Phase versus amplitude sorting
of 4D-CT data
Nicole M. Wink,1 Christoph Panknin,2 and Timothy D. Solberg1,3
Department of Radiation Oncology,1 David Geffen School of Medicine at University of California, Los Angeles, Peter Morton Medical Building, Suite B265, Los Angeles, California 90095 U.S.A.; Siemens AG, Medical Solutions,2 Henkestrasse 127, D-91052 Erlangen, Germany; Department of Radiation Oncology,3 University of Nebraska Medical Center, 987521 Nebraska Medical Center, Omaha, Nebraska 68198 U.S.A.
nwink@mednet.ucla.eduReceived 3 October 2005; accepted 5 October 2005
Image quality of CT scans suffers when objects undergo motion. Respiratory motion causes artifacts, which prevents adequate visualization of anatomy. Four-dimensional CT (4D-CT) is a method in which image reconstruction of moving objects is retrospectively gated according to the recorded phase information of the monitored motion pattern. Although several groups have investigated the use of 4D-CT in radiotherapy, little has been detailed with regard to the sorting method. We present a new retrospective gating technique with sorting based on the amplitude of the motion trace. This method is compared to previously developed methods that sort based on phase. A 16-slice CT scanner (Sensation 16, Siemens Medical Solutions, Erlangen, Germany) was used to acquire images of two phantoms on a motion platform moving in two dimensions. The motion was monitored using a strain gauge inserted inside an adjustable belt. A 180° interpolation was used for reconstruction after gating. Significant improvement using the amplitude-sorting technique was observed, particularly when testing nonperiodic motion functions.
PACS numbers: 87.59.Fm, 87.53.Kn, 87.57.Ce
Key words: 4D-CT, RCCT, gated radiotherapy, respiratory motion, computed tomography
I. INTRODUCTION
Motion due to respiration causes artifacts in CT
scanning, which results in a skewed size, shape, and density of
objects in the image.(1-4) In
diagnostic imaging, patients are often instructed to hold their
breath for the duration of the scan to alleviate these problems.
This approach has two shortcomings. First, a long breath-hold places
increased burden on patients, many of whom are unable to do so for
an extended length of time.(5) Second,
the position of anatomical structures in the lung and diaphragm
region changes with respiration, and it is different during a full
inspiration breath-hold than it is during normal respiration.
Further, the administration of radiation therapy cannot be performed
under breath-hold due to the duration of the procedure. Therefore,
the ability to acquire images that reflect the position of
anatomical structures as a function of respiration during free
breathing is highly desirable.
In
four-dimensional CT (4D-CT), the patient is allowed to breathe
freely while the respiration trace is recorded for the duration of
the CT scan.(5-10) The time of image
acquisition is recorded along with the respiration trace for
subsequent correlation of image data with respiratory phase. This
correlation allows images that were acquired at the same phase of
respiration for all Z positions (the superior/inferior
patient axis) to be combined into a phase-specific volume. Doing
this for all phases of the respiratory cycle yields complete
information about the respiratory motion pattern of the entire
volume.
The order and method in
which the image data are reconstructed and sorted into proper phase
groups is important. Several investigators have proposed a method in
which the image data are first reconstructed and subsequently
sorted.(6-10) However, with a finite
sampling, this will yield reconstructed slices that were not
acquired precisely at the desired phase of respiration and the
desired Z position, but slightly off in either space or time
or both. This necessitates combining adjacent slices that were
acquired at slightly different phases of respiration, producing
spatial misalignment throughout the volume, and compromising image
quality. One group has proposed reconstructing the image data after
the sorting process.(5) This method is
not restricted to the location of the reconstructed slices and
therefore eliminates phase misalignment throughout the image volume.
The majority of the published
4D-CT techniques have used a time-based sorting method for
retrospectively gating the image data.(5-10) In this method, the phase position is
determined by specifying a percentage of the period of each cycle.
As will be demonstrated, data sorted in this manner are subject to
misalignment due to varying slopes, periods, and amplitudes in the
respiration trace. Rietzel et al.(9)
have proposed an amplitude-sorting method in which a constant
amplitude on the respiratory curve is chosen to designate the phase
position. Although this method eliminates the possibility of
misalignment, it also increases the potential for missing phase
sets. If a certain amplitude is not reached in all cycles of the
recorded respiration trace, then that phase set cannot be created
without a gap in the volume.
We
present a new amplitude-based sorting method in which the reference
phase is determined by specifying a percentage of the amplitude of
each cycle. This method is compared to the previously published
time-based sorting method. The ability to obtain phase-specific
image volumes for all phases of respiration with minimal
misalignment is beneficial for viewing CT images of freely breathing
patients with minimal motion artifacts.
