An inhomogeneity correction
algorithm for irregular fields
of high-energy photon beams
based on Clarkson integration
and the 3D beam subtraction
method
Sotirios Stathakis,1 Constantin Kappas,2 Kiki Theodorou,2 Nikos Papanikolaou,3 and Jean-Claude Rosenwald4
Radiation Oncology Department,1 Fox Chase Cancer Center, Philadelphia, Pennsylvania 19111 U.S.A.; Medical Physics Department,2 Medical School, University of Thessalia, Larisa, Hellas; Department of Radiation Oncology,3 University of Arkansas for Medical Sciences, Little Rock, Arkansas 72205 U.S.A.; Institut Curie,4 Section Medicale, 26 Rue d' Ulm, 7500 Paris, France
sotirios.stathakis@fccc.edu, kappas@med.uth.grReceived 16 August 2004; accepted 11 July 2005
A number of treatment-planning systems still use conventional correction methods for body inhomogeneities. Most of these methods (power law method, tissue-air ratio (TAR), etc.) consider only on-axis points, rectangular fields, and inhomogeneous slabs covering the whole irradiating field. A new method is proposed that overcomes the above limitations. The new method uses the principle of the Clarkson method on sector integration to take into account the position and lateral extent of the inhomogeneity with respect to the point of calculation, as well as the shape of the irradiating field. The field is divided into angular sectors, and each sector is then treated separately for the presence of inhomogeneities using a conventional correction method. Applying this method, we can predict the correction factors for Co-60 and 6-MV photon beams for irregular fields that include inhomogeneities of lower or higher densities relative to water. Validation of the predicted corrections factors was made against Monte Carlo calculations for the same geometries. The agreement between the predicted correction factors and the Monte Carlo calculations was within 1.5%. In addition, the new method was able to predict the behavior of the correction factor when the point of calculation was approaching or moving away from the interface between two materials.
PACS number(s): 87.53.Bn, 87.53.Wz
Key words: inhomogeneity correction, radiation therapy, photons
I. INTRODUCTION
Traditionally, the clinical experience with
radiation therapy has been based on the tissue responses to the
planned doses in a homogeneous body even if the irradiated volume
contains inhomogeneous volumes (mainly lungs, air cavities, and
bones).
The dramatic increase in
computing power at affordable prices has greatly enhanced several
technical advances in radiotherapy. The radiotherapy
treatment-planning system (RTPS) that uses 3D patient data is a
reality. Several algorithms have been proposed to implement some
sort of inhomogeneity correction, from the simplified tissue-air
ratio (RTAR), which yields a correction factor for water-based
calculations, to superposition/convolution and Monte Carlo methods,
which include the inhomogeneity in the calculation of patient dose.
At present, the majority of the current commercial RTPS offer the
equivalent TAR (ETAR)(1) method and
other conventional methods developed more than 20 years ago(2-4) as inhomogeneity correction
algorithms. The shortcomings of these earlier methods are well
known, and sometimes calculation differences of 10% from
measurements are not uncommon.
The
conventional methods, such as the power law method, RTAR, and
differential TAR (DTAR),(5-8) of
inhomogeneity corrections in high-energy X-ray beams assume that
- the points where the primary dose is altered by the presence of inhomogeneity are considered to be centered on the beam axis, and the lateral dimensions of the inhomogeneity being larger than the field dimensions;
- for the points where the primary dose is not altered, the correction factor is taken as unity.
In addition, these methods
are not able to directly take into account the shape of the
irradiating field (irregular fields).
The most common inhomogeneity
method is the Batho method for which several improvements have been
proposed. Such improvements were focused on correcting the extent of
the inhomogeneity as well as the position of the point of interest.
Wong and Purdy(9) have shown that
modification of the Batho power law method into an additive form
would improve its application. El-Khatib and Battista(8) replaced TAR values by tissue maximum
ratio (TMR) values in the power law Batho method for cobalt-60 and
found marked improvement by nearly 5% in the accuracy of dose
calculated within the lung. This was confirmed by numerical
comparison of the Batho expression with an analytic solution of the
primary and first-scattered radiation. In addition, Yuen and
Kornelsen(10) have shown that the
differential Batho method can give good results for circular fields
where annulus slabs of inhomogeneity are inserted. Woo et al.(11) proposed a new method of primary
scatter separation that improves the dose calculations when used
with the ETAR method. Kappas and Rosenwald(12) proposed a method that takes into
account the relative position of the point of calculation to the
inhomogeneity as well as the lateral extent of the inhomogeneity.
