Dosimetric evaluation of
GAFCHROMIC® XR type
T and XR type R films
Sharifeh A. Dini, Rafiq A. Koona, John R. Ashburn, and Ali S. Meigoonia
Department of Radiation Medicine, University of Kentucky Medical Center, 800 Rose Street, Lexington, Kentucky 40536 U.S.A.
meigooni@yahoo.com, rafiq.koona@gmail.com, ashburn@uky.edu, alimeig@uky.edu
(Received 10 September 2004; accepted 27 September 2004)The high spatial resolution of radiochromic film makes it ideal for dosimetric measurements and dose distributions in regions of high dose gradient. Intensity-modulated radiation therapy, intravascular brachytherapy, and eye-plaque radiation therapy demand precise spatial dosimetric calculations. Such precision is not possible with conventional dosimeters, such as thermoluminescent dosimeters and ionization chambers. Recently, new GAFCHROMIC® XR type T and type R films have been developed for radiation dosimetry, specifically in interventional radiology procedures. Dosimetric characteristics (i.e., linearity, post-exposure density growth, energy dependence, dose-rate dependence, and UV light sensitivity) of these new films were investigated. To evaluate the clinical applications of these films, their characteristics were compared with other commercially available film models. GAFCHROMIC® XR type T and type R films were found to be more sensitive to low-energy doses as compared with GAFCHROMIC® MD-55 films.
PACS numbers: 87.66-a, 87.53-j
Key words: dosimetry, GAFCHROMIC® film, XR type T film, XR type R filma Corresponding author: alimeig@uky.edu
I. INTRODUCTION
Dosimeters with high spatial resolution are
desirable for measurement and quality control of radiation in
high-dose gradient areas. These detectors could provide accurate
dose distribution in the close vicinity of brachytherapy sources.
Two procedures in particular that require higher resolution
dosimeters are intravascular brachytherapy,(1,2) for the treatment of restenosis, and
eye-plaque radiation therapy,(3) for
the treatment of ocular melanoma. Moreover, the quality assurance of
the intensity-modulated radiation therapy technique could also
benefit from these types of detectors.(4-7) Conventional measuring systems such as
thermoluminescent dosimeters, ionization chambers, and
semiconductors are not capable of such spatial resolution. Although
radiographic silver halide film provides high spatial resolution, it
is not a tissue-equivalent detector due to its high
z-component. Moreover, the high z-element of this film
type increases its energy dependence, particularly at
low-energy radiation levels. Conversely, the organic-based
radiochromic film dosimeters offer superior tissue-equivalent
properties and high spatial resolution that could fulfill the
dosimetric requirements of the latest advancements in the field of
radiation
therapy.
GAFCHROMIC®, or
radiochromic, films were originally designed for industrial
applications by International Specialty Products (ISP), Inc. (Wayne,
NJ). In radiochromic films an organic-based dye is used, which
changes color due to polymerization when exposed to radiation.
Various superior features of this film compared with silver halide
film include the following: relatively low-energy response,(8) insensitivity to visible light,
self-developing characteristics, and dose-rate independence.(9) These features have attracted the
attention of medical physicists for the use of
GAFCHROMIC® films in medical applications.(10) Detailed studies were performed by
McLaughlin et al.(11-13) to
determine the dosimetric properties of the original models of
radiochromic films. The recent improvements in film sensitivity,
accuracy, precision of film manufacturing, and ease of use have
further extended the medical applications of these film types for
different treatment modalities, such as brachytherapy,(8,14,15) interface studies,(16,17) stereotactic radio surgery,(13,18-20) dosimetry in the penumbra
region,(21) and proton dosimetry.(22) Butson et al. have extensively
investigated the effects of polarization and light sources on the
response of MD-55-2(23) and HS(24-26) GAFCHROMIC® media.
Applications of these films have been discussed in detail in AAPM
Task Group 55 (TG-55).(27)
Recently,
ISP developed GAFCHROMIC® XR type R and type T
films for the measurement of high radiation dose in interventional
radiological procedures. This film is very useful for direct
measurements of skin dose from fluoroscopic procedures. The
sensitivity of these films to photons with energies less than
200 keV is substantially greater than HD-810 and MD-55 films.
Thomas et al. discussed various means of measuring the response
of the XR type R films using different commercially available
densitometers and flatbed scanners.(28)
The characterization of the XR type T films was presented as an
abstract by Haque et al.(29) in
the World Congress on Medical Physics and Biomedical Engineering in
2002. Also, characterization of the HS and XR type R film was
presented as an abstract by Ashburn et al.(30) in the 2001 annual meeting of the
American Association of Physicists in Medicine (AAPM). However, to
our knowledge there is no publication to date on the dosimetric
characteristics of the XR type T and type R
media.
The main objective of this
project was to determine the dosimetric characteristics
(sensitivity, energy dependence, linearity, post-exposure density
growth, dose-rate dependence, and UV light sensitivity) of the XR
type T and type R films. This report offers a complete
dosimetric evaluation of XR type T and type R films, along
with a comparison of these films with commonly available film models
for both low- and high-energy X-ray beams.
II. MATERIALS AND METHOD
A. GAFCHROMIC® XR type T and type R films
Two different types of GAFCHROMIC® XR films-type T and type R-were studied in these investigations. In both film types, the active layer is sandwiched between the two sheets of polyester. In the type T film, both polyester substrates are transparent, whereas in type R one is transparent and the other is opaque. These film types are designed and manufactured by ISP and are available in sheet sizes of 14 in. × 17 in. (type R) and 5 in. × 5 in. (type T). The color of XR type R film turns from amber to dark greenish-black, and XR type T films turn from orange to brownish-black, depending on the level of exposure.
