Comparison of six phantoms for
entrance skin dose evaluation in
11 standard X-ray examinations
Gaetano Compagnone, Laura Pagan, and Carlo Bergamini
Medical Physics Department, S. Orsola-Malpighi Hospital, Via Massarenti 9, 40138 Bologna, Italy
gcompa@orsola-malpighi.med.unibo.it
(Received 1 July 2004; accepted 14 September 2004)Entrance skin dose (ESD) is an important parameter for assessing the dose received by a patient in a single radiographic exposure. The most useful way to evaluate ESD is either by direct measurement on phantoms using an ionization chamber or using calculations based on a mathematical model. We compared six phantoms (three anthropomorphic, two physical, and one mathematical) in 11 standard clinical examinations (anterior-posterior (AP) abdomen, posterior-anterior (PA) chest, AP chest, lateral (LAT) chest, AP lumbar spine, LAT lumbar spine, LAT lumbo-sacral joint, AP pelvis, PA skull, LAT skull, and AP urinary tract) for two reasons: to determine the conversion factors to use for ESDs measured on different phantoms and to validate the mathematical model used. First, a comparison was done between the three anthropomorphic phantoms (Alderson Rando, chest RSD-77SPL, and 3M skull) and the two physical phantoms (Uniform and AAPM 31); for each examination we obtained "relative entrance skin dose factors." Second, we compared these five phantoms with the mathematical phantom: the overall accuracy of the model was better than 14%. Total mathematical model and total ionization chamber uncertainties, calculated by quadratic propagation of errors of the single components, were estimated to be on the order of ±12% and ±3%, respectively. To reduce the most significant source of uncertainty, the overall accuracy of the model was recalculated using new backscatter factors. The overall accuracy of the model improved: better than 12%. For each examination an anthropomorphic phantom was considered as the gold standard relative to the physical phantoms. In this way, it was possible to analyze the variations in phantom design and characteristics. Finally, the mathematical model was validated by more than 400 measurements taken on different phantoms and using a variety of radiological equipment. We conclude that the mathematical model can be used satisfactorily in ESD evaluations because it optimizes available resources, it is based on direct measurements, and it is an easy dynamic tool.
PACS number(s): 87.66.Xa
Key words: anthropomorphic phantom, diagnostic radiology, entrance dose measurements, entrance skin exposure, phantom study
I. INTRODUCTION
Entrance skin dose (ESD) is an important parameter
in assessing the dose received by a patient in a single radiographic
exposure. The European Union has identified this physical quantity
as one to be monitored as a diagnostic reference level in the hopes
of optimizing patient dose.(1,2) It is
possible to evaluate ESD either by direct measurements (on suitable
phantoms using ionization chambers or on patients using
thermoluminescent dosimeters, TLDs) or using mathematical model
calculations based on the X-ray tube output.(1) Using TLDs is time-consuming in large
hospitals. Therefore, in this paper ESDs were evaluated using both
measurements taken by ionization chambers and values calculated by a
mathematical model; this allowed us to study the accuracies inherent
in different experimental setups. To evaluate the ESD, it is
necessary to use "standard phantoms,"(1,3) and it is important to know the
difference between them because some may be bought commercially or
"home-made."(4,5) There is no advice in
the literature to help with phantom selection in different clinical
situations.(6) Therefore, two or more
similar phantoms are often available in medical physics departments
for dosimetric measurements in conventional radiology. Moreover,
these phantoms are not always available simultaneously (for
instance, one phantom may be being used by someone else); it is also
necessary to have conversion factors between different
phantoms.
This paper reports a
comparison between ESDs measured by five phantoms in 11 standard
clinical examinations (anterior-posterior (AP) abdomen,
posterior-anterior (PA) chest, AP chest, lateral (LAT) chest, AP
lumbar spine, LAT lumbar spine, LAT lumbo-sacral joint, AP pelvis,
PA skull, LAT skull, and AP urinary tract) in order to have
"relative ESD factors" (REFs) between each phantom and the others.
These REFs can be used in a normal dosimetry routine where the ESD
measurements have been done using different phantoms for the same
kind of examination, performed either after long time intervals with
the same radiological system (one X-ray tube and one generator in a
well-identified radiological room) or using similar clinical
technique factors with different radiological equipments (different
X-ray tubes and generators but similar types of radiological
apparatus). In addition, a comparison with ESDs calculated by a
mathematical model (which can be considered as a sixth phantom) is
made, because this is another possibility-easier but less
accurate-for this kind of evaluation.
