Effects of field parameters on
IMRT plan quality for
gynecological cancer: A case study
Albert Y.C. Fung, Charles A. Enke, Komanduri M. Ayyangar, Robert B. Thompson, Weining Zhen, Natarajan V. Raman, David Djajaputra, Sicong Li, Ramasamy M. Nehru, Sushakumari Pillai, Paul Sourivong, Mary Headley, and Ann L. Yager
Department of Radiation Oncology, Nebraska Medical Center, 987521 Nebraska Medical Center, Omaha, Nebraska 68198-7521 U.S.A.
afung@unmc.eduReceived 19 November 2004; accepted 22 April 2005
Traditional external beam radiotherapy of gynecological cancer consists of a 3D, four-field-box technique. The radiation treatment area is a large region of normal tissue, with greater inhomogeneity over the treatment volume, which could ben-efit more with intensity-modulated radiation therapy (IMRT). This is a case report of IMRT planning for a patient with endometrial cancer. The planning target vol-ume (PTV) spanned the intrapelvic and periaortic lymph nodes to a 33-cm length. Planning and treatment were accomplished using double isocenters. The IMRT plan was compared with a 3D plan, and the effects of field parameters were stud-ied. Delineated anatomical contours included the intrapelvic nodes (PTV), bone marrow, small bowel, bladder, rectum, sigmoid colon, periaortic nodes (PTV), spinal cord, left kidney, right kidney, large bowel, liver, and tissue (excluding the PTVs). Comparisons were made between IMRT and 3D plans, 23-MV and 6-MV energies, zero and rotated collimator angles, different numbers of segments, and opposite gantry angle configurations. The plans were evaluated based on dose-volume histograms (DVHs). Compared with the 3D plan, the IMRT plan had superior dose conformity and spared the bladder and sigmoid colon embedded in the intrapelvic nodes. The higher energy (23 MV) reduced the dose to most critical organs and delivered less integral dose. Zero collimator angles resulted in a better plan than "optimized" collimator angles, with lower dose to most of the normal structures. The number of segments did not have much effect on isodose distribu-tion, but a reasonable number of segments was necessary to keep treatment time from being prohibitively long. Gantry angles, when evenly spaced, had no notice-able effect on the plan. The patient tolerated the treatment well, and the initial complete blood count was favorable. Our results indicated that large-volume tu-mor sites may also benefit from precise conformal delivery of IMRT.
PACS numbers: 87.53.Kn, 87.53.Tf
Key words: gynecological, radiation, IMRT, treatment planning
I. INTRODUCTION
It is estimated that there are 83 000 new cases of
gynecological cancer in the United States in 2004 and 29 000 deaths
per year.(1) Half the gynecological
cancer cases originate at the uterine corpus. The American Cancer
Society estimated that "a malignant endometrial tumor will de-velop
in approximately 700,000 of the 48 million women in the United
States aged 35 or older.(2) For
late-stage disease with metastasis to the intrapelvic and
para-aortic lymph nodes, treatment often involves surgery and
postoperative radiation therapy and/or chemotherapy. The five-year
survival rate for lymph node positive disease ranges from 20% to
60%.(2) The radiation therapy regimens
typically consist of external beam radiation and/or intracavitary
brachytherapy. Traditional external beam technique uses a 3D,
four-field box with shielding by multileaf collimators (MLCs).(3) This four-field-box-technique includes a
substantial amount of normal tissue in the treated field, causing
morbidity and complication.
Intensity-modulated radiation
therapy (IMRT) is one of the most significant technological advances
in radiation therapy in the past decade. A majority of radiation
clinics in the United States have implemented IMRT or plan to
implement IMRT in the near future. The common treatment sites
related to IMRT include head and neck, prostate, and intracranial
tumors. Gyne-cological treatment has not been a common area for
IMRT. A recent survey indicated that IMRT was being performed on 88%
of head and neck patients, while only 15% of gynecologi-cal cancer
radiation therapy patients were treated with IMRT.(4) The conventional thinking is that IMRT
is more suitable for small-volume tumors, while late-stage
gynecological cancer, with a large volume of lymph nodes as target
volume, is not an appropriate candidate for IMRT. While there is
some truth to this, the opposite argument is equally valid.
