Target localization for post-
prostatectomy patients using CT
and ultrasound image guidance
Kamen Paskalev, Steven Feigenberg, Rojymon Jacob, ShawnMcNeeley, Eric Horwitz, Robert Price, Charlie Ma, and Alan Pollack
Department of Radiation Oncology, Fox Chase Cancer Center, 333 Cottman Avenue,Philadelphia, Pennsylvania 19111
U.S.A.K_Paskalev@fccc.eduReceived 4 May 2005; accepted 17 August 2005
We conducted a study comparing B-mode acquisition and targeting (BAT) ultra-sound alignments based on CT data in the postoperative setting. CT scans were obtained with a Primatom CT-on-rails on nine patients. Two CT scans were obtained each week, while setup error was minimized by BAT ultrasounds. For the first three patients, a direct comparison was performed. For the next six patients, a template based on the shifts from the week 1 CT during treatment was used for subsequent setup. Comparison of isocenter shifts between the BAT ultrasound and CT was made by the difference, absolute difference, and improvement (using CT alignments as the reference technique). A total of 90 image comparisons were made. The average interfraction motion was 3.2 mm in the lateral, 3.0 mm in the longitudinal, and 5.1 mm in the AP direction. The results suggest that the CT-based ultrasound templates can improve the localization of the prostate bed when the initial displacements are greater than 4 mm. For initial displacements smaller than 4 mm, the technique neither improved nor worsened target localization. However, ultrasound alignments performed without the use of a template deteriorated patient positioning for two out of three patients, demonstrating that the use of a CT template was beneficial even at small initial displacements.
PACS numbers: 87.53.-j, 87.53.Kn, 87.53.Xd
Key words: post-prostatectomy, localization, organ motion, CT, ultrasound
I. INTRODUCTION
Improvements in radiotherapy treatment techniques
and delivery have led to improvements in treatment outcomes. Perhaps
the best example of this is in the management of prostate cancer
where the development of 3D conformal radiotherapy has improved
freedom from biochemical failure due to the ability to safely
escalate radiation dose.(1) As doses
are increased, the precision of dose delivery becomes crucial, so
the treatment must be performed with great accuracy. However, daily
uncertainties regarding patient setup reproducibility (i.e., daily
setup error and internal organ motion) diminish the ability to
achieve this goal.(2) In the past, the
frequently changing nature of these variations has made corrections
difficult and dictated that a wider margin of normal tissue be
included within the radiation field to ensure tumor coverage.
Various approaches have been used
to reduce these uncertainties including daily imaging prior to
treatment using abdominal ultrasound(3-9) and CT scans,(10-12) immobilization devices(13) (i.e., rectal balloon), and placement
of fiducial markers(7,8,14-18) in the
prostate gland. Unfortunately, the majority of the data with these
devices are in the definitive setting where the target or prostate
is well visualized. In the postoperative setting, the target is not
easily visualized, which potentially reduces the effectiveness of
these devices.
Currently at our
institution, daily uncertainties are reduced with the use of the
B-mode acquisition and targeting (BAT) ultrasound system for
definitive and postoperative patients. In October 2002, a CT was
placed in the LINAC vault. The CT-on-rails system has been used,
like the BAT ultrasound system, to reduce daily uncertainties.(12) This report describes our initial
experience in implementing this device in the postoperative setting.
II. METHODS
In June 2003, a Primatom sliding CT gantry (Siemens
Medical Solutions, Concord, CA) was first used for localization of a
post-prostatectomy patient at our institution. Since then, it became
the department's policy to obtain CT scans twice per week to verify
localization of post-prostatectomy treatment volume for patients
currently undergoing daily localization using BAT ultrasound system
(Nomos Corporation, Cranberry Township, PA). This study describes
our initial experience based on the first nine patients evaluated.