II. MATERIALS AND METHODS
As explained by Moores,(7) the phantoms for dose assessment can be anthropomorphic (they possess aspects of anatomical structure), physical (they do not attempt to reproduce anatomical details directly and may range from a single block of material to more sophisticated structures), or mathematical (they may be simple mathematical models to represent the interaction of X-ray beams with biological tissue in order to assess ESD or other dosimetric parameters). In this study, we first compared three anthropomorphic and two physical phantoms. Then these five phantoms were compared with one mathematical phantom, as described below.
A. Anthropomorphic and physical phantom comparison
The standard examinations, clinical technique factors used (with respective standard deviations), and number of both radiographic systems and measurements considered in this study are shown in Table 1. The technique factors reported for each examination are the following: the average kilovolt peak (kVp), milliampere × seconds (mAs), and focus-to-phantom surface distance (FSD) values used clinically in our hospital. Nearly all the X-ray generators were three-phase (6 or 12 pulse) models or high-frequency generators. All radiographic systems were controlled by an ISO 9001-2000 certified quality assurance program, which provides good equipment performance according to acceptance, status, and constancy tests. In particular, to control the reproducibility and the linearity of the X-ray tube output, three measurements free-in-air at two different kVp values and at five different mAs settings (total: 30 output measurements, 15 at 80 kVp and 15 at 100 kVp, for each X-ray tube) were taken. These measurements were taken yearly or when a tube was replaced. The acceptability limit for both reproducibility and linearity was 10%.
TABLE 1. The 11 examinations, averages, and standard deviations of technique factors, number of radiological systems, and number of measurements considered in this study |
The direct ESD
measurements were made on five different phantoms, the first three
anthropomorphic (Fig. 1): Alderson Rando (ALD); chest RSD-77SPL
(CHE), Radiology Support Device, Long Beach, USA; skull 3M (SKU);
uniform (UNI), 25 cm × 25 cm × 20 cm polymethyl methacrylate (PMMA);
American Association of Physicists in Medicine (AAPM) phantom as
described in AAPM Report No. 31(A31),(8) which combines PMMA with aluminum sheets
and air gaps, in order to simulate various anatomical parts.
FIG. 1. The five anthropomorphic and physical phantoms used in this study: (a) Alderson Rando; (b) chest RSD–77SPL and uniform; (c) skull 3M and AAPM31. |
||
For each measurement,
the phantoms were positioned as a standard patient, and for each
phantom the same technical parameters (kVp, mAs, FSD, and X-ray
field size as indicated by senior radiologists) were selected. The
ESDs were measured as indicated in other papers,(9) by positioning at the surface of each
phantom on the beam central axis an ionization chamber model 90X6-6
connected to a Radiation Monitor Controller model 9010 (Radcal
Corporation,
Monrovia, CA).
The
experimental setup was chosen so that the variables possibly
affecting the results were carefully controlled: the detector did
not significantly perturb the photon fluence on the phantom surface
underneath it; the cross-sectional area of the detector was
significantly less than the area of the irradiated phantom; and the
FSD was measured by taking into account the position of the focal
spot inside the X-ray tube.
All
instruments are calibrated yearly, with the calibration traceable to
an SIT (National Calibration Service in Italy) center. To verify kVp
accuracy, direct measurements during the exposures were taken with a
noninvasive kVp meter model Mult-O-Meter 510 (Unfors Instruments,
Billdal, Sweden).
B. Mathematical versus anthropomorphic and physical phantom comparison
Every ESD
measured on anthropomorphic and physical phantoms was compared with
the corresponding ESD calculated with the mathematical
phantom.
To determine the output K of a diagnostic X-ray tube (in terms of absorbed dose to
air or exposure free-in-air), many mathematical models have been
suggested.(10-14) In this study, the
model proposed by Harpen(14) was
adopted:
(1) |
where parameters α and β depend on the type of
X-ray generator, anode material, FSD, and X-ray tube
filtration.
Equation (1) gives K as a function of kVp and mAs, by taking only two X-ray tube
output measurements at two different voltages. To this end, the
average of the 15 measurements taken at 80 kVp and the average
of the 15 measurements taken at 100 kVp during the
reproducibility and linearity quality controls mentioned above were
used as the two output
values.