B. Dosimetric evaluation
The dosimetric characteristics of the XR
type T and type R films were determined by cutting sheets
of these films into small 4 cm × 4 cm squares. These
pieces were exposed to low-energy X-ray beams ranging from
10 kVp to 150 kVp from an Oldeft Therapax HF 150T
superficial unit (Oldeft/Nucletron, Columbia, MD), and high-energy
X-ray beams of 6 MV and 18 MV from a Varian LINAC (Clinac
2100C/D, Varian Oncology Systems, Palo Alto, CA). The effective
energies and half value layers of the beams from a superficial unit
are shown in Table 1.
The
low-energy measurements were performed in air by placing the film on
an empty box with a 14 cm × 14 cm cutout on its top.
A layer of clear transparency was taped on top of the cutout area in
order to support the film. Another transparency was placed on the
film to hold the film in place and to provide a similar setup used
for the calibration of the superficial unit using the Markus chamber
(PTW-New York, Hicksville, NY). All the low-energy exposures were
performed using a 15-cm diameter cone. For high-energy exposures, an
appropriate buildup material of Solid WaterT phantom
(Radiation Measurements Inc., Middleton, WI) was used.
FIG. 1(a) Comparison of XR type T (solid symbols) and MD-55-2 (open symbols) film sensitivities (optical density/cGy) as a function of dose for the 100-kVp X-ray beam. The solid lines are a second-order polynomial fit through the data points. |
The radiochromic films used were XR type T
(lot number 30189-1B), XR type R (lot number 30198-2), and MD
55-2 (lot number 37175). These films were handled in accordance with
the precautions recommended in TG-55.(27)
The XR
type T films were scanned before and after exposure with a
LUMISCAN-50 laser film scanner (Lumisys, Sunnyvale, CA) using a
spatial resolution of 0.35 mm. The optical densities of these
films along the two orthogonal axes were corrected for background
(reading from pre-exposed film) and averaged for each piece of film
using the PTW software Mephysto v6.30 (PTW-New York, Hicksville,
NY). Optical densities of XR type R films were read by a
GretagMacbeth D19C reflection densitometer (GretagMacbeth, New
Windsor, NY). This densitometer provides readings in four different
colors: black, yellow, cyan, and magenta. While ISP recommends using
cyan for this densitometer, the other colors were also studied in
these investigations. The optical densities taken from five
different points on the film, using the center and one point from
each quadrant, were averaged and corrected for background to obtain
the net optical density. This procedure was followed throughout
these investigations. The MD-55-2 films were read with a
GretagMacbeth D19C reflection densitometer for comparisons with XR
type R films. A white paper was used as a backing material for
reading the MD-55-2 films with this densitometer. For comparison of
XR type T with MD-55-2 films, a LUMISCAN laser scanner was used
to scan both film types. To minimize the UV exposure, these films
were kept in black envelopes when not in use and were also stored in
normal room temperature and humidity conditions.(11,31-33) The measured optical densities
were used to determine the dosimetric characteristics of the XR
type T and type R films as described in the following
sections.
B.1 Film sensitivity
The sensitivities (i.e., response per unit absorbed dose) of XR type T, type R films, and MD-55-2 films were measured and compared in this study. The MD-55-2 data was also obtained along with XR type T and type R films in order to have an identical experimental setup for comparisons of these film types. This process eliminates any uncertainties that may arise when comparing the new films with the published data for MD-55-2 films measured with different scanning techniques. Four pieces of each film (XR-type T and MD-55-2 films) were exposed to a 100-kVp beam, and four more pieces of each film were exposed to a 6-MV beam for the doses ranging from 186 cGy to 1856 cGy. These films were scanned with the LUMISCAN-50 laser film scanner after 24 h, and their results were analyzed to obtain the film sensitivity. Similarly, four pieces of each film (XR-type R and MD-55-2 films) were exposed to a 50-kVp X-ray beam for the following doses: 184 cGy, 460 cGy, 920 cGy, and 1840 cGy. After 24 h, these films were read with the GretagMacbeth D19C reflection densitometer. A white paper was used as a backing material for reading the MD-55-2 films with this densitometer. The sensitivity of each film type for different beam qualities was determined by measuring the optical density of the films as a function of absorbed dose. The sensitivities of different film types were compared as a function of scanning techniques.
B.2 Energy dependence
The energy dependence of the XR type T
film was evaluated by measuring the optical densities of
13 pieces of film exposed to 800 cGy, with different beam
energies ranging from 20 kVp to 18 MV. A dose of
800 cGy was selected to optimize the signal-to-noise ratio and
improve the accuracy of the data. One piece of film was exposed to a
6-MV X-ray beam and another to an 18-MV X-ray beam from a LINAC.
Nine other pieces of film were exposed to low-energy beams ranging
from 11 keV to 79 keV (i.e., 20 kVp to 150 kVp)
from a superficial unit. One piece of film was exposed to the 192Ir HDR (high dose rate) source
(VariSource, Varian Oncology Systems, Palo Alto, CA) with an average
energy of 380 keV, and another piece of film was exposed to
Tc99m (Cardinal Health, Lexington, KY)
with an average energy of 140 keV. For 192Ir HDR source irradiation, the film was
placed on a 10 cm × 20 cm × 20 cm piece of
Solid WaterT phantom material. A Styrofoam block with a
circular cutout at the center was used to hold the 192Ir source, in a single dwell position,
using an endobronchial catheter. The film was kept at a distance of
4.8 cm directly beneath the HDR source. The irradiation with 99mTc was performed by placing a vial
containing 438 mCi of 99mTc on the
film, which was located on the surface of a 10-cm thick Solid
WaterT phantom material. The initial dose rate of this
vial was measured using an NIST calibrated Markus ionization chamber
prior to the film exposure. Both the calibration and experiment were
performed with identical geometric setups. The exposure time was
adjusted to correct for the decay of the source during the
irradiation. These films were scanned after 24 h with the
LUMISCAN laser scanner. The variation of XR type T film
response as a function of beam energy was used to evaluate the
energy dependence of the film response. In addition, three pieces of
MD-55-2 film were exposed to a fixed total dose of 800 cGy with
different beam energies of 50 kVp, 100 kVp, and 6-MV X-ray
beams. These films were read with the LUMISCAN laser scanner after
24 h, and the results were compared with those of XR
type T films.
