A feasibility study of the Dynamic
Phantom scanner for quality
assurance of beam profiles at
various gantry angles
Yunkai Zhang, Wen C. Hsi, James C.H. Chu, Damian B. Bernard,
and Ross A. Abrams
Departments of Medical Physics and Radiation Oncology, Rush University Medical Center, 1653 West Congress Parkway, Chicago, Illinois 60612 U.S.A.
whsi@rush.eduReceived 1 November 2004; accepted 12 January 2005
The effect of gantry rotation on beam profiles of photon and electron beams is an important issue in quality assurance for radiotherapy. To address variations in the profiles of photon and electron beams at different gantry angles, a Dynamic Phantom scanner composed of a 20 × 12 × 6 cm3 scanning Lucite block was designed as a cross-beam-profile scanner. To our knowledge, differences between scanned profiles acquired at different gantry angles with a small size Lucite block and those acquired a full-size (60 × 60 × 50 cm3) water phantom have not been previously investigated. We therefore performed a feasibility study for a first prototype Dynamic Phantom scanner without a gantry attachment mount. Radiation beams from a Varian LINAC 21EX and 2100C were used. Photon beams (6 MV and 18 MV) were shaped by either collimator jaws or a Varian 120 Multileaf (MLC) collimator, and electron beams (6 MeV, 12 MeV, and 20 MeV) were shaped by a treatment cone. To investigate the effect on profiles by using a Lucite block, a quantitative comparison of scanned profiles with the Dynamic Phantom and a full-size water phantom was first performed at a 0° gantry angle for both photon and electron beams. For photon beam profiles defined by jaws at 1.0 cm and 5.0 cm depths of Lucite (i.e., at 1.1 cm and 5.7 cm depth of water), a good agreement (less than 1% variation) inside the field edge was observed between profiles scanned with the Dynamic Phantom and with a water phantom. The use of Lucite in the Dynamic Phantom resulted in reduced penumbra width (about 0.5 mm out of 5 mm to 8 mm) and reduced (1% to 2%) scatter dose beyond the field edges for both 6 MV and 18 MV beams, compared with the water phantom scanner. For profiles of the MLC-shaped 6 MV photon beam, a similar agreement was observed. For profiles of electron beams scanned at 2.9 cm depth of Lucite (i.e., at 3.3 cm depth of water), larger disagreements in profiles (3% to 4%) and penumbra width (3 mm to 4 mm out of 12 mm) were observed. Additional profiles with the gantry at 90° and 270° were performed for both MLC- and jaw-shaped photon beams and electron beams to evaluate the effect of gantry rotation. General good agreement is seen (less than 1 % variation) at all field sizes for collimator-shaped 6 MV and 18 MV photon beams. Similar variations observed for MLC-shaped photon beams indicate that the uncertainty in MLC position is similar to that for the collimator jaws. We conclude that the Dynamic Phantom scanner is a useful device for the routine quality assurance on beam profiles of photon beams and for constancy check on electron beams at various gantry angles. Caution should be taken when using this device to acquire basic electron dosimetry data.
PACS number: 87.53.-j
Key words: Dynamic Phantom, profile scanner, quality assurance
I. INTRODUCTION
There are many advantages associated with the use
of a multileaf collimator(1) (MLC),
including enhanced precision and ease of dose conformality to
designated targets, and protection of critical normal organs and
tissues for both 3D conformal and intensity-modulated radiation
therapy. However, a comprehensive process of quality assurance (QA)
for the MLC has not yet been fully established due to the complexity
of this task. By dividing the MLC QA process into routine
LINAC-based and patient-specific components, the challenge of MLC QA
is more readily approached. During patient-specific dosimetric
verification, the calculated dose distribution for each treatment
field with reassigned 0˚ gantry angle can be verified against one
measured from film or MAPCHECK, a diode-based dose measurement
device available in many institutions, including our own. In this
verification approach, one assumes that dose profiles and output
factors for each treatment field do not vary with gantry angle. One
problem associated with this approach is that it does not account
for gravitational pulling effects on an individual leaf's location
as the gantry angle varies. Therefore, it is essential to consider
and measure the gantry-angle-dependent dosimetric
variation.