Harpen's formula gives the
absorbed dose to air, free-in-air; therefore, to determine ESD, some
corrections must be made for backscatter factors (BSF). European
guidelines(1) propose a simple and
generic value for conventional radiography:
1.35.
For each phantom j and
examination k, the integral accuracy
Aphantom j of the mathematical model
relative to a single phantom and the differential accuracy
Asinglepoint relative to a single measurement were
calculated. The Aphantom j relative to
examination k is defined to be
(2) |
where Nj is the number of
measurements made with the phantom j in each examination
k, MOD is the ESD value calculated by the mathematical model,
and PHA is the ESD value measured on the phantom j. In this
way, it was possible to calculate AALD,
ACHE, ASKU,
AUNI, and AA31 for each specific
phantom and examination.
The Asinglepoint is defined by the equation
(3) |
This is analogous to but slightly different than
Eq. (2): the numerator of Eq. (2) uses the absolute value
of the difference in the ESDs, whereas the numerator in Eq. (3)
does not.
To achieve better
accuracy of the mathematical model, for each examination every MOD
value was recalculated using the more accurate BSF values given in
Harrison(15) and Grosswendt(16) (see Table 2 for examples of
typical values used), obtaining new integral accuracies NEWAALD,
NEWACHE,
NEWASKU,
NEWAUNI, and
NEWAA31 for each
phantom.
The "overall accuracy of
the mathematical model" AOVE relative to all the
phantoms and all the examinations was also calculated. It is defined
to be
(4) |
where NjTOT is the number of measurements made with the phantom j in all examinations, and M is the number of examinations in which the phantom j was used. When the more accurate BSF values are used, a new overall accuracy NEWAOVE is obtained.
TABLE 2. Typical measured values of ESDs at different phantoms, values obtained from the mathematical model and typical α, β, and backscatter factors values |
III. RESULTS
All the factors presented in this study are
"relative" values. Therefore, to obtain a general idea and a
comparison of the "absolute" values, consider Table 2. The
typical measured values of ESDs at different typical anthropomorphic
phantoms, the values obtained from the mathematical model (using the
BSF values shown), and typical values of α and β of Eq. (1) are
listed in Table 2 based on values in Table 1. The value of
K depends linearly on the α value, which is related to a
clinical technique, in particular, to the inverse of FSD, and
exponentially on parameter β , which is more sensitive to
intermachine variations and has almost a constant
value.
The mean values of REFs
measured on anthropomorphic and physical phantoms are given in
Table 3; they too are based on Table 1. For each
measurement the ratio between the ESD measured on Phantom1 to the
ESD measured on the Phantom2 (in the columns Phantom1/Phantom2) was
calculated. For each examination, the averages of the ratios taken
with all radiological systems used are shown in Table 3. When a
phantom is not appropriate for an examination (e.g., a chest phantom
for an AP abdomen examination), N/A is indicated.
The reproducibility and the linearity
of the output were better than 10% (typically, about 5%) for all
radiographic systems. The kVp accuracy was ±3% for all
measurements.
In Table 4 (fourth
column) the accuracies AALD,
ACHE, ASKU,
AUNI, and AA31 of the
mathematical model relative to the other phantoms are given, based
on Table 1. In this case, the BSF value used is 1.35. The
averages of all accuracies are also provided to show how different
phantoms compare in general, even though not all phantoms are
suitable for all examinations.
The
AOVE is better than 14%. This value is of the same
order reported in other works.(11,13,14) Figures 2(a) to (e)
show the accuracies Asinglepoint of the
mathematical model relative to the other phantoms plotted as a
function of the kVp used. Using Eq. (3) to calculate this
parameter should better visualize and take into account the
displacements around the 0 value (which would mean perfect
correspondence between phantoms).
TABLE 3. REF (relative ESD factors). The factors are the averages of the ratios of the ESDs measured on different phantoms based on Table 1. |
| N/A means that the phantom in the denominator is not appropriate for that examination. |
TABLE 4. Accuracies of the mathematical model relative to the other phantoms based on Table 1: AALD, ACHE, ASKU, AUNI, and AA31 with BSF = 1.35; NEWAALD, NEWACHE, NEWASKU, NEWAUNI, and NEWA31 recalculated using the BSF from previous works(15,16) |
FIG. 2. Mathematical model accuracy Asinglepoint relative to the anthropomorphic and physical phantoms: (a) Alderson Rando; (b) chest RSD-77SPL; (c) 3M skull; (d) uniform; (e) AAPM 31. |
||
Total mathematical
model uncertainty, estimated on the order of ±12%, was calculated by
quadratic propagation of errors of the single model components, that
is, using only one BSF and the possible variation in X-ray tube
output owing to the time elapsed from the latter quality control.