Similarly,
13 pieces of XR type R film and 6 pieces of MD-55-2
film were exposed to 400 cGy with energy beams in the range
11 keV to 6000 keV. The XR type R film is more
sensitive, so a dose of 400 cGy was found sufficient to
optimize the signal-to-noise ratio and hence to improve the accuracy
of the experimental data. These films were read with the
GretagMacbeth D19C reflection densitometer 24 h after the time
of exposure. The responses of XR type R and MD-55-2 films for
four different color spectra of the densitometer were used to
determine the energy dependence of these film types.
B.3 Linearity of the films
The linearity of XR type T film response was examined by exposing seven pieces of film to doses ranging from 46 cGy to 1856 cGy with a 100-kVp X-ray beam, and another eight pieces were exposed to doses ranging from 10 cGy to 3000 cGy with a 6-MV X-ray beam. These films were then scanned 24 h after the time of exposure using the LUMISCAN-50 laser film scanner. Variation of the film response as a function of the absorbed dose was used to determine the linearity of the film response. As suggested by Zhu et al.,(34) deviation of the measured net optical density of the film from the linear approximations, with zero y intercept, was introduced as the nonlinearity correction factor for this film. Similarly, the linearity of XR type R film was determined by exposing 21 pieces of the film to doses ranging from 10 cGy to 400 cGy and 7 pieces each with 50-kVp, 100-kVp, and 150-kVp X-ray beams. These films were read with the GretagMacbeth D19C reflection densitometer 24 h post-exposure. The optical densities of this film were plotted as a function of the absorbed dose to determine their nonlinearity correction factors. The linearity responses of XR type T and type R films were compared with each other as well as with the linearity of model MD-55-2 film.
B.4 Time dependence of the film response
Three pieces of XR type T film were exposed to doses of 842 cGy, 1684 cGy, and 3368 cGy with a 6-MV X-ray beam and were scanned 1 h, 3 h, 5 h, 15 h, 21 h, 44 h, 97 h, and 314 h after the exposure. Similarly, three pieces of XR type R film were exposed to a dose of 400 cGy with 50 kVp, 100 kVp, and 150 kVp beams and read at different time periods ranging from 1 h to 336 h. The responses of these films were plotted and analyzed. Variation of the film responses as a function of time were used to determine their time dependence.
B.5 Dose-rate dependence
Responses of seven pieces of XR type T and type R films, exposed to the same total dose using various dose rates ranging from 16.5 cGy/min to 755.9 cGy/min, were evaluated to determine the dose-rate dependence of these films. The films were irradiated to a dose of 400 cGy for type T films and 368 cGy for type R films using the Therapax HF 150T superficial unit. Different dose rates were generated by increasing the film-to-source distance and calibrating the beam using an NIST traceable calibrated Markus ionization chamber. The exposed films were read after 24 h, and the data was analyzed in order to determine the dose-rate dependence of these film types.
B.6 UV light sensitivity
The effect of UV light on the XR type T and type R films was measured by exposing them to the light emitted by fluorescent tubes. One piece of each type of film was placed at a distance of approximately 1.5 m from the light source (two 40-W tubes). The film responses were measured prior to exposure and after 10 days, 30 days, 50 days, and 60 days. The 100-kVp calibration curves were used to convert the net film response to absorbed dose (cGy). The measured absorbed dose (Gy) was divided by the illumination of the light bulbs (lux) for each film type and plotted as a function of time. The illumination of the fluorescent tubes was determined by using the following equation(35)
where watt is the total power of the fluorescent light tubes, assuming a luminous efficiency(36) of 80 lm/W.
III. RESULTS AND DISCUSSIONS
The dosimetric characteristics (i.e., sensitivity, energy dependence, linearity, time dependence, dose-rate dependence, and UV light sensitivity) of the GAFCHROMIC® XR type T and type R films were experimentally determined in air and Solid WaterT phantom material, and the results are discussed below.
A. Film sensitivity
The responses of the XR type T film and MD-55-2 model, exposed to 100-kVp and 6-MV X-ray beams, are compared in Figs. 1(a) and 1(b), respectively. These results indicate that for a dose of 928 cGy from a 100-kVp X-ray beam, the XR-type T film is approximately 14 times more sensitive than the MD-55-2 film. However, for the 6-MV beam, the sensitivity of the XR type T film is approximately (within ±10%) the same as the MD-55-2 film for doses lower than 1000 cGy, but the differences increase at larger doses.