All patients were treated in the supine position using an alpha
cradle cast as previously described.(19) Target volumes and critical normal
structures were contoured using the combined fused images from the
CT and MR simulations as previously described.(20) CT and MR simulations were obtained
using axial images obtained every 3 mm through the pelvis. The
target volume was based on presurgery CT scans of the pelvis and the
pathologic specimen in order to cover the prostatic fossa and
periprostatic tissues, specifically including the bladder neck,
anastomosis, seminal vesicles, and surgical clips. Some of the
regional lymph nodes were included if the initial surgical lymph
node dissection was not done or was considered inadequate. All
patients were treated with intensity-modulated radiotherapy (IMRT)
to a total dose of 64 or 66 Gy to 70 Gy for either adjuvant or
salvage, respectively. IMRT was planned using the Corvus
treatment-planning system (Nomos Corporation, Cranberry Township,
PA).
To reduce organ motion and
setup error, the BAT ultrasound system is used at our institution.
The BAT ultrasound system calculates the daily shifts based on an
alignment of cross sections of the volumes drawn on the simulation
CT scan with ultrasound images of the target region in the treatment
position. The procedure used to acquire a daily CT scan and to
perform ultrasound-based alignment is described as follows:
- The patient is immobilized in a treatment position, and the skin marks from simulation are aligned with the room lasers. Radio-opaque fiducial markers are placed on the skin. Markers coincide with the lasers and are used to define the daily isocenter in the localization CT scan.
- The treatment table is rotated 180° (scanning position), and a CT scan is obtained with the moving gantry.
- The table is rotated back to the treatment position.
- Alignment of the fiducial markers with the room lasers is verified. If the fiducial markers have moved more than 1 mm during the scan, the displacement is recorded and is later taken into account when calculating the daily shifts. If the displace ment is greater than 3 mm or one of the fiducials is not visible, the scan is discarded from the study.
- BAT ultrasound alignment is performed, and the treatment table is shifted in the x, y, z direction prior to treatment.
The pretreatment CT (i.e., localization CT) scans were fused with the simulation CT scan in order to calculate the daily CT shifts. The shifts were determined using the Coherence Dosimetrist workstation (Siemens Medical Solutions, Concord, CA). The main fusion landmarks were the posterior bladder wall and the anterior rectal wall. After the fusion, the daily CT shift was measured as the distance between the fiducial markers, which are visible on both simulation and localization scans (Fig. 1).
Fig. 1. CT shift calculation. After the targets on the simulation and the localization CT scans have been aligned, the cross hair is positioned at the isocenter of the simulation scan (left). Then the distances between the cross hair and the fiducial markers in the localization scan give the CT shifts. |
||
The CT shifts were
assumed to be more accurate than the BAT ultrasound shifts and were
used as the reference shifts. The uncertainty of our image fusions
for the two CT scans was estimated to be 2 mm based on the 3-mm
slice thickness and associated pixel size. The uncertainty in the
position of the fiducial markers was 1 mm. Therefore, the total CT
shift uncertainty was estimated to be 2.5 mm. This was verified by
testing both the intrauser and interuser variability of the CT
shifts, which was within 2 mm (data not shown).
The first three patients underwent
treatment with BAT ultrasound localization alone. The CT shifts were
calculated without any change in the ultrasound alignment. The
variability appeared higher than expected (higher than 10 mm on some
treatment days). This was thought to be related to difficulties in
aligning the postoperative anatomy, which is more difficult to
visualize in comparison to the definitive setting. To rectify this
problem, special templates for BAT alignments were created during
the first week of treatment for every patient, based on CT shifts
calculated in real time (with the patient in the treatment
position). The shifts from the CT were programmed into the BAT
localization system with the 3D representation of a template (Fig.
2). The goal of using a template was to create a BAT alignment that
corresponds to the prostate bed position the same as the position
during CT simulation. Once a template was created, it was used at
every subsequent BAT alignment for the given patient. The template
was expected to help the therapists find the relative location of
the treatment-planning structures with respect to the patient
anatomy in the ultrasound image.
