Experimental verification of the
application of lateral buildup ratio
on the 4-MeV electron beam
James C.L. Chow1,2,a and Scott Newman1
Medical Physics Department,1 Grand River Regional Cancer Center, Grand River Hospital, P.O. Box 9056, 835 King Street West, Kitchener, Ontario N2G 1G3; Department of Physics,2 University of Waterloo, 200 University Avenue, Waterloo, Ontario N2L 3G1 Canada
james.chow@rmp.uhn.on.caReceived 17 June 2005; accepted 12 December 2005
The lateral buildup ratio (LBR) used to estimate the depth dose distribution of electron beams for an irregular cutout field was obtained for a 4-MeV energy beam from a Varian 21 EX linear accelerator. The depth-dose curves for a group of circular cutout fields starting from a 2-cm diameter were measured. Electron diodes were used in a large water tank to measure the LBR values for 6, 9, 12, and 16 MeV electron beam energies and a 10 × 10 cm2 applicator. The results agreed with the published data. When the same equipment, setup, and technique were used to determine the LBR values for the 4-MeV energy beam, the values were only reasonable, being within the clinical treatment range (i.e., LBR <1) for the smallest 6 × 6 cm2 applicator. The calculated LBR values were clinically unacceptable for the circular cutout fields with a diameter larger than 2 cm with the 10 × 10 cm2 applicator. The difficulty in the LBR measurement may be due to the significant contribution of scattered electrons from the beam defining system. This study also focused on how well the sigma values for the 4-MeV beam can predict depth-dose curves for other field sizes and whether the values are applicator-dependent.
PACS numbers : 87.53.Fs; 87.53.Hv; 87.66.Jj
Key words: electron therapy, dosimetry, lateral buildup ratio
a Permanent address: Radiation Medicine Program, Princess Margaret Hospital, 610 University Avenue, Toronto, Ontario M5G 2M9 Canada
I. INTRODUCTION
It is well known that the elementary pencil beam
algorithms(1-5) have been used to
calculate the dose distribution for electron beam treatment
planning. These algorithms required measured beam data of depth
doses and beam profiles which varied with the collimation systems of
different accelerators. For that, the dose per monitor unit of
irregularly shaped cutout fields has individually been measured in
clinical practice. These patient-specific measurements were
time-consuming, so different models based on the pencil beam
approach were proposed to calculate the output factor in irregularly
shaped electron fields.(6-9) Khan et
al.(10-12) introduced a semi-empirical
model based on the lateral spread of the pencil beam and proposed a
function called the lateral buildup ratio (LBR). It is defined as
the ratio of dose at a point of depth for a given circular field to
the dose at the same point for a reference broad field, which is
large enough to provide the lateral scatter equilibrium for the
incident fluence and profile. The LBR represents the fractional
change in dose at a field point due to the loss of scatter relative
to the large reference field and can be calculated from the
percentage depth dose (PDD) normalized near to the surface. The
ratio is related to the lateral pencil beam spread function,
σ(z, E), where z is the depth, and E is
the energy of a circular electron beam of radius r.(11,12) By using the σ r(z, E) values derived from the
measured LBRs for a small circular reference field (usually with a
2cm diameter), sector-type integration can be used to calculate the
average LBR at any depth of an irregular field.(10,12-14)
Khan
et al. have demonstrated the convenience and accuracy of using the
LBR to calculate the electron beam output factors for 6, 9, 12, 16,
and 20 MeV electron beams,(10-12) and
Higgins et al.(15) have further
evaluated the application of this model to calculate dose outside
the field edge and in a heterogeneous media. Although the above five
electron beam energies are more frequently used by most centers for
clinical treatment, 4 MeV is an available option.(16) This energy, with its relatively
shorter practical range and depth of maximum dose (d m), is particularly useful for treating
superficial lesions when the superficial X-ray treatment unit is not
available.(17-20) However, 4-MeV beam
data such as depth-dose and beam profiles are more unstable, due to
difficult beam tuning, and are difficult to measure accurately
compared to higher energies.(21-23) The
aim of this study is to investigate whether the LBR approach is
beneficial to the 4-MeV electron beam through measurements.