However, it should be mentioned that the literature concerning bulk
correction methods for inhomogeneities lying inside irregular fields
is limited.
The majority of the
RTPS offer the option of the ETAR method as the primary
inhomogeneity correction method. It has been shown, though, that the
ETAR method yields results similar to the Batho method for single
photon beams traversing geometries with inhomogeneities except with
small inhomogeneities, where the ETAR method is more accurate than
the Batho method. Moreover, there are several older RTPS still in
use that offer the option of using only conventional methods. Hence,
we propose a new approach based the principles of the Clarkson
method of scatter integration(13) and
the 3D beam subtraction method (3D-BSM),(12) which could be used for irregular
fields and any shape of inhomogeneous structures. The basis of this
new method is to use the Clarkson principle so that each sector will
be a circular field that has or does not have an inhomogeneity
embedded inside, and the point of calculation is always at the
central axis of the circular field in order to fulfill the
requirements of the conventional methods. Unlike the original
Clarkson method, in our approach we sum both primary and scatter
dose in the sector integration. Then we apply the 3D-BSM principle,
which uses a mathematical combination of on-axis conventional
correction factors, each of which is calculated according to the
actual size and position of the inhomogeneity relative to the
calculation point.
Splitting the
irradiating field into sectors according to the Clarkson principle
(for each calculation point) and applying the concept of the 3D-BSM
method of computing the dose for each sector as a sum of
"theoretical fields" leads to a "general correction factor"
(combination of the corrections of all the individual sectors).
Unlike most conventional methods, with this approach we consider (1)
the projection of the shape of the inhomogeneity to the calculation
plane and the size and position of the inhomogeneity inside the
irradiating field, (2) the position of the calculation point, and
(3) the shape of the irradiating field.
II. MATERIALS AND METHODS
A. Review of the 3D-BSM
The 3D-BSM was developed by Kappas and Rosenwald.(12) Their method is based on the Day method(1) for rectangular fields. Consider the rectangular field in the beam's-eye view (BEV) (Fig. 1). Inside the irradiating volume is a parallelepiped of inhomogeneous volume and a calculation point, P. According to this method, the algebraic distances from the point of calculation to the limits of the inhomogeneity are calculated. Then the initial rectangular field can be split into sections (theoretical beams), where each section is made from the above calculated distances so that
- each individual section contains the point P, and
- each section is made only of one specified density (either water equivalent or inhomogeneity).
Fig. 1. Case of a rectangular inhomogeneity in a rectangular field of dimensions 2x and 2y. Ui and Wj: distances from point P to the inhomogeneity limits x1, x2 and y1, y2 Ui = xi – xp, Wj = yj – yp |
||
Then the dose to the point of calculation can be expressed as a sum of these theoretical beams using the Day method. This leads to a general correction factor (GCF):
(1) |
where εij is the
sign of the dot product (Ui, Wj),(12)
Cij is the correction
factor for each of the theoretical beams created, D ow is the dose to
water for the each field of dimensions defined in the parentheses,
Ui and Wj are the algebraic distances from the
point of calculation to the inhomogeneity limits, and
Xi and Yj are the distances from the point of
calculation to the field boundaries.
The correction factor Cij could be calculated
with a standard method such as the power law Batho method, which
gives good results.(8,9)
B. 3D-BSM applied to irregular fields and irregular inhomogeneities
The proposed method is based on the principle of the 3D-BSM where the dose at a point is calculated as the algebraic summation of theoretical beams. The combination with the Clarkson method of sector integration allows the calculation of the dose at any point in the irradiating volume, taking into account the position and shape of the inhomogeneity relative to the point of calculation, as well as the shape of the field (Fig. 2). Consider Fig. 2, where the irradiating field, in BEV, is divided into angular sectors of φ degrees (the origin of the sectors is the calculation point P). Such sectors may contain a portion, all, or none of the inhomogeneity, and the dose to the point of calculation will be the sum of the contributions of all the sectors.
