Verifying monitor unit calculations
for tangential breast fields
Ian Kay and Peter Dunscombe
Tom Baker Cancer Centre, 1331 29th Street NW, Calgary, Alberta T2N 4N2; Departments of Oncology and Physics and Astronomy, University of Calgary, 2500 University Drive NW, Calgary Alberta T2N 1N4 Canada
iankay@cancerboard.ab.ca; peterdun@cancerboard.ab.caReceived 19 August 2005; accepted 11 April 2006
An essential component of quality assurance in radiation therapy is verifying the accuracy of monitor unit calculations. Differences between sophisticated algorithms using 2.5D or 3D calculations and simpler monitor unit check algorithms or hand calculations assuming a flat water phantom must be expected. For many anatomical sites, such differences are small and of little or no consequence in the context of expected clinical impact. However, for tangential breast fields the discrepancies are considerably larger than those that would generally be considered acceptable. A simple model to reconcile the differences between sophisticated and simple algorithms is presented, based on replacing the breast contour with a triangular or elliptical contour and using this to estimate an equivalent rectangular prism providing equivalent scatter to the prescription point. The elliptical approximation reconciles the observed differences in calculated monitor units. The analysis we present can assist the treatment planning physicist in selecting a method and tolerance window for verifying monitor unit calculations for tangential breast fields.
PACS: 87.53.Kn, 87.53.Tf, 87.53.Xd
Key words: radiotherapy, breast, tangent field, treatment planning, quality assurance
I. INTRODUCTION
External beam therapy quality assurance (QA)
requires that plans be validated,(1)
and an independent monitor unit calculation is an essential part of
that QA process. At the Tom Baker Cancer Centre our policy in this
regard requires all plans (generated by ADAC PinnacleT) to be
downloaded to RadCalcT for the independent monitor unit check. If
disagreement between PinnacleT and RadCalcT calculated monitor units
is larger than the greater of 2% or 2 monitor units (MU) for any
field, the plan must be referred to a physicist for final approval.
This policy provides adequate assurance of the accuracy of the
calculated monitor units while limiting the workload on the
physicists. As will be seen below, fewer than 25% of beams fail to
meet this established criterion for all anatomical sites except the
breast, where no beams display agreement between PinnacleT and
RadCalcT of 2% or better.
Missing
tissue, field "flash," and an oblique irradiation of the surface
combine to make the dosimetry of breast irradiation using tangent
fields complex. Modern treatment-planning systems (such as
PinnacleT) meet this challenge by accounting for the external body
surface (hence field "flash" and missing tissue) and oblique
incidence inherently in their algorithm.
Validation of the monitor unit
calculation is performed either using hand calculations or one of
many software packages, both "home-grown"(2) and commercial. These methods usually
either assume or use data measured for radiation beams perpendicular
to a flat water surface and with full scatter conditions. These
conditions are not realized in tangential breast fields; hence,
discrepancies arise when comparing the monitor unit calculations.
Due to the approximations in
monitor check routines, strategies for validating breast plans have
been developed. Many of these involve some adjustment of the field
size to an "equivalent square" field and are frequently based on
experience. For example, Ayyangar et al.(3) describe a method of correcting the
"second check" for the effects of flash and missing tissue. This
method requires a measurement of a "missing tissue factor" as a
function of a "percent missing tissue index," and an estimated
equivalent square based on the beam's-eye-view extent of field
flash. They report 1.5% agreement of monitor units over 15 fields
examined, with a standard deviation of 2.0%.
We have investigated alternative
schemes for correcting monitor units calculated for tangential
breast fields under simple geometrical conditions. The schemes are
based on the breast profile and the location of the prescription
point. The notion of equivalent square is extended, replacing the
breast contour with an "equivalent rectangular parallelepiped,"
which would provide similar scatter to the prescription point.