II. MATERIALS AND METHODS
A. Beam profiles and PDD
The beam profiles and PDD curves of the electron beams
were measured using a scanning water tank system (RFA 300,
Scanditronix Medical AB with Omni Pro 6 software). A waterproof
high-doped p-type silicon diode (Scanditronix Medical AB, EFD-3G)
was used to measure both the beam profiles and PDD at the central
beam axis. The thickness of the silicon chip was 0.5 mm, and the
diameter of the active area was 2 mm. Since some of the circular
cutout fields were too small to put a reference detector on the beam
path, the reference dose signal for the measurements was obtained
from the internal monitoring ionization chamber within the gantry
head. A Varian 21 EX linear accelerator with 4, 6, 9, 12, and 16 MeV
clinical electron beams was used in the measurement. The 4-MeV
electron beam uses a scattering foil designed for the specific
energy. The central axis for the PDD curve was located according to
the peak position of the profile for the measurement. This was
particularly important when performing the PDD measurement for a
very small circular cutout field (close to 2 cm in diameter), where
the beam penumbras were relatively large for such a low energy due
to electronic disequilibrium.
The
diode was positioned vertically, perpendicular to the water surface.
This setting made the sampling resolution dependent on the thickness
of the diode sensitive volume (~0.5 mm). The position of the
sensitive region from the detector front surface was provided by the
manufacturer and verified in this study, and was considered the
effective point of measurement. A depth ionization curve was scanned
first to determine the position of d m before the beam profile scanning. The
sensitive volume of the diode was then positioned there, and the
beam profiles along the in- and cross-plane directions were scanned.
Both the 6 × 6 cm2 and 10 × 10 cm2 applicators were used in the measurement.
For the 10 × 10 cm2 applicator,
circular cutouts with diameters of 2, 3, 4, 6, 8, and 10 cm and a
square cutout of 10 × 10 cm2 were made.
For the 6 × 6 cm2 applicator, circular
cutouts with diameters 2, 2.5, 3, 4, 5, and 6 cm and a square cutout
of 6 × 6 cm2 were made. The thickness
of cutout in this study was 15 ± 1 mm, and the field edges were
sharp and not divergent. The beam profiles for all cutouts of the 10
× 10 cm2 applicator with 4, 6, 9, 12,
and 16 MeV were measured. For the cutouts of the 6 × 6 cm2 applicator, only the 4-MeV beam profiles
were measured.
The position of the surface was determined by noting the dose variation in the diode reading at the water-air interface. Percentage depth ionization (PDI) curves were measured with the highest sampling resolution and the slowest speed. Central beam axis PDI curves for the different cutouts of the 10 × 10 cm2 applicator as mentioned above were measured for all five energies. For the cutouts of the 6 × 6 cm2 applicator, only the 4-MeV PDI curves were measured. All measurements were taken using a source-to-surface distance (SSD) of 100 cm with an air gap of 5 cm. These measurements were carefully repeated one by one within the same day. It was found that the repeated scan agreed with the original results within ±0.5%. The SSD and zero water level were checked frequently in order to prevent any physical effects, such as evaporation, from introducing measurement error. In addition, the actual cutout dimensions used for the measurements were checked to ensure that the center of the circular cutout was positioned at the central beam axis within ±2 mm. The radiation characteristics of the diode were verified with the ionization chamber to confirm that the depth ionization curve obtained by the diode could be used as the depth-dose curve without correction.
B. Formalism and LBR calculation
The LBR is defined as(10)
(1) |
where D is the dose, r is the radius
of the field defined at the water surface (in our study, it was SSD
= 100 cm), z is the depth, and E is the incident
electron beam energy. Φ is the incident fluence, and r∞ is the broad field radius, that is, the
radius of field large enough to provide lateral scatter equilibrium.