Fig. 2. BAT template alignment. The BAT system is forced to align the cross sections of the volumes (rectum, bladder, and prostate bed) to the ultrasound image according to the CT shifts on the particular day. |
||
All shifts were
recorded as isocenter shifts (opposite to table shifts). The signs
of the shifts were chosen in the following way: (1) positive shifts
were in posterior, left, and inferior directions; (2) negative
shifts were in the anterior, right, and superior directions.
In our statistical analysis the
following parameters were used to quantify our results: (1) the
absolute CT shift, (2) the difference between the BAT and the CT
shift, (3) the absolute difference between the BAT and CT shift, and
(4) a parameter that we defined as "improvement." Each of these
quantities was averaged over a number of observations for each
patient. Each parameter was calculated separately in the AP,
longitudinal, and lateral directions. The days on which the
templates were created were excluded from the statistical analysis.
Statistical significance was calculated using a correlated
two-tailed t-test.
The
"absolute CT shift" is defined as the magnitude of the actual shifts
for a particular patient. The absolute CT shift is equal to the
initial displacement of the target from its ideal position (i.e.,
interfraction target motion and setup error).
The "difference between the BAT
and the CT shift" (difference = BAT - CT) represents the systematic
error between the CT and the BAT shifts. The systematic errors
represent tendencies in BAT alignment. This information is
meaningful and is used to improve the BAT localization (i.e.,
residual shift after BAT alignment). The residual shift is assumed
to be a normal distribution; therefore, its average value for a
particular patient along with the standard deviation can be used to
estimate how often the target is within a certain margin.
The "absolute difference between
the CT and BAT shift" (abs. difference = |BAT - CT|) is a
characterization of the discrepancy between the CT and BAT shift.
Averaged over the number of observations, this quantity represents
both the systematic and random errors (i.e., the absolute residual
shift after BAT alignment).
"Improvement" illustrates the role
of BAT localization assuming that the CT shifts are ideal and is
defined by the following equation:
(1) |
The CT shifts also demonstrate setup error and target motion if no daily localization was performed (in Fig. 3, parameter a). The difference between the CT shift (i.e., ideal position) and the BAT shift represents the discrepancy between BAT localization and the ideal position (in Fig. 3, parameter b). Improvement is defined as the difference between a and b (Fig. 3), which represents the effect of BAT localization (beneficial or detrimental). When the difference between a and b is greater than zero, BAT localization improves the target positioning. A difference of less than zero demonstrates the BAT localization system is detrimental and worsens localization.
Fig. 3. Definition of the parameter “improvement”: (A) when the BAT shift is in the correct direction but is smaller than the CT shift, the improvement is equal to the BAT shift (see |a| – |b| in A); (B) when the BAT shift is greater than the CT shift, the improvement is still defined as |a| – |b|, but it is no longer equal to the BAT shift. |
||
III. RESULTS
A total of 90 image comparisons (CT versus ultrasound) were recorded for nine patients. Comparisons for each patient are presented separately in the AP, lateral, and longitudinal directions (Table 1).
| Table 1. Comparison between CT and BAT localization |
| *Standard Deviation |
The average absolute
CT shift of the target (i.e., interfraction motion) was relatively
small in the lateral and longitudinal directions. The CT shifts
averaged over all patients were 3.2 mm (range of patient averages
1.6 mm to 6.0 mm) in the lateral direction and 3.0 mm (range of
patient averages 1.9 mm to 5.2 mm) in the longitudinal direction. In
the AP direction, the CT shifts were more substantial, with an
average displacement of 5.1 mm (range of patient averages 2.1 mm to
9.4 mm). The largest single target displacement in the AP direction
was 16.9 mm. A careful examination of the localization CT scans
showed that the daily variation in the filling of both the bladder
and the rectum was significant, resulting in shift of the prostate
bed in the AP direction with minor rotation causing lateral and
longitudinal shifts.
