Fractionated stereotactic
radiotherapy for pediatric
patients with retinoblastoma
Patrick D. Higgins,1,a Bruce J. Gerbi,1 Mark Macedon,2 and Kathryn E. Dusenbery1
University of Minnesota Department of Therapeutic Radiology-Radiation Oncology,1 University of Minnesota, Minneapolis, Minnesota; Marshfield Clinic,2 Marshfield, Wisconsin, U.S.A.
higgi010@umn.eduReceived 30 June 2005; accepted 26 January 2006
In this report, we discuss the application of a modified Gill-Thomas-Cosman (GTC) relocatable head frame to enable fractionated stereotactic radiotherapy (SRT) of infants under anesthesia. This system has been used to treat two infants, ages 12 and 18 months, for bilateral retinoblastoma on a Varian 6/100 linear accelerator. The GTC head frame was used to reproducibly position and treat the orbits of these children to between 2520 cGy and 3960 cGy in 180-cGy fractions. A standard head and neck tray, with accompanying thermoplastic mask, was adapted to mount to the head frame to enable these treatments. We found the maximum average deviation in the repeat fixations, as compared with the initial fitting data, to be ±2 mm. The overall average difference and standard deviation in measurement was 0.47 ± 0.63 mm for the first case and 0.19 ± 0.94 mm for the second case, with a combined average of 0.35 ± 0.79 mm overall from a total of 381 point measurements. The stereotactic treatment plan (Radionics®) incorporated a single isocenter for each orbit and 3 or 4 arcs per isocenter. An intercomparison has been made between this technique and a standard lateral field technique, designed using the stereotactic radiosurgery (SRS) planning system. Dose-volume histograms and corresponding normal tissue complication probabilities (NTCP) based on pediatric bone growth inhibition have been calculated for each method for the orbital bone areas. We found that the NTCP is reduced from 95% or more in the standard treatment method to 16% or less with SRT. Use of the modified head frame provides excellent setup reproducibility, facilitates access to patients for anesthesia, and reduces the chances of a poor cosmetic result in these growing children.
PACS number: 87.53.Ly
Key words: pediatric radiotherapy, stereotactic radiotherapy, patient fixation
aCorresponding author: Patrick D. Higgins, Ph.D., Department of Therapeutic Radiology-Radiation Oncology, University of Minnesota Medical School, Mayo Mail Code 494, Minneapolis, MN 55455 U.S.A.
I. INTRODUCTION
Retinoblastoma is an uncommon disease that affects
600 children per year in the United States. It is the most common
orbital tumor in children and accounts for 5% of childhood
blindness. Usually manifesting at a very early age (95% diagnosed
before age 5), it is unilateral in 2/3 of patients and bilateral in
1/3 of patients. Bilateral disease is nearly always a sign of a
hereditary mutation, while the unilateral disease is more commonly
sporadic.
The treatment of
intracranial tumors in children poses several unique challenges that
require a different approach from adults. Children may be more prone
to develop certain specific adverse sequelae as a result of
radiation therapy. Neuropsychological sequelae such as memory
problems and learning disabilities have been observed.(1) Endocrine dysfunction unrelated to
underlying malignancy may also be seen.(1) Second malignancies are also a concern
because patients successfully treated as children survive well into
adulthood.(1) Finally, radiation to
developing bones may stunt growth, resulting in possible facial
deformities.(2) In an effort to limit
long-term adverse effects by minimizing bone dose, wherever
possible, clinicians have increasingly been turning to more
conformal methods of delivering radiation, including stereotactic
radiosurgery (SRS), in which large doses (10 Gy to 20 Gy) are
delivered in a single treatment, and stereotactic radiotherapy
(SRT), which more closely follows conventional radiation therapy,
where a course of treatment is fractionated into a number of daily
treatments (1.5 to 2.5 Gy/fraction). Proton therapy has also been
explored as a treatment option, along with chemotherapy,
thermo-chemotherapy, hyperthermia, and photocoagulation.(3)
Compared with
single-fraction SRS, which is the only possible method of treatment
on Gamma Knife-type treatment units, fractionated stereotactic
radiation therapy also has a considerable advantage for the
treatment of intracranial tumors in children. As with conventional
radiation therapy, daily fractionation is better tolerated by normal
tissue structures.(1,4) In addition,
repeat fixation devices are much less traumatic than the invasive
frames required for the SRS method and may be removed much more
quickly, if the need arises.