FIG. 1(a) Comparison of XR type T (solid symbols) and MD-55-2 (open symbols) film sensitivities (optical density/cGy) as a function of dose for the 100-kVp X-ray beam. The solid lines are a second-order polynomial fit through the data points. |
||
FIG. 1(b) Comparison of XR type T (solid symbols) and MD-55-2 (open symbols) film sensitivities (optical density/cGy) as a function of dose for the 6-MV X-ray beam. The solid lines are a second-order polynomial fit through the data points. |
||
Similarly, Fig. 1(c) compares the responses of the XR type R film and model MD-55-2 films exposed to a 50-kVp X-ray beam and measured in the cyan color spectrum. This figure indicates that for the 920 cGy absorbed dose from the 50-kVp X-ray beam, the XR type R film is approximately five times more sensitive than the MD-55-2 model. Table 2 shows the ratio of sensitivity of XR type R film to that of MD-55-2 film for the four different color spectra of the densitometer as a function of absorbed dose. This table indicates that for the 50-kVp X-ray beam the XR type R film is approximately 6.5 ± 0.97 times more sensitive than MD-55-2 film.
FIG. 1(c) Comparison of XR type R (solid symbol) and MD-55-2 (open symbols) film sensitivities (optical density/cGy) for the cyan color spectrum of the densitometer as a function of dose for the 50-kVp X-ray beam. The solid lines are a second-order polynomial fit through the data points. |
||
TABLE 2. Ratio of sensitivity of XR type R film to that of MD-55-2 film as a function of absorbed dose for the 50-kVp X-ray beam, measured for the four different color spectra of the densitometer |
B. Energy dependence
Energy dependence of the XR type T and MD-55-2 films was measured for the energies ranging from 20 kVp to 18 MV. Figure 2(a) shows that the net optical densities of these films, normalized to the value at 6 MV, are nearly constant (within ±10%) for the photon energies greater than 300 keV. For energies less than 300 keV, the MD-55-2 film shows a decrease in optical density with decreasing beam energy. However, the relative optical density of XR type T films increases and then decreases as energy decreases, with a maximum value at 100 kVp. The MD-55-2 film results are consistent with the published data.(8,25,27,37,38) This figure indicates that the response of the XR type T film at 41 keV (i.e., 100 kVp) is approximately 7.5 times larger than that of 6-MV photons. However, the variation of the MD-55-2 film response in the energy range of 41 keV to 6 MV was found to be negligible (less than 5%). Table 3(a) shows relative optical densities of the XR type T film as a function of the effective beam energy.
FIG. 2(a) Energy dependence of XR type T and MD-55-2 film responses shown as the variation of the relative optical density versus the effective beam energy. These films were exposed to a fixed total dose of 800 cGy, with different effective beam energies ranging from 11 keV to 6000 keV. The solid lines connect the data points. |
||
TABLE 3(a). The relative optical densities of the XR type T film as a function of beam energy ranging from 11 keV to 6000 keV, normalized to the 6 MV (i.e., 2000 keV) X-ray beam |
Similarly, the energy dependence of XR type R film was measured for the different energies ranging from 20 kVp to 18 MV and compared with those of the MD-55-2 film. Figure 2(b) shows the relative optical density of the XR type R and MD-55-2 films for the cyan color spectrum of the densitometer as a function of effective beam energy. This figure indicates that the relative response of both MD-55-2 and XR type R films, for energies greater than 650 keV, is nearly constant (within ±10%). For energies less than 650 keV, the relative optical density of XR type R films increases and then decreases as energy decreases, with a maximum value at 100 kVp, whereas the optical density of MD-55-2 decreases monotonically with decreasing beam energy. The MD-55-2 film results are consistent with the published data.(8,27,37,38) Figure 2(b) indicates that at 41 keV (i.e., 100 kVp) the response of the XR type R film is approximately six times larger than that of 6-MV photons while the MD-55-2 film, measured with the same technique, showed approximately 30% smaller response than a 6-MV beam. Figure 2(c) shows the net optical density of the XR type R film as a function of beam energy and color of the densitometer. This figure indicates that the absolute value of the responses is largest for the cyan color spectrum, which is the manufacturer's recommended spectrum for use of fluoroscopic skin dose measurements. Table 3(b) shows relative optical densities of the XR type R film as a function of the effective beam energy.
FIG. 2(b) Energy dependence of XR type R and MD-55-2 film responses, for the cyan color of the densitometer, shown as the variation of the relative optical density versus the effective beam energy. These films were exposed to a total fixed dose of 400 cGy with different effective beam energies ranging from 11 keV to 6000 keV. The solid lines connect the data points. |
||
FIG. 2(c) The net optical density of XR type R film as a function of effective beam energy for the cyan, black, magenta, and yellow colors of the densitometer. These films were exposed to a total fixed dose of 400 cGy with different effective beam energies ranging from 11 keV to 6000 keV. The solid lines connect the data points. |
||
TABLE 3(b). The relative optical density of XR type R film as a function of beam energy ranging from 11 keV to 6000 keV for four different color spectra of the densitometer, normalized to the 6-MV (i.e., 2000 keV) X-ray beam |
C. Linearity of the film response
Figure 3(a) shows the measured optical density of XR type T films as a function of absorbed dose in the range 10 cGy to 3368 cGy for 100-kVp and 6-MV X-ray beams. Although both curves appear more linear at lower optical densities, they show some nonlinearity as the optical density increases. The solid lines show the second-order polynomial fit through the measured data points. The dashed lines with open symbols in the figure represent the nonlinearity correction factor calculated as a ratio of the measured optical density and linear approximation with no y-intercept, as suggested by Zhu et al.(34) With a second-order polynomial fit, the agreement between the fitted data and measured data is within 5% throughout the range, except at the lowest dose of 46 cGy, where 35% disagreement has been observed. Conversely, a linear fit shows a poorer agreement across the entire range: within 10% for doses above 500 cGy, and degenerating to an approximate 40% difference at 93 cGy for 100-kVp X-ray beam. The film uncertainty was obtained by measuring the film response at different segments of the film and comparing that with the average value. For high-energy beams, the statistical fluctuation of the film response was found to be less than ±5% for doses greater than 800 cGy, and ranged from ±22% at 200 cGy to ±1.6% at 3200 cGy. However, for low-energy beams the statistical fluctuation was less than ±5% for doses above 50 cGy.