Radiation for gynecological malignancies delivers extensive dose to
a large region of normal tissue, which may benefit from the
technical capabilities of IMRT.
There are several reports in the
literature on IMRT planning for gynecological malignancies. Mundt et
al.(5) compared conventional whole
pelvic radiation therapy plans to IMRT plans for 10 patients (5
cervical, 5 endometrial). They found that IMRT resulted in more
con-formal dose and a reduction of irradiated volume in small bowel,
rectum, and bladder. Heron et al.(6)
did a similar comparison on 10 gynecological patients and arrived at
the same conclusion. Adli et al.(7)
investigated the IMRT planning of 16 patients and discovered that
prone position-ing with a belly board decreased the small bowel even
more. The published articles on gynecological IMRT have used one
single isocenter for all treatment beams.
This report is a case study of
IMRT planning in a patient with endometrial cancer. The planning
target volume (PTV) spanned the periaortic and the pelvic lymph
nodes to a length of 33 cm. This was larger than the IMRT capability
of MLCs in most of the LINACs. Planning and treatment were
accomplished using two separate isocenters. All critical organs in
the ab-dominal and pelvic regions were delineated, and their dose
distributions were investigated. The IMRT plan was compared with a
traditional 3D plan. The field parameters in IMRT plan-ning were
also studied: beam energy, collimator angles, the number of
segments, and gantry angles, to observe the effect of these
parameters on the final dose distribution.
II. METHODS
The patient is a 53-year-old female who had
FIGO(8,9) stage IIIC, grade 3
endometrial adenocarcinoma. The patient had undergone total
abdominal hysterectomy with bilateral salpingo-oophorectomy and
pelvic as well as periaortic lymphadenectomy and pelvic washings.
One of the 8 lymph nodes from left pelvic node dissection and one of
12 from the right were positive for metastatic carcinoma. Four
periaortic lymph nodes were recovered, all of which were negative
for malignancy spread. It was decided to proceed with postoperative
radiation therapy followed by systemic chemotherapy.
The patient was simulated for
radiation treatment with CT. Simulation was done in a supine posture
with the arms above the head. A Vac Lok (MED-TEC Inc., Orange City,
IA ) cradle was used for immobilization, and a styroform block was
placed between the taped feet. CT contrast was not used. The
Pinnacle radiation planning system (Philips Medical Systems,
Bothell, WA) was used for the IMRT and 3D planning, and radiation
was delivered with segmental MLCs of a Siemens Primus LINAC (Siemens
Medical Solution, Erlangen, Germany). The goals of radiation therapy
were to treat both the pelvic and periaortic lymph nodes to a dose
prescription of 45 Gy in 1.8-Gy fractions, while minimizing the risk
for small bowel morbidity and sparing bone marrow. The bone marrow
would be important for chemotherapy afterward. Thirteen anatomical
contours in the pelvic and abdominal regions were delineated in the
plan. The planning target volumes, as defined in ICRU Report 50,(10) were intrapelvic and periaortic nodes.
The tissue structure included all the tissue in the CT image series
with the target vol-umes (intrapelvic and periaortic nodes)
excluded. The dose to tissue would give a measure of the "integral
dose" delivered.
We studied the
effect on the plan quality by various treatment parameters: energy,
collima-tor angle, number of segments, and gantry angles. Table 1
lists the parameters of the "original" plan (the original parameters
were used for the actual treatment of the patient), as well as the
alternative parameter settings for plan comparison in this study.
The whole PTV, including intrapelvic and periaortic nodes, traversed
33 cm in the superior-inferior direction and 15 cm in the lateral
direction. The total target volume was 1293 cm3. Since the target volumes were lymph
nodes, it was not possible to separately distinguish the gross (GTV)
or clinical (CTV) target volume; hence, the physician directly drew
the PTV. Although the Primus MLCs permit-ted a maximum field size of
40 cm at the isocenter, only the middle 27 cm of a 1-cm leaf width
could be used for IMRT delivery. Therefore, it was necessary to
split the treatment field into two isocenters: one at the pelvis and
the other at the abdomen, 15 cm from each other. The original IMRT
plan had a total of 14 beams, that is, 7 beams centered at each
isocenter, with evenly spaced gantry angles starting at 0°
(anterior).