The residual
shift was reduced when a template was implemented (after the first
three patients). When using a template (patients 4 to 9), the
residual shifts ranged from 0.2 mm to 4.6 mm (AP), 0.7 mm to 2.9 mm
(lateral), and 0.2 mm to 2.6 mm (longitudinal), while for the
nontemplate patients (patients 1 to 3), the residual shift was
higher [(3.3 mm to 3.9 mm (AP), 1.2 mm to 3.9 mm (lateral), and 2.7
mm to 3.1 mm (longitudinal)]. The template minimized random errors
in BAT alignments by reducing the standard deviations of the
residual shifts over the course of treatment (especially in the AP
direction: nontemplate from 4.3 mm to 6 mm; template from 1.7 mm to
3.5 mm). The template also reduced the systematic error (average
residual sifts) for 4 of 6 template patients. For patients 4 and 5
the rectal filling on the day when the templates were created was
different than the rectal filling during the rest of the treatment,
resulting in high systematic errors in the AP direction (4.3 mm and
4.6 mm). This suggests that for some patients a second template may
be needed during the course of treatment.
The template reduced the total
errors (including systematic and random errors) by reducing the
average absolute residual shifts (template: AP 1.3 mm to 5.0 mm,
lateral 1.7 mm to 3.2 mm, and longitudinal 1.8 mm to 3.3 mm;
nontemplate: AP 4.1 mm to 5.8 mm, lateral 2.8 mm to 4.8 mm, and
longitudinal 3.6 mm to 4.1 mm).
Figure 4 illustrates the benefit
of having a BAT ultrasound template. BAT alignments for patient 7
were based on a template, while no template was used for patient 2.
The residual shifts for patient 2 were much higher, exceeding 10 mm
in several instances. For patient 7, all residual shifts were within
5 mm, and most of them were within 3 mm, which is similar to the
uncertainty of the CT shifts. Patient 5 had significant day-to-day
variations in rectal filling resulting not only in an anterior
displacement of the prostate bed, but also in change of its shape
(or rotation of the prostate bed). In such cases it is impossible to
perfectly align the planning target to the daily target using linear
shifts only. A good alignment in the inferior region (bladder neck)
of the target will ultimately lead to misalignment in the superior
region and vice versa. The alignment then becomes a matter of
clinical decision made by the treating therapist on a daily basis,
which is guided by the physician. This issue was the main reason for
the results seen for patient 5 compared to the other template
patients.
Fig. 4. Daily residual target shift after BAT localization. One patient from each group (with and without using a template) is presented. |
||
Target motion in the
AP direction is a serious obstacle when treating post-prostatectomy
patients. A margin of 8 mm in the AP direction is used at our
institution. Based on the residual shifts for nontemplate patients,
the target was within the margin in less than 83% of the cases
(patient 1: 82.7%; patient 2: 75.3%; patient 3: 77.6%), while other
template patients were all covered in greater than 83% (range 83.4%
to 99%). The low percentage values for patients 4 (87.6%) and 5
(83.4%) were mostly due to the high systematic component of their
residual shifts. However, the standard deviations (random component)
of the residual shifts for these patients were considerably lower
than the standard deviations for the nontemplate patients, meaning
that further improvement was possible, as mentioned above.
Improvement is limited by the
initial displacement of the target. For example, if the target
happens to be in a perfect position before the daily BAT alignment,
further improvement is impossible. The average improvements in each
direction for all nine patients are presented as a function of the
initial target displacement in Fig. 5. The figure illustrates that
in both the longitudinal and in the lateral directions, the
improvements are small when the initial target displacements are
small (less than 4 mm). In these directions, improvements are either
positive or negative for the template patients, and mostly negative
(meaning deterioration of the positioning) for the nontemplate
patients. These values are insignificant when the initial
displacement is small. However, in the AP direction a statistically
significant deterioration in positioning was demonstrated for two of
the three patients treated without the aid of a template even at
small initial displacements of the target (average improvement of
-3.0 mm with p < 0.002).
