Study of the influence of
phantom material and size
on the calibration of ionization
chambers in terms of absorbed
dose to water
Mehenna Arib,1 Toufik Medjadj,1 Youcef Boudouma2
Secondary Standard Dosimetry Laboratory,1 Nuclear Research Centre of Algiers,2 Bd Frantz Fanon BP 399 Alger Gare, Alger 16000, Algérie; Laboratoire Sciences Nucléaires et Interactions Rayonnement-Matière,2 Faculté de Physique, USTHB, BP 32 El-Alia, 16123 Bab Ezzouar, Algiers, Algeria
Mehenna.arib@comena-dz.orgReceived 18 March 2006; accepted 26 May 2006
In the International Atomic Energy Agency's (IAEA) code of practice (TRS 398) and the American Association of Physicists in Medicine's dosimetry protocol (TG-51), full-scatter water phantoms are recommended for the determination of the absorbed dose for both photon and electron beams and, consequently, for the calibration of the user's ionization chambers. This procedure is applied in the Secondary Standard Dosimetry Laboratory, where the calibration is performed on a 60Co gamma beam, in comparison with reference chambers whose absorbed dose-towater calibration coefficients, ND,w, are known. In this work, we present the results of the calibration of 10 Farmer-like ionization chambers calibrated in three water phantoms (sizes 20 × 20 × 15 cm3, 30 × 30 × 30 cm3, and 35 × 35 × 37 cm3) and two plastic phantoms (size 20 × 20 × 20 cm3) polymethyl methacrlyate (PMMA) and polystyrene). Calibrations are performed by the substitution method using an ionization chamber whose ND,w has been supplied by the IAEA's reference laboratory. It is shown that the results, expressed as the percentage ratio of the calibration coefficient in a given phantom to that of the standard IAEA phantom, is less than 0.35% for all investigated chambers, and that the standard deviation of the mean of the ND,w calibration coefficients determined in all five phantoms is less than 0.06%, except for one nylon-walled ionization chamber, where the observed 0.34% value could be explained by the hygroscopic properties of nylon. Furthermore, a chamber-to-chamber dependence of the calibration coefficient has been shown to vary by up to 2.8%. These results emphasize that the phantom dimensions and its material are not sensitive criteria for the calibration of cylindrical ionization chambers in terms of absorbed dose to water. The results also show that generic calibration coefficients could not be considered for a given type of chamber.
PACS number: 87.53.Dq
Key words: absorbed dose to water, calibration coefficient, ND,w, water phantom, solid phantom
I. INTRODUCTION
It has been established that the required accuracy
of the absorbed dose delivered to the target volume in radiotherapy
is considerably improved if the calibration of a clinical beam is
performed with ionization chambers calibrated in terms of absorbed
dose to water.(1) In this context, the
recent dosimetry protocols published by the International Atomic
Energy Agency(2) (IAEA) and the
American Association of Physicists in Medicine(3) (AAPM) are based on the use of an
ionization chamber calibrated in terms of absorbed dose to water in
a standard laboratory's reference quality beam, generally
taken as a 60Co gamma ray beam.
Both cylindrical and
plane-parallel ionization chambers are recommended as reference
instruments for the calibration of the user's 60Co gamma ray beam, and water is
recommended as the reference medium for absorbed dose measurements.
The accuracy and traceability of
the calibration of radiotherapy dosimeters are of great concern,
since absorbed dose cannot be obtained with non-calibrated
instruments. The IAEA's code of practice recommends that
the calibration coefficient used to determine the absorbed dose be
obtained in a water phantom. The recommended water phantoms should
be full-scatter ones extending at least 5 cm beyond all four sides
of the largest field size used at the depth of measurement and with
at least 5 cm beyond the maximum depth of measurement.
In this work, we investigate the
influence of the phantom size and its material over the calibration
coefficient. For this purpose, 10 Farmer-like ionization chambers
were calibrated in five different phantoms using the substitution
method, and the calibration coefficients were compared to those
obtained in the IAEA standard water phantom.
II. MATERIALS AND METHODS
A. Experimental equipment and reference
conditions
A.1 The Secondary Standard Dosimetry
Laboratory (SSDL) 60Co
unit
Calibrations were performed in 60Co gamma beams using an Eldorado 78 unit.
The output in terms of absorbed dose to water was 1120 mGy/min, at a
source-to-surface distance (SSD) of 80 cm, a depth of 5 cm, and a
field size of 10 × 10 cm2 at the
phantom surface (Fig. 1). The beam flatness at these conditions is
about 0.3% within ±4 cm of the beam axis.