The dose "objectives"
used in the optimization that generated the original IMRT plan are
listed in Table 2. As stated, it was possible to put in more than
one objective for each region of interest (ROI). The critical organs
included in the optimization objectives were all in the pelvic
region, except for the bone marrow, which extended from the abdomen
to the pelvis. The same objectives were used in all the IMRT plans
for fair comparison. The number of iterations used in optimization
was 25 because the total cost function was observed to be stabilized
(i.e., not decreasing any more) well before the 25th iteration. For
quality assurance, a hybrid plan com-posed of the treatment beams
irradiating the virtual water phantom was generated, and the doses
at the two isocenters were verified with ion chamber measurement.
Table 1. Parameters of the “original” plan (the original parameters were used for the actual treatment of the patient), as well as the alternative parameter setting for plan comparison in this study |
| Table 2. Objectives used in the optimization that generated the original IMRT plan |
Test 1 compared the
IMRT plan with a 3D non-IMRT plan. The typical 3D plans had fewer
numbers of radiation beams than the IMRT plans. Our alternative 3D
plan had 4 × 2 = 8 fields, giving 4-field box distributions. The MLC
apertures of the 3D portals were designed based on the beam's-eye
views (BEVs) of the PTVs. The collimator settings for the 3D plan
were differ-ent than the IMRT plan, especially at the junction
between the pelvic beams and the abdominal beams to ensure proper
field matching.
In Test 2, the
IMRT optimization was a rerun from iteration one with the same
settings and objectives. The only change was the energy, which
changed from 23 MV to 6 MV. Test 3 examined the effect of the
collimator angle rotation. By rotating the collimator angle and
mak-ing the PTVs more "diagonal" in the BEV, the abdominal beams
could encompass more of the intrapelvic nodes (Fig. 1), and the
pelvic beams could encompass more of the periaortic nodes. We
studied whether this would result in a better plan. We used another
planning system, the Nomos Corvus system (Sewickley, PA), to
automatically determine the "optimal" collimator angle for each
field relative to the shape of the PTVs. For each beam, the
algorithm found the collimator angle so that the MLC leaves were
perpendicular to the longest dimension of the combined PTVs. These
rotated collimator angles were put in the ADAC Pinnacle system for
IMRT optimization.
Fig.1. Beam’s-eye view (BEV) of field #7 at the abdominal isocenter in the collimator-rotated plan. By rotating the collimator angle to make the PTV more “diagonal” in the BEV, the abdominal beams could encompass more of the pelvic PTV, and the pelvic beams could encompass more of the periaortic PTV. |
||
Test 4 investigated the effect of the total number of MLC segments. The number of MLC segments determined the radiation-on time, and we aimed at keeping it within 20 min. The original result from the Pinnacle optimization showed the intensity profile of each beam, which was then converted to MLC segments. Depending on the conversion criteria, the resulting MLC segments might give intensities slightly different than the ideal solutions, with fewer segments resulting in larger differences. All plans in the tests were reoptimized, resulting in slightly different segments. For Test 4, the alternative plan was converted to about three times the number of segments of the original plan. This would indicate how much improvement to expect from increasing the number of segments, although we were aware that the alternative plan would prohibitively take longer to deliver. Test 5 was a plan with all gantry angles in-verted 180° and opposite to the original. The seven gantry angles are depicted in Fig. 2.