Total ionization chamber uncertainty, calculated in the same way,
was estimated to be of the order of ±3% due to both the calibration
factor given by the SIT center and the experimental FSD
measurements. The use of only one BSF value was the most significant
source of error. The idea of reducing it to a wider set of BSF
values has been adopted.
Many
papers have been published regarding the calculated or measured
BSF.(15-21) Listed in the fifth column
of Table 4 are the new accuracies and their averages relative
to the anthropomorphic and physical phantoms
(NEWAALD,
NEWACHE,
NEWASKU,
NEWAUNI, and
NEWAA31) recalculated using the BSF
from Harrison(15) and Grosswendt,(16) which are listed as a function of both
the field size in the various examinations and the kVp values used.
The NEWAOVE has now improved, better
than 12%.
IV. DISCUSSION
The backscatter factor for a simple water phantom can be written as
(5) |
where X(w) is the exposure at the
surface of the water phantom, X(free) is the
exposure at the same point in space without the phantom, and ,
are
the ratios of the mass-energy transfer coefficients for water and
air in the presence of scatter medium and in free space,
respectively.
In Eq. (5), the
ratios of the energy transfer coefficients are for the same media,
but they cannot be cancelled because they are determined for
different photon spectra: is averaged over
the spectral energy fluence distribution of the beam at the phantom
surface, and
is averaged over the spectral energy fluence
distribution of the primary beam without the phantom. If phantoms
different from water are used, an analogous relationship of course
can be applied. These theoretical considerations can explain why the
backscatter factors and the ESDs are very strongly dependent on the
material, shape, and size of the different
phantoms.
Three of six phantoms
used in this study are anthropomorphic. The ALD phantom is used more
in radiotherapy, but it can be used at diagnostic energies as
well.(22) The CHE and SKU phantoms are
more useful in radiology because they are optimized for image
quality studies. The two physical phantoms can be more easily
available than the anthropomorphic phantoms, but the latter better
simulates the full scatter properties of human tissue. The ALD, CHE,
and SKU phantoms have soft-tissue-equivalent material, which almost
exactly duplicates water (in radio-absorptive and scatter
properties), and synthetic skeletons with cortical bone, which is
radio-equivalent to natural bone and matches the volumetric electron
densities and the mass attenuation coefficients of ICRU-44(23) across the entire energy range of
diagnostic energies. It is therefore interesting to select for every
examination an anthropomorphic phantom to be used as a "gold
standard" relative to the physical phantoms and to analyze the
physical versus anthropomorphic phantoms
comparisons.
The gold standard for
the PA chest, AP chest, and LAT chest is the CHE phantom. In every
measurement it gives lower ESDs than physical phantoms. A possible
explanation for this result is that the backscattered radiation is
less when lungs are involved, and the nonanthropomorphic phantoms
are not able to take into account this physical process. Moreover,
when automatic exposure controls (AECs) are used, the phantoms can
have a significantly different "equivalent thickness"(24) and, therefore, a different ESD value.
In AEC systems we investigated, there are usually three detectors
located as the vertices of a triangle. Every detector can be used
alone or together with the others, depending on the examination. For
example, in PA chest the two lateral detectors are used; in LAT
chest the central detector is used; and in AP pelvis all three
detectors are usually used. Therefore, in chest examinations, if in
AEC systems lung-centered detectors are used, the physical phantoms,
which are soft-tissue-equiva-lent in terms of primary transmission,
will transmit less radiation to a "lung-field detector" than would
anthropomorphic phantoms, and the AEC will terminate exposure too
late. This effect will lead to (ESD physical phantoms)/(ESD CHE)
ratios being higher than one for chest examinations in
Table 3.