Performance of this method appears similar to that of Ayyangar et
al.(3)
Currently, our clinic plans
in 2.5D using a single digitized breast profile (in PinnacleT) and
checked either with software (RadCalcT) or a hand calculation,
resulting in the disagreements described here. We anticipate that
full 3D planning will make reconciliation of the monitor unit
calculations more difficult. This initial study showed that an
equivalent scattering area approximation works well in 2.5D, and
extension to an equivalent scattering volume in 3D is a promising
strategy for using current second check tools and reconciling the
differences observed.
II. METHODS
Eighty-six tangential breast plans, each with a medial and a lateral field, and each treated at 6 MV, were culled at random from our clinical database. These 2.5D plans were prepared by acquiring a 2D contour of the breast and using ADAC PinnacleT to calculate both the dose distribution and the monitor units for each field. According to our planning protocol, 2 cm of flash is added to each field, and the dose is prescribed to a point approximately midway between the beam entry points and one-third of the distance from the posterior field border to the anterior breast surface (Fig. 1). The fields included in this study were asymmetric, with wedges and half-blocking in most cases, although in some cases 1 cm or 2 cm are allowed on the chest wall side of the central axis.
Fig. 1. Histogram of the percent disagreement in monitor units calculated by RadCalc™ compared to Pinnacle™ for 86 tangential breast fields showing an average difference of 4% due to differences in the computational algorithm. MU = monitor units |
||
The resulting plans
were exported to RadCalcT for a second check. Comparing monitor
units from PinnacleT and RadCalcT, we observe a difference of -4.4%
and a standard deviation of 1.4%. (RadCalcT prescribed fewer monitor
units than ADAC PinnacleT.)
As a
point of comparison, 405 fields treating sites other than
breast/chest wall were examined. The mean difference between the
monitor units calculated by RadCalcT and PinnacleT is -0.7%, with a
standard deviation of 1.6%, consistent with the work of Leszczynski
and Dunscombe.(4) The largest
deviations (~8%) occurred for fields treating the arm, axilla, or
sinus, where contour, flash, and inhomogeneity are expected to
produce discrepancies. It should be noted that these other fields
are almost all full 3D conformal plans and usually included
corrections for tissue heterogeneity, whereas our breast tangent
fields are usually calculated assuming homogeneous tissue density.
Yet for these other 405 fields, agreement is much better than for
the breast fields, indicating that it is this particular geometry
that is pushing RadCalcT beyond the valid limits of its assumptions.
Plots of the PinnacleT versus
RadCalcT monitor units are highly linear and unrevealing; R2 is in excess of 0.999 for
both the breast and nonbreast data. The slope of the regression line
is about 1.04 for the breast data versus 1.007 for the nonbreast
site. More revealing are the histograms of the percentage
disagreement of RadCalcT's monitor units from that of PinnacleT's
(Figs. 1 and 2).
Fig. 2. Histogram of the percent disagreement in monitor units calculated by RadCalc™ compared to Pinnacle™ for 405 fields applied to sites other than the breast. On average, the agreement is better and the distribution is tighter, demonstrating that tangential breast fields pose a special challenge. MU = monitor units |
||
The first and easiest
correction to the geometrically simple approach (RadCalcT) is to
reduce the field size for calculating phantom scatter by the amount
of flash that we will denote by f. We have simply assumed
that planners have rigorously held to the prescribed 2 cm of flash.
We next need to account for an
overestimation of tissue providing scatter to the prescription
point. Consider an audacious assumption, that the amount of scatter
to the prescription point would remain constant if the breast tissue
were rearranged into a rectangular prism. The corrections we
investigated are based on either a triangular or an elliptical
approximation of the breast contour as shown in Fig. 3. The degree
to which either approximates the breast contour will clearly vary
for any given patient and field. Also note that we use the depth, d, to the prescription point as half of one dimension of the
rectangle as shown in Fig. 4.