II. METHODS AND MATERIALS
A. Acquisition
A
16-slice CT scanner (Siemens Medical Systems, Erlangen, Germany) was
used with a helical scanning protocol for 4D-CT. The scanner is
equipped with a direct anode cooling X-ray tube, enabling scanning
for extended periods of time. A new protocol was installed on the
scanner to facilitate scanning at a pitch of 0.1 for heavily
overlapped data acquisition. Sorting and reconstruction of all 4D-CT
data was performed off-line using software developed inhouse.
Two phantoms were placed on a
platform (Newmark Systems, Mission Viejo, CA) that moves under
computer control using high-precision stepper motors.(11) In this study, the platform was
programmed to move simultaneously in the superior-inferior
(z-axis) and anterior-posterior (y-axis) patient axes.
Motion was monitored using a strain gauge inside an adjustable belt
(Anzai Medical, Tokyo, Japan) placed around the motion platform
(Fig. 1). A differential RS422 signal was sent from the CT scanner
to the Anzai system to indicate X-ray start and stop times, thus
allowing correlation of image acquisition with the respiration
trace.
Fig. 1. Motion platform used to simulate breathing patterns with the motion-monitoring system. Adjustable belt is placed around the platform to monitor the AP motion direction. |
||
B. Gating
A new gating method was implemented for the helical 4D-CT procedure. The location of the midpoint of the desired phase image data was determined by taking a percentage of the peak-to-peak distance in each cycle. Each cycle was separated at the trough position on the respiration curve. In this context, a 50% phase would correspond to the peak amplitude, or end inhale bin. An illustrative example of the phase sorting and amplitude based methods on a nonperiodic motion pattern is shown in Fig. 2.
Fig. 2. Pictorial description of percentage-amplitude and percentage-time gating methods on a nonperiodic, constant amplitude motion pattern. Reference points separate each cycle in the motion function. |
||
The method of time-based sorting was compared to the percentage-amplitude gating method by plotting the phase position (midpoint of image data) on every cycle of the recorded motion function for every phase. The consistency in relative amplitude was observed because of its direct correlation with image set alignment.
C. Phantom Study
Two different phantoms were imaged for analysis and comparison of each 4D-CT sorting method. A high-contrast radiosurgery QA phantom consisting of a sphere, a box, a cylinder, and a cone inside an artificial skull is shown in Fig. 3. The American College of Radiology CT accreditation phantom consisting of two identical but separated spherical beads (both 0.28 mm in diameter) was also used.(12) The adjustable belt on the system monitored primarily the anterior-posterior (AP) direction of motion.
Fig. 3. Radiosurgery QA phantom including conical, cylindrical, rectangular, and spherical test objects. This phantom was used for gating method comparisons. |
||
Scans were acquired using a periodic motion function consisting of a peak-to-peak superior-inferior distance of 4 cm, AP distance of 1 cm, and period of 4.2 s to accurately mimic physiologically relevant motion patterns.(1,3,13-19) Scans of the stationary phantom were acquired with the same scanning parameters for comparative purposes. In addition, a nonperiodic motion profile was used, with periods ranging from 2 s to 4 s, and peak-to-peak amplitudes from 0.5 cm to 4 cm. The three motion profiles are plotted in Fig. 4.
Fig. 4. Plot of the three different motion functions programmed for gating method comparison. Periodic and nonperiodic patterns with and without varying amplitudes are diagrammed. |
||
A standard technique of 100 kV and 400 effective milliampere seconds (mAs) was used for each scan. Detector collimations of 0.75 mm and 1.5 mm were used with a 0.5-s rotation time. In every case, the maximum scan length, limited by a raw data file size of 2 GB, was acquired. For 4D-CT imaging, scans of moving phantoms were acquired at a pitch of 0.1 and were reconstructed with a 180° interpolation algorithm.(20) For nongated imaging, scans of stationary and moving phantoms were acquired at a pitch of 0.5, the lowest pitch available for nongated scans. The 0.5 pitch scans were reconstructed with a 2.0-mm slice width and 1.0-mm increment, while the 0.1 pitch scans used a 1.2-mm increment and a 2.0-mm slice width. The scans performed on each phantom are listed in Table 1.
| Table 1. Scan protocol for all phantom studies |
III. RESULTS
The largest difference between the phase and
amplitude-based gating methods was observed in the constant
amplitude, nonperiodic motion patterns. As expected, no significant
differences were observed in scans performed using the periodic
motion pattern with varying amplitudes. Therefore, the following
results comparing the two gating methods are from those scans using
the constant amplitude, nonperiodic motion
pattern.
The midpoint of the block
of image data used to construct each phase-specific image set is
plotted for every cycle of a recorded respiratory curve (Fig. 5).