Fig. 2. The 7 gantry angles overlaid on top of the pelvic isodose distribution for (a) the original IMRT, gantry 0 plan, (b)
the 3D (gantry 0) plan, and (c) test 5 IMRT plan with opposite gantry angles. The bladder (purple) and the sigmoid colon
(gray) are at the center and are deeply embedded in the intrapelvic nodes (red). Both IMRT plans tailored the prescription
isodose (yellow) around the bladder and sigmoid colon, excluding these two organs from the highest dose. The 3D plan,
as expected, delivered a convex uniform dose for the whole area, and the doses to bladder and sigmoid colon were as high |
||
III. RESULTS AND DISCUSSION
The plans were primarily evaluated by comparing the dose volume histograms (DVHs). Fig-ures 3 to 7 show the DVHs of tests 1 to 5, respectively. For each test, (a) contained the structures in the pelvic regions, while (b) contained those in the abdominal regions. Bone marrow was included in the pelvic DVH graph, and tissue was included in the abdominal DVH graph arbi-trarily. The symbols O, 3, E, C, and G stand for the original, 3D, energy 6 MV, collimator rotated, and gantry 180° plans, respectively. For the sake of this study, all the plans were nor-malized so that the DVHs of the intrapelvic and the periaortic nodes each had 90% of the volume receiving more than the prescribed dose of 45 Gy, that is, D90 = 45 Gy. Identical normalization ensured clinically relevant comparison. Table 3 summarizes the differences of each plan in every organ delineated. Table 4 lists the D10 (dose delivered to at least 10% of the volume) for every organ in each plan. D10 was chosen as a robust indicator of the maximum dose for each anatomical structure. The ranking in Table 3 might not agree with the numbers in Table 4, since Table 3 is based on the whole DVH curves, while Table 4 essentially represents single points on the DVH curves.
Fig. 3. Test 1: DVH comparison between the original IMRT plan (O) and the 3D plan (3): (a) pelvic structures, (b) abdominal structures. |
||
Fig. 4. Test 2: DVH comparison between the original 23 MV plan (O) and the energy 6 MV plan (E): (a) pelvic structures, (b) abdominal structures. |
||
Fig. 5. Test 3: DVH comparison between the original zero collimator plan (O) and the collimator-rotated plan (C): (a) pelvic structures, (b) abdominal structures. |
||
Fig. 6. Test 4: DVH comparison between the original 165-segment plan (O) and the 509-segment plan (S): (a) pelvic structures, (b) abdominal structures. |
||
Fig. 7. Test 5: DVH comparison between the original gantry-start-at-0° plan (O) and the gantry-opposite plan (G): (a) pelvic structures, (b) abdominal structures. |
||
| Table 3. Summary of the differences of each plan in every organ delineated |
√ = better than the original plan; 0 = similar to the original plan; x = worse than the original plan; xx = much worse than the original plan |
| Table 4. D10 (Gy) of each plan in every organ delineated |
A. 3D plan
The
3D plan delivered much higher dose to all the critical structures in
the pelvic region, including the small bowel, bladder, rectum, and
sigmoid colon. (Refer to the DVHs in Figs. 3 to 7 and the numbers in
Table 4.) Figure 2 compares the isodose distributions at the pelvic
level of (a) the original 23-MV IMRT plan, (b) the 3D 23-MV plan
(test 1), and (c) the 6-MV IMRT plan (test 2). The bladder (purple)
and the sigmoid colon (gray) are at the center and are deeply
embedded in the intrapelvic nodes (red). Both IMRT plans tailored
the prescription isodose (yellow) around the bladder and sigmoid
colon to exclude these two organs from the highest dose. The 3D
plan, as expected, delivered a convex uniform dose for the whole
area, and the doses to bladder and sigmoid colon were as high as
that of the PTVs. Hence, the IMRT plan was more conformal than the
3D plan. It was found that the dose uniformity of the PTVs was
better in the 3D than the IMRT plan. The dose to tissue, indicating
the integral dose, was in fact less with the 3D plan. This
represented a workable compromise to achieve IMRT conformity. The
unexpected result here was that large bowel had lower dose in the 3D
plan. IMRT, while keeping most of the critical organs in lower
doses, had to deliver the radiation somewhere. In this case, higher
dose was delivered to the large bowel with IMRT. Nevertheless, the
3D plan was overall considered inferior to the IMRT plan.
The 3D planning process has its
own problem of field matching at the pelvic-abdominal junction.
Figure 8 shows the coronal view of isodose distribution of two 3D
plans with slightly different Y1 jaw settings of the abdominal
fields: (a) Y1 = 8.9 cm, and (b) Y1 = 9 cm. When the field size was
too small, a cold spot developed at the junction, while a
one-millimeter increase in field size created a hot spot. The DVHs
of the 3D plan in Fig. 3 were generated using Y1 = 8.9 cm. This
demonstrates the importance of careful field matching in 3D
planning. In contrast, no manual field matching was necessary for
double isocenter IMRT. Dose nonuniformity at the junction was taken
into account with the IMRT optimization objectives. The algorithm
auto-matically generated a solution including field sizes and IMRT
segments, and the resulting IMRT plans showed no problem at the
pelvic-abdominal junction. This is one of the key ad-vantages of
double isocenter IMRT.