Fig. 5 Average improvements for all nine patients. Some p values are also presented. |
||
In all cases where the initial target displacement was greater than 5 mm, the improvements were positive for the template patients. Statistically significant improvements increase as the target displacement increases. The CT-based template is most crucial for the BAT alignment in the AP directions. The maximum average improvement of a template patient is also in the AP direction and has a value of 6.3 mm (p = 5 × 10-6).
IV. DISCUSSION
Radiation therapy with or without androgen
deprivation following a radical prostatectomy is considered for
patients with pathologic T3 disease, positive surgical margins, or a
slowly rising PSA profile. Contrary to the treatment of prostate
cancer in the definitive setting, the target volume in the
postoperative setting is not well-defined. The most common sites of
clinical failure are the peri-anastomotic site, bladder neck,
retrovesical space, and, rarely, the regional lymph nodes.(21) Similar to patients treated in the
definitive setting, precision radiotherapy requires reducing daily
uncertainties in order to ensure target coverage while minimizing
radiation exposure to critical surrounding structures.
In this study, the interfraction
motion was measured using the average CT shifts over the course of
treatment for each patient (called initial target displacement). As
mentioned above, the uncertainty of the CT shifts was 2.5 mm. The
average interfraction motions in the lateral, longitudinal, and AP
directions were 3.0 mm, 3.2 mm, and 5.1 mm, respectively; these
results are similar to the values in the treatment of definitive
patients.(5,22,23) The CT scans were
then used to evaluate the accuracy of BAT ultrasound localization in
the postoperative setting. With the lack of a clearly defined
target, such as the prostate, systematic errors in BAT ultrasound
alignment were found. ImplnXentation of a template BAT ultrasound
based on CT significantly reduced systematic errors and led to
improved target localization, especially in the AP direction.
Specifically looking at the AP
direction, an 8-mm margin was added to the clinical target volume to
account for daily uncertainties. The target of the nontemplate
patients was fully within this margin for between 75.3% and 82.7% of
the scans, while the aid of the template improved the tumor coverage
to 83.4% to 99%. Actually, all but two of the patients treated with
the aid of the template exhibited greater than 94.5% coverage. The
fact that the target was fully covered in 87.6% of the treatments
for patient 4, and 83.4% of the treatments for patient 5, suggests
that probably a planning tumor volume (PTV) margin of greater than 8
mm is needed. However, positioning errors for these patients were
mostly due to their high average residual shifts, whereas the
standard deviations of the residual shifts were considerably lower
than the standard deviations for the nontemplate patients. The
implication is that the discrepancies between the BAT and CT shifts
are primarily systematic for the template patients, and therefore
these discrepancies may be reduced by creating a second template
during the course of treatment. If the template technique fails to
reduce the systematic errors, then an increased patient specific
margin should be considered.
The
reduction of both systematic and random discrepancies for the
template patients resulted in lower absolute residual shifts after
BAT alignment (Table 1). This applies to all template patients
except patients 4 and 5, which, for example, had average absolute
residual shifts in the AP direction of, respectively, 4.5 mm and 5.0
mm. However, these residual shifts were primarily due to systematic
differences between the BAT and CT shifts, and the systematic
differences could be decreased as previously discussed. In the
future, when more data on template assisted alignments is collected,
further analysis could be performed, in order to correlate the
magnitude of the residual shifts with patient specific parameters,
such as weight, for example. Such a study would help develop
guidelines on choosing a patient specific PTV margin.