FIG. 3(a) The net optical density and nonlinearity correction factors for XR type T films as a function of absorbed dose to water for 100-kVp and 6-MV X-ray beams. The solid lines are a second-order polynomial fit through the measured optical densities. The dashed lines connect the calculated nonlinearity correction points. |
||
The response of XR type R films as a function of absorbed dose, ranging from 10 cGy to 400 cGy for 50-kVp, 100-kVp, and 150-kVp X-ray beams for the cyan color of the densitometer, is plotted in Fig. 3(b). Also, the response of this film to a 6-MV X-ray beam for doses in the range of 10 cGy to 3368 cGy for the cyan color of the densitometer is shown in Fig. 3(c). These figures indicate that the type R film response is nonlinear for both the low-energy and high-energy beams. With a second-order polynomial fit, the agreement between measured and fitted data for low-energy X-ray beams was found to be within 10% for doses above 10 cGy. However, at 10 cGy the fitted curve disagrees with the measured data by approximately 128%. Similarly, with the polynomial fit, the 6-MV X-ray beam showed an agreement within 9% for doses more than 50 cGy and degenerated to 140% at 10 cGy. Conversely, a linear fit to the measured data showed a poorer agreement across the entire range. Tables 4(a) and 4(b) show the nonlinearity correction factor calculated as a ratio of the measured optical density to linear approximation, with no y-intercept, for four different colors of the densitometer.(34) No significant difference in percentage deviation from linearity was observed for 50-kVp, 100-kVp, and 150-kVp X-ray beams; their average values for the four colors of densitometer are shown in Table 4(a). These results indicate a nonlinearity correction factor greater than 10% for low-energy X-ray beams with doses less than 200 cGy and high-energy X-ray beams with doses less than 1700 cGy. For high-energy beams, the statistical uncertainty of this film was found to be less than ±5.6% for doses greater than 200 cGy, and ranged from 15% at 25 cGy to 2% at 3200 cGy. However, for low-energy beams this uncertainty was found to be less than ±5% for doses above 25 cGy.
FIG. 3(b) The net optical density of XR type R film as a function of absorbed dose to water for 50-kVp (open triangles), 100-kVp (open circles), and 150-kVp (closed triangles) X-ray beams for the cyan color spectrum of the densitometer. The solid lines are a second-order polynomial fit through the data points. |
||
FIG. 3(c) The net optical density of XR type R film for the cyan color spectrum of the densitometer as a function of absorbed dose to water for the 6-MV X-ray beam. The solid line is a second-order polynomial fit through the data points. |
||
TABLE 4(a). Nonlinearity correction factors for XR type R film in the dose range 10 cGy to 400 cGy for low-energy X-ray beams for the four colors of densitometer. Since no significant differences were observed between the 50-kVp, 100-kVp, and 150-kVp X-ray beams, an average value of the three energies is shown. |
TABLE 4(b). Nonlinearity correction factors for XR type R film in the dose range 11 cGy to 3368 cGy for the 6-MV X-ray beam, for different color spectra of the densitometer |
D. Time dependence
Optical densities of the GAFCHROMIC® films as a function of time after the exposure were used to evaluate their time dependence. Figure 4(a) shows the post-irradiation film response for the XR type T films for the absorbed doses of 842 cGy, 1684 cGy, and 3368 cGy. This figure indicates that the film response for a given dose changes by approximately 16% within the first 24 h after exposure and an additional 4% for the next 24 h. However, no significant changes (less than 2%) have been observed up to 314 h after exposure.
FIG. 4(a) Variation of net optical densities of XR type T film as a function of time after exposure for doses of 3368 cGy, 1684 cGy, and 842 cGy. The solid lines connect the data points. |
||
Similarly, Fig. 4(b) shows the post-irradiation film response of the cyan color spectrum of the XR type R film for a fixed total dose of 400 cGy with 50-kVp, 100-kVp, and 150-kVp X-ray beams. An increase of approximately 16% in optical densities was noted between the first hour and 24th hour of exposure, and a growth of approximately 4% for the next 24 h and approximately 2% for the following 300 h. A similar trend was observed for the other color spectra of the densitometer. Tables 5(a) and 5(b) provide detailed information regarding the time dependence of both XR type T and type R films (cyan color), respectively.
FIG. 4(b) Variation of net optical densities (cyan color) of XR type R film as a function of time after exposure for a dose of 400 cGy using a 50-kVp, 100-kVp, and 150-kVp X-ray beam. The solid lines connect the data points. |
||
TABLE 5(a). Post-exposure density growth of XR type T films as a function of time and absorbed dose, relative to the value at 1 h after exposure |
TABLE 5(b). Post-exposure density growth of XR type R films as a function of time and absorbed dose, relative to the value at 1 h after exposure, for the cyan color of the densitometer |
E. Dose-rate dependence
To accommodate a wide range of clinical applications using the radiochromic film, a detailed knowledge of the film's dose-rate dependence is required. Dose-rate dependence of the XR type T radiochromic films for a given total dose was measured using different dose rates in the range 16.5 cGy/min to 755.9 cGy/min. Figure 5(a) shows the variation of the film response as a function of dose rate. The solid line in this figure represents a linear fit to the measured data. As seen in this figure, the approximately zero slope of this line indicates no dose-rate dependence of this film, within an experimental uncertainty of ±5%. Similarly, the responses of XR type R films to different dose rates ranging from 30 cGy/min to 699 cGy/min, with a total dose of 368 cGy, are shown in Fig. 5(b). This figure shows a random variation of the film response around a linear fit with approximately zero slope, to within ±5%, indicating that the film is dose-rate-independent.