The reference for PA
skull and LAT skull is the SKU phantom. In every measurement it
gives lower ESDs than physical phantoms. In Figs. 2(a) to(e) at
lower kVp values there is a large spread of the Asinglepoint values because the dependence of ESDs
on different atomic numbers of phantoms is stressed. The kVp values
used in clinical practice for skull examinations are the lowest in
all examinations considered. At these low energies the interactions
of photons in the phantoms are modulated differently by the atomic
numbers of soft tissue and bone. In the case of soft tissue, the
Compton effect predominates; in the case of bone, the Compton and
photoelectric effects are approximately equivalent.(25) For this reason, the backscattered
radiation is more in the soft tissue than in the bone. Therefore,
the ESD in physical phantoms (almost entirely made in PMMA) is
higher than in SKU phantoms (made of rubber and bone equivalent).
Another factor that affects the different responses is this: a
rounded surface such as that on the SKU phantom reduces the amount
of scattered radiation, contributing to
ESD.
For the remaining examinations
(AP abdomen, AP lumbar spine, LAT lumbar spine, LAT lumbo-sacral
joint, AP pelvis, and AP urinary tract), the gold standard is the
Alderson Rando phantom: it gives (except two cases) higher ESDs than
physical phantoms. A possible explanation is that in these
examinations, where the energies are higher, the more accurate
composition of anthropomorphic phantoms takes better into account
the higher backscatter radiation. Between the two physical phantoms,
A31 is slightly better because of its structure, more sophisticated
relatively to the simpler uniform
phantom.
Note in Table 3 that,
in general, for every couple of phantoms, the REFs Phantom1/Phantom2
are not always equal to the inverse of the corresponding REFs
Phantom2/Phantom1. The explanation is that the REFs reported in
Table 3 are averages of ratios of ESDs measured on many
different radiological systems, where the same kind of examination
is performed with different technique factors, and the phantom's
response is nonlinear at different energies. From a strictly
theoretical point of view, this is a limitation because all the
examinations should be performed with a clinical protocol that gives
exactly the same ESD to the standard patient. However, from a more
practical and realistic point of view, that aim is seldom reached
clinically because it is dependent on radiological equipment
characteristics as well as the working procedures of the operators.
Therefore, the above reported "theoretical limitation" becomes an
interesting "working hypothesis." The number of radiological
systems, reported in Table 1, includes a wide range of
equipment typologies used in clinical practice, and, together with
the high number of measurements made, it should give a consistent
reliability in all operative clinical conditions to comparisons
presented in this paper.
Since the
individual radiographic systems had reproducibility and linearity
values better than 5%, the variability of ESD and then of REF values
is most likely attributable to variations in phantom material,
design, and characteristics, which are exactly the distinctive
features to study.
As far as the
mathematical phantom is concerned, the model has been validated by
more than 400 ESD calculations compared with measurements taken on
different phantoms and radiological equipment. An overall accuracy
better than 12% was obtained, comparable with data reported in
literature. Therefore, in ESD calculations, this model can be used
satisfactorily because it presents many advantages. First, it allows
the optimization of available resources by using data already taken
during quality controls. Second, the ESDs are calculated
theoretically with a general formula but are also based on direct
measurements, so that every radiological system keeps its own
characteristics. Finally, the model is very easy to implement using
an electronic spreadsheet and to use as a dynamic tool.
V. CONCLUSIONS
It is difficult to make comparisons between patient
dose results in different studies because different measuring
phantoms are used. In this respect, the comparison performed in the
present study could be useful in order to have data comparable with
those taken both at different times and with different phantoms, in
the same or in other radiology departments. In this way, the phantom
used most often for ESD measurements in diagnostic radiology-the
uniform phantom can be compared with other more sophisticated but
less available phantoms, and each of them can be compared with a
simple and easy-to-use mathematical
model.
This study shows that in
measuring ESD values, the phantoms are not as "standard" as the
medical physicist wishes, but that it is possible to take into
account the relative differences in order to have more comparable
and consistent data.
ACKNOWLEDGMENTS
We wish to express our thanks to Marina Benati and Luciano Selleri for their assistance in data collection.
REFERENCES
- European Commission. European guidelines on quality criteria for diagnostic radiographic images. EUR 16260 EN. Brussels; 1996.
- Muhogora WE, Nyanda AM, Kazema RR. Experiences with the European guidelines on quality criteria for radiographic images in Tanzania. J Appl Clin Med Phys. 2001;2(4):219-226.