Traditionally, films and
the water tank scanner are used to measure profiles. Using film
dosimetry, many factors need to be considered to obtain accurate
relative dose distributions. When using a water-tank scanner at
nonzero gantry angles, one faces an unavoidable setup complexity
with a sealed and reduced size water tank that may introduce
measurement errors. Therefore, a Dynamic Phantom consisting of a 20
× 12 × 6 cm3 Lucite block was designed
as a cross-beam-profile scanner by Advanced Radiation Measurements
Inc.(2) With a gantry attachment mount,
required radiotherapy beam profiles at different gantry angles for
routine beam profile QA can be performed with an easy setup. Due to
its finite size, the Dynamic Phantom may underestimate the scatter
dose near and outside of the field edge, especially for large field
sizes. However, based on Bragg-Gray theorem, the main contribution
to the measured dose comes from secondary electrons from the small
volume surrounding the radiation detector in a charged-particle
equilibrium condition. Therefore, profiles measured at a certain
depth in Lucite with the Dynamic Phantom should be similar to those
measured at an effective depth in water with a water phantom. The
effective depths of scanned profiles with a water tank scanner were
calculated according to the electron density of Lucite. For a
feasibility study on a first prototype Dynamic Phantom scanner,
quantitative comparison of scanned profiles for 5 cm, 10 cm, 15 cm,
20 cm, and 30 cm field sizes with the Dynamic Phantom at 1.0 cm and
5.0 cm depths in Lucite and a full-size water phantom at 1.1 cm and
5.7 cm effective depth were performed at a 0° gantry angle for the 6
MV and 18 MV photon beams. For a constancy check on the effect of
gantry rotation, additional scanned profiles with the gantry at 90°
and 270° with the same field sizes at 1.0 cm and 5.0 cm depths in
Lucite were also performed with 6 MV and 18 MV photon beams. To
examine the use of the Dynamic Phantom with electron beams,
measurements of various field sizes (10 cm, 15 cm, and 20 cm) at
different depths with 6 MeV, 12 MeV, and 20 MeV electrons were
performed at 0°, 90°, and 270° gantry angles. Based on quantitative
comparisons of photon/electron beam profiles at different gantry
angles, we evaluate the suitability of the Dynamic Phantom scanner
to determine the effect of gantry rotation on beam profiles in
routine QA.
II. MATERIALS AND METHODS
A. Dynamic Phantom and water tank scanners
The Dynamic Phantom scanner used in this study is shown in the top panel of Fig. 1. A Lucite block is moved along a threaded steel rod, driven by a stepping motor controlled by scanning software. The maximum scan range is 45 cm with a 1 mm scanning interval. The electron density of Lucite(3) is 1.137 with respect to water, and its physical density is 1.1364 g/cm3. An insertion tunnel at one side of the Lucite slab is used to place an ionization chamber for measurements. The depth of each scanned profile includes the thickness of Lucite in the beam path of the detector inserting slab.
|
||
The water-tank scanner used in this study has 3D scanning capability as shown in the bottom panel of Fig. 1. An ionization chamber is mounted on the detector holder, which is attached to a moving arm. The dimensions of the water tank are 60 × 60 × 50 cm3.
B. Profile measurements
Profiles for several field sizes (5 × 5, 10 ×
10, 15 × 15, 20 × 20, and 30 × 30 cm2 at 100 cm source-to-axis distance (SAD)) were acquired with the
Dynamic Phantom using 6 MV and 18 MV photon beams delivered by
Varian 21EX and 2100C LINACs, respectively, with the gantry at 0°.
Radiation fields for the photon beams were defined by collimator
jaws instead of a MLC to reduce beam edge uncertainty to less than 1
mm. Two scanned depths (1.0 cm and 5.0 cm in Lucite for 6 MV, 2.9 cm
and 5.0 cm in Lucite for 18 MV) were acquired for each field size.
Corresponding scanned effective depths (1.1 cm and 5.7 cm in water
for 6 MV, 3.3 cm and 5.7 cm in water for 18 MV) were acquired for
each field size with the water scanner. Profiles of the same set of
field sizes and depths for the 6 MV and 18 MV beams were acquired
with Dynamic Phantom with the gantry angles at 90° and
270°.
To study the effect of gantry
rotation on MLC-shaped photon beams delivered by a Varian 21EX with
120 MLC, beam profiles in the direction along the MLC motion for
field sizes of 5 × 5 and 10 × 10 cm2 (at 100 cm SAD with retracted collimator jaws) were acquired with
the Dynamic Phantom at 0°, 90°, and 270°gantry angles. Beam profiles
for the same field sizes were also acquired with the water tank at a
0°gantry angle. Scans at two depths, 1.0 cm and 5.0 cm in Lucite,
and 1.1 cm and 5.7 cm in water, were acquired for each field size
and gantry angle. For each field size and depth, profiles along
(X) and perpendicular to (Y) the MLC leaf direction
were acquired at a 0° gantry angle in both the Dynamic Phantom and
the water tank. Profiles in the X direction were acquired
only for a 90° gantry angle, and in the Y direction only for
a 270° gantry angle.