In this study, the variation of the incident fluence factor in Eq.
(1) was factored out by normalizing the depth-dose data of the
circular field to the dose near to the surface. The broad fields for
the 6 × 6 cm2 and 10 × 10 cm2 applicator were selected to be the open
square cutout fields of 6 × 6 cm2 and
10 × 10 cm2, respectively.
Since LBR can be written as
(2) |
and therefore
(3) |
provided that the LBR value is smaller than one, it is also possible to determine the σ r(z) values with respect to the depth, z, or normalized depth, z/R p, for each electron beam energy.
III. RESULTS
Figure 1 is a plot of σ r as a function of the normalized depth for the 4, 6, 9, 12, and 16 MeV energies. The values were calculated using the LBR data for the 2-cm diameter circular cutout as suggested by Khan et al.(10) According to the ICRU Report 35 (ICRU 1984),(24) measuring the dose at a depth of 0.5 mm instead of at the surface was suggested as a normalization to ensure proper detector positioning and to avoid measurements in the unstable buildup region near the surface. Figures 2(a) and 2(b) show the 4-MeV PDD curves using the circular cutouts with 2, 2.5, 3, 4, 5, and 6 cm diameters for the 6 × 6 cm2 and 10 × 10 cm2 applicators, respectively. The PDD curves were normalized to their doses at 0.5 mm from the water surface. The LBR curves against the normalized depth, z/R p, calculated from Figs. 2(a) and 2(b), are shown in Figs. 3(a) and 3(b), respectively. All LBR values larger than one were eliminated in Fig. 3.
Fig. 1. σr plotted against the normalized depth, z/Rp, for the 4 to 16 MeV electron beam. The values were calculated from the LBR data for the 2-cm diameter field for the 10 × 10 cm2 applicator. |
||
Fig. 2. 4 MeV PDD curves of circular cutouts with diameters 2, 2.5, 3, 4, 5, and 6 cm for the (a) 6 × 6 cm2 applicator and the (b) 10 × 10 cm2 applicator. All curves are normalized to the dose near the water surface (0.5 mm depth of water). |
||
Fig. 3. LBR curves calculated from (a) Fig. 2(a) and (b) Fig. 2(b), respectively. The curves are plotted against the depth normalized to the practical range, Rp, of the 4-MeV beam. |
||
IV. DISCUSSION
In Fig. 1, the σ r values were calculated using the 2-cm
diameter cutout for the 10 × 10 cm2
applicator. These values were verified and agree with the
similar published results(10) except
for the 4-MeV energy. It was found that, as previously reported,(12) the PDD calculated using our σ r values can reproduce the PDD for other
field sizes with energies from 6 to 16 MeV within ±1% error bar,
while for the 4-MeV, the error in reproducing the PDD is ±4%.
Moreover, the σ r versus
z/R p and LBR versus
z/R p are independent of
the cutout shape as reported.(10) There
is no problem in the application of LBR in the above energy range.
In Fig. 2(a), showing the PDD
curves of the 6 × 6 cm2 applicator, it
was found that beyond 1.5 cm, the depth doses of the circular
cutouts with diameters ≥5 cm are very close or even very slightly
larger (about 3% on average) than that of the broad field. Such
uncertainty is larger than the reproducibility error bar of the
measurement (±0.5%). In Fig. 2(b), there was a more significant
increase in the depth doses of cutouts (> 5 cm) than in the broad
field, in the bremsstrahlung tail range for the 10 × 10
cm2 applicator. In the two figures,
when the electronic equilibrium condition had been reached for those
sufficiently large circular fields (i.e., > 5 cm diameter),
measurement uncertainty of the depth dose near to the
bremsstrahlung tail became significant in the LBR
calculation. Or it is understood that those large circular fields
can also be recognized as broad fields. For the 4-MeV energy,
neglecting the consideration for the physical effects from the
measurement, the PDD curve is very difficult to predict through
measurements, especially for depths larger than 2 cm because the
scattered electrons from the applicator and cutout contribute
significantly to the doses, due to its relatively large electron
angular scattering cross section. Such dose contribution from the
scattered electrons is difficult to predict because it depends on
the material and geometry of the beam defining system. The scattered
electrons from the applicator have a different energy spectrum and
angular scattering cross section than the primary incident
electrons. The difference of field size dependence between the 6 × 6
cm2 and 10 × 10 cm2 applicator is also probably due to the
different beam defining system used, although both applicators have
the same photon jaws setting (20 × 20 cm2). For highly accurate results, Monte Carlo
investigations would be useful but are beyond the scope of this
study.