- Conway BJ, Butler PF, Duff JE, et al. Beam quality independent attenuation phantom for estimating patient exposure from x-ray automatic exposure controlled chest examinations. Med Phys. 1984;11(6):827-832.
- Havukainen R, Pirinen M. Patient dose and image quality in five standard x-ray examinations. Med Phys. 1993; 20(3):813-817.
- Servomaa A, Tapiovaara M. Patient equivalent phantoms in chest radiography. In Moores BM, Stieve FE, Eriskat H, Schibilla H, editors. BIR Report 18-Technical and physical parameters for quality assurance in medical diagnostic radiology. London: British Institute of Radiology; 1989: 65-67.
- Pernicka F, Carlsson GA, Dance DR, DeWerd LA, Kramer HM, Ng KH. Development of an international code of practice for dosimetry in x-ray diagnostic radiology. In Proceedings of an international conference-Radiological protection of patients in diagnostic and interventional radiology, nuclear medicine and radiotherapy. Malaga, March 2001. Vienna: IAEA; 2001: 93-98.
- Moores BM. The role of phantoms in standardisation of the radiological process. Radiat Prot Dosim. 1993;49(1/3): 19-26.
- American Association of Physicists in Medicine (AAPM). Standardized methods for measuring diagnostic X-ray exposures. AAPM Report no. 31. New York; 1990.
- Mah E, Samei E, Peck DJ. Evaluation of a quality control phantom for digital chest radiography. J Appl Clin Med Phys. 2001;2(2):90-101.
- McGuire EL, Dickson PA. Exposure and organ dose estimation in diagnostic radiology. Med Phys. 1986;13(6): 913-916.
- Zamenhof RG, Shahabi S, Morgan HT. An improved method for estimating the entrance exposure in diagnostic radiographic examinations. AJR. 1987;149:631-637.
- David G, Price SC. Performing JCAHO dose calculations with the aid of a microcomputer spreadsheet program. AJR. 1990;154:863-866.
- Société Française des Physiciens d'Hôpital (SFPH). Evaluation of radiation dose during radiodiagnostic examinations. J Radiol. 1991;72(8-9):403-420.
- Harpen MD. A mathematical spread sheet application for production of entrance skin exposure nomograms. Med Phys. 1996;23(2):241-242.
- Harrison RM. Backscatter factor for diagnostic radiology (1-4 mm Al HVL). Phys Med Biol. 1982;27(12): 1465-1474.
- Grosswendt B. Dependence of the photon backscatter factor for water on source-to-phantom distance and irradiation field size. Phys Med Biol. 1990;35(9):1233-1245.
- Stanton L, Brattelli SD, Day JL. Measurements of diagnostic x-ray backscatter by a novel ion chamber method. Med Phys. 1982;9(1):121-130.
- The Institute of Physical Sciences in Medicine (IPSM). Patient dosimetry techniques in diagnostic radiology. IPSM Report No. 53. York; 1988.
- Klevenhagen SC. Experimentally determined backscatter factors for x-rays generated at voltages between 16 and 140 kV. Phys Med Biol. 1989;34(12):1871-1882.
- Tomljenovic II, Ninkovic MM, Bek-Uzarov D, Stankovic SJ, Kovacevic M. Water phantom backscatter factors for x-rays in the 60 kV to 300 kV region. Phys Med Biol. 1999;44(9):2193-2200.
- Behrman RH. The impact of increased Al filtration on x-ray tube loading and image quality in diagnostic radiology. Med Phys. 2003;30(1):69-78.
- Shrimpton PC, Wall BF, Fisher ES. The tissue-equivalence of the Alderson Rando anthropomorphic phantom for x-rays of diagnostic qualities. Phys Med Biol. 1981;26(1):133-139.
- International Commission on Radiological Units and Measurements (ICRU). Tissue substitutes in radiation dosimetry and measurement. ICRU Report no. 44. Bethesda; 1989.
- Conway BJ, Duff JE, Fewell TR, Jennings RJ, Rothenberg LN, Fleischmann RC. A patient-equivalent attenuation phantom for estimating patient exposure from automatic exposure controlled x-ray examinations of the abdomen and lumbo-sacral spine. Med Phys. 1990;17(3):448-453.
- Dendy PP, Heaton B. Physics for diagnostic radiology. 2nd ed. Bristol: Institute of Physics Publishing; 1999.
© 2005 Am. Coll. Med. Phys.