Comparison of measured and
computed portal dose for IMRT
treatment
Savino Cilla,1 Pietro Viola,1 Luigi Azario,3 Luca Grimaldi,1 Maurizio Craus,1 Guido D'Onofrio,1 Andrea Fidanzio,3 Francesco Deodato,2 Gabriella Macchia,2 Cinzia Digesù,2 Alessio G. Morganti,2 and Angelo Piermattei1,3
U.O. Fisica Sanitaria Centro di Ricerca e Formazione ad Alta Tecnologia nelle Scienze Biomediche,1 Università Cattolica S. Cuore, Campobasso Italy; U.O. Radioterapia Centro di Ricerca e Formazione ad Alta Tecnologia nelle Scienze Biomediche,2 Universita Cattolica S. Cuore, Campobasso Italy; Istituto di Fisica,3 Universita Cattolica S. Cuore, Roma, Italy
a.piermattei@rm.unicatt.itReceived 5 April 2006; accepted 26 May 2006
A new 2D array Seven 29T model (PTW, Freiburg), equipped with 729 vented plane-parallel ion chambers, projected for pretreatment verification of radiotherapy plans, was used as a detector for the transmitted or portal dose measurements below a Rando phantom. The dosimetric qualities of the 2D array make it attractive for measuring transmitted dose maps from step-and-shoot intensity-modulated radiotherapy (IMRT). It is well known that for step-and-shoot IMRT beams that use a small number of monitor units (MUs) per sequence, the early and recent electronic portal imaging devices (EPIDs) present a different response at X-ray start-up that affects the accuracy of the measured transmitted dose. The comparison of portal doses measured to those calculated by a commercial treatment-planning system (TPS) can verify correct dose delivery during treatment. This direct validation was tested by irradiating a simulated head tumor in a Rando anthropomorphic phantom by step-and-shoot IMRT beams. The absolute transmitted doses on a plane orthogonal to the beam central axis below the phantom were measured by the 2D array calibrated in terms of dose to water and compared with the computed portal dose extracted by custom software. In a previous paper, the comparison between the IMRT portal doses, computed by a commercial TPS and measured by a linear array that supplied a 1 mm spatial dose resolution, was carried out. The γ-index analysis supplied an agreement of more than 95% of the dose point with acceptance criteria, in terms of dose difference, ΔDmax, and distance agreement, δdmax, equal to 4% and 4 mm, respectively. In this paper, we verify the possible use of the PTW 2D array for measurements of the transmitted doses during several fractions of head and neck tumor radiotherapy. There are two advantages in the use of this 2D array as a portal dose device for the IMRT quality assurance program: first is the ability to perform absolute dose comparisons for hundreds of measurement positions to verify the correct dose delivery in several fractions of the therapy; second is the efficiency in time to detect these kinds of dose distributions within the field of view area of the CT scanner.
PACS number: 87.53.Xd
Key words: portal dose, IMRT, radiotherapy
I. INTRODUCTION
Inverse planned intensity-modulated radiotherapy
(IMRT) treatment is capable of producing complex dose distributions
that can conform to even a concave volume. However, because of the
production of steep dose gradients associated with IMRT, the in
vivo verification of delivered dose in several fractions of the
therapy is prudent for the assurance of patient safety.
Although sophisticated means to
calculate and deliver modulated dose distributions have been
developed, means to verify their actual delivery by radiographic
film, 2D arrays, and electronic portal imaging devices (EPIDs) are
often relatively cumbersome. In addition, these detectors are
generally used for standard tests in pretreatment conditions to
verify the dose map in homogeneous standard phantoms.
The EPID seems to be a valuable
tool for in vivo quality assurance purposes,(1-4) in particular, to assure beam
centering by visual inspection. However, there does not exist any
direct and practical method to check the correct positioning of the
patient, the correct leaf position, and the dosimetry during
treatment. In some institutions, methods to determine the
transmitted or portal dose distribution below the patient have been
developed to obtain a comparison between planned and reconstructed
dose distributions in patient.(5-7)
Indeed, if accurate electron density information in the CT scanners
truly represents the patient in the treatment position, and the dose
calculation of the treatment-planning system (TPS) below the patient
is accurate, it is possible to compare computed and measured online
portal doses within the field of view (FOV) of the CT, to detect
incorrect dose delivery or changes in the centering of the patient.
Unfortunately, early generation liquid-filled matrix ion chambers
(LiFi) and the camera-based fluoroscopic EPID generally produced
images with low contrast and limited stability, the latter due to
temperature fluctuations and radiation damage. Moreover, their
nonlinear dose response and field-size-dependent spreading of
optical photons make them difficult to calibrate and use clinically.