The
delivery of fractionated stereotactic treatment in young children
poses additional challenges. Small children are unable to remain
motionless during radiation therapy and thus require anesthesia.
Most conventional immobilization techniques requiring both patient
cooperation and teeth (for bite block systems) cannot be used in
their standard configuration. In this report, we detail a simple
modification that may be applied to a standard fixation device that
circumvents these problems and allows us to take advantage of this
powerful technique for the treatment of infants.
The most common noninvasive frames
for SRT are the Gill-Thomas-Cosman (GTC) frame (used here at the
University of Minnesota) and the Laitinen system (LS) frame,(5,6) although variants also exist. The
original GTC frame of Gill et al.(7)
modified by Kooy et al.(8) uses a rigid
ring, split to allow an opening for patient access. Attached to the
ring, as an assembly, is a patient-specific oral appliance (dental
plate) custom-fitted to the upper teeth and palate of the patient. A
head support plate (back plate) is attached to the back of the ring
and filled with a quick-setting compound, which is formed to the
patient's occiput, providing both support while the patient lies
down and additional immobilization. A three-point Velcro strap is
attached to the lateral sides of the ring and the back plate.
Pulling this tightly together is the final immobilization step. The
ring and immobilization assembly is finally attached to a flange,
which is connected to the end of the CT or treatment couch. Variants
of these frames have been used for pediatric cases, specifically, a
smaller version of the LS frame and an adaptation of the GTC frame
at the Boston Children's Hospital (BCH), also reviewed by Kooy et
al.(8) The BCH frame uses some of the
features of the LS frame system and replaces the oral immobilization
appliance with earplugs and a thermoplastic mask, which includes
only the anterior face (nose, orbits, forehead, and temples). The
BCH frame allows anesthesia access but is limited in the sense that
a significant amount of material is added around the cranium, which
may reduce access or otherwise affect the dosimetry of
retinoblastoma treatment.
Confirmation of position is
performed using a depth confirmation helmet. Reproducibility of
setup has been satisfactory with Kooy et al.(8) reporting on data from 20 patients a
mean measurement error of 0.71 ± 0.06 mm (range 0.31 mm to 1.22 mm),
mean lateral movement of 0.35 ± 0.06 mm (range 0.07 mm to 0.79 mm),
mean superior movement of 0.52 ± 0.09 mm (range 0.00 mm to 0.94 mm
with one outlier at 1.77 mm), and mean occipital movement of 0.34 ±
0.09 mm (range 0.00 mm to 1.30 mm). More recently, Burton et al.(9) reviewed the value of using the depth
helmet to evaluate repositioning accuracy.
Of two patients treated in our
clinic with fractionated stereotactic radiotherapy, one was 18
months old; he presented at age 6 months with decreased red reflexes
in both eyes. He was diagnosed with bilateral retinoblastoma and
subsequently underwent several cycles of chemotherapy and
cryotherapy. He was referred to our clinic when he developed a
recurrence in his left retina only months after completing
chemotherapy. The tumor in the right eye remained quiescent, but the
decision to treat both eyes stereotactically was made. Virtually the
same clinical scenario was true for our second patient (12 months
old) with bilateral retinoblastoma. For each patient, we used our
modified GTC frame. We discuss that construction as well as the
quality of positioning accuracy and reproducibility in the
following.
To further review the
advantages of SRT over conventional treatment techniques for
pediatric bilateral orbit treatment, we constructed and analyzed
dose-volume histograms for the orbital bones, comparing SRT with
simulated, parallel-opposed lateral fields. There are more than a
few ways of designing bone-sparing treatment fields beyond a simple
bilateral technique. For example, wedged pairs (for single orbits)
and 3D-conformal field designs would be expected to better reduce
bone dose.(3) For the SRT case, there
is certainly no question that penumbra is much better limited, even
if 3D or intensity-modulated radiotherapy approaches are used for
comparison. We originally planned to compare a variety of
techniques, but would have had to use different planning systems,
requiring the targets and bones to be redrawn and dose-volume
analyses to be, potentially, slightly different between them. By
doing all the planning on the same treatment-planning system
(Radionics®), we felt the comparison
would be the most self-consistent. Further, the comparison
demonstrates access to advanced 3D treatment planning using this
immobilization system.