FIG. 5(a) Variation of the optical density of the XR type T film as a function of dose rate ranging from 16 cGy/min to 756 cGy/min. The solid line is a linear fit through the data points. |
||
FIG. 5(b) Variation of the optical density (cyan color) of the XR type R film as a function of dose rate ranging from 30 cGy/min to 700 cGy/min. The solid line is a linear fit through the data points. |
||
F. UV light sensitivity
The sensitivities of the GAFCHROMIC® XR type T and type R films to UV light were measured to determine the change in the film optical density due to room light during radiation dosimetry. The impact of room light (i.e., fluorescent tubes) was measured and converted to equivalent absorbed dose using a calibration curve from a 100-kVp X-ray beam. Figure 6(a) shows the variation of the equivalent absorbed dose per unit illuminance (lux) after 10, 31, 52, and 60 days of exposure to the room light for the XR type T film. The dose received by this film for 10 days of continuous exposure to two 40-W tubes at a distance of 1.5 m was measured to be approximately 6.74 cGy. Figure 6(b) shows the variation of equivalent absorbed dose after 3, 8, 15, and 30 days of exposure to room light for the XR type R film. These results indicate that a short exposure of these films to room light during the measurements will not have a significant impact on the radiation dosimetry. A one-day exposure to room light at a distance of 1 m or a 4-day exposure at a distance of 2 m may add a dose of 0.8 cGy to both XR type T and type R films, which is insignificant (less than 3%) for measuring doses greater than 25 cGy.
FIG. 6(a) Ultraviolet light dose contamination per lux at a distance of 1 m for dosimetry with XR type T film. The solid line represents a linear fit through the measured data points. |
||
FIG. 6(b) Ultraviolet light dose contamination per lux at a distance of 1 m for dosimetry with XR type R film read with the cyan color of the densitometer. The solid line represents a linear fit through the measured data points. |
||
IV. CONCLUSION
The dosimetric characteristics of both XR
type T and type R films were determined experimentally for
both low- and high-energy X-ray beams. The color of XR type R
and T films changes from amber to brownish-green and orange to
blackish-green, respectively, depending on the level of exposure.
The results of these two film types were compared with each other as
well as with model MD-55-2 film. These results indicated that
type T films were approximately 14 times more sensitive
than MD-55-2 film, for 928 cGy from a 100-kVp beam. However,
for high-energy beams their sensitivities were found to be within
±10%. Table 2 shows that for a dose of 920 cGy from a
50-kVp beam, the XR type R film is approximately 5.42 times
more sensitive than MD-55 film for the cyan color of the
densitometer. Also, this result indicated that the average ratio of
sensitivities of type R to MD-55 film, for all four colors of
the densitometer, was found to be 6.5 ± 0.97. In addition to
the sensitivity, the type T and R films have a lower
threshold than the MD-55-2 film (100 cGy for type T,
25 cGy for type R and 200 cGy for MD-55-2 film,
reported in this investigation). Therefore, type T and R
films are superior to MD-55-2 film for the measurements of low
absorbed doses in the diagnostic energy
range.
Energy dependence of XR
type T and R film response was measured and compared with
that of MD-55-2 film. As shown in Figs. 2(a) and 2(b), the
optical density of MD-55-2 film is nearly constant for high-energy
photons. However, at lower energies the optical density of this film
decreases with decreasing energy. At 18 keV, the optical
density of MD-55-2 film is approximately 40% smaller than that of a
6-MV beam. These results are consistent with the published data.(8,25,27,37,38) In addition, Figs. 2(a)
and 2(b) show that the variation of measured optical density of XR
type T and type R films, for photon beams greater than
300 keV and 650 keV, respectively, is within ±10%.
However, for lower energies the relative optical densities of XR
type T and R films increase to a value of approximately 7 at
41 keV (i.e., 100 kVp) and then decrease to a value of
approximately 1.5 at 11 keV (20 kVp). Therefore, for
high-energy beams, the energy dependence of the new films is similar
to that of MD-55-2, but at low energies new films present a larger
variation of optical density compared with MD-55-2
film.
The results of these
investigations show that the responses of XR type R and T films
as a function of absorbed dose can be approximated better with a
second-order polynomial than with a linear function. The deviation
from linearity for these films was introduced as a nonlinearity
correction factor, similar to that of MD-55 film as suggested by Zhu
et al.(34) As shown in
Fig. 3(a), the nonlinearity corrections for the XR type T
film are within 10% for doses above 500 cGy and degenerate to
approximately 40% at approximately 93 cGy. Table 4(a)
shows the nonlinearity correction factors for type R film for
the low-energy photons as a function of the color of the
densitometer. These results indicate that for cyan, the nonlinearity
corrections of the type R film are in the range 1.5 to 0.9 for
doses ranging from 10 cGy to 400 cGy. However, for
high-energy photons, the nonlinearity corrections for cyan are in
the range 1.7 to 0.9 for the doses ranging from 11 cGy to
3368 cGy, as shown in Table 4(b). The nonlinearity
corrections for high-energy beams were found to be highest for the
yellow color of the densitometer. The statistical fluctuation of XR
type T film response, for high-energy beams, was found to be
less than ±5% for doses greater than 800 cGy, while for
low-energy beams, the fluctuation was less than ±5% for doses above
50 cGy. For XR type R film, the statistical fluctuation
was found to be less than ±5.6% for doses greater than 200 cGy,
while for low-energy beams, the fluctuation was within ±5% for doses
above 25 cGy.