Chinnaiyan et al.(24) recently reported a similar series
using an optically guided 3D ultra-sound guided localization. They
reported average absolute ultrasound shifts of 5±4 mm, 3±3 mm, 3±4
mm, over the entire course of treatment for 6 patients treated in
the postoperative setting in the AP, lateral, and CC directions,
respectively. These values are similar to 10 patients treated in the
definitive setting. In their study, the bladder neck was used as the
primary reference to minimize systematic errors. In addition, a
rectal balloon was used on a daily basis for internal
immobilization. In the same study the ultrasound shifts were
compared with shifts based on portal films and digitally
reconstructed radiographs from the time of simulation. However, the
purpose of this comparison was not to evaluate the accuracy of the
ultrasound shifts but only to differentiate the setup errors from
the actual internal target motion. Although steps were taken to
limit these uncertainties with the use of the rectal balloon, better
daily imaging (CT) could be used to evaluate the ultrasound shifts.
The rectal balloon has also been shown to reduce the volume of
rectum receiving a significant radiation dose,(25) but this is counterbalanced by the
discomfort of the patient, the daily reproducibility of the balloon,
and the potential hypoxia caused by the pressure of the balloon,
which could alter the effectiveness of the radiation.(26)
This
study proposes a localization technique based on CT and ultrasound
image guidance that reduces uncertainties of organ motion and setup
error in the postoperative setting. However, clearly there are
serious concerns when considering reduction of treatment margins in
cases where the target volumes are not well-defined. In addition,
there are also concerns whether the target volumes move in
relationship to each other and whether uniform margins are
appropriate. In the definitive setting, similar issues exist but are
typically less significant. The distal seminal vesicles relative to
the prostate move more than the proximal seminal vesicles.
Furthermore, the movement of the lymph nodes relative to the
prostate and seminal vesicles is unknown. In the postoperative
setting, the bladder neck and retrovesical space move relative to
the filling of the bladder and rectum, although the peri-anastomotic
site does not. In addition, there is the issue of the movement of
the pelvic lymph nodes. Currently, we are using serial CT scans in
the definitive and postoperative setting to further evaluate these
issues. Daily imaging to reduce daily uncertainties should be
beneficial, but further studies are necessary to ensure adequate
margins to account for target motion and whether immobilization of
the tumor bed is necessary and/or useful.
V. CONCLUSIONS
The presented results suggest that when the initial displacement of the prostate bed is greater than 4 mm, daily ultrasound localization can be improved with the aid of a CT-based template. When the initial target displacements are smaller than 4 mm, the presented technique is neither beneficial nor detrimental. Patient positioning performed using ultrasound alignments without the aid of a CT-based template may potentially worsen the localization of the prostate bed.
REFERENCES
- Pollack A, Zagars GK, Starkschall G, et al. Prostate cancer radiation dose response: Results of the M. D. Anderson phase III randomized trial. Int J Radiat Oncol Biol Phys. 2002;53:1097-1105.
- Ten Haken RK, Forman JD, Heimburger DK, et al. Treatment planning issues related to prostate movement in response to differential filling of the rectum and bladder. Int J Radiat Oncol Biol Phys. 1991;20:1317-1324.
- Lattanzi J, McNeeley S, Donnelly S, et al. Ultrasound-based stereotactic guidance in prostate cancer-Quantification of organ motion and setup errors in external beam radiation therapy. Comput Aided Surg. 2000;5:289-295.
- Lattanzi J, McNeeley S, Pinover W, et al. A comparison of daily CT localization to a daily ultrasound-based system in prostate cancer. Int J Radiat Oncol Biol Phys. 1999;43:719-725.
- Chandra A, Dong L, Huang E, et al. Experience of ultrasound-based daily prostate localization. Int J Radiat Oncol Biol Phys. 2003;56:436-447.
- Morr J, DiPetrillo T, Tsai JS, et al. Implementation and utility of a daily ultrasound-based localization system with intensity-modulated radiotherapy for prostate cancer. Int J Radiat Oncol Biol Phys. 2002;53:1124-1129.
- Van den Heuvel F, Powell T, Seppi E, et al. Independent verification of ultrasound based image-guided radiation treatment, using electronic portal imaging and implanted gold markers. Med Phys. 2003;30:2878-2887.
- Langen KM, Pouliot J, Anezinos C, et al. Evaluation of ultrasound-based prostate localization for image-guided radiotherapy. Int J Radiat Oncol Biol Phys. 2003;57:635-644.