The y-axis represents the relative amplitude of the monitored
motion pattern. The legend distinguishes each phase of motion (i.e.,
50% is end inhale). The x-axis represents the elapsed cycles
on the recorded motion trace. To construct a 4D dataset, the blocks
of image data from one phase of motion for every cycle recorded
would then be combined. It is important to note that the relative
amplitude of each phase from cycle to cycle should be consistent in
order to avoid spatial misalignment in the images.
Fig. 5. Comparison of (a) percentage-amplitude and (b) percentage-time gating methods. Each point represents the midpoint of a block of image data used to construct a phase specific image set. The y-axis is the relative amplitude of the monitored motion pattern, while the x-axis represents the elapsed cycles on the recorded motion trace. The legend distinguishes each phase of motion (i.e., 50% is end inhale). It is important to note the consistency in relative amplitude from cycle to cycle. |
||
The plots show that the percentage-amplitude gating method (Fig. 5(a)) results in a more consistent phase location over time, particularly at the phases of maximum motion (midinhale/ midexhale). The correlation coefficient (R2) to a linear fit of the midpoints for each phase is much higher (representing linearity) when reconstructing with the amplitude-sorting method than with the phase-sorting method (Table 2). The differences between the gating methods are smallest at peak and trough positions since the least distance is traversed at these points.
| Table 2. Coefficient of correlation for linear fit on midpoints for each phase and sorting method |
It is also apparent that each mirror phase set (i.e., midinhale and midexhale) corresponds more accurately with its counterpart when using the amplitude-sorting method. An example of the 30% and 70% phase positions for each sorting method is shown in Fig. 6. It is evident that the phase-sorting method is much more susceptible to inhale and exhale slope differences than the amplitude-sorting method.
Fig. 6. Comparison of (a) percentage amplitude and (b) percentage time gating methods on the 30% and 70% phase sets only, showing the phase location over nine cycles on a nonperiodic, constant amplitude motion pattern. These plots demonstrate the effect of inhale and exhale slope changes on phase location. |
||
IV. DISCUSSION
Retrospective gating of image data
before reconstruction has many advantages. Unlike other techniques,
it is not limited in terms of the location of the reconstructed
slices. Rather, each image is reconstructed at the exact desired
phase and spatial location. It is not dependent on the
reconstruction increment or spacing between slices. If gating is
performed after reconstruction, the slices must be close together to
reduce the potential for large phase position shifts from cycle to
cycle.
The percentage-amplitude
sorting method is a more reliable alternative to other published
techniques. The amplitude-sorting method detailed by Rietzel et al.
is optimal for phase location consistency from cycle to cycle, but
may not provide an image set for all phases of respiration in a
variable-amplitude breathing pattern. The phase position for each
cycle in the percentage-time sorting method is affected
significantly by amplitude, period, and inhale and exhale slopes,
which makes this technique susceptible to misalignment in image
location throughout a phase-specific image set. Although the
percentage-amplitude method presented here is also susceptible to
irregularities in amplitude from cycle to cycle, it adequately
handles changes in period and slope. It is important to note that
any misalignments in image location would directly affect the
quality of phase-specific image sets.
The programmed motion functions
were chosen to most accurately mimic physiologically relevant motion
patterns based on various published abdominal motion studies.(3,14-19,21,22) While a wider range of
motion speed could be easily be accommodated by the platform, a
4.2-s period was implemented. If the detector rows move past the
image plane before one full respiratory cycle plus one fan angle is
complete, a gap in the image dataset occurs. This situation,
encountered when scanning in helical mode, enforces a period limit
on the monitored motion function.(5)
The 4D-CT approach used in
this paper is based on external monitoring; however, internal motion
may not correlate directly with the external monitor.(23) The incorporation of internal target
motion into future respiratory motion correction techniques would
reveal the degree to which external monitoring can accurately
reflect internal motion. Monitoring internal motion in patients
would require fluoroscopy or surgically implanted markers; however,
it would provide precise answers for 4D verification.
V. CONCLUSIONS
The use of 4D information has numerous applications
in furthering radiotherapy treatment planning. Based on the
experiments shown, it potentially reduces motion blurring in CT
images of therapy patients that are unable to hold their breath. In
addition, monitoring the location of the target provides the extent
of movement in any direction and the trajectory it follows. This
information can be used for target volume definition to ensure
complete coverage of the target. Knowing the trajectory can also be
helpful when tumor tracking or gated therapy is desired.
Several advances have been made to
ameliorate the process of acquiring a 4D image dataset. The results
presented here demonstrate a novel retrospective sorting method
based on the full respiration curve that provides an image set at
every phase. The benefits of this method over established phase
sorting techniques have been demonstrated on multiple phantoms.
ACKNOWLEDGMENTS
This work was supported by grant #03-028-01-CCE from the American Cancer Society and Siemens.