For electron
beams delivered by a Varian 2100C LINAC with the energies of 6 MeV,
12 MeV, and 20 MeV, profiles of several field sizes (10 × 10, 15 ×
15, and 20 × 20 cm2 defined at 100 cm
source-to-surface distance (SSD)) at a 0°gantry angle were acquired
with the Dynamic Phantom. The field size of electron beams was
defined by a trimmer bar positioned at 95 cm SSD. Two scanned depths
(1.0 cm and 1.4 cm in Lucite for 6 MeV, 1.0 cm and 2.9 cm in Lucite
for 12 MeV and 20 MeV) were acquired for each field size.
Corresponding scanned effective depths (1.1 cm and 1.6 cm in water
for 6 MeV, 1.1 cm and 3.3 cm in water for 12 MeV and 20 MeV) are
acquired for each field size with the water-tank scanner. Profiles
with the same field sizes and depths for the 6 MeV, 12 MeV, and 20
MeV electron beams were acquired with the Dynamic Phantom with the
gantry angles at 90°and
270°.
Measurements of photon and
electron beams with both the water tank and the Dynamic Phantom were
performed with a PTW Freiburg 0.1 cm3 sealed chamber and a 1 mm scanning interval at 100 cm SAD and 100 cm
SSD, respectively. During scanning with the Dynamic Phantom, the
scanner was placed directly on the treatment table and was carefully
aligned with a level. The uncertainty on the LINAC gantry position
was less than 1°.
Since the
Lucite/water-to-air ratios of the average mass energy absorption
coefficient, , for the megavoltage photon beams do not vary as a
function of depth, the distribution of measured ionization readings
is proportional to the relative dose profiles of the photon beams.
Therefore, measured ionization distributions were used to present
measured dose profiles in this study. To assure that the spectral
distribution and the fluence of primary and scattered photons at a
certain scanned depth of Lucite is the same as that at a comparable
depth in water (i.e., at the corresponding effective depth in
water), the scaling factor(4)
is
used to obtain the effective depth in water. This is adopted from
AAPM TG21. Since
is proportional to the electron density of
Lucite with respect to water, the relative electron density (1.137)
is used as the scaling factor to calculate the effective depth in
water used for photon beam in this study.
Since the
ionization-to-dose conversion when using an ionization chamber for
electron beam dosimetry depends strongly upon the electron's energy
through the effects of mass stopping-power ratios and other
perturbation factors,(5) caution should
be taken to apply the appropriate correction factors. However, at
depths less than the depth of I50 (the
depth at which the ionization reading is at the 50% of its maximum
reading), the energy of electrons at the same depth does not vary
significantly, since the primary electrons contribute the most
ionization. Therefore, the ionization readings were acquired at
depths less than I50 depth to examine the
effect of gantry rotation on electron beam profiles. The formula
, based on the scaling factor by the HPA (Hospital
Physicists'Association 1985),(6) is
used to calculate the effective depth in water for electron beams.
The Zw and ZLuc are the scanning depths of water and
Lucite,
is the ratio of physical densities of Lucite to water,
and
is the mass stopping power ratio of Lucite to water.
For profiles scanned at depths close to dmax, the
is
approximately 1.0 with small variations. Therefore, the ratio of
physical densities (1.1363 g/ cm3) of
Lucite to water is used for calculating effective depth in this
study.
III. RESULTS
A. Profiles at 0° gantry angle
Figure 2 shows the comparisons between profiles acquired with the Dynamic Phantom at 5 cm depth in Lucite and with the water phantom at 5.7 cm depth in water with the gantry angle at 0°for 6 MV and 18 MV photon beams. The symmetry of each measured profile was within 1% for all fields acquired with the Dynamic Phantom and water tank. Results obtained by comparing the scanned profiles just inside of the field edges for both 6 MV and 18 MV at 5 cm depth of Lucite gave good agreement (<1% variance). The field edges of profiles scanned with the Dynamic Phantom and the water tank were aligned within 1 mm. Similar agreement was observed for 6 MV and 18 MV at a 1 cm depth in Lucite. However, lack of scattered radiation due to the finite size of the Dynamic Phantom produced lower doses outside the field edge and became more prominent (about 2%) for a 30 × 30 cm2 field of a 6 MV beam. For a quantitative comparison, a summary of the measured penumbra widths for all measured profiles is listed in Table 1. The penumbra width is defined as the width between 20% and 80% of central axis dose. Penumbra widths range from 5 mm to 8 mm for the measured photon beam profiles. For 6 MV and 18 MV, penumbra widths of a given field size increase by about 0.5 mm as the depth increases. This increase of the penumbra width is a result of the divergence of primary beam. For a profile at a given depth in the same beam, about 1.0 mm increase in penumbra width is observed for an increase in field size from 5 cm to 30 cm. This increase is due to a larger number of scattered photons generated in a larger field. The penumbra widths at a given depth for the 18 MV beam are about 1.5 mm wider than those for the 6 MV beam. In addition, as shown in Fig. 3, the use of the Dynamic Phantom reduced the penumbra width by about 0.5 mm for both 6 MV and 18 MV beams in comparison with a standard water phantom scans.