- Serago CF, Chungbin SJ, Buskirk SJ, et al. Initial experience with ultrasound localization for positioning prostate cancer patients for external beam radiotherapy. Int J Radiat Oncol Biol Phys. 2002;53:1130-1138.
- Lattanzi J, McNeely S, Hanlon A, et al. Daily CT localization for correcting portal errors in the treatment of prostate cancer. Int J Radiat Oncol Biol Phys. 1998;41:1079-1086.
- Hua C, Lovelock DM, Mageras GS, et al. Development of a semi-automatic alignment tool for accelerated localization of the prostate. Int J Radiat Oncol Biol Phys. 2003;55:811-824.
- Paskalev K, Ma CM, Jacob R, et al. Daily target localization for prostate patients based on 3D image correlation. Phys Med Biol. 2004;49:931-939.
- Teh BS, McGary JE, Dong L, et al. The use of rectal balloon during the delivery of intensity modulated radio-therapy (IMRT) for prostate cancer: More than just a prostate gland immobilization device? Cancer J. 2002;8:476-483.
- Crook JM, Raymond Y, Salhani D, et al. Prostate motion during standard radiotherapy as assessed by fiducial markers. Radiother Oncol. 1995;37:35-42.
- Litzenberg D, Dawson LA, Sandler H, et al. Daily prostate targeting using implanted radiopaque markers. Int J Radiat Oncol Biol Phys. 2002;52:699-703.
- Vigneault E, Pouliot J, Laverdiere J, et al. Electronic portal imaging device detection of radioopaque markers for the evaluation of prostate position during megavoltage irradiation: A clinical study. Int J Radiat Oncol Biol Phys. 1997;37:205-212.
- Balter JM, Lam KL, Sandler HM, et al. Automated localization of the prostate at the time of treatment using implanted radiopaque markers: Technical feasibility. Int J Radiat Oncol Biol Phys. 1995;33:1281-1286.
- Pang G, Beachey DJ, O'Brien PF, et al. Imaging of 1.0-mm-diameter radiopaque markers with megavoltage X-rays: An improved online imaging system. Int J Radiat Oncol Biol Phys. 2002;52:532-537.
- Horwitz EM, Hanlon AL, Pinover WH, et al. Defining the optimal radiation dose with three-dimensional conformal radiation therapy for patients with nonmetastatic prostate carcinoma by using recursive partitioning techniques. Cancer 2001;92:1281-1287.
- Mah D, Steckner M, Hanlon A, et al. MRI simulation: Effect of gradient distortions on three-dimensional prostate cancer plans. Int J Radiat Oncol Biol Phys. 2002;53:757-765.
- Connolly JA, Shinohara K, Presti JC, Jr, et al. Local recurrence after radical prostatectomy: Characteristics in size, location, and relationship to prostate-specific antigen and surgical margins. Urology 1996;47:225-231.
- Langen KM, Jones DT. Organ motion and its management. Int J Radiat Oncol Biol Phys. 2001;50:265-278.
- Huang E, Dong L, Chandra A, et al. Intrafraction prostate motion during IMRT for prostate cancer. Int J Radiat Oncol Biol Phys. 2002;53:261-268.
- Chinnaiyan P, Tomee W, Patel R, et al. 3D-ultrasound guided radiation therapy in the post-prostatectomy setting. Technol Cancer Res Treat. 2003;2:455-458.
- Patel RR, Orton N, Tome WA, et al. Rectal dose sparing with a balloon catheter and ultrasound localization in conformal radiation therapy for prostate cancer. Radiother Oncol. 2003;67:285-294.
- Movsas B, Chapman JD, Hanlon AL, et al. Hypoxic prostate/muscle pO2 ratio predicts for biochemical failure in patients with prostate cancer: Preliminary findings. Urology 2002;60:634-639.
© 2005 Am. Coll. Med. Phys.