A third and more recent class of EPIDs uses amorphous silicon and is
more efficient.(4,8)
The limitations of the EPIDs used
as detectors for X-ray transmitted dosimetry were analyzed, and
complex solutions have been proposed to take into account the
following:
- duty cycle (how much of the "beam-on" time is actually integrated by the imager)
- sensitivity (how large the noise is on the final imager)
- stability (how much the detector gain varies with time, temperature, and gantry angle)
- dose response (relationship between camera signal and dose to water)
Moreover, for step-and-shoot IMRT applications,
many beam sequences with a small number of monitor units (MUs) are
often used, and the deviations occurring at beam start-up
significantly affect the accuracy of the EPID portal dosimetry.(2,9) In other words, the EPID phantom,
having dose-deposition properties that differ significantly from
those of a simple water phantom (which could be easily simulated by
a TPS), presents a complicated dose. response relationship.
Another dosimetric device that
presents high spatial resolution is radiographic film. Radiographic
film requires many hours of work and extreme care in film
processing. Indeed, generally the films are oversensitive to
low-energy photons, which may be important in penumbra regions.
Moreover, due to the absence of chemical processing of radiographic
films in our center (as in many other hospital centers), film
dosimetry was not considered.
Recently, the authors performed
measurements of portal doses for square fields on the Rando phantom
by a calibrated linear (1D) array, LA48 PTW,(10) that uses 47 liquid ion chambers(11) that can be moved in a plane to obtain
dose distributions with a spatial resolution of 1 mm. These
measurements were compared with the computed portal dose profiles
obtained by three different TPSs, using Low et al.'s
analysis.(12) Generally, TPSs do not
supply accurate absolute or relative dose calculations outside the
body contours and, in particular, at the distance where the EPID
operates. The TPS Plato version Sunrise supplied the best
results(10) in terms of agreement
between computed and measured absolute portal dose profiles on a
plane orthogonal to the beam central axis within the FOV area of the
CT scanner. On the basis of these results, the acceptance criteria
for the Plato portal dose computation for head and neck tumors
irradiated by step-and-shoot IMRT were determined again using LA48
measured profiles.
A 2D array
(PTW, Freiburg, Germany), consisting of 729 air ion chambers, was
used in this work as a device for direct and online measurement of
transmitted dose on a plane. The 2D array can feasibly verify
accurate dose delivery of complex fluence modulated fields in a
standard plastic water-equivalent phantom for pretreatment quality
control. Recently, the suitability of the PTW 2D array was
analyzed(13) in terms of
reproducibility in the short, medium, and long term. Moreover, the
array successfully detects positional movements at the millimeter
scale of the LINAC secondary collimators and presents good water
equivalence (which can be easily simulated in the TPS). The
measurements reported in this paper are encouraging for use as a
device for measuring transmitted doses below a Rando head phantom
during the IMRT treatment, to compare with the TPS computation.
II. MATERIALS AND METHODS
A. LINAC
A 6-MV X-ray beam, delivered by
an Elekta Precise LINAC, was used in this work. The 6-MV X-ray beam
generally operates for radiotherapy treatments with a nominal 330 MU
min-1, corresponding to 3.3 Gy min-1 at the reference depth of 10 cm for a 10
× 10 cm2 field size at a phantom
source-to-surface distance (SSD) of 100 cm. The Elekta Precise LINAC
is equipped with a standard multileaf collimator (MLC), which
consists of two leaf banks of 40 leaves each, is 1 cm wide at the
isocenter distance of 100 cm, and can operate in step-and-shoot mode
for IMRT.
The portal imager of the
Elekta Precise is an early generation camera-based fluoroscopic
EPID, and for the reasons reported in the introduction, it was not
used as a portal dose detector.
B. PTW 2D array
The PTW 2D array model
L981359 is equipped with 729 vented plane-parallel ion chambers with
a 0.6 g/cm2 graphite wall. Every ion
chamber, 5 × 5 mm2 surface area and 5
mm thick, is polarized with 400 V; the reference point is located
0.5 cm from the 2D array surface. Spaced 1 cm apart, the chambers
are located in a 27 × 27 matrix. The 2D array external dimensions
are 30 × 42 × 2.2 cm3, and the
surrounding material is polymethyl methacrylate (PMMA). The
measuring system (mass 2.4 kg) consists of the chamber array itself,
which also accommodates part of the electronic device, the array
interface, and a data acquisition board for the personal computer.
As specified by the manufacturer, the 2D array supplies dose
measurements with high reproducibility in a range between 0.2 Gy and
10 Gy and dose rates in a range between 0.5 Gy min-1 and 8 Gy min-1, with a resolution of 1 mGy min-1 (using a display cycle that can be
selected from between 400 ms and 999 ms).