FIG. 2. Profiles of field sizes (10 cm, 20 cm, and 30 cm) scanned at 5.0 cm depth in Lucite with the Dynamic Phantom (lines) and at 5.7 cm depth in water with water (symbols) are plotted for 6 MV and 18 MV photon beams in the top and bottom panels, respectively. All profiles were measured at a gantry angle of 0°. |
||
TABLE 1. Penumbra widths of various field sizes and depths of scanned profiles for 6 MV and 18 MV photon beams at a 0° gantry angle with the Dynamic phantom and water phantom |
FIG. 3. Extracted penumbra widths for profiles of 6 MV and 18 MV photon beams with the Dynamic Phantom (depths in Lucite) and water phantom (depths in water). |
||
Due to the limited available thickness of Lucite slabs, profiles acquired at 1 cm and 2.9 cm depths in Lucite were used. Since the 2.9 cm depth in Lucite is deeper than the depth of I50 for a 6 MeV electron beam, profiles of the 6 MeV beam were scanned at a 1.4 cm depth in Lucite instead of 2.9 cm to avoid the scanned depth deeper than I50. The symmetry of each measured profile was within 2% for all fields and all energies acquired with the Dynamic Phantom and water phantom. Profiles of the 20 MeV beam at both 1.0 cm and 2.9 cm depths in Lucite as shown in Fig. 4 reveal good agreement (less than 2%) between the Dynamic Phantom and water phantom. Although similar good agreement between profiles scanned at a 1 cm depth in Lucite for the 12 MeV beam was observed, large disagreement (about 3%) was observed between profiles scanned at a 2.9 cm depth in Lucite as shown in Fig. 4. For a quantitative comparison, a summary of measured penumbra widths for all measured field sizes and electron energies is listed in Table 2. Penumbra widths fall between 5 mm and 15 mm for the measured profiles of electron beams. Penumbra widths increase with increasing depths due to multiple scattering. At the scanned 1.0 cm depth in Lucite, penumbra widths of profiles with the Dynamic Phantom agree to within 1 mm with those at effective depths with the water phantom. The penumbra increases with the depth for all field sizes and energies. The difference is more than 2 mm for 12 MeV electron beam penumbra measured at 2.9 cm depth in Lucite and at 3.3 cm depth in water.
FIG. 4. Left panels: Profiles of field sizes (10, 15 cm, and 20 cm) scanned with the Dynamic Phantom (lines) and with water (symbols) are plotted for the 20 MeV electron beam. Scanned depths at 1.0 cm and 2.9 cm in Lucite (i.e., 1.1 cm and 3.3 cm in water) are shown in the top and bottom panels, respectively. Right panels: Same field sizes and depths for the 12 MeV electron beam. |
||
TABLE 2. Penumbra widths for various field sizes and depths of scanned profiles for 6 MeV, 12 MeV, and 20 MeV electron beams at 0° gantry angle with the Dynamic Phantom and water phantom |
B. Profiles at various gantry angles
Figure 5 shows profiles scanned at gantry angles of 0°, 90°, and 270° for 6 and 18 MV collimator-shaped photon beams at a 5 cm depth in Lucite for a constancy check on the effect of gantry rotation. The symmetry of each measured profile was within 1% for all fields at gantry angles of 90° and 270°. General good agreement is seen (less than 1%) for all field sizes of 6 MV and 18 MV photon beams. Similar agreement was observed for profiles scanned at 1.0 cm depth in Lucite with 6 and 18 MV photon beams.