II. METHODS
We began by adapting a commercially available head
and neck tray (Thermoplastic Masks, MedTec, Orange City, IA) to the
GTC ring by simply cutting it to size and screwing it onto the two
rear stanchions that normally hold the GTC back plate. These trays
are made in-house but are available commercially, in slightly
different forms. They typically include a machined slot with at
least two mounting pins for head rests and additional pins and
clamps for guiding and retaining Aquaplast® immobilization masks. The dental fixation
plate and Velcro straps from the original design are not needed.
For the two patients treated we
used a 1/4-in. plastic (acrylic) plate. A more rigid plate, such as
one made with stainless steel or aluminum, could also be used. In
these instances, it was determined that the infants' small size and
weight did not incur significant flexing of the plastic. The ring
and support plate are attached to the end of the simulation room
table using a standard bracket. The ring assembly may be split above
the mounting bracket by releasing two screws and sliding the ring
upward along the revealed 4-in. to 5-in. long guide pins. With the
ring open, the patient may be inserted through the ring and the head
cradled in a selected, angled head support. The ring is then closed
and secured. A thin, thermoplastic mask, which has the property of
being highly elastic when warm and rigid on cooling, is stretched
over the upper half of the face and attached to the plate with nylon
wing nuts (Fig. 1). Care is taken to open the mesh over the
patient's tragi and fleshy canthus to allow a clear view of these
anatomical structures to aid patient positioning. A measurement is
taken from the nasion to the edge of the ring to fix
forward-backward tilt. Right and left lateral level and alignment
marks are also made, aligned with the tragi, and checked prior to
each treatment with the lateral treatment room alignment lasers. An
advantage of this system over the BCH frame is that the immobilizing
mask is thin, it is standard for any department, and it is simply
attached to a standard, albeit modified, baseplate without
forfeiting anesthesia access.
Fig. 1. Initial setup. GTC ring (closed) with thermoplast mask being screwed into the baseplate. This combination is the primary difference with the BCH design. |
||
A key component of the GTC repeat fixation system is a secondary method for evaluating the accuracy of repositioning.(9) This is accomplished using a Radionics® depth confirmation helmet (Fig. 2). This helmet mounts to the base ring and includes 25 holes spaced into three rings. A posterior view, showing the locations of the holes for comparison with Fig. 2, is given in Fig. 3. A depth gauge having millimeter markings is used to measure the depths from the top of each hole to the patient's skin (mask in some instances). These depths provide an aid in evaluating repositioning problems prior to each treatment. Discrepancies of more than 2 mm are reviewed and may require that the patient be repositioned.
Fig. 2. Depth helmet verification step. Depth helmet shown attached to GTC ring. Measurements are taken using a depth gauge of the skin-to-helmet distances. |
||
Fig. 3. Display of depth helmet geometry as seen from directly behind the helmet. |
||
Following the initial fitting, patients receive a CT scan. (The frame is not MRI compatible.) A fiducial marker cage is attached to the base ring, and a series of 2 mm × 2 mm axial scans is acquired through the entire head. Treatment planning is performed using the Radionics® X-Knife 4T software. In both preliminary cases, bilateral orbits and disease extension (with 4-mm to 5-mm margins) were treated using single isocenters for each orbit and 3.0-cm to 3.5cm diameter inserts. Three or four arcs (approximately 300° total arc angle) were delivered to each orbit. Figure 4 displays the arc arrangement for patients 1 (Fig. 4(a)) and 2 (Fig. 4(b)). The corresponding dose distributions in which we see the prescription isodose volumes (90%) (orange, transparent region) are shown in Fig. 5.