Fig. 2. The BSM applied for each sector for a rectangular field with blocks shaping the field |
||
The goal is to correct the dose contribution from
all sectors that include inhomogeneities. For every sector that
intersects the inhomogeneities, the distances from the point of
calculation to the proximal and distal boundaries of the
inhomogeneities are calculated. This way, concentric fields are
created that can be thought as if they are irradiating uniform media
of density ρ. The density ρ of these media is associated with the
corresponding radius of the field. For example, in Fig. 2 for the ith sector, the field of radius R1 will irradiate the homogeneous medium of
density ρ1, the field of radius
R2 will irradiate the homogeneous
medium of density ρ2, and the field of
radius R3 will irradiate the
homogeneous medium of density ρ1. For
such concentric fields, the dose can be corrected using any bulk
correction method such as the power law (Batho), RTAR, etc. From now
on, we will refer to these correction methods as "internal
correction methods" and the corresponding correction factors as
"internal correction factors," since they are applied to each sector
individually. The contribution of each such sector is proportional
to the total number of sectors, given that all sectors have the same
angle φ . For points of calculation that do not lie along the beam
central axis, we assume that the field is flat and no off-axis
ratios are considered in the calculations. For each sector one can
algebraically add/subtract the circular fields created for any given
sector in order to account for the perturbation of the dose for this
sector. The same procedure should be followed for all the sectors
until the completion of a 360° rotation. It should be clear that
more than one inhomogeneity could be included in any sector. In such
cases, the number of radii that have to be computed for these
sectors will be increased accordingly.
Two cases are presented below to
demonstrate the principles of the algorithm.
B.1 Case 1
Let us assume the rectangular field given in BEV (Fig. 2). The irradiated medium is water equivalent (density 1 g/cm3) containing an inhomogeneity slab of density ρ g/cm3 relative to water. For simplicity, assume that the inhomogeneity is included only in the ith sector. The dose from each sector to the point of calculation P can be computed using a Clarkson integration technique. The contribution of the dose from the sector i that intersects the inhomogeneity is given as
(2) |
where D(ρ,R)i is the dose contribution from the ith sector to the point of calculation P, D(ρ,R1) are D(ρ,R2) are the doses of the circular fields of radii R1 and R2 containing inhomogeneity of density ρ relative to water, D(w,R1), D(w,R2), and D(w,R3) are the doses from the circular fields of radii R1, R2, and R3 in water (w), and φ is the angle of the sector in degrees. The same can be applied for the rest of the sectors not intersecting an inhomogeneity; therefore, the total dose to the point PD(ρ,R)Total can be calculated as
(3) |
Dp(w,R)Total is the total dose from all sectors to
the same point P in water. (In Eq. (3) we assume that one sector
encloses the entire inhomogeneity.) C(ρ,R1) and C(ρ,R2) are the correction factors for the circular
fields of radius R1 and
R2, respectively, irradiating
medium of ρ density.
If we try to
generalize the above and get a GCF for the point P from all sectors,
(4) |
We can write Eq. (4) using Eq. (3) and get the GCF for this example:
(5a) |
We can write Eq. (5a) as
(5b) |
Equation (5b) is valid when the inhomogeneity is anywhere in the field and is included in one sector of φ/360°.
B.2 Case 2
Now let us assume the irregular field in Fig. 3, where there are two inhomogeneity slabs in the field. For simplicity, we assume that both inhomogeneities are intersected by only one sector. The medium is water, and the two slabs have densities ρ1 and ρ2 relative to water, respectively. The dose from each sector to the point of calculation P can be calculated as in the previous example. The contribution of the dose from the sector i that intersects both the inhomogeneous slabs is given as
(6) |
where D(ρ,R)i is the dose contribution from the ith sector to the point of calculation P, D(ρ,R4), D(ρ,R3), D(ρ,R2), and D(ρ,R1) are the doses from the circular fields of radii R1, R2, R3, and R4 containing inhomogeneity of relative density ρ1 or ρ2, and D(w,R1), D(w,R2), and D(w,R3) are the doses of the circular fields of radii R1, R2, and R3 of a water-like medium, respectively. φ is the angle of the sector in degrees.
Fig. 3. Case 2 beam’s-eye view of the BSM applied for a sector containing two inhomogeneities with the last one adjacent to the edge of the field |
||
Following the same algorithm as in the previous example and taking into account the geometry of our example, we can calculate the total dose to the point P and therefore the GCF, assuming that only one sector intersects the geometry, as
(7) |
B.3 Generalization of the correction factor
Let us consider Eqs. (5) and (7). If the inhomogeneities span over two, the second parts of Eqs. (5b) and (7) would be written as a sum of two terms so that each of them would take into account each sector. In a generalized form, the second part of Eqs. (5b) and (7) can be represented as a sum, where the number of the terms is equal to the number of the sectors that intersect the inhomogeneity. Equations (5b) and (7) can be written in a more general form for all cases:
(8) |
where n is the number of sectors
intersecting the inhomogeneities, and m is the number of
intersections between the point of calculation and the boundary of
the irradiating field due to the presence of inhomogeneities.