The dosimetric characterization of
the 2D array was recently published.(13) The extensive measurements performed in
that work show that the 2D array is reproducible in the short,
medium, and long term, and it presents a good output factor response
for small radiation fields. However, the most important
characteristic is that the array is very sensitive to positional
changes on the millimeter scale of the collimation system. The
limitations of the array-sampling capabilities due to the detector
design with a 1-cm spacing of ion chambers was analyzed, comparing
dose profiles with high-dose gradient regions, obtained with a
pinpoint ion chamber. Nonetheless, the agreement between the two
datasets was very good. In other words, the 2D array data accurately
matched the reference ion chamber profiles when the collimation
system was moved 1 mm at a time.
In this work, the dosimetric
accuracy of the 2D array was confirmed using a linear array, LA48
PTW. This device contains 47 liquid ion chambers (spaced 8 mm apart,
4 × 4 mm2 in size, and 0.5 mm thick),
and it can supply dose distributions with a spatial resolution of 1
mm. Martens et al.(11) and the present
authors(10) observed that the dose
profiles supplied by the LA48 presented differences with the
profiles obtained with a PTW natural diamond detector (type 60003)
well within a ΔDmax of 1.5% and a
δdmax of 1 mm for square fields of
2 cm, 5 cm, 10 cm, 15 cm, and 20 cm side length and
intensity-modulated (IM) beams. For these measurements the diamond
detector (3 × 3 mm2 in size and 0.27 mm
thick) was oriented for the maximum spatial resolution in the scan
direction. As mentioned above, in this work the comparison between
the dose profiles of open and wedge fields obtained by the LA48 in
water phantom and those measured by the 2D array in PMMA phantom was
carried out using custom software called Cilgam, described in
section C and developed to verify the agreement between computed and
measured doses. The dose profiles were measured for different square
fields, 3 cm, 5 cm, 15 cm, 20 cm, and 25 cm side length. Some
comparisons were carried out for the IM beams used in this work for
the head and neck tumor.
The
commissioning of the 2D array in a PMMA phantom was examined in this
work to verify the signal reproducibility for doses as low as 0.01
Gy, with the aim of using the device for transmitted absolute dose
measurements. A PTW thimble ion chamber type (model TM31002; active
volume 0.125 cm3 air) was used as a
reference detector for the absolute dose measurements in a water
phantom, the latter carried out using the practical code IAEA TRS
398(14) to determine the absolute dose
values.
When the 2D array was used
as a portal dose detector, the 2D array was positioned at the depth
of maximum dose (buildup) for the Elekta 6-MV X-ray beam (1.55 cm of
water equivalent material), obtained with an adjunctive PMMA 0.8 cm
slab (Fig. 1).
Fig. 1. Experimental setup of a Rando head phantom for the simulation of an IMRT beam at 252°. A PMMA slab is placed over the dosimetric array. In a previous work,(10) the dosimetric array was an LA48 linear array; in this work, it is a 2D array. The distance of the isocenter, P, from the source, S, is 100 cm; the distance PP′ is 20 cm. |
||
|---|---|---|
The signal, M,
of the 2D array central ion chamber was at first corrected for
standard pressure and temperature, and calibrated in terms of dose
to water using a 10 × 10 cm2 field size
at SSD = 100 cm and 330 MU min-1. A
range of dose values between 1 cGy and 10 Gy was used.
The dose rate dependence of the 2D
array central ion chamber was analyzed measuring dose rates between
0.1 Gy min-1 and 5 Gy min-1, obtained by changing the source detector
distance and the MU rate. In this range the correction factor for
the ion recombination of the PTW reference ion chamber ranged
between 1.001 and 1.002. These measurements were performed with the
PMMA buildup slab on the top of the 2D array.
The manufacturer of the 2D array
supplied normalization factors for each chamber relative to the
central ion chamber. These factors were verified by irradiating the
array with a 30 × 30 cm2 field and
comparing dose profiles extracted from the array to those measured
by the thimble chamber.
Using an
SSD of 100 cm, the signal of the central ion chamber was analyzed to
obtain the total scatter correction factor (Scp) at the buildup using square fields,
between 2 × 2 cm2 and 20 × 20 cm2. This measurement was carried out to
verify the presence of the fluence perturbation due to the
neighboring ion chambers as a function of field size.
Generally, the dose calibration in
terms of centigrays per monitor unit is done for a number equal to
or greater than 100 MUs, once the beam has stabilized. Due to the
low number of MUs used for the sequences of a step-and-shoot IMRT
beam, a set of measurements with the central ion chamber of the 2D
array was carried out at buildup. A range of MUs between 1 and 100
with 330 MU min-1 and beam of 10 × 10
cm2 at phantom surface was used to
check the stability of the dose per MU of the Elekta Precise LINAC.
PTW software called VeriSoft was
used to convert 2D dose maps from the TPS to gray scale virtual
films. These virtual films can be presented on the screen of a
personal computer with the locations of the 729 ion chambers
superimposed over the gray scale image.
In this work, we investigated the
feasibility of the PTW 2D array to detect some delivery errors,
simulating ±5 mm shifts of the phantom isocenter or rotating the
beam collimators of 5°.
