Limited accuracy of dose
calculation for large fields at
deep depths using the BrainSCAN
v5.21 treatment-planning system
Wen C. Hsi, Yunkai Zhang, Michael C. Kirk, Damian Bernard, and James C.H. Chu
Departments of Medical Physics and Radiation Oncology, Rush University Medical Center, 1653 West Congress Parkway, Chicago, Illinois 60612 U.S.A.
whsi@rush.eduReceived 24 May 2004; accepted 20 October 2004
The Varian 120 multileaf collimator (MLC) has a leaf thickness of 5 mm projected at the isocenter plane and can deliver a radiation beam of large field size (up to 30 cm) to be used in intensity-modulated radiotherapy (IMRT). Often the dose must be delivered to depths greater than 20 cm. Therefore, during the commissioning of the BrainSCAN v5.21 or any radiation treatment-planning (RTP) systems, extensive testing of dose and monitor unit calculations must encompass the field sizes (1 cm to 30 cm) and the prescription depths (1 cm to 20 cm). Accordingly, the central-axis percent depth doses (PDDs) and off-axis percentage profiles must be measured at several depths for various field sizes. The data for this study were acquired with a 6-MV X-ray beam from a Varian 2100EX LINAC with a water phantom at a source-to-surface distance (SSD) of 100 cm. These measurements were also used to generate a photon beam module, based on a photon pencil beam dose-calculation algorithm with a fast-Fourier transform method. To commission the photon beam module used in our BrainSCAN RTP system, we performed a quantitative comparison of measured and calculated central-axis depth doses and off-axis profiles. Utilizing the principles of dose difference and distance-to-agreement introduced by Van Dyk et al. [Commissioning and quality assurance of treatment planning computers. Int J Radiat Oncol Biol Phys. 1993; 26:261--273], agreements between calculated and measured doses are <2% and <2 mm for the regions of low- and high-dose gradients, respectively. However, large errors (up to ~5% and ~7% for 20-cm and 30-cm fields, respectively, at the depth 20 cm) were observed for monitor unit calculations. For a given field size, the disagreement increased with the depth. Similarly, for a given depth the disagreement also increased with the field size. These large systematic errors were caused by using the tissue maximum ratio (TMR) in BrainSCAN v5.21 without considering increased field size as depth increased. These errors have been reported to BrainLAB.
PACS number: 87.53.-j
Key words: radiation treatment-planning system, BrainSCAN, BrainLAB, quality assurance
I. INTRODUCTION
The commercial BrainSCAN v5.21 (BrainLAB AG,
Munich, Germany) radiation treatment-planning (RTP) system has been
routinely used in association with the BrainLAB micro multileaf
collimator (MLC) (3 mm minimum leaf width and 10 cm maximum leaf
position projected at isocenter) for treatment planning with narrow
radiation beams (typically 9 cm field size and 20 cm patient
thickness) for intensity-modulated radiosurgery (IMRS).(1) With a Varian 120 MLC (5 mm minimum leaf
width and 14.5 cm maximum leaf position project at isocenter) in the
6-MV X-ray beam of a Varian 2100EX LINAC (Varian Medical Systems,
Palo-Alto, CA), radiation beams of large field sizes (up to 30 × 40
cm2) can be used in intensity-modulated
radiotherapy (IMRT) or 3D radiation therapy for delivering dose to
prescription points at depths greater than 20 cm. Therefore, during
commissioning of the BrainSCAN v5.21 RTP system for treatment
planning with a Varian 120 MLC, extensive dosimetry data for various
field sizes (1 cm to 30 cm) and prescription depths (1 cm to 20 cm)
in a water phantom were acquired to ensure the accuracy of the dose
and monitor unit (MU) calculations for the BrainSCAN RTP
system.
The central-axis percent
depth doses (PDD) and off-axis profiles were measured with a water
phantom at a source-to-surface distance (SSD) of 100 cm.