Di(w,Rj) and Ci(ρj,Rj) are the dose to the point of
calculation and the correction factor of that dose, respectively.
The subscript i denotes the sector, and the subscript
j denotes the number of intersections from the point of
calculation to the field boundaries. The first term in the
parentheses accounts for the medium that the dose or the correction
factor is applied for, and the second term denotes the radius of the
field. The subscript "max" in Eq. (8) is used for the largest
radius, which is the distance from the point of calculation to the
field limits, and for the density (ρ max)
of the medium that is lastly traversed from that radius.
As one can see, if the
inhomogeneity forms an "island" inside the irradiating volume (in
BEV) of water and the point of calculation is not below or inside
it, then the last term of Eq. (8) is zero because Ci(ρ max, R
max) = 1 (no correction is necessary
when the medium is water). In the example of Fig. 2 we should have
that Ci(ρ max, R max)
= 1, since the medium after the inhomogeneity is water. In this
example we also have n = 1, m = 3. Then application of
Eq. (8) yields
(9) |
The correction factors for the fields of radii
R1 and R2 correcting for the first and third
inhomogeneity are C1(ρ1,R1) and
C1(ρ3,R2). In
this case, ρ1 and ρ3 are water; therefore, the correction factors
are equal to 1. Hence, Eq. (9) becomes Eq. (5b).
For the second case of Fig. 3,
application of Eq. (8) for the point P inside a water-like medium (ρ
= 1 g/cm3), assuming that only one
sector intersects the inhomogeneities, we have n = 1, m
= 4, and ρ1, ρ2 are the densities of the inhomogeneities
relative to water, yielding again Eq. (7). From now on, we will call
the new method 3D BSM IRREG.
C. Development of test code
We implemented a test code for the method described above
in order to validate Eq. (8) in various conditions. The
implementation was made using C++. A simple user interface was
developed where the user can enter the points that define the shape
of the irregular field at source-to-axis (SAD) = 100 cm in Cartesian
coordinates having the center of the field at the point (0, 0). The
user also enters the points that determine the inhomogeneity shape
in BEV, the depths of the upper and lower surface of the
inhomogeneity, and the depth of the point of measurement. For any
given depth of the point of interest, its x- and y-coordinates are
required. Points outside the field dimensions are not considered.
The code is able to take into account the increase of the distance
from the point of measurement to the source as the point of interest
moves away from the origin (0, 0). The Batho method is the internal
correction method in our implementation (Ci(ρj,Rj)). Two sets of depth data were used: one for
a cobalt-60 machine taken from the British Journal of Radiology
tables(14) and one from measurements in
water of our department's 6-MV Philips SL75-5 LINAC. Different depth
dose data depending on the energy of the beam can be loaded by
choosing the corresponding databases.
It should be mentioned that
methods other than the Batho method of inhomogeneity correction
could be applied as internal correction factors, since these
routines are independent from the geometry routines. In the future,
the code will have the option of using more correction methods, such
as the DTAR, RTAR, etc., and more sets of dosimetric data for
various energies.
III. RESULTS
A. Benchmark simulation
First, the new method was compared against the original
BSM method with the same experimental setup (for rectangular fields)
as described in the paper by Kappas and Rosenwald,(12) and the results were in agreement with
a minor discrepancy (0.5%), which can be attributed to round-up
errors in the interpolation routines.
Then, in order to verify the
validity of Eq. (8), the setup geometry of Fig. 4 was simulated
using the BEAM-OMEGA(15) code. An MDS
Nordion Co-60 machine and the 6-MV Philips SL75-5 LINAC were
simulated in order to provide the necessary phase spaces for dose
calculations.
Fig. 4. Setup geometry for the experiment. The dimensions of the blocks when projected to the level of the point of interest are 10 cm × 10 cm. |
||
First, the phase space files for the 10 × 10 cm2 were created for both machines, and the
dose distributions were calculated in water in order to verify our
simulations against measurement data. Bremsstrahlung splitting of 20
and Russian roulette were used as global variance reduction
techniques. The ECUT value was set to 0.521 MeV, and the PCUT was
set to 0.01 MeV. For the cobalt machine, the measurement dataset
used was the one from Ref. 14. For the LINAC, the measured data were
acquired during commissioning. Agreement between measured and
calculated data was obtained within less than 1% uncertainty.