Fig. 4. Stereotactic arc arrangement for patients 1 (a) and 2 (b). |
||||
Fig. 5 Isodose volumes (shown as orange, transparent surfaces) for 90% isodose (prescription isodose) for patients 1 (a) and 2 (b). |
||||
The first patient
(Figs. 1 and 2) (18 months old) received 3600 cGy in 20 fractions
over 26 days to his right eye and 2520 cGy in 14 fractions over 18
days to his left. The second patient (12 months old) received 3960
cGy in 22 fractions over 28 days to her right eye and 3600 cGy in 20
fractions over 26 days to her left.
The Radionics® X-Knife 4T software was also used to
simulate a parallel-opposed lateral treatment technique. To
accomplish this, 1° right and left lateral arcs were set up and
centered at midline, between the orbits. Five-centimeter diameter
cutouts were shaped, using asymmetric jaws, into rectangular fields
having a margin of approximately 1 cm around the indicated targets.
Figure 6(a) displays a typical beam's-eye view of the lateral
fields, and Fig. 6(b) the resultant 90% volume dose distribution.
Fig 6. Right-lateral beam’s-eye view of patient 1 lateral field (a) and 90% isodose volume (b). |
||||
Dose-volume histograms were constructed for each treatment method. The effective volume method of Kutcher et al.(10) was used to generate effective doses and volumes that apply Lyman's(11) theory with Burman et al.'s(12) parameterization to calculate normal tissue complication probabilities (NTCP). Classically, this model is written as follows:
where, for effective volume v, t =
(D - TD50(v))/(m ×
TD50(v)) and TD50(v) = TD50(1) × v -n
.
The parameters m
and n were estimated to be 0.1 and 0.25, respectively,
consistent with typical values obtained by Burman et al. The 50%
probability of bone growth arrest (TD50),
used in the calculation, was estimated as 30 Gy, based on Eifel et
al.'s(13) assessment of complication
probabilities in growing bone. They noted that the region of
steepest dose effect was 15 Gy to 30 Gy, with little significant
growth abnormalities reported below 25 Gy.
III. RESULTS AND DISCUSSION
Each of the two patients received between 20 and 22 total treatments. During this time, positioning accuracy was monitored by performing measurements using the depth helmet before each treatment. Using the initial setup measurements taken prior to CT data acquisition as our baseline, we have computed the deviations between those measured depths and those taken each day prior to treatment. In Fig. 7 we have summarized those measurements in graphical form as the average deviation from baseline as a function of location (see Fig. 3). We found the maximum average deviation to be ±2 mm (locations B and H), due to difficulty in obtaining depths to the skin (through the mask) at these locations. The overall errors averaged to 0.47 ± 0.63 mm for the first case, 0.19 ± 0.94 mm for the second case, and 0.35 ± 0.79 mm overall from a total of 381 point measurements. Not all points were sampled before each treatment, with an average of 11/25 checkpoints taken for the first patient and 7.5/25 for the second patient. It was felt that these randomly distributed points provided adequate sampling of the setup precision.
Fig. 7. Summary graph of the average setup variations on a point-by-point basis, as assayed with the depth helmet, for locations mapped out in Fig. 3. |
||
Results of the dose-volume reduction analyses are summarized in Table 1 (with the corresponding dose-volume histograms displayed in figures 8 and 9) for our two patients and for the compared treatment methods. We find that for the single orbit, treated to 25.2 Gy, both techniques predict less than about 2% complication probability. For orbits treated to 36 Gy, the SRT technique keeps complication probability to 3% or less for both patients, while the lateral technique is predicted to yield between 87% and 95% chance of bone growth arrest. For the single orbit treated to 39.6 Gy, use of SRT results in an NTCP prediction of 6.6% and the lateral field method almost 99%. Presumably, other 3D treatment techniques offer intermediate levels of improvement over the simple lateral technique.