C. γ-Index analysis
Custom software,
called Cilgam, was implemented to verify the agreement between the
profiles measured and computed by the TPS Plato Sunrise. Following
the Low et al. analysis,(12) all
computed dose values on a "calculation plane" were converted to
percent values by dividing them by the maximum dose measured on the
same plane.(15) Quantitative
comparisons between the measured and calculated dose at the
vectorial spatial location and
, in both the dose and the
physical distance, scaled as a function of the acceptance criteria,
were carried out using the γ-index defined by the following
expression(12):
(1) |
where r() is the distance between the
measured and calculated points; δ(
) is the dose difference
between the same points; ΔDmax is
the dose-difference criterion; and Δdmax is the distance-to-agreement (DTA)
criterion. The regions where γ < 1 correspond to locations where
the experimental dose confirmed the computed dose within the
acceptance criteria.
In a previous
paper,(10) a 1D array LA48 with high
spatial resolution was used to determine the dose profiles for IMRT
beams, and the passing criteria ΔDmax = 4% and δdmax = 4 mm for the TPS Plato Sunrise dose
computation were selected so that Cilgam supplied more than 95% of
the γ-values less than 1. The same software was used in this work to
compare the measured doses obtained by the 2D array and the computed
dose by the TPS Plato.
D. Treatment-planning system
The Plato
inverse treatment planning (ITP) module is contained in the Plato
TPS version Sunrise 2.6.3 (Nucletron), which optimizes IMRT beams
individually for each patient and tumor location. By means of
absolute dose limits for overdosage of organs at risk (OARs) and for
overdosage and underdosage of the tumor, the oncologist is able to
control the optimization to a wide extent. Weight factors allow an
additional ranking of these dose limits according to clinical
importance. The beam setup parameters (gantry angle, energy) and
fractionation schemes are defined in the Plato radiation therapy
planning software (RTS) module (a conventional 3D treatment-planning
and calculation module). The RTS is implemented with dose profiles
measured by a PTW thimble ion chamber model TM31002 (mentioned in
section B). A natural diamond (model 60003)(16) was used to improve the modeling of the
penumbra regions for profiles of beam sizes ranging between 1 × 1
cm2 and 3 × 3 cm2. The implementation of the RTS for beams
up to 2 × 2 cm2 supplied an accuracy of
the computed profiles of 3% in dose difference and 3 mm in DTA. The
disagreement is essentially due to the modeling of the penumbra
regions of the Plato TPS for small fields. This is the reason why,
generally, segments less than 2 × 2 cm2
are eliminated(17) during the planning
process. For beam sizes equal or greater than 3 × 3 cm2, the accuracy was 2% in dose and 2 mm in
DTA.(15)
To treat the head tumor shown in
Fig. 2, five 6-MV X-ray beams located at the angles of 36°, 108°,
180°, 252°, and 324° were selected. The beam setup parameters
defined in the RTS module were then passed to the ITP module. This
module is comprised of an optimization, a sequencer, and a dose
display module. The ITP calculation algorithm is a 3D single pencil
beam convolution. The program takes into account the restrictions of
the Elekta MLC (minimum distance between opposing leaves,
transmission of the backup jaw, tongue, and groove). To start with,
the user defines dose or dose-volume constraints to the target and
to the OARs by entering the optimization parameter table or by
defining points for a desired dose-volume histogram (DVH). During
the optimization process, isodose lines are displayed on transverse
and reformatted CT images, and DVH curves and fluence maps are
updated. A good plan optimization was found using a total of 45
sequences. In particular, to obtain a dose of 1.8 Gy with five
beams, the sequence number per beam ranged between 5 and 15, while
the MUs per sequence ranged between 10 and 40.
Fig. 2. Tomography of a Rando head section. The concave target volume and a small OAR (circular in shape) are reported (white lines), together with some relative isodoses. The arrows show the direction of the five selected incident IMRT beams at 36°, 108°, 180°, 252°, and 324°; the five external lines are indicative of the “calculation plane” where the dose distributions have been computed. |
||
|---|---|---|
After the dose
optimization and the dose calculation are completed in ITP, the
sequences and relative MUs are then sent to the RTS module, which
gives the final dose calculation by using the more precise 3D pencil
beam algorithm. The dose was calculated throughout every CT section
with a 2-mm dose grid resolution. The calculation plane was in a
PMMA slab, 2 cm thick, at a depth of 1.3 cm (1.55 cm of
water-equivalent material), simulating the position of the 2D array
at SDD = 120 cm. The 20 cm distance from isocenter to calculation
plane is ideal because all five calculation planes fit within the
FOV area (48 cm in diameter) of the Aura Philips CT scanner used in
this work. For each of the five treatment beams, the PMMA slab was
drawn within the FOV with the Rando head sections.(10)
The
3D dose distribution contained in the TPS dose file was interpolated
by the custom software Azabu to obtain the dose values on the
calculation planes at different angular orientations (Fig. 2). The
dose profiles along the lines reported in Fig. 2 were computed with
a 2-mm resolution, while the distances between the profiles parallel
to such lines (along the longitudinal axis of the phantom) were 2.5
mm.