Fig. 3. A typical breast contour (solid) can be approximated as a triangle (long dashes) or as an ellipse (short dashes). A half-blocked tangential field is indicated incident from the left side. pp = prescription point; iso = isocenter |
||
Fig. 4. Replacing the triangle or ellipse in Fig. 3 with a rectangle of “equivalent mass.” pp = prescription point; iso = isocenter |
||
The rectangle width w′ is chosen such that the area w′d is equal to the area of the triangle or the half ellipse in Fig. 3. The difference between d and the depth to the isocenter is ignored. This "equivalent rectangle" would be similar to the field illuminated by a beam of width w′ on a flat water phantom, ignoring the divergence of the field's upper border. For our two approximations:
(1) |
(2) |
The original rectangular field of length L and width w + f had an equivalent square field size F = 2L(w + f)/(L + w + f). The equivalent rectangular scatterer has a width w′ as given above. The equivalent square field size F is then replaced by F′, calculated using the original field length L, but the revised field width w′ so that F′ = 2Lw′/(L + w′); hence
(3) |
When calculating our second check (using RadCalcT, hand calculation, or other based on full scatter conditions), we replace the (w + f) by L field of equivalent square size F, with a new field w′ by L and equivalent square size F′. This change of field size should account for most of the missing scatter, by changing the values of S p and tissue phantom ratio (TPR) used in the calculation. This can be carried out with a simple multiplicative correction factor:
(4) |
with F′ related to F as given above. This factor is greater than 1, and dividing the number of monitor units calculated for a flat water phantom by this factor should approximate the correct number of monitor units (as given by PinnacleT or other modern planning systems) within the geometrical assumptions of the triangle and ellipse.
III. RESULTS
Monitor units were recalculated in RadCalcT for our selected 86 pairs of tangential breast fields, reducing the field width by 2 cm to correct for field flash. This leaves the effect of surface contour uncorrected. The results are summarized in the histogram in Fig. 5 and in Table 1.
Fig. 5. Histogram of the percent disagreement in monitor units after RadCalc™’s field dimension is altered to account only for excess field flash |
||
| Table 1. Performance of correction strategies |
The proposed correction factor in Eq. (4) was then applied, using both the triangle and elliptical approximations, yielding the results in Table 1 and the histograms in Figs. 6 and 7.
Fig. 6. Histogram of the percent disagreement in monitor units after RadCalc™’s field dimension is altered to account both for excess field flash and a triangular estimate of the missing tissue |
||
Fig. 7. Histogram of the percent disagreement in monitor units after RadCalc™’s field dimension is altered to account both for excess field flash and an elliptical estimate of the missing tissue |
||
The sample data have
a range of depths to the prescription point of 9.1 ± 1.6 cm and
field widths of 10.2 ± 3.1 cm. (Results are quoted with 1 SD,
corresponding to 68% CI.)
As a
further simplification, consider approximating the superior/inferior
length of the treatment field to be a constant nominal length of 20
cm (our data have an average of 19.0 ± 2.2 cm). This alters the data
only slightly to that shown in Table 2.
| Table 2. Performance of correction strategies with the further assumption of a constant 20 cm field length |
IV. DISCUSSION
Flash reduction alone provides a partial
reconciliation of the monitor unit disagreement, but leaves an
average disagreement of 2.7% unresolved.
From Table 1 and Fig. 6, it
appears that the triangle approximation overcorrects the monitor
units in most cases. This is expected because it likely
overestimates the amount of missing tissue as can be appreciated in
Figs. 2 and 3. The increase in standard deviation also indicates
that this approximation is not performing well. In contrast, the
elliptical approximation performs quite well.
Assuming a constant field length
as in Table 2 preserves the same pattern; the triangle approximation
overcorrects, and the ellipse appears to perform well. Making this
further simplifying assumption allows us to produce a 2D table of
correction factors that are functions of only field size and depth.