Fig. 8. Coronal view of isodose distribution of two 3D plans with slightly different Y1s of the abdominal fields: (a) Y1 = 8.9 cm, (b) Y1 = 9.0 cm. When the field size was too small, a cold spot developed at the junction, while a one-millimeter larger field size created a hot spot. |
||
B. Energy 6 MV
Most anatomical structures received less desirable doses with 6 MV than 23 MV. The maxi-mum doses in the intrapelvic and periaortic nodes were higher with 6 MV at the same normalization of D90 = 45 Gy, meaning worse PTV homogeneity with 6 MV. The doses to the small bowel, bladder, sigmoid colon, large bowel, and liver were also higher with 6 MV. Some critical organs, such as bone marrow, spinal cord, and left kidney, received lower doses with 6 MV. The integral dose in general, as indicated by the dose to tissue, was higher with 6 MV. Dose in the 6-MV plan tended to spread out more into the surrounding normal structures, which seemingly made the 6-MV plan inferior. However, recall that the plans only considered dose from photons and ignored the neutron dose. It is well known that high-energy X-rays (above 10 MV) generate neutrons from the LINAC, which may increase the possibility of secondary malignancy.(11-13) In addition to the neutron dose to the patients with higher-energy photon beams, there is another health risk by the neutron-activated products(14): the residual radiation after the treatment of higher-energy IMRT can be considerable to therapists who enter the treatment room often. The physician should decide whether the photon dose advan-tage of higher energy is more significant than the neutron dose disadvantage..
C. Collimator angle rotation
The rotated collimator angles as determined by the Corvus system are listed in Table 5. The plan with "optimized" collimator rotation was uniformly worse than the original plan with zero collimator angle. Bone marrow, small bowel, bladder, rectum, right kidney, and large bowel all received higher doses in the collimator rotated plan. There was not a single anatomical struc-ture in which the collimator-rotated plan fared better. We postulated that although the pelvic-centered fields covered some of the periaortic nodes with rotated collimator, the 6.5-cm wide superior end leaf of the MLC did not facilitate precise dose delivery, and the same for the inferior end leaf of the abdominal-centered fields.
| Table 5. Rotated collimator angles (in degrees) as determined by the Nomos Corvus system |
D. Number of segments
Optimization created different intensity profiles for
each beam. A conversion from intensity profiles to MLC segments was
necessary, and the converted MLC segments might not repro-duce
exactly the ideal intensity profiles. The Pinnacle conversion
algorithm allowed criteria to discard segments that might not have a
significant effect on total dose distribution. This kept the number
of segments within a practical level. The original IMRT plan was
generated with the following settings: error tolerance = 7%, minimum
segment area = 5 cm2, and minimum
segment monitor units (MUs) = 4. That plan had a total of 165
segments for 14 fields. From the same optimization result, we also
converted the MLC segments with error tolerance of 3%, no minimum
segment area, and no minimum segment MUs. This gave a plan with 509
segments.
It was found that
permitting more MLC segments resulted in better uniformity of the
two PTVs. However, the plan with a higher number of segments
delivered a higher dose to the abdominal critical organs, such as
the spinal cord, right kidney, large bowel, and liver. It also
resulted in a higher integral dose as indicated by the dose to
normal tissue. A plan with more segments should be able to get
better optimization results than one with fewer segments in general.
The abdominal organ dose increased by 509 segments may be attributed
to random occurrence. For this particular patient case, it was not
possible to clearly conclude whether the plan with more or fewer
segments was better. Given that a larger number of segments
pro-longed treatment time at the LINAC, an attempt to increase the
number of segments in the hope of a better plan was determined
impractical.
Incidentally, when
the intensity profiles were different, the same conversion criteria
did not guarantee exactly the same number of segments. Our test
plans generated the following num-ber of segments: original, 165;
energy 6 MV, 164; collimator rotated, 150; segment test, 509; gantry
180°, 168.