Then, using the BEAM code, the
phase-space files of a 30 × 30 cm2 were
produced at the bottom surface of the secondary collimators for both
machines. The phase-space files for the irregular fields were
created by simulating the blocks. (The same variance reduction and
the same parameters were used as for the 10 × 10 cm2 fields). The phase-space files were used
with DOSXYZ(16) to calculate the dose
distributions in two phantoms each of resolution 5 mm along the
x- and y-axes, and 10 mm along the z-axis. The
first phantom was made of water, and the second phantom had a slab
of inhomogeneity of dimensions 10 cm × 10 cm × 5 cm located 2 cm
below the surface. The inhomogeneity represented either the lung
equivalent or bone material. The results were analyzed at various
depths, inside and below the inhomogeneity, for both simulations
(lung and bone equivalent materials) and presented as correction
factor graphs along the axis. The number of histories simulated gave
a standard error of less than ±1% for all simulations.
The Batho method was used as the
internal correction method in Eq. (8). Using the developed test
code, we obtained the correction factors for the setup geometry of
Fig. 4 at various depths along the x- and y-axes for
both beam energies. Our results were compared against those from the
Monte Carlo calculations, with results obtained from the PLATO RTPS
(Nucletron B.V., Veenendaal, the Netherlands) using ETAR and with
results of the Batho not taking into account the field shape.
In Figs. 5 to 8 the correction
factors are calculated along the x- and y-axes for
both photon beam energies at the depth of 10 cm for both lung and
bone equivalent materials. The irradiating field is 30 × 30 cm2 defined at SAD = 100 cm for both 6-MV and
Co-60 photon beams. The SAD technique was used in our case; that is,
the source-to-skin distance was set to 90 cm for all cases. The
inhomogeneity was represented as a slab of dimensions 10 × 10 × 5
cm3 placed symmetrically in the central
beam axis and 2 cm under the surface. As mentioned before, the
off-axis effect is not taken into account because the beams are
assumed to be flat and symmetrical over the entire irradiating
field. The correction factors were obtained every 1 cm along both
axes. The 3D BSM IRREG method correctly predicts the correction
factor below the inhomogeneity and also at points where no
inhomogeneity lies directly above them. When moving away from the
interface and no inhomogeneity is above the calculation point, the
correction factor is affected by the presence of the inhomogeneity.
This is due to a lack or an excess of scatter from the inhomogeneity
slab to points close to the interface, depending on whether ρ is
lower or higher than 1 g/cm3. Moving
farther from the interface the phenomenon is less pronounced, and
the correction factor approximates unity. Using the ETAR as the
inhomogeneity correction algorithm in our commercial TPS, we found
that there is no significant difference in our case when compared to
the Batho method. This is in agreement with the results of du
Plessis et al.,(17) who found that the
difference between Batho and ETAR is minimal for large fields. Also,
we can see that Batho and ETAR do not predict the effect on the
correction factor when the points are close to the "interface" of
the inhomogeneity and the medium (Figs. 5 to 8). Furthermore, the
results from ETAR appear to be the same as the those of the Batho
method; the difference between the two methods is negligible. The
difference between ETAR, Batho, and the 3D BSM IRREG is about 2.5%
and is more obvious in the Co60 case (Figs. 5 and 6). Overall, we
can see that the BSM IRREG results are in very good agreement with
the Monte Carlo results; the discrepancy is less than 1.0% in most
cases, which can be attributed to the limitations of the internal
correction factor and to the Monte Carlo uncertainties.