REFERENCES
- Allen AM, Siracuse KM, Hayman JA, Balter JM. Evaluation of the influence of breathing on the movement and modeling of lung tumors. Int J Radiat Oncol Biol Phys. 2004;58(4):1251-1257.
- Aruga T, Itami J, Aruga M, et al. Target volume definition for upper abdominal irradiation using CT scans obtained during inhale and exhale phases. Int J Radiat Oncol Biol Phys. 2000;48(2):465-469.
- Ekberg L, Holmberg O, Wittgren L, Bjelkengren G, Landberg T. What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer? Radiother Oncol. 1998;48(1):71-77.
- Gagne IM, Robinson DM. The impact of tumor motion upon CT image integrity and target delineation. Med Phys. 2004;31(12):3378-3392.
- Keall P. 4-dimensional computed tomography imaging and treatment planning. Semin Radiat Oncol. 2004;14(1):81-90.
- Ford EC, Mageras GS, Yorke E, Ling CC. Respiration-correlated spiral CT: A method of measuring respiratory-induced anatomic motion for radiation treatment planning. Med Phys. 2003;30(1):88-97.
- Low DA, Nystrom M, Kalinin E, et al. A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing. Med Phys. 2003;30(6):1254-1263.
- Pan T, Lee T-Y, Rietzel E, Chen GTY. 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT. Med Phys. 2004;31(2):333-340.
- Rietzel E, Pan TS, Chen GTY. Four-dimensional computed tomography: Image formation and clinical protocol. Med Phys. 2005;32(4):874-889.
- Vedam SS, Keall PJ, Kini VR, et al. Acquiring a four-dimensional computed tomography dataset using an external respiratory signal. Phys Med Biol. 2003;48(1):45-62.
- Hugo GD, Agazaryan N, Solberg TD. An evaluation of gating window size, delivery method, and composite field dosimetry of respiratory-gated IMRT. Med Phys. 2002;29(11):2517-2525.
- McCollough CH, Bruesewitz MR, McNitt-Gray MF, et al. The phantom portion of the American College of Radiology (ACR) computed tomography (CT) accreditation program: Practical tips, artifact examples, and pitfalls to avoid. Med Phys. 2004;31(9):2423-2442.
- Giraud P, De Rycke Y, Dubray B, et al. Conformal radiotherapy (CRT) planning for lung cancer: Analysis of intrathoracic organ motion during extreme phases of breathing. Int J Radiat Oncol Biol Phys. 2001;51(4):1081- 1092.
- Hof H, Herfarth KK, Munter M, et al. The use of the multislice CT for the determination of respiratory lung tumor movement in stereotactic single-dose irradiation. Strahlenther Onkol. 2003;179(8)542-547.
- Lujan AE, Larsen EW, Balter JM, Ten Haken RK. A method for incorporating organ motion due to breathing into 3D dose calculations. Med Phys. 1999;26(5):715-720.
- Seppenwoolde Y, Shirato H, Kitamura K, et al. Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. Int J Radiat Oncol Biol Phys. 2002;53(4):822-834.
- Shirato H, Shimizu S, Kitamura K, et al. Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor. Int J Radiat Oncol Biol Phys. 2000;48(2):435-442.
- Sixel KE, Ruschin M, Tirona R, Cheung PCF. Digital fluoroscopy to quantify lung tumor motion: Potential for patient-specific planning target volumes. Int J Radiat Oncol Biol Phys. 2003;57(3):717-723.
- Stevens CW, Munden RF, Forster KM, et al. Respiratory-driven lung tumor motion is independent of tumor size, tumor location, and pulmonary function. Int J Radiat Oncol Biol Phys. 2001;51(1):62-68.
- Flohr T, Ohnesorge B, Bruder H, et al. Image reconstruction and performance evaluation for ECG-gated spiral scanning with a 16-slice CT system. Med Phys. 2003;30(10):2650-2662.
- Allen AM, Siracuse KM, Hayman JA, Balter JM. Evaluation of the influence of breathing on the movement and modeling of lung tumors. Int J Radiat Oncol Biol Phys. 2004;58(4):1251-1257.
- Giraud P, De Rycke Y, Dubray B, et al. Conformal radiotherapy (CRT) planning for lung cancer: Analysis of intrathoracic organ motion during extreme phases of breathing. Int J Radiat Oncol Biol Phys. 2001;51(4):1081- 1092
- Koch N, Liu HH, Starkschall G, et al. Evaluation of internal lung motion for respiratory-gated radiotherapy using MRI: Part I-Correlating internal lung motion with skin fiducial motion. Int J Radiat Oncol Biol Phys. 2004;60(5):1459-1472.
© 2005 Am. Coll. Med. Phys.