Before transferring the plan
via DICOM-RT for later delivery on the Elekta Precise LINAC, the
five gantry angles were changed to 0° to allow the irradiation of
the Rando head as described in the following section.
E. Portal dose determinations
In a
previous work,(10) the PTW LA48 linear
array, positioned at SDD = 120 cm, was used to determine the
acceptance criteria of the computed portal dose supplied by the TPS
Plato. In the first step, the Rando female phantom was irradiated
with antero-posterior square fields. Every transmitted dose profile
was determined in current mode acquisition with 1 mm spatial
resolution, obtained using the automatic shift of the linear array
along the latero-lateral direction of the Rando phantom during the
beam irradiation. The agreement between computed and measured portal
dose profiles of square beams was quantified by Cilgam. The
resulting percentage of points with γ < 1 was greater than 95%
(Pγ<1 > 95%), when the
acceptance criteria ΔDmax = 3% and
δdmax = 3 mm were selected. In a
second step, the LA48 array was used to test the accuracy of the
computed portal dose for IMRT irradiation. The Rando head sections
were locked between two PMMA slabs with a mechanism that allowed the
phantom head to be rotated around the isocenter point and five IMRT
beams were delivered, maintaining the LINAC gantry at 0° (Fig. 1).
The dose distributions were computed by the TPS Plato and extracted
by Azabu for comparison with the experimental data obtained by the
LA48. In this case, the LA48 worked in integral mode acquisition,
and 50 iterations of the IMRT beam were used to obtain, at 120 cm
from the source, a portal dose matrix with a 2-mm spatial resolution
along the profile and 1 cm between the profiles. Figure 3 shows the
comparison between a measured dose profile along the principal axis
(which intercepts the beam central axis) and three computed dose
profiles. The first computed profile is along the principal axis,
and the other two are at ±2.5 mm laterally. Using Cilgam for the
γ-index analysis, a percentage Pγ<1 > 95%, with the acceptance criteria
ΔDmax = 4% and δdmax = 4 mm, was obtained for the five IMRT
beams.(10)
Fig. 3. Comparison between a measured dose profile by an LA48 linear array along the principal axis, (O),and three computed dose profiles, the profile along the principal axis (continuous line) and other two at +2.5 mm (-----) and–2.5 mm (……) laterally. The γ-index (—+—), computed with the acceptance criteria ∆Dmax = 4% and ∆dmax = 4 mm, is reported on the right scale. |
||
|---|---|---|
It is interesting to
note that using the LA48 device in water phantom for pretreatment
verification of the computed IMRT plan, a Pγ<1 > 95% was obtained with acceptance
criteria of 3% in dose difference and 3 mm in DTA. These results are
in agreement with others reported in the literature,(18) where it is emphasized that the
disagreements are generally due to the level of accuracy of the
calculation in high-dose gradient regions for small segments.
In this work, the same five IMRT
beams used in the previous work(10) were used, and the computed IMRT portal dose distributions were
compared with the measured portal doses obtained by the 2D array.
All the transmitted dose measurements below the Rando phantom were
carried out with the gantry at 0° and 330 MU min-1, rotating the phantom head around the
isocenter point (Fig. 1). Using the five IMRT beams, a dose of 1.8
Gy at the isocenter was delivered. The total time to deliver the
plan to the phantom was approximately 15 min. Two minutes were
required for irradiation, and the rest was used to modify sequences
and phantom angles between beams.
III. RESULTS
A. 2D array calibration
The dose
measurements carried out in this work for the 2D array commissioning
show that the signal reproducibility was estimated to be better than
0.1% (1ρ) for a number of MUs ≥ 5 (≥ 0.05 Gy) and better
than 0.5% (1ρ) for MUs ≥ 2 (≥ 0.02 Gy). Figure 4 shows, for
the central axis ion chamber of the 2D array, the trend of the
signal per MU (normalized at 100 MUs) obtained when the MUs changed
between 1 and 100. This ratio was within 0.5% for MUs greater than
10. The bars show the reproducibility (1ρ) of the normalized
signal.
Fig. 4. Signal per MU normalized to 100 MUs obtained by the central ion chamber of the 2D array when the MU changed between 1 and 100. The bars indicate measured reproducibility (1σ). |
||
|---|---|---|
The 2D array does not
allow correction of ion recombination by two voltage techniques.
However, ratios between the signals obtained by the central 2D array
ion chamber and the reference PTW thimble ion chamber were constant
within the experimental reproducibility (0.2% 1ρ) in the
range of dose rate between 0.1 Gy min-1
and 5 Gy min-1. This means that the
correction factor for the ion recombination of the 2D array ion
chambers was of the same order as that of the PTW reference chamber
(1.002) and could be negligible for our purposes.