Fig. 5. Comparison of the correction factors for lung inhomogeneity at a depth of 10 cm along the x- and y-axes for the Co- 60 30 × 30 cm2 field |
||
Fig. 6. Comparison of the correction factors for bone inhomogeneity at a depth of 10 cm along the x- and y-axes for the Co-60 30 × 30 cm2 field |
||
Fig. 7. Comparison of the correction factors for lung inhomogeneity at a depth of 10 cm along the x- and y-axes for the 6- MV 30 × 30 cm2 field |
||
Fig. 8. Comparison of the correction factors for bone inhomogeneity at a depth of 10 cm along the x- and y-axes for the 6- MV 30 × 30 cm2 field |
||
We should also note that the addition of the two blocks in the field (as shown in Fig. 4) affects the correction factor by approximately 2%, depending on the location of the point of calculation and on the extent of the inhomogeneity. The original BSM method would fail to forecast this, since it is not able to account for the irregularity of the field. ETAR and Batho also fail to forecast this phenomenon, but 3D BSM IRREG predicts the correction factor more accurately than the other methods (x-axis graphs in Figs. 5 to 8). This effect is more pronounced when considering the points in or under the inhomogeneity slab and toward the side of the field where the block is.
IV. DISCUSSION
It is generally accepted that Monte Carlo
algorithms can achieve the optimum dose calculation inside and/or in
the proximity of heterogeneity, where there is analytical
calculation of primary and scatter photons and electrons. Taking
into account the complexity of calculations for clinical photon
beams as well as the time limitation, there is only one TPS that
directly uses Monte Carlo code for clinical cases, the PEREGRINE(18) RTPS. Furthermore, according to our
knowledge, the majority of the commercial RTPS use the ETAR method
or another conventional heterogeneity correction algorithm. Thus, we
believe that there is room for improvement in the existing
conventional methods.
The original
3D-BSM method was able to accurately predict the dose perturbation
for rectangular fields when the inhomogeneity was not intersected by
the beam axis. We propose a new method that is based on the 3D-BSM
and is able to predict the correction to the dose perturbation due
to the presence of inhomogeneity for any given field shape. This is
achieved by adapting the Clarkson method of sector integration to
the 3D-BSM algebraic summation of theoretical fields. The method
cannot be applied under conditions of electronic disequilibrium,
since it is dependent on the internal correction methods (such as
the power law (Batho), etc.). It is in the same category as the
original 3D-BSM method of local deposition (no electron transport)
with the ability to use 3D density sampling.
As we can see from Figs. 5 to 8,
the ETAR and Batho methods break down at the interface where they
could administer significant clinical underdosage in the case of
lung and overdosage in the case of the bone. The proposed method
provides an alternative to computing the dose more accurately with
minimal input. It is also easily implemented.
Since 3D BSM IRREG is largely
dependent on the internal correction method, Ci(ρj,Rj), the current version using the Batho
method of inhomogeneity correction can be applied to energies of
photon beams up to 6 MV. This choice of energies was made because
the majority of treatment techniques involving heterogeneities
(e.g., lung) include low-energy beams.(19) Correction methods acceptable for
higher energies (above 6 MV) such as the improved power law (Batho)
method, which uses TPR instead of TAR or DTAR, etc., can be used and
are expected to give good results.
The results of the method are
found to be in good agreement with Monte Carlo simulations, and in
most cases the discrepancy is less than 1.5%. Measurements were made
using a solid water phantom and cork to mimic the geometry for a few
selected points, in order to verify the Monte Carlo results. Again,
the agreement was within the Monte Carlo uncertainty. The largest
discrepancies are observed at the points that are close to the
interfaces. This is mostly due to the lack of dosimetric data for
very small fields (less than 4 × 4 cm2)
and very large fields (larger than 40 × 40 cm2). Very small or very large fields are
substituted by the smallest and largest available in the dataset,
respectively.
The current
implementation of the method can compute the correction factors for
a plane of calculation at a time. For resolution of 1 × 1 mm (at the
plane of the point of interest) and for a 20 × 20 cm2 field it takes approximately 20 s on a
2-GHz PC. Further development of the code should improve the
calculation speed, introduce more "internal correction methods,"
provide more dosimetric data, and allow the user to introduce his or
her own dosimetric dataset. Also, improvement of the interpolation
routines, for fields smaller and larger than those in the dosimetric
dataset, will eliminate the source of error that appears mostly near
the inhomogeneity interface. Furthermore, improvement in the
dosimetric dataset in order to better treat the smaller field as
proposed by Woo et al.(11) would
increase the accuracy of the algorithm, especially for the cases
where small fields have to be calculated.