Fig. 1. The geometrical conditions used for the calibrations of the ionization chambers in terms of absorbed dose to water. (The figure shown is for a water phantom.) |
||
|---|---|---|
A.2 Ionization chambers
Five types
of commonly used ionization chambers were provided by radiotherapy
centers for the present work, giving an overall total of 10
chambers. The characteristics of these chambers are shown in Table
1. Among these chambers, six are the widely used Farmer NE 2571
chamber. The WDIC70 ionization chamber, manufactured by Wellhofer,
is inherently waterproof. However, for measurements in water, all
the chambers are used with a 1-mm PMMA waterproofing sleeve. The
reference chamber WDIC70 #141 was calibrated at the IAEA's
reference laboratory. The calibration coefficient (ND,w = 47.9 ± 0.2 mGy/nC), established at the
conditions T = 20 °C and P = 101.325 kPa, is thus
traceable to the Bureau International des Poids et Mesures
(International Bureau of Weights and Measures).
Table 1. Characteristics of the cylindrical ionization chambers used in the present study |
|---|
A.3 Current measurements
The
reference ionization current was measured with a Keithley 6517
electrometer, and the chambers listed in Table 1 were calibrated
with a PTW UNIDOS 10002 electrometer. The time required by the
measuring assembly to stabilize was investigated with the Keithley
electrometer along with all the chambers. Current measurements were
collected at regular intervals after switching on the system, by
means of a GPIB computer interface. The results showed that within
the first 30 min, the signal varied up to 0.5% for all chambers and
could reach 1% for the NE 2581 chamber. Therefore, the measuring
assemblies were left at least 2 h before carrying out any
measurements. The leakage currents were found to be less than 0.01%
of the typical measured currents. The calibration factors of the
electrometers were checked with a current source of the Keithley 263
type and were found to be within 0.03% of unity for both
electrometers. The recombination and polarity effects were checked
and were found to be negligible for all the studied chambers. Thus,
the measured currents were not corrected for these effects.
A.4 Calibration phantoms
The
calibrations were carried out in three water phantoms of different
sizes and two plastic phantoms (PMMA and polystyrene). The
characteristics of these phantoms are given in Table 2.
Table 2. Characteristics of the phantoms used for the calibrations in terms of absorbed dose to water |
|---|
A.5 Geometrical conditions
The
calibrations were carried out with the geometrical axis of the
chambers situated at a depth of 5 cm. The SSD was 80 cm, and the
field size at that distance was 10 × 10 cm2 (Fig. 1). All the calibrations were
performed using a horizontal beam except for the PMMA phantom, which
consists of a homogeneous parallelepipedic block fixed on a
supporting plate that can be accurately fixed to the collimator of
the 60Co Eldorado 78 unit using four
rigid metallic stems, and in which the ionization chambers can be
irradiated vertically in reproducible conditions.
For the IAEA standard phantom and
the Mylar window phantom, the chambers were used with the same 1-mm
thick PMMA waterproofing sleeve. For these two phantoms, special
attention was paid to the outward bowing of their windows owing to
the water pressure on the inner surface. This effect, which occurs
as soon as the phantom is filled and which depends on the size of
the phantom and the window, can change the chamber depth and the SSD
considerably. Indeed, the water pressure is important with larger
phantoms, and the deformation is maximal with larger and thinner
windows. The actual deformation was monitored with a 0.01-mm
precision comparator and evaluated to be 0.06 mm and 3 mm for the
IAEA and Mylar window phantoms, respectively. The SSD and the
chamber depth were corrected accordingly; in addition, to avoid any
fluctuations, the water level of the two phantoms was controlled and
kept constant during the entire period of the work.
The positioning of the chambers in
the plastic phantom is done thanks to a bar, made of the same
phantom material, and fitting into existing holes centered at a
depth of 5 cm in the phantoms. These bars are machined in order to
precisely accommodate the ionization chambers, bringing their
geometric center at this depth.
B. Calibration method
The chambers were
calibrated in terms of absorbed dose to water using the substitution
method. According to this method, the reference chamber is placed at
the reference point in the beam, and a set of readings is taken,
leading to the mean reading Mref.
It is then replaced by the chamber to be calibrated and a similar
set of readings is taken, in the same conditions; the mean reading
is then Mu.
The calibration coefficient
ND,w is given by
(1) |
where Dw is the
absorbed dose to water, determined by the reference chamber, and is the
corrected reading of the chamber to be calibrated.
According to the TRS398 IAEA code
of practice,(2) the absorbed dose to
water, DW, is determined according
to
(2) |
where is the absorbed dose-to-water calibration coefficient of
the reference chamber obtained from a standard laboratory, and
is its reading
corrected for temperature and pressure according to the following
equation:
(3) |
T and P are, respectively, the
temperature in degrees Celsius and the pressure in kilopascals. The
temperature is measured with a classic thermometer inserted in a
waterproofing sleeve identical to that of the cylindrical chamber.