The linear fit of the 2D array
central chamber signal, M, corrected for standard pressure
and temperature as a function of the dose value, D, was
allowed to obtain a calibration factor N = 0.2024 Gy/nC and a
regression coefficient of 0.999. This calibration factor was
initially associated with all 729 ion chambers. The comparisons
between the dose profiles for a 30 × 30 cm2 beam, obtained by the reference ion
chamber and the 2D array positioned at the buildup of water and PMMA
phantoms, respectively, allowed the correction of the calibration
factors. Due to the small differences in size of the chamber
volumes, the variations between these factors, normalized at the
central ion chamber, were within ±1%.
The dosimetric data obtained in
standard phantoms by the 2D array, compared with the dose profiles
obtained by the LA48 (ion chambers with 8 mm3 in volume), showed discrepancies with
ΔDmax = 2% and δdmax = 2 mm. Figure 5 shows an example of the
comparison between profiles of open and wedged fields measured with
the 2D array and those obtained by the LA48 linear array. These data
were in good agreement with those obtained by Spezi et al.,(13) who observed a maximum discrepancy of
ΔDmax = 3% for the IM beam
profiles obtained by the 2D array and a pinpoint ion chamber with a
volume of 16 mm3.
Fig. 5. Comparison between dose profiles of the (a) open and (b) wedged fields, 20 × 20 cm2 in size, obtained with the LA48 linear array (•) and the 2D array (☐). |
||
|---|---|---|
The deviations
between the Scp values obtained by
the 2D array and the reference PTW ion chamber were less than 1%,
justifying the negligible effect of the neighboring ion chambers.
Comparing these results with those
obtained by the dose characterization of MapCHECK,(18) it is possible to conclude that the PTW
2D array presents minor dose rate dependence as compared to 2% or 3%
for n-type or p-type diodes. However, in a recent
paper(19) a comparison of the
dosimetric characteristics of the 2D array PTW, MapCHECK (Sun
Nuclear, Melbourn, FL), and the more recent PIX (Pixel ionization
chamber prototype)(20) shows that these
2D arrays are useful tools for the quality assurance and
verification of the IMRT plans in the pretreatment step.
B. Portal dose determinations
Figure 6
shows a zoom of the 2D array panels (21 × 21 ion chambers) as
presented by VeriSoft, with the shadowed areas showing the virtual
film of the five intensity beams. The closed square symbols
represent the ion chambers where the resulting γ-index was greater
than 1, for acceptance criteria ΔDmax = 4% and δdmax = 4 mm as computed by Cilgam. With
accurate Rando phantom centering, the number of the closed symbols
resulted less than 5% for all five beams. The measurements were
repeated five times obtaining the same result.
Fig. 6. Zoom of the 2D array panels (21 × 21 chambers instead of the 729) with the shadowed areas showing the modulation of the five IMRT computed beams obtained with 6-MV X-ray beams. The closed, square symbols represent the ion chambers where the resulting γ-index was greater than 1 for acceptance criteria ∆Dmax = 4% and ∆dmax = 4 mm. |
||
|---|---|---|
For the IMRT beam located at 36°, Fig. 7 reports 16 dose profiles computed by the TPS on the calculation plane, to compare them with the measured portal dose by the 2D array. Two hundred and seventy-two ion chambers received a dose of at least 5% of the maximum dose value (0.3 Gy) and were considered in this analysis. Each of the five beams was irradiated between 150 and 280 ion chambers with at least this threshold value.
Fig. 7. Dose profiles computed by the TPS (continuous lines) at different Y distances, along the lines of ion chambers of the 2D array (☐) for the IMRT beam with incidence angle of 36°. The closed square symbols represent the points where the differences between computed and measured doses are outside the acceptance criteria. |
||
|---|---|---|
Figure 8 reports, for the beam at 36°, an example of the comparison between measured portal dose profiles at Y = -1 cm and Y = 0 cm obtained by the 1D-array LA48 and the 2D array, with the profile computed by Plato. The agreement between the dose profiles obtained by the linear detector and 2D detector determined by Cilgam is within ΔDmax = 2% and δdmax = 2 mm. The closed symbol is relative to the ion chamber of the 2D array where the γ-index is greater than 1. Considering that the acceptance criteria of the TPS portal dose calculation were estimated to be 4% and 4 mm, the loss of agreement between the dose calculation and the 2D array measurements in less than 5% of the chambers (Fig. 6) is a good result, considering that about 1000 dose points were checked.