The optical
densities of XR type T and R films were found to increase by
approximately 16% within the first 24 h after exposure and an
additional 4% for the next 24 h. No significant (less than 2%)
changes were found approximately 360 h after the exposure. These
results are consistent with the data obtained for XR type films
by Giles and Murphy(39) and ISP.(40) In addition, Ali et al.(41) and Meigooni et al.(33) have shown that the optical density of
MD-55 films increases by approximately 16% after 24 h from the
time of exposure. Therefore, no time dependence correction is needed
if the patient and calibration films are exposed and read
simultaneously. However, if the patient film is read immediately
after exposure, but the calibration is read 24 h after
exposure, the optical density of the patient film must be increased
by approximately 16%. Responses of both XR type T and
type R films were found to be dose-rate-independent. The
results of these investigations showed a variation of ±5% for the
dose rates ranging from 16 cGy/min to 755 cGy/min. Saylor
et al.(9) showed a ±5% variation
in optical densities of HD-810 film for dose rates ranging from
2 cGy/min to 20 000 cGy/min. In addition, Niroomand-Rad
et al.(27) state that MD-55 film
response is independent of dose rates in the clinical range of
200 cGy/min to 400 cGy/min. More recently, Giles and
Murphy(39) and ISP(40) showed that XR type films are
dose-rate-independent within ±5%. These published results support
the findings of our
investigation.
Several
investigators have noted that GAFCHROMIC® films are
sensitive to UV light and hence suggest that proper care and
handling can eliminate UV contamination.(27,32,33,40) The results of our
investigation show an insignificant effect of UV light on the
sensitivities of GAFCHROMIC® XR type T and
type R films. These results indicate that the dose received by
XR type T and type R films from a continuous exposure to
two 40-W tubes for a period of 4 days, at a distance of
2.0 m, is equivalent to approximately 0.8 cGy dose from a
100 kVp X-ray beam. This amount of UV contamination is
insignificant (less than 3%) for clinical applications, which
require a dose greater than
25 cGy.
In summary, the higher
sensitivity and lower threshold of the XR type R and
type T films, at low photon energies, is superior to other film
models. This can be helpful for radiation dosimetry, particularly
for low-dose radiation measurements during interventional
radiological procedures. However, the higher energy dependence of
the new films relative to the MD-55-2 films should be taken into
consideration for absolute dose measurements. For the new films, use
of a previously calibrated film-strip without a clear identification
of the beam quality may introduce an erroneous result. Therefore,
the new film types should be either calibrated with the same beam
quality as the measured beam or corrected for the differences in the
film response using the data in Figure 2. No significant
differences in linearity, time dependence, and dose-rate dependence
of the new films to that of MD-55-2 films have been observed.
Because of the opaque backing material on the XR type R films,
they must be read with a reflective densitometer, while the XR
type T film can either be read with a laser scanner or a white
light scanner with appropriate filtration.
ACKNOWLEDGMENT
The authors would like to thank Dr. David Lewis from ISP for providing the GAFCHROMIC® films used in these investigations, as well as Dr. Walter Miller for providing the Tc99m for this project. The authors note their appreciation for Mr. Shahid B. Awan, Mr. Curtis Baker, Mr. Venkata Rachabatthula, and Mrs. Jenny Clark, for their valuable editorial comments and suggestions during the preparation of this manuscript.
REFERENCES
- Duggan DM, Coffey CW, Levit S. Dose distribution for a 32P-impregnated coronary stent: Comparison of theoretical calculations and measurements with radiochromic film. Int J Radiat Oncol Biol Phys. 1998;40(3):713-720.
- Duggan DM, Coffey CW, Lobdell JL, Schell MC. Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapy. Med Phys. 1999;26(11):2461-2464.
- Krintz AL, Hanson WF, Ibbott GS, Followill DS. Verification of plaque simulator dose distribution using radiochromic film. Med Phys. 2002;29(6):1220-1221.
- A. Mack, G. Mack, D. Weltz, S. Scheib, H. Bottcher, V. Seifert. High precision film dosimetry with Gafchromic® films for quality assurance when using small fields. Med Phys. 2003;30(9):2399-2409.
- Bayouth JE, Wendit D, Morill SM. MLC quality assurance techniques for IMRT applications. Med Phys. 2003; 30(5):743-750.
- Bayouth JE, Morill SM. MLC dosimetric characteristics for small field and IMRT applications. Med Phys. 2003; 30(9):2545-2552.
- Graves MN, Thompson AV, Martel MK, McShan DL, Fraass BA. Calibration and quality assurance for rounded leaf-end MLC systems. Med Phys. 2001;28(11):2227-2233.
- Muench PT, Meigooni AS, Nath R, McLaughlin WL. Photon energy dependence of the sensitivity of radiochromic film compared with silver halide and LIF TLDs used for brachytherapy dosimetry. Med Phys. 1991;18(4):769-775.
- Saylor MC, Tamargo TT, McLaughlin WL, Khan HM, Lewis DF, Schenfele RD. A thin film recording medium for use in food irradiation. Radiat Phys Chem. 1988;31:529-536.
- 10. Klassen NV, van der Zwan L, Cygler J. GafChromic MD-55: Investigated as a precision dosimeter. Med Phys. 1997;24:1924-1934.
- McLaughlin WL, Chen YD, Soares CG, Miller A, Van Dyke G, Lewis DF. Sensitometry of the response of a new radiochromic film dosimeter to gamma radiation and electron beams. Nucl Instr Methods Phys Res. 1991;A302: 165-176.