Figure 3 shows the LBR
values calculated from Fig. 2. In Fig. 3(a), the LBR values are seen
to be nonsense beyond 15 mm or normalized depth (z/R
p) of 0.88. However, the values are
reasonable if the focus is restricted to the clinical treatment
range between 7 mm (d m) and 14 mm
(i.e., 100 to 50% isodose contour), or normalized depth between 0.41
and 0.82, and excludes the LBR calculated from the large circular
cutouts equivalent to the broad field. The discontinuities near R
p in Fig. 3(a) are due to the
measured depth-dose uncertainty as explained in Fig. 2(a). However,
in Fig. 3(b) for the 10 × 10 cm2
applicator, it is seen that only the LBR values of circular cutouts
with 2 cm and 3 cm diameter are reasonable. It seems that the LBR
was difficult to measure and calculate for the 4-MeV electron
energy, using an applicator larger than 6 × 6 cm2 within the clinical treatment range. In
the discontinuity range of the figures, the sigma values calculated
using the PDD data of the 2-cm cutout could not accurately predict
the PDD data for other field sizes. This is different for higher
energies from 6 to 16 MeV, where the PDD calculated using sigma
values for a 2-cm diameter cutout could reproduce the PDD for other
field sizes as reported.(10) Since the
PDD data for higher energies is applicator-independent, according to
our measurement, the sigma values and LBR versus z/R
p are also applicator-independent;
this has been verified for the 6-MeV electron beam. However, this is
not the case for the low-energy 4-MeV beam, as shown in Figs. 2 and
3. Different LBR versus z/R p should be plotted for different applicators.
V. CONCLUSION
The LBR values of the 4-MeV electron energy using the Varian 21 EX accelerator were measured and calculated. It was found that, for a small applicator size of 6 × 6 cm2, the calculated LBR values were reasonable and within the clinical treatment range. However, when a larger 10 × 10 cm2 applicator was used, the LBR values with circular fields larger than 2 cm in diameter were not reasonable compared to those of higher energies. This is because the PDD for a range of cutouts was higher than the 10 × 10 cm2 applicator broad beam PDD at certain depth ranges, and the LBR model breaks down under these conditions. It was found that the depth-dose uncertainty increased when an insert and applicator larger than 6 × 6 cm2 was used. Another reason is that the lateral scatter/electronic equilibrium is more easily reached by increasing the circular cutout size from 2 cm diameter compared to higher energies. Since, in the output (dose per monitor unit), calculation of an irregular field requires sector-type integration of LBR with a variation of radii data, the uncertain LBR values in the larger circular fields cause significant error in the output estimation of the 4-MeV electron beam compared to those at higher energies. Based on the measured results in this study, it can be concluded that the LBR could not be used to model the 4-MeV electron beam well.
REFERENCES
- Perry DJ, Holt JG. A model for calculating the effects of small inhomogeneities on electron beam dose distribution. Med Phys. 1980;7:207-215.
- Brahme A, Lax I, Andreo P. Electron beam dose planning using discrete Gaussian beams, mathematical background. Acta Radiol Oncol. 1981;20:147-158.
- Hogstrom KR, Mills MD, Almond PR. Electron beam dose calculations. Phys Med Biol. 1981;26:445-459.