E. Gantry angle
The test plan used gantry angles complementary to the
original plan. Hence, 0° became 180°, 205° became 25°, etc. It was
generally believed that as long as the fields were evenly spaced
around the treatment area, the exact gantry angles did not create a
significant difference in achievable dose distribution. Our results
confirmed this belief. The DVH curves of both gantry configurations
were essentially similar for every anatomical structure. The two
plans could be considered identical. The choice of gantry angles
might be based on other considerations, such as couch interference
during radiation delivery.
The
above findings were based on evaluating each plan and all the DVHs
as a whole. We were careful in our evaluation to not read too much
into the specific DVH of each anatomical structure in this study
(i.e., if bone marrow preservation is of top priority for a patient,
among our plans the 6-MV plan gave the best DVH, but that did not
conclude that 6 MV was the method of choice in sparing bone marrow).
Individual DVH results were probably due more to random chance and
not a result of the absolute merit of that treatment parameter. The
ranking of each structure in Table 3 was rather subjective,
particularly on whether the pairs of DVHs were similar (0), so it
would not be surprising if some readers disagree with a few of the
rankings. Another limitation of our study was that the treatment
plans did not explicitly take into consideration the motion of the
various organs. The magnitude of patient movement in the abdomen and
the pelvis likely had a significant effect on degrading the quality
of radiation delivery.(15) Tissue
heterogeneity was ignored in all plans, that is, dose calculation
was per-formed assuming unit density in all the tissue.
Nevertheless, it was not expected that the inclusion of patient
motion or tissue heterogeneity would alter the relative quality of
the alternative plans. We chose not to investigate the effect of
optimization objectives on isodose distribution, since any result
might be peculiar to the Pinnacle planning system or the geometry of
this particular patient.
Our
results are in agreement with the findings from other investigators.
Application of IMRT on a large abdominal target has been previously
reported. Hong et al.(16) studied 10
patients with whole-abdomen radiation and discovered that IMRT
resulted in significant dose reduc-tion to the bones and improved
PTV coverage as compared with conventional treatment. The high-dose
regions within the PTV increased slightly. Mundt et al.(5) studied 10 patients with pelvic
treatment and found that IMRT dose distribution (compared with 3D)
was more confor-mal to the PTV, with better DVHs for small bowel,
rectum, and bladder. They also found IMRT plans resulted in more
dose inhomogeneity within the PTV. Heron et al.(6) studied 10 patients treated with pelvic
fields and concluded that the doses to small bowel, bladder, rectum
were lower from IMRT versus 3D. Ahmed et al.(17) studied 5 patients with abdominal
treatment and demonstrated reduced doses to bone marrow, bowel,
spinal cord, and both kidneys.
Our
study was unique in several ways: the treated PTV (encompassing the
intrapelvic and periaortic nodes with 33 cm total length) was among
the largest of IMRT planning reported in the literature. This
necessitated two isocenters and a total of 14 radiation fields. Our
study was also meticulous in terms of delineating all critical
structures involved in the abdominal and pelvic regions. Finally,
while most investigations in IMRT gynecological treatment focused on
comparison with the 3D plans, we also studied the effects of various
IMRT parameters on isodose distribution. The discussion in
literature of large-volume IMRT treatment often in-cluded the
limitation in field width of some models of LINAC (maximum 14.5 cm),
and the consequent splitting of IMRT fields into twice the number of
smaller subfields. The Primus LINAC had a limitation in field width
of 20 cm, and we did not need to split any of our IMRT fields into
smaller fields.
Since IMRT was
developed in the past decade, most of the treatment delivered has
been for small-volume tumor sites, such as head and neck, prostate,
and intracranial locations. There was an unspoken assumption within
the radiation oncology community that large-volume tu-mor sites were
not suitable candidates for IMRT. We believe that this may be
misplaced dogma and that large treatment fields encompassing
extensive normal tissue may also benefit from precise conformal
delivery of IMRT. IMRT allows for higher than conventional dose
delivered to the target volume, while sparing critical organs from
radiation toxicity. The patient tolerated this high-dose treatment
with remarkable condition, and the initial lab finding on complete
blood count was favorable. It is expected that more research effort
will be directed to this area, with adequate patient numbers and
follow-up to evaluate the clinical outcome of IMRT in gynecological
treatment.