The absorbed dose-to-water
calibration coefficient of the calibrated ionization chambers is
thus given by
(4) |
where is calculated according to the same procedure described
above.
III. RESULTS AND DISCUSSIONS
A. Results of the calibrations
The
results of the calibrations are summarized in Table 3. As can be
seen, the relative standard deviations of the mean of the
calibration coefficients are less than 0.06% (relative standard
deviation 0.14%), except for the NE 2581 ionization chamber, where
it is 0.34% (relative standard deviation 0.67%).
Table 3. Results of the calibrations in all phantoms. The last column represents the standard deviation of the mean, which is obtained by dividing the standard deviation by the square root of N, where N is the number of individual values of ND,w (in this case, N = 4 or N = 5). |
|---|
Furthermore, the results are expressed as the relative deviation of the calibration coefficients obtained in a given phantom to the ones obtained in the IAEA phantom according to the following equation:
(5) |
where and
are, respectively, the calibration coefficients obtained in the
standard IAEA water phantom and in the other phantoms.
As can be seen in Fig. 2, these
deviations, in absolute values, lie between 0.01% and 0.35%. We have
calculated that 35% and 87% of the deviations are less than 0.05%
and 0.2%, respectively.
Fig. 2. Deviations of the calibration coefficients from the values obtained in the IAEA cubic water phantom. |
||
|---|---|---|
The results are less satisfactory with the NE 2581 ionization chamber, which has not been included in Fig. 2. The wall of this chamber is made of nylon, and, as pointed out by several authors,(4,5) the response of such chambers is strongly affected by environmental conditions due to the hygroscopic properties of nylon.
B. Chamber-to-chamber variation
The
chamber-to-chamber variation of the calibration coefficients is
illustrated in Fig. 3 for the NE 2571 chamber. As can be seen, the
relative standard deviation of the ND,w mean is 0.15%. This appears to be a good
result; however, if we consider the mean values of the calibration
coefficients obtained in the four phantoms, the deviation between
the maximum and the minimum values is around 2.8%. This stresses the
need for individual calibration of each chamber of the same type and
allows us to recommend not using generic calibration coefficients.
For the polystyrene phantom, only four chambers have been made
available; thus the chamber-tochamber study could not be extended
for this solid phantom.
Fig. 3. Chamber-to-chamber variations of the NE 2571 ionization chamber. The dashed line represents the overall mean calibration coefficient, calculated for all the ionization chambers. |
||
|---|---|---|
C. Uncertainty analysis
Many factors
affect the absorbed dose-to-water calibration coefficient. Some of
these factors have been identified, and their effect, evaluated in
terms of uncertainties, is analyzed. The overall uncertainty is
calculated according to the recommendations of the ISO guide for the
expression of uncertainty.(6)
C.1 Type A and type B
uncertainties
According to the ISO guide,(6) the uncertainties are classified as type
A or type B: A type A uncertainty can be evaluated using statistical
techniques. This uncertainty, which measures the repeatability of a
result under constant conditions, is assumed to have a normal
probability distribution and can be determined by a series of
measurements (yi) in which an
estimate, s(ρ), of the standard deviation, ρ,
is obtained by a series of n measurements, applying
(6) |
where ym is the
mean value of y.
The
standard deviation of the mean, , is the value to be used for
the summation in quadrature for obtaining the combined standard
uncertainty (see below).
All
uncertainties that cannot be determined by a series of repeated
measurements are type B uncertainties. They arise from a variety of
sources, and the probability distributions may take a variety of
shapes. Practical guidance on evaluating type B components is given
in Ref. 6.
C.2 Combined standard
uncertainty
The combined standard uncertainty of the
output quantity, u(y), is derived by the summation in
quadrature of all type A and type B standard uncertainties due to
the input parameters:
(7) |
C.3 Expanded uncertainty
The
overall uncertainty to be quoted for the calibration coefficient is
the expanded uncertainty, U, which represents the total
uncertainty for a specific level of confidence. It is derived by
multiplying the combined standard uncertainty, u(y),
by a coverage factor, k, which is selected to give the
desired level of confidence for a normal distribution. Typical
choices for k are the integers 1, 2, or 3, which correspond to
confidence levels of 67.7%, 95.5%, and 99.7%, respectively. For most
radiological applications, a 95% confidence level, for which
k = 1.96, is recommended. For convenience, this is rounded
upwards to a value of exactly 2.0.
C.4 Uncertainty over the ND,w calibration coefficient
Table 4
shows the uncertainty budget for the calibration process. Using Eqs.
(3) and (4), the following uncertainty contributors can be
identified.