- McLaughlin WL, Soares CG, Puhl JM, et al. Novel radiochromic films for clinical dosimetry. Radiat Prot Dosim. 1996;66:263-271.
- McLaughlin WL, Soares CG, Sayeg JA, et al. The use of a radiochromic detector for the determination of stereotactic radiosurgery dose characteristics. Med Phys. 2001;21(3):2227-2233.
- Sayeg JA, Gregory RC. A new method for characterizing beta-ray ophthalmic applicator sources. Med Phys. 1991;18(3):453-461.
- Chiu-Tsao S-T, de la Zerda A, Lin J, Ho Kim J. High sensitivity GafChromic film dosimetry for 125I seed. Med Phys. 1994; 21(5):651-657.
- Farahani M, Eichmiller FC, McLaughlin WL. Metal-polysiloxane shields for radiation therapy of maxillo-facial tumors. Med Phys. 1991;18: 273-278.
- Niroomand-Rad A, Razavi R, Thobejane S, Harter K. Radiation dose perturbation at tissue-titanium dental interfaces in head and neck cancer patients. Int J Rad Biol Phys. 1996;34(2):475-480.
- Mack A, Scheib SG, Rieker M, et al. A system for quality assurance in radiosurgery. In: Kondziolka D, editor. Radiosurgery. Basel: Karger, 5; 2004. 236-246.
- Bjarngard E, Tsai JS, Rice RK. Doses on the central axes of narrow 6-MV X-ray beams. Med Phys. 1990;17: 794-799.
- Somigliana A, Cattaneo GM, Fiorino C, et al. Dosimetry of gamma knife and linac-based radiosurgery using radiochromic and diode detectors. Phys Med Biol. 1999;44:887-897.
- Galvin JM, Smith AR, Lally B. Characterization of a multi-leaf collimator system. Int J Radiat Oncol Biol Phys. 1993;25:181-192.
- Vatnitsky SM, Schulte RW, Galindo R, Meinass HJ, Miller DW. Radiochromic film dosimetry for verification of dose distribution delivered with proton-beam radiosurgery. Phys Med Biol. 1997;42:1887-1898.
- Butson M, Yu P, Metcalfe P. Effects of readout light sources and ambient light on radiochromic film. Phys Med Biol. 1998;43:2407-2412.
- Butson M, Yu P, Cheung T, Inwood D. Polarization effects on a high-sensitivity radiochromic film. Phys Med Biol. 2003;48:N207-N211.
- Butson M, Cheung T, Yu P. High sensitivity radiochromic film dose comparisons. Phys Med Biol. 2002;47: N291-N295.
- Butson M, Cheung T, Yu P. Corresponding dose response of radiographic film with layered gafchromic film. Phys Med Biol. 2002;47:N285-N289.
- Niroomand-Rad A, Blackwell CR, Coursey BM, et al. Radiochromic film dosimetry. Recommendations of AAPM Radiation Therapy Committee Task Group 55. Med Phys. 1998;25(11):2093-2115.
- Thomas G, Chu RY, Rabe F. A study of GafChromic XR type R film response with reflective-type densitometers and economical flatbed scanners. J Appl Clin Med Phys. 2003; 4(4):307-314.
- Haque M, Ralston A, Suchowerska N, Andretic A. Characterisation of GAFCHROMIC® XR type-T fluoroscopic film for kilovoltage X-ray therapy beams. Proceedings of the World Congress on Medical Physics and Biomedical Engineering, Aug 24-29, 2003, Sydney, Australia [CD-ROM]. ISBN 1877040142.
- Ashburn JR, Al-Otoomi A, Sowards K, Tamimi MM, Meigooni AS. Investigation of the new highly sensitive Gafchromic HS and XR films. Med Phys. 2001;28:1244 (abstract).
- Janovsky I, Mehta K. The effects of humidity on the response of radiochromic dosimeters FWT-60-00 and GafChromic DM-1260. Radiat Phys Chem. 1994;43:407-409.
- Reinstein LE, Gluckman GR, Amols HI. Predicting optical densitometer response as a function of light source characteristics for radiochromic film dosimetry. Med Phys. 1997;24(12):1935-1942.
- Meigooni A, Sanders M, Ibbott G, Szeglin S. Dosimetric characteristics of an improved radiochromic film. Med Phys. 1996;23(11): 1883-1888.
- Zhu Y, Khirov AS, Mishra V, Meigooni AS, Williamson JF. Quantitative evaluation of radiochromic film response for two-dimensional dosimetry. Med Phys. 1997;24(2):223-231.
- Smith FG, King TA. Optonics and photonics: An introduction. John Wiley & Sons, Ltd; 2000. 415 p.
- Mumford R. Improving visual efficiency with selected lighting. J Optom Vision Devel. 2002;33(3).
- Chiu-Tsao S-T, Duckworth T, Zhang C, et al. Dose response characteristics of new models of GAFCHROMIC films: Dependence on densitometer light source and radiation energy. Med Phys. 2004;31(9);2501-2508.
- Das IJ, Cheng CW. Dosimetric characteristics of new Gafchromic-HS film. Med Phys. 2001;28(6):1244.
- Giles ER, Murphy PH. Measuring skin dose with radiochromic dosimetry film in the cardiac catheterization laboratory. Health Phys. 2002;82(6):875-880.
- GAFCHROMIC® Radiochromic Dosimetry film background information and characteristic performance data (www.ispcorp.com/products/dosimetry/index.html).
- Ali I, Costescu C, Vicic M, Dempsey JF, Williamson JF. Dependence of radiochromic film optical density post-exposure kinetics on dose and dose fractionation. Med Phys. 2003;30(8):1958-1967.
© 2005 Am. Coll. Med. Phys.