IV. CONCLUSIONS
In summary, we have planned and treated a patient with endometrial carcinoma using IMRT. The PTV including the intrapelvic and periaortic lymph nodes had a length of 33 cm, and treatment was accomplished with two isocenters. In this case study, we compared the effect of various field parameters on the resultant plan. The findings were the following: the IMRT plan delivers a smaller amount of dose to critical organs compared with the 3D plan. High energies (23 MV) produce a more desirable plan because of lower total photon integral dose (ignoring neutron dose). Collimator rotation is unnecessary and, in fact, leads to a slightly inferior plan. Including more MLC segments does not improve plan quality significantly and certainly in-creases the treatment time. Switching the gantry angles to opposite directions does not change the isodose distribution in any meaningful way.
ACKNOWLEDGMENTS
The authors would like to thank Philip M. Bruch, M.S., for constructive advice in preparing the manuscript.
REFERENCES
- Cancer facts and figures 2004. American Cancer Society; Atlanta; 2004.
- Perez CR, Brady LW, Halperin EC, Schmidt-Ullrich RK. Principles and practice of radiation oncology. Philadel-phia (PA): Lippincott Williams & Wilkins; 2004.
- Khan FM, Potish RA. Treatment planning in radiation oncology. Baltimore (MD): Williams & Wilkins; 1998.
- Mell LK, Roeske JC, Mundt AJ. A survey of intensity-modulated radiation therapy use in the United States. Cancer. 2003;98(1):204-211.
- Mundt AJ, Roeske JC, Lujan AE. Intensity-modulated radiation therapy in gynecologic malignancies. Med Dosim. 2002;27(2):131-136.
- Heron DE, Gerszten K, Selvaraj RN, et al. Conventional 3D conformal versus intensity-modulated radiotherapy for the adjuvant treatment of gynecologic malignancies: A comparative dosimetric study of dose-volume histograms. Gynecol Oncol. 2003;91(1):39-45.
- Adli M, Mayr NA, Kaiser HS, et al. Does prone positioning reduce small bowel dose in pelvic radiation with intensity-modulated radiotherapy for gynecologic cancer? Int J Radiat Oncol Biol Phys. 2003;57(1):230-238.
- Zaino R. Pathologic indices of prognosis in endometrial cancer: Selected aspects encompassing the gynecological oncology group experience. In: Rosen PF, editor. Pathology manual for 1995, part I, vol 30. Norwalk (CT): Appleton & Large; 1995: 1-28.
- International Federation of Gynecology and Obstetrics. Classification and staging of malignant tumors in the female pelvis: Annual report on the results of treatment in gynecological cancer. Int J Gynaecol Obstet. 1989;28:189.
- ICRU Report 50. Prescribing, recording, and reporting photon beam therapy. Bethesda: International Commission on Radiation Units and Measurements; 1993.
- Vanhavere F, Huyskens D, Struelens L. Peripheral neutron and gamma doses in radiotherapy with an 18 MV linear accelerator. Radiat Prot Dosim. 2004;110(1-4):607-612.
- Difilippo F, Papiez L, Moskvin V, et al. Contamination dose from photoneutron processes in bodily tissues during therapeutic radiation delivery. Med Phys. 2003;30(10):2849-2854.
- Waller EJ. Neutron production associated with radiotherapy linear accelerators using intensity modulated radiation therapy mode. Health Phys. 2003;85(5 Suppl):S75-S77.
- Guo S, Ziemer PL. Health physics aspects of neutron activated components in a linear accelerator. Health Phys. 2004;86(5 Suppl):S94-S102.
- Langen KM, Jones DT. Organ motion and its management. Int J Radiat Oncol Biol Phys. 2001;50:265-278.
- Hong L, Alektiar K, Chui C, et al. IMRT of large fields: Whole-abdomen irradiation. Int J Radiat Oncol Biol Phys. 2002;54(1):278-289.
- Ahmed RS, Kim RY, Duan J, Meleth S, De Los Santos JF, Fiveash JB. IMRT dose escalation for positive para-aortic lymph nodes in patients with locally advanced cervical cancer while reducing dose to bone marrow and other organs at risk. Int J Radiat Oncol Biol Phys. 2004;60(2):505-512.
© 2005 Am. Coll. Med. Phys.