- Werner BL, Khan FM, Deibel FC. A model for calculating electron beam scattering in treatment planning. Med Phys. 1982;9:180-187.
- Jette D. The application of multiple scattering theory to therapeutic electron dosimetry. Med Phys. 1983;10:141-146.
- Dutreix A, Briot E. The development of a pencil-beam algorithm for clinical use at the Institut Gustave Roussy. In: Nahum AE, editor. The computation of dose distributions in electron beams, radiotherapy. Sweden: University of Umea; 1990): 242-270.
- Mills MD, Hogstrom KR, Almond PR. Prediction of electron beam output factors. Med Phys. 1982;9:60-68.
- Bruinvis IAD, Amstel AV, Elevelt AJ, Van der Laarse R. Dose calculation for arbitrarily shaped electron fields. Acta Radiol Suppl. 1983;364:73-79.
- McParland BJ. A method of calculating the output factors of arbitrarily shaped electron fields. Med Phys. 1989;16:88-93.
- Khan FM, Higgins PD, Gerbi BJ, Deibel FC, Sethi A, Mihailidis DN. Calculation of depth dose and dose per monitor unit for irregularly shaped electron field. Phys Med Biol. 1998;43:2741-2754.
- Khan FM, Higgins PD. Calculation of depth dose and dose per monitor unit for irregularly shaped electron fields: An addendum. Phys Med Biol. 1999;44:N77-N80.
- Khan FM, Higgins PD. Field equivalence for clinical electron beams. Phys Med Biol. 2001;46:N9-N14.
- Jones D, Andre P, Washington JT, Hafermann HD. A method for the assessment of the output of irregularly shaped electron fields. Br J Radiol. 1990;63:59-64.
- Wu RK, Wang W, El-Mahdi AM. Irregular field output factors for electron beams. In: Bruvinvis IAD, van der Giessen PH, van Kleffens HJ, and Wittkamper FW, editors. Proceedings of the 9th international conference on the use of computers in radiation therapy. Sweden: Scheveningen; 1987: 453-456.
- Higgins PD, Gerbi PD, Khan FM. Application of measured pencil beam parameters for electron beam model evaluation. Med Phys. 2003;30:514-520.
- Followill DS, Davis DS, Ibbott GS. Comparison of electron beam characteristic from multiple accelerators. Int J Radiat Oncol Biol Phys. 2004;59:905-910.
- Walker C, Wadd NJ, Lucraft HH. Novel solutions to the problems encountered in electron irradiation to the surface of the head. Br J Radiol. 1999;72:787-791.
- Scholten AN, Griep C, Davelaar J, Chin A, Leer JW. Electron beam irradiation is effective in the treatment of skin carcinomas; a comparison with superficial roentgen therapy. Ned Tijdschr Geneeskd. 1996;148:428-431.
- Amdur RJ, Kalbaugh KJ, Ewald LM, et al. Radiation therapy for skin cancer near the eye: Kilovoltage x-rays versus electrons. Int J Radiat Oncol Biol Phys. 1992;23:769-779.
- Das IJ, Kase KR, Copeland JF, Fitzgerald TJ. Electron beam modifications for the treatment of superficial malignancies. Int J Radiat Oncol Biol Phys. 1991;21:1627-1634.
- van Battum LJ, Huizenga H. On the initial angular variances of clinical electron beams. Phys Med Biol. 1999;44:2803-2820.
- McParland BJ, Cunningham JR, Woo MK. The optimization of pencil beam widths for use in an electron pencil beam algorithm. Med Phys. 1988;15:489-497.
- Li XA, Rogers DWO. Electron mass scattering powers: Monte Carlo and analytical calculations. Med Phys. 1995;22:531-541.
- Radiation dosimetry: Electron beams with energy between 1 and 50 MeV. ICRU Report 35 1984; Bethesda, MD: ICRU.
© 2005 Am. Coll. Med. Phys.