Fig. 8. Comparison between measured portal dose profiles for the IMRT beam at 36° at (a) Y = –1 cm and (b) Y = 0 cm obtained by an LA48 (o) and a 2D array (☐), with the profile computed by the Plato (continuous line). The closed symbol is relative to ion chamber of the 2D array where the γ-index is greater than 1. |
||
|---|---|---|
Six 5-mm shifts of the isocenter point were simulated moving the Rando phantom (±5 mm) along the three principal axes. The percentage Pγ<1 ranged between 85% and 90%, maintaining the acceptance criteria of 4% and 4 mm. For the beam at 324°, Fig. 9 presents the zoom of the 2D array panel as presented by VeriSoft, when the beam collimators were misaligned by a 5° rotation around the central axis. The percentage Pγ<1 decreased to 85%, and the result can be compared with that reported in Fig. 6 for the same beam.
Fig. 9. Zoom of the 2D array panels (21 × 21 chambers instead of the 729) with the shadowed areas showing the modulation of the IMRT computed beam obtained with a 6-MV X-ray beam at 324°. The measurement was carried out with the beam collimators misaligned by 5° around the central axis. The closed square symbols (about 15%) represent the ion chambers where the γ-index resulted greater than 1 for acceptance criteria of ∆Dmax = 4% and ∆dmax = 4 mm. |
||
|---|---|---|
IV. DISCUSSION AND CONCLUSIONS
A dosimetric quality assurance (QA) program in
complex radiotherapy treatment plans such as IMRT requires advanced
technology. However, for IMRT, besides a complete program for the QA
of the LINAC, the simulator, the TPS, and independent checks of
radiotherapy treatment parameters (distances, field size, etc.), a
pretreatment control of the computed dose in standard phantom for
every beam is mandatory.
In many
centers, pretreatment verification of the dose calculation for the
beams selected for patient therapy is carried out by 2D arrays as
the PTW here reported. However, this verification requires a lot of
work before treatment, which interrupts the activity of the
radiotherapy department for 15 min or 30 min, depending on the
number of fields. To avoid such interruptions, this verification is
not generally repeated in the following fractions of the therapy. It
is our opinion that the first pretreatment verification is mandatory
for the control of dose computation, while it could also be useful
to check the following fractions of the therapy to verify the
constancy of (1) the correct dose delivery and (2) correct patient
positioning.
EPIDs are probably
the most promising devices for IMRT verification and in vivo
dosimetry.(8,21,22) These detectors can
provide high resolution and highly efficient planar dose maps.
However, further developments are underway for the deployment of
this technique into routine clinical practice.(23-25) In the meantime, we think that 2D
online dose detectors could be used together with EPIDs for complete
in vivo QA purposes.
The
PTW 2D array here examined is a practical detector for radiotherapy
verification even if the ion chamber spacing of 1 cm results in a
limited sampling of the radiation beam. However, this limitation
does not seem to affect its possible use to accurately check several
dose points. The dose calibration for the 2D array is easy and
stable. For a small number of MUs (≥10) the step-and-shoot sequences
of our IMRT offer a high level of dosimetric accuracy (Fig. 4).
Moreover, this array seems to be a water-equivalent detector.
The 2D array was simulated in a
TPS for the IMRT portal dose computation using a slab of
water-equivalent material. This way, custom software was realized to
extract from the TPS the maps of portal doses on specific planes in
the FOV.
In irradiation of a Rando
phantom with step-and-shoot IMRT beams, and measuring portal doses
with a 1D array with high spatial resolution, we have determined
that the TPS Plato Sunrise supplies a dose computation with a
Pγ<1 > 95% of portal dose
points with acceptance criteria of ΔDmax = 4% and δdmax = 4 mm. Figure 6 shows the comparison
between the computed and measured portal dose values by the 2D
array. The number of chambers that found a dose computation within
the acceptance criteria is greater than 95%. In other words, the
good agreement observed between the measured dose profiles obtained
in standard phantoms by the 1D array LA48 and the 2D array
(ΔDmax = 2% and δdmax = 2 mm) justifies the 2D array's
Pγ<1 > 95%. We think that
the discrepancies are due both to the accuracy level of the portal
dose calculation as well as the grid (2.5 mm) used by Azabu in the
sampling of the dose points between the different portal dose
profiles.
In conclusion, even if
the 2D array presents a limited sampling capability, it is our
opinion that the observation of about 1000 points (200 per beam) is
enough for a detailed evaluation of the portal dose reproducibility
in many fractions of the therapy. This kind of verification can be
carried out during the treatment, positioning the 2D array in a
particular jig that can follow the rotation of the portal vision.
The effect of 5-mm shifts of the
Rando head allows one to estimate a decrease of Pγ<1 to 85% of points inside the acceptance
criteria of ΔDmax = 4% and
δdmax = 4 mm. This result can be
useful for the in vivo QA. Moreover, the procedure allows the
observation of an incorrect angular position of the multileaf
collimator. Figure 9 shows a drop of Pγ<1 to 85% when an error of 5° in the
rotation of the beam collimators was simulated.