| Table 1. Comparison of calculated normal tissue complication probabilities for stereotactic versus bilateral radiation techniques |
Fig. 8. Dose-volume histogram for SRT and right-left lateral field treatments of patient 1. LOB and ROB indicated the left orbital bone and right orbital bone, respectively. |
||
Fig. 9. Dose-volume histogram for SRT and right-left lateral field treatments of patient 2. LOB and ROB indicated the left orbital bone and right orbital bone, respectively. |
||
IV. CONCLUSIONS
We have developed a modification of the
Gill-Thomas-Cosman relocatable head frame that is compatible with
anesthesia needs and suitable for treatment of infants of about 3
years old or less. We have found patient setup to be simple and
reproducible, with repositioning accuracy of the order of ±1 mm or
better on average. The modified immobilization system uses a
thermoplastic mask that is thinner than that used for the BCH frame
and a simple attachment system adapted from a standard head and neck
device. We believe that intracranial radiation treatment,
particularly of retinoblastoma, is the method of choice for
pediatric cases, where sparing of bone growth centers is extremely
important for the long-term normal development of the
child.
This immobilization system
allows the choice of a variety of treatment approaches including
SRT, as well as 3D conformal therapy and intensity-modulated
radiotherapy. We have, as an example, demonstrated the order of gain
that can be achieved using fractionated stereotactic radiosurgery
versus a simple, parallel-opposed lateral field technique for
bilateral orbit irradiation. Using dose-volume histogram reduction,
we illustrated the kinds of differences to be expected and the level
of dose sparing one might hope to achieve with SRT capability. The
simplicity of this head frame modification extends stereotactic
treatment to pediatric cases that might otherwise have more limited
treatment options.
ACKNOWLEDGMENTS
We would like to thank our chief therapist, Kristin Robinson, and departmental engineer, Roy Erickson, for their contributions to this project.
REFERENCES
- Saran FH, Baumert BG, Khoo VS, et al. Stereotactically guided conformal radiotherapy for progressive low-grade gliomas of childhood. Int J Radiat Oncol Biol Phys. 2002;53:43-51.
- Cormack RA, Kooy HM, Bellerive MR, et al. A stereotactic radiation therapy device for retinoblastoma using a noncircular collimator and intensity filter. Med Phys. 1998;25:1438-1442.
- Deegan WF. Emerging strategies for the treatment of retinoblastoma. Curr Opin Ophthalmol. 2003;14:291-295.
- Dunbar SF, Tarbell NJ, Kooy HM, et al. Stereotactic radiotherapy for pediatric and adult brain tumors: Preliminary report. Int J Radiat Oncol Biol Phys. 1994;30:531-539.
- Kalapurakal JA, Ilahi Z, Kepka AG, et al. Repositioning accuracy with the Laitinen frame for fractionated stereotactic radiation therapy in adult and pediatric brain tumors: Preliminary report. Radiology 2001;218:157-160.
- Miranpuri AS, Tome WA, Paliwal BR, et al. Assessment of patient-independent intrinsic error for a noninvasive frame for fractionated stereotactic radiotherapy. Int J Cancer 2001;20:320-325.
- Gill SS, Thomas DGT, Warrington AP, Brada M. Relocatable frame for stereotactic external beam radiotherapy. Int J Radiat Oncol Biol Phys. 1991;20:599-603.
- Kooy HM, Dunbar SF, Tarbell NJ, et al. Adaptation and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. Int J Radiat Oncol Biol Phys. 1994;30:685-691.
- Burton KE, Thomas SJ, Whitney D, et al. Accuracy of a relocatable stereotactic radiotherapy head frame evaluated by use of a depth helmet. Clin Oncol. 2002;14:31-39.
- Kutcher J, Burman C, Brewster L, et al. Histogram reduction method for calculating complication probabilities for three-dimensional treatment planning evaluations. Int J Radiat Oncol Biol Phys. 1991;21:137-146.
- Lyman JT. Complication probability as assessed from dose-volume histograms. Radiat Res. 1985;104(S):13-19.
- Burman C, Kutcher GJ, Emami B, Goitein M. Fitting of normal tissue tolerance date to an analytic function. Int J Radiat Oncol Biol Phys. 1991;21:123-135.
- Eifel PJ, Donaldson SS, Thomas PR. Response of growing bone to irradiation: A proposed late effects scoring system. Int J Radiat Oncol Biol Phys. 1995;31:1301-1307.
© 2006 Am. Coll. Med. Phys.