare the anisotropy function of the ADVANTAGE source with those
of several other commercially available sources at 1 cm and 5 cm respectively. From the
anisotropy
function of the ADVANTAGE source, the anisotropy factors
φan(r) and the anisotropy constant
have been extracted. Table 3 shows the anisotropy functions, anisotropy
factors, and anisotropy constant of the ADVANTAGE source in liquid water. Table 4 compares
the anisotropy factors and constant with those of other commercially available sources. The
calculated anisotropy constant for the ADVANTAGE source in water was 0.880 ± 0.040.
Fig. 3. Variation of the Monte Carlosimulated F(r,q) of the ADVANTAGE 103Pd source in liquid water at distances ranging from 0.5 cm to 7 cm. |
||
Fig. 4. Comparison in liquid water at 1 cm radius of the Monte Carlosimulated anisotropy functions of the ADVANTAGE 103Pd source with the anisotropy functions of other commercially available sources. |
||
Fig. 5. Comparison in liquid water at 5 cm radius of the Monte Carlosimulated anisotropy function of the ADVANTAGE 103Pd source with the anisotropy functions of other commercially available sources. |
||
Table 3. Monte Carlosimulated two-dimensional anisotropy function of the ADVANTAGE 103Pd brachytherapy source in liquid water |
Table 4. Comparison of the Monte Carlosimulated anisotropy factors and constant of the ADVANTAGE 103Pd source and other commercially available sources in liquid water |
IV. DISCUSSION AND CONCLUSIONS
The dose rate constant of the ADVANTAGE 103Pd source was found to be 0.709 ± 0.014 cGy·h1·U 1. This value is in good agreement with other commercially available 103Pd sources. Table 1 compares the dose rate constant of the ADVANTAGE source with the dose rate constants calculated for several other commercially available 103Pd sources such as the Model 200 source ,by Williamson(13); previous work on the ADVANTAGE source calculated by Meigooni et al.(12); the NAS MED3633 calculated by Li and Palta(15); and the Best 103Pd calculated by Meigooni et al.(14)
Fig. 2 compares the g(r) for the new source with those for other commercially available sources such as the Model 200 source determined by Williamson,(13) previous work on the ADVANTAGE source determined by Meigooni et al.,(12) NAS MED3633 determined by Li and Palta,(15) and the Best 103Pd determined by Meigooni et al.(14) The figure shows that the radial dose function of the ADVANTAGE 103Pd source is in good agreement with those of the other sources.
Fig. 3 shows the variation of the anisotropy function with radial distance in water. The F(r,θ) of the ADVANTAGE source in water was compared with those of the Model 200 source determined by Weaver,(16) previous work on the ADVANTAGE source determined by Meigooni et al.,(12) the NAS MED3633 determined by Li and Palta,(15) and the Best 103Pd determined by Meigooni et al.(14) (Figs. 4 and 5). The figures show good agreement between the ADVANTAGE source and other 103Pd sources for angular ranges of 20 degrees to 90 degrees. Below 20 degrees, differences in endcap construction yield significant deviations in the plotted anisotropy functions.
Table 3 presents the values of the measured and calculated anisotropy functions, anisotropy factors, and anisotropy constants for the ADVANTAGE source in liquid water. Table 4 compares the anisotropy factors and constant of the ADVANTAGE source with other commercially available 103Pd sources. The anisotropy constant of the ADVANTAGE 103Pd source in water was found to be 0.880 ± 0.040.
The dosimetric characteristics of the ADVANTAGE source were theoretically determined based on TG-43(U1) recommendations. These characteristics were found to be comparable to the values reported for other commercially available 103Pd sources. As per TG-43(U1), the parameters determined in liquid water are recommended for clinical applications.
ACKNOWLEDGMENT
This project was supported by IsoAid LLC. Special thanks are due to Ali Meigooni for his assistance in the preparation of this manuscript.
REFERENCES
1. Wallner K, Merrick G, True L, Cavanaugh W, Simpson C, Butler W. 125I versus 103Pd for low-risk prostate cancer: modality outcomes from a prospective randomized multicenter trial. Cancer J. 2002;8(1):6773.
2. Dattoli M, Wallner K, True L, Cash J, Sorace R. Long term outcomes after treatment with external beam radiation therapy and 103Pd for patients with higher risk prostate carcinoma. Cancer J. 2003;97(4):979983.
3. Porrazzo MS, Hilaris BS, Moorthy CR, et al. Permanent interstitial implantation using 103palladium: The New York Medical College preliminary experience. Int J Radiat Oncol Biol Phys. 1992;23(5):10331036.
4. Blasko JC, Wallner K, Grimm PD, Ragde H. Prostate specific antigen based control following ultrasound guided 125I implantation for stage T1/T2 prostate carcinoma. J Urol. 1995;154(3):10961099.
5. Williamson JF. Monte Carlo evaluation of kerma at a point for photon transport problems. Med Phys. 1987;14(4):567576.
6. Dillman LT, Van der Lage FC. Radionuclide Decay Schemes and Nuclear Parameters for Use in Radiation-Dose Estimates. NM/MIRD pamphlet No. 10. Revised edition. New York (NY): Society of Nuclear Medicine, Medical Internal Radiation Dose Committee; 1975.
7. Williamson JF. Monte Carlo simulation of photon transport phenomena. In: Morin RL, editor. Monte Carlo Simulation in the Radiological Sciences. Boca Raton (FL): CRC Press; 1988: 53102.
8. Nath R, Anderson LL, Luxton G, Weaver KA, Williamson JF, Meigooni AS. Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. Med Phys. 1995;22(2):209234.
9. Meigooni AS, Williamson JF, Nath R. Single source dosimetry for interstitial brachytherapy. In: Williamson JF, Thomadson BR, Nath R, editors. Brachytherapy Physics. Madison (WI): Medical Physics Publishing Corporation; 1995: 210233.
10. Rivard MJ, Coursey BM, DeWerd LA, et al. Update of AAPM Task Group No. 43 report: a revised AAPM protocol for brachytherapy does calculations. Med Phys. 2004;31(3):633674. [Erratum in: Med Phys. 2004;31(12):35323533]
11. Williamson JF, Coursey BM, DeWerd LA, Hanson WF, Nath R, Ibbott GS. Dosimetric prerequisites for routine clinical use of new low energy photon interstitial brachytherapy sources. Med Phys. 1998;25(12):22692270.
12. Meigooni AS, Dini SA, Dou K, Awan SB, Gopalakrishnan G. Experimental and Monte Carlo dosimetric characterization of ADVANTAGE 103Pd. Appl Radiat Isot. 2006;64(8):881887.
13. Williamson JF. Monte Carlo modeling of the transverse-axis dose distribution of the model 200 103Pd interstitial brachytherapy source. Med Phys. 2000;27(4):643654.
14. Meigooni AS, Bharucha Z, YoeSein M, Sowards K. Dosimetric characteristics of the Best double-wall 103Pd brachytherapy source. Med Phys. 2001;28(12):25682575.
15. Li Z, Palta JR, Fan JJ. Monte Carlo calculations and experimental measurements of dosimetry parameters of a new 103Pd source. Med Phys. 2000;27(5):11081112.
16. Weaver KA. Anisotropy functions for 125I and 103Pd sources. Med Phys. 1998;25(12):22712278.