Table 4. Uncertainty budget for the absorbed dose-to-water calibration coefficient |
|---|
C.4.1 D
For the
reference measurements, a series of 20 readings is taken, and the
standard deviation of the mean obtained is of the order 0.01%. The
resolution of the reference instruments is equal to 0.02%. Assuming
a rectangular distribution, the uncertainty is 0.01%. This
uncertainty is taken to be of type B.
C.4.2 U
The uncertainty
over the calibration coefficient of the SSDL working standard, taken
from the calibration certificate, is reported as a type B
uncertainty.
C.4.3 C
Constancy of the secondary standard is obtained
by evaluating the long-term stability of the secondary standard
system. This uncertainty, estimated from a series of output rate
measurements, is dominated by a component having a type A evaluation
(standard deviation of the mean of the measurements). Because this
component is not directly determined during each measurement, it is
treated as a type B evaluation.
C.4.4 C
Typically, the resolution of the therapy
level dosimeters is within ±0.1%. This is evaluated as a type B
component. Assuming a rectangular distribution, the relative
uncertainty is 0.06%.
C.4.5 L
The
leakage is evaluated before each calibration and is subtracted from
the instrument reading. The uncertainty over this factor is
evaluated to be 0.02%.
C.4.6 E
For corrections using Eq. (3), the temperature
readings are done with a thermometer placed in water or plastic
phantom. The difference between the temperature measured by the
thermometer and the temperature inside the air cavity is estimated
to be less than 0.1%. Assuming a rectangular distribution, the
uncertainty (taken as type B) is evaluated to be 0.06%. Another type
B component coming from the thermometer resolution (0.1 °C) is
included for both measurements (reference and user). Assuming a
rectangular distribution, the relative uncertainty is 0.02%.
C.4.7 E
In Eq. (3), the pressure is measured with a
classic barometer. When compared with a mercury absolute barometer,
its readings are within ±0.1%. Assuming a rectangular distribution,
the type B relative uncertainty is 0.06%. The uncertainty component
related to the resolution of the barometer is negligible.
C.4.8 R
This uncertainty, taken as type A, is
evaluated to be 0.04%.
As can be
seen in Table 4, the overall uncertainty over the calibration
coefficient is 1.05%.
IV. CONCLUSION
Compared to the overall uncertainty over the
absorbed dose-to-water calibration coefficient, which is around 1.1%
(given at 95% confidence level), the difference between the
calibration coefficients obtained in this study for the different
phantoms and those for all the investigated ionization chambers is
not significant. The most important discrepancy is observed for the
NE 2581 chamber, whose response is strongly affected by
environmental conditions due to the composition of its wall (nylon).
Therefore, the deviations observed in this study for this particular
chamber was expected.
Our study
shows that the type and dimensions of the phantom used are not
critical parameters. The absorbed dose-to-water calibration
coefficients can be determined in any water or solid phantom whose
dimensions are at least 20 × 20 × 20 cm3. Smaller phantoms were not available
during our study; thus, no conclusions could be drawn for these
phantoms. The results concerning plastic phantoms do not seem to be
consistent with modern dosimetry protocols, which recommend using
water phantoms for the calibration of ionization chambers in terms
of absorbed dose to water. However, we would like to point out that
with the results obtained in this work, we are not advocating the
calibration of ionization chambers in plastic phantoms. This should
be performed according to the internationally recognized codes of
practice.
Furthermore, we have
shown an important chamber-to-chamber dependence of the calibration
coefficients obtained in different phantoms, for the widely used NE
2571 ionization chamber, which demonstrates the individual physical
properties of each ionization chamber of the same type and stresses
the need of carrying out individual calibrations for each type of
ionization chamber.
AKNOWLEDGMENT
This work was carried out in the framework of an IAEA Coordinated Research project (ALG 11623).
REFERENCES
- Reich H. Choice of the measuring quantity for therapy-level dosimeters Phys Med Biol. 1979;24(5):895-900.
- Andreo P, Burns DT, Hohlfeld K, et al. Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water. IAEA Technical Report Series No 398. Vienna: International Atomic Energy Agency; 2000.
- Almond PR, Biggs PJ, Coursey BM, et al. AAPM's TG-51 protocol for clinical reference dosimetry of highenergy photon and electron beams. Med Phys. 1999;26(9):1847-1870.
- Mijnheer BJ. Variations in response to radiation of a nylon-walled ionization chamber induced by humidity changes. Med Phys. 1985;12(5):625-626.
- Mijnheer BJ, Wittkamper FW. Comparison of recent codes of practice for high-energy photon dosimetry. Phys Med Biol. 1986;31(4):407-416.
- ISO (International Organization for Standardization). Guide to the expression of uncertainty in measurement. 2nd ed. Geneva: ISO; 1995. (Published by ISO in the name of BIPM, IEC, IFCC, IUPAC, IUPAP and OIML.)