The elliptical approximation is
surprisingly good at correcting the monitor unit calculation, better
than we expected given the very crude approximations being made. The
range of differences (max to min) in the original uncorrected data
is 0.0% to -7.7%. After correction with the elliptical model the
range is 3.4% to -4.2%. The mean is being corrected for lack of
scatter, contour, etc., while preserving the range of values,
increasing our faith in the results.
TG-40(1) recommends that primary and secondary
monitor unit calculations should agree to within ±2%. The data in
Table 1 suggest that uncorrected tangential breast fields would meet
this criterion only 4.3% of the time. Using the elliptical
correction factor, even with the residual -1.2% difference, 68% of
the plans would pass this test. As a point of comparison, for the
405 fields delivered to sites other than breast 75% would meet the
2% tolerance level, and 25% would require checking by a physicist.
However, the question does remain
as to whether or not one needs to apply a correction factor to check
monitor units for tangential breast fields. If the analysis above,
together with the experimental validation of PinnacleT by Ayyangar
et al.,(3) can be used to infer both
that the physical reason for the observed discrepancies is
understood and that PinnacleT (or any other modern algorithm) is
inherently accurate under the relevant geometric conditions, then it
would be sufficient to specify that secondary monitor units be in
the range of -6.4% to -2.4% to pass the checking procedure,
encompassing 80% of the breast plans. Implementation of such a
policy provides the same assurance of accuracy as for other
anatomical sites and results in a site-independent workload for
physicists performing plan checks.
If one is confident that the
increase of standard deviation from 1.4% to 1.7% in Table 2 was due
to variation in flash and field length, and not related to any
serious errors in planning, one might increase the acceptance
criteria to 1.7/1.4 × 2% = 2.4% or 1.4 standard deviations. This
would pass about 84% of plans, and 16% would still be checked. Or
one could select a range that conforms to one's comfort level based
on an estimation of the frequency of a serious error. The approach
presented here allows each center to make an informed decision.
V. CONCLUSIONS
We have considered the problem of disagreement in
monitor unit calculations between a primary and a second check for
2.5D planning of tangent breast fields. The disagreement arises when
the second check is performed with a simplified algorithm that does
not account for the absence of full scatter conditions.
We have described a simplified
method of correcting for the absence of full scatter conditions in
the case of tangential breast fields based on a geometrical estimate
of an equivalent scattering rectangular parallelepiped. The
assumptions are justified on physical grounds and supported by the
statistics demonstrating the efficacy of the method. The results
presented give the clinical physicist the option of correcting
monitor units for better agreement with more accurate 2.5D
calculations or simply shifting the tolerance window to accommodate
the approximations made during checking. The analysis can, and
should, be performed by any user by a retrospective examination of
plans to ensure that they get acceptable performance before
implementing it as a solution in their clinic.
Extension of this method to the
case of full 3D planning is obvious in principle; an estimate of the
treated volume of breast tissue could be used to calculate an
equivalent rectangular parallelepiped. Extension of our ellipse to a
half ellipsoid is tempting, but may prove to be a poor approximation
to real anatomy. As our clinic moves to 3D breast planning, we hope
to collect the data that would allow us to test extensions to this
simple model.
ACKNOWLEDGMENT
Thanks to Mr. S. Morgan for his assistance collating and analyzing this data.
REFERENCES
- Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994;21:581-618.
- Knoos T, Johnsson SA, Ceberg CP, et al. Independent checking of the delivered dose for high-energy X-rays using a hand-held PC. Radiother Oncol. 2001;58:201-208.
- Ayyangar KM, Saw CB, Gearheart D, et al. Independent calculations to validate monitor units from ADAC treatment planning system. Med Dosim. 2003;28:79-83.
- Leszczynski KW, Dunscombe PB. Independent corroboration of monitor unit calculations performed by a 3D computerized planning system. J Appl Clin Med Phys. 2000;1:120-125.
© 2006 Am. Coll. Med. Phys.