In conclusion, the procedure here
tested, based on step-and-shoot IMRT irradiation of the Rando head,
can be used for head and neck tumors where the computation of the
dose is well within the FOV area of the CT scanner. We intend to use
it for the IMRT in vivo treatments to supply a contribution
to the selection of the percentage of points Pγ<1 and the passing criteria significant
for the judgment of the reproducibility of therapy fractions.
ACKNOWLEDGMENTS
We are grateful to Nuclital s.r.l., Dr. A. Bursi, and Dr. G. Alberta for their technical assistance.
REFERENCES
- Boellaard R, van Herk M, Uiterwaal U, Mijnheer B. First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose. Radiother Oncol. 1998;47:303-312.
- Curtin-Savard AJ, Podgorsak EB. Verification of segmented beam delivery using a commercial electronic portal imaging device. Med Phys. 1999;26(5):737-742.
- Louwe RJW, Tielemburg R, van Ingen KM, Mijnheer BJ, van Herk MB. The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes. Med Phys. 2004;31:(4):819-827.
- Chang J, Ling CC. Using the frame averaging of aS500 EPID for IMRT verification. J Appl Clin Med Phys. 2003;4:287-299.
- Ying X, Geer LY, Wong JW. Portal dose images II: Patient dose estimation. Int J Radiat Oncol Biol Phys. 1990;18:1465-1475.
- McNutt TR, Mackie TR, Reckwerdt P, Paliwal BR. Modeling dose distributions from portal dose images using the convolution/superposition method. Med Phys. 1996;23(8):1381-1392.
- McDermott LN, Wendling M, Sonke J, et al. First clinical experience with pre-treatment and in vivo IMRT verification using EPID dosimetry. Med Phys. 2005;32(6):2090.
- Van Esch A, Depuydt T, Huyskens DP. The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. Radiat Oncol. 2004;71:223-234.
- Hansen VN, Evans PM, Budgell GJ, et al. Quality assurance of the dose delivered by small radiation segments. Phys Med Biol. 1998;43:2665-2675.
- Piermattei A, Cilla S, Pepe D, et al. Verification of computed portal doses by a linear array of liquid ion-chambers. Phys Med. 2004;XX(3):111-119.
- Martens C, De Wagter C, De Neve W. The value of the LA48 linear ion chamber array for characterization of intensity-modulated beams. Phys Med Biol. 2001;46:1131-1148.
- Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25(5):656-661.
- Spezi E, Angelini AL, Romani F, Ferri A. Characterization of a 2D ion chamber array for the verification of radiotherapy treatments. Phys Med Biol. 2005;50:3361-3373.
- IAEA (International Atomic Energy Agency). Absorbed dose in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water. IAEA TRS 398; 2001.
- Van Dyke J, Barnett RB, Cygler JE, Shragge PC. Commissioning and quality assurance of treatment planning computers. Int J Radiat Oncol Biol Phys. 1993;26:261-273.
- Piermattei A, Azario L, Fidanzio A, Arcovito G. Quasi water-equivalent detector for photon beams that present lateral electron disequilibrium. Phys Med. 1998;24(1):9-17.
- Letourneau D, Gulam M, Yan D, Oldham M, Wong JW. Evaluation of a 2D diode array for IMRT quality assurance. Radiat Oncol. 2004;70:199-206.
- Jursinic PA, Nelms BE. Diode array for verification of IMRT delivery. Med Phys. 2003;30(5):870-879.
- Cirio R, Garelli E, Giordanengo S, Marchetto F, Peroni C. Preliminary results from a comparison of 2D verification systems for IMRT. Estro 23 Annual Meeting, Amsterdam, 2004 October 24-28. Radiother Oncol. 2004;73(Suppl 1):256:S116.
- Amerio S, Boriano A, Bourhabel F, et al. Dosimetric characterization of a large area pixel segmented ionization chamber. Med Phys. 2004;31:414-420.
- Van Esch A, Vanstraelen B, Verstraete J, Kutcher G, Huyskens D Pre-treatment dosimetric verification by means of a liquid-filled electronic portal imaging device during dynamic delivery of intensity modulated treatment fields. Radiother Oncol. 2001;60(2):181-190.
- Warkentin B, Steciw S, Rathee S, Fallone BG. Dosimetric IMRT verification with a flat-panel EPID. Med Phys. 2003;30:3143-3155.
- Louwe RJW, McDermott LN, Sonke JJ, et al. The long-term stability of amorphous silicon flat panel imaging device for dosimetry purposes. Med Phys. 2004;31:2089-2095.
- Chin PW, Lewis DG, Spezi E. Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry angle. Phys Med Biol. 2004;49:N93-N103.
- McDermott LN, Louwe RJ, Sonke JJ, van Herk MB, Mijnheer BJ. Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device. Med Phys. 2004;13:115-122.