V. CONCLUSION
The combination of the 3D-BSM and Clarkson methods
of sector integration allows us to propose a new method, which can
generalize any conventional inhomogeneity correction method. In
fact, the appropriate choice of a bulk method (in our case, the
power law/Batho method) assures an acceptable correction for
situations where the point of calculation lies off the beam axis,
and the lateral extent of the inhomogeneity is smaller than the
field size. This approach improves the basic inhomogeneity
correction method by taking into account that (1) the point of
calculation could be anywhere in the irradiating volume, (2) the
primary arriving to the point of calculation may not be affected by
the presence of the inhomogeneity, and (3) the shape of the field
could be altered by the presence of shielding blocks. It is clear
that the correction method proposed works also for cases where the
point of interest lies under the inhomogeneity, and the lateral
extent of the inhomogeneous medium is larger than the field size,
that is, for all cases. In most of the cases investigated, the
agreement between Monte Carlo results and the results from the
computer application of the proposed method is less than 1.5%.
This new method can improve the
accuracy in the prediction of the correction factor in the presence
of irregular fields by a factor of 2% to 3% compared with the
original BSM method. That is mostly because the original method
fails to take into account the shape of the irradiating field. The
3D BSM IRREG method gives the same results as its ancestor and can
be used instead.
ACKNOWLEDGMENTS
We would like to thank the MDS-Nordion for providing the data of the cobalt machine for our Monte Carlo calculations.
REFERENCES
- Sontag MR, Cunningham JR. The equivalent tissue-air ratio method for making absorbed dose calculations in a heterogeneous medium. Radiology 1978;129:787–794.
- Engelsman M, Damen EMF, Koken PW, van t' Veld AA, van Ingen KM, Mijnher BJ. Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumors. Radiat Oncol. 2001;60:299–309.
- Kappas C, Rosenwald JC. Quality control of inhomogeneity correction algorithms used in treatment planning systems. Int J Radiat Oncol Biol Phys. 1995;15:32(3):847–858.
- Kosunen A, Jarvinen H, Vatnitskij S, et al. Intercomparison of radiotherapy treatment planning systems for external photon and electron beam dose calculations. Radiat Oncol. 1993;29(3):327–335.
- Batho MF. Lung corrections in Co-60 beam therapy. J Can Assoc Radiol. 1964;15:79–83.
- Kappas K, Rosenwald JC. Theoretical and experimental analysis of scatter from inhomogeneous slabs in a 60Co beam: The differential tissue-air ratio method (DTAR). Phys Med Biol. 1986;11:1211–1228.
- O'Connor JE. The Variation of scattered x-rays with density in an irradiated body. Phys Med Biol. 1957;1:352-369.
- El-Khatib E, Battista JJ. Improved lung dose calculation using tissue-maximum ratios in the Batho correction. Med Phys. 1984;11(3):279–286.
- Wong JW, Purdy JA. On methods of inhomogeneity corrections for photon transport. Med Phys. 1990;17(5):807–814.
- Yuen K, Kornelsen RO. Practical application of the differential Batho method for inhomogeneity correction on kerma in a photon beam. Med Phys. 1988;15(1):74–77.
- Woo MK, Cunningam JR, Jezioranski JJ. Extending the concept of primary and scatter separation to the condition of electronic equilibrium. Med Phys. 1990;17(4):558–595.
- Kappas K, Rosenwald JC. A 3-D beam subtraction method for inhomogeneity correction in high energy X-ray radiotherapy. Radiother Oncol. 1986;5:223–233.
- Clarkson JR. A note on depth dose in fields of irregular shape. Br J Radiol. 1941;14:265.
- Central axis depth dose data for use in radiotherapy. Br J Radiol. 1983;Suppl 11.
- Rogers DWO, Faddegon BA, Ding GX, Ma C-M, Wei J, Mackie TR. BEAM: A Monte Carlo code to simulate radiotherapy treatment units. Med Phys. 1995;22:503–524.
- Walters BRB, Rogers DWO. DOSXYZnrc Users manual. Ionizing radiation standards. National Research Council of Canada, Ottawa, NRCC Report PIRS-7942; 2003.
- du Plessis FC, Willemse CA, Lotter MG, Goedhals L. Comparison of the Batho, ETAR and Monte Carlo dose calculation methods in CT based patient models. Med Phys. 2001;28:582–589.
- Hartmann Siantar CL, Walling RS, Daly TP, et al. Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom. Med Phys. 2001;28:1322–1337.
- Yorke E, Harisiadis L, Wessels B, Aghdam H, Altemus R. Dosimetric considerations in radiation therapy of coin lesions of the lung. Int J Radiat Oncol Biol Phys. 1996;34:481–487.
© 2005 Am. Coll. Med. Phys.