Fig. 5. Profiles of field sizes (10 cm, 20 cm, and 30 cm) of collimator-shaped photon beams scanned at a 5.0 cm depth in Lucite with the Dynamic Phantom at gantry angles of 0°, 90°, and 270° are plotted for 6 MV and 18 MV photon beams in the top and bottom panels, respectively. |
||
Figure 6 shows X (transverse) and Y (radial) profiles for the 6 MV MLC-shaped photon beams. The symmetry of each measured profile was less than 1%. At a 0° gantry angle, Y profiles measured with the Dynamic Phantom for agreed with the profiles measured with the water tank for both scanning depths to within 2%. The Y profiles measured at a 270° gantry angle with the Dynamic Phantom also agreed with the profiles measured at a 0° gantry angle. A good agreement was observed between the X profiles measured with the Dynamic Phantom at gantry angles of 0° and 90°. Similar gantry angle variations for both X and Y profiles of the MLC-shaped photon beams scanned using the Dynamic Phantom indicate that the uncertainty in the MLC position is similar to that for the collimator jaws.
FIG. 6. X and Y profiles of the field sizes (5 cm and 10 cm) for the MLC-shaped 6 MV photon beams are plotted, respectively, in the left and right panels. Both the X and Y profiles at the gantry angle of 0° are plotted for 1 cm and 5 cm depths of Lucite with the Dynamic Phantom and 1.1 cm and 5.7 cm depth of water with the water tank. X profiles at the gantry angle of 90° and Y profiles at the gantry angle of 270° are also plotted. The X and Y are in the directions of along and perpendicular to the MLC leaf, respectively |
||
Figure 7 shows profiles scanned at gantry angles of 0°, 90°, and 270° for 6 MeV, 12 MeV, and 20 MeV electron beams at 1.0 cm depth in Lucite for the effect of gantry rotation. The symmetry of each measured profile was within 1% for all fields at gantry angles of 90° and 270°. A difference between the 90 and 270 profiles of up to 4% was observed near the field edge for the 20 ′ 20 cm2 field of the 20 MeV electron beam. To investigate observed disagreement for profiles scanned at gantry angles of 0° and 90°/270° is beyond the scope of this paper. Similar results were obtained for profiles scanned at larger depths of 1.4 cm in Lucite for 6 MeV electron beam and of 2.9 cm in Lucite for the 12 MeV and 20 MeV electron beams.
FIG. 7. Profiles of field sizes (10 cm, 15 cm, and 20 cm) scanned at a 1.0 cm depth in Lucite with the Dynamic Phantom at gantry angles of 0°, 90°, and 270° are plotted for 6 MeV, 12 MeV, and 20 MeV electron beams. |
||
IV. Discussion and Conclusion
Using the Dynamic Phantom scanner for measuring
profiles of electron beams produces large differences in measured
penumbra widths with respect to a water phantom scanner. Since the
scaling factor used to calculate effective depths does not take into
account differences in the mass angular scattering power(7,8) between Lucite and water, the observed
differences in penumbra width may be mainly caused by this variation
in angular scattering power. Therefore, the Dynamic Phantom scanner
should not be used to measure primary electron beam profiles for
dosimetry purposes.
Based on the
good agreement between profiles with the Dynamic and water phantom
for collimator-shaped 6 MV and 18 MV at a 0° gantry angle, Dynamic
Phantom scan measurements are acceptable for beam flatness and
symmetry QA checks. Additional good results were obtained for
profiles of a collimator-shaped beam at different gantry angles
scanned with the Dynamic Phantom scanner. Observed similar
variations between profiles of a MLC-shaped photon beam at gantry
angles of 0°, 90°, and 270° indicate that the uncertainty in MLC
position is similar to that for the collimator jaws. We conclude
that the Dynamic Phantom scanner has the potential to be a useful
device for the routine quality assurance on beam profiles of photon
beams and constancy check of electron beams at various gantry
angles.
REFERENCES
- Battista JJ, G.S. Bauman GS. The future of IMRT. Intensity-modulated radiation therapy: The State of the Art. Madison (WI): Medical Physics Publishing; 2003. 843-874.
- Advanced Radiation Measurements Inc. Product description, http://www.armx2000.com.
- Wong VYW. Quality assurance devices for dynamic conformal radiotherapy. J Appl Clin Med Phys. 2004; 5:8-15.
- AAPM Task Group 21. A protocol for the determination of absorbed dose from high-energy photon and electron beams. Med Phys. 1983;10:741-771.
- Khan FM, Doppke KP, Hogstrom KR, et al. Clinical electron-beam dosimetry: Report of AAPM Radiation Therapy Committee Task Group No. 25. Med Phys. 1991;18:73-109.
- HPA (Hospital Physicists'Association). Code of practice for electron beam dosimetry in radiotherapy. Phys Med Biol. 1985;30:1169-1194.
- Werner BL, Khan FM, Deibel FC. A model for calculating electron beam scattering in treatment planning. Med Phys. 1982;9:180-187.
- ICRU Report 35. Radiation dosimetry: Electron beams with energies between 1 and 50 MeV; 1984.