Quantitative comparisons between the measured and calculated
distributions were performed using the concept of dose difference
and distance-to-agreement criteria (<2% and <2 mm for the
regions of low- and high-dose gradients, respectively), introduced
by Van Dyk et al.(2) The MUs calculated
from the BrainSCAN RTP system for delivering 100 cGy to various
prescription depths at 100 cm source-to-axis distance (SAD) with
different field sizes were compared with extracted values obtained
from measured output factors. Measured output factor is defined as
the ratio of the chamber reading of a given field size and depth to
the reading of a 10 × 10 cm2 field and
a depth at dmax with the same
delivered MUs at 100 cm SAD setup for both measurements. Our LINAC
was calibrated to deliver 100 cGy per 100 MUs to the depth of
dmax in water of a 10 × 10 cm2 field. Therefore, to deliver 100 cGy to a
prescription point with a defined field size and depth, the required
MU value is equal to 100 divided by the measured output factor for
the defined field size and
depth.
The results of quantitative
comparisons of percent depth doses, off-axis profiles, and MU
calculations are presented. Systematic errors in the dose and MU
calculations were analyzed to identify the implementation error in
the BrainSCAN dose calculation algorithm.
II. MATERIALS AND METHODS
A. Beam data measurements
With the Advanced Radiation Measurements,
Inc., water phantom scanner, the central-axis PDDs of 1 cm, 2 cm, 4
cm, 6 cm, 8 cm, 10 cm, 14 cm, 20 cm, and 30 cm square fields at 100
cm SSD were measured at depths up to 28 cm with a water tank phantom
(60 × 60 × 50 cm3) in a 6-MV photon
beam. The field size was defined with the same opening for the MLCs
and secondary collimator jaws. The field size for the MLC or the jaw
was calibrated against the light field. A PTW Freiburg sealed
chamber and an Exradin A16 micro chamber were used for the
measurements. To reduce the influence of detector size(3) for small fields, the PDDs for fields
10 cm were measured with the micro chamber. To avoid stem/cable
effects for large fields, the PDDs for fields >10 cm were
measured with the sealed
chamber.
With the same phantom
setup, the transverse off-axis percentage profiles for 6-cm and
20-cm square fields at depths of 1.4 cm, 5 cm, 10 cm, and 20 cm in
directions along and perpendicular to the leaf movement of the MLC
were measured. Each field size in the profile measurements was
defined by the MLC, while the jaws were set at 24 cm. Similar to the
PDD measurements, the smaller and larger fields were measured with
micro and sealed chambers, respectively. To obtain the radial factor
function used to correct the off-axis dependence of photon fluence,
diagonal cross-field profiles of a 40-cm square field, defined by
the jaws with retracted MLC, were measured with the sealed chamber
at depths of 0.5 cm, 1.4 cm, 2.5 cm, 5 cm, 10 cm, 20 cm, and 30
cm.
The output calibration of the
LINAC, that is, the delivered dose (in units of cGy) at the depth of dmax in water per MU by a
radiation of 10 × 10 cm2 field size,
was performed according to the AAPM TG-51 protocol. Total scatter
factors, including the components of the collimator and phantom, at
a depth of 5 cm with the water phantom at 100 cm SSD, were measured
for various square fields defined by the MLC or jaws. Measurements
for square fields 6 cm were measured with the micro chamber, while
measurements for fields >6 cm were measured with the sealed
chamber. To evaluate the MU calculation from the BrainSCAN RTP
system, the output factors were measured with a water tank (60 × 60
× 50 cm3) and a solid water phantom
(Gammex RMI 457, 30-cm thick of 30 × 30 cm2 slabs) for square fields of 1 cm, 2 cm, 4
cm, and 6 cm at depths of 1.4 cm, 5 cm, 10 cm, and 20 cm. Although
the physical density of this solid water phantom is 1.04 g/cm3, its electron density is only 1.011 ±
0.005 g/ cm3 with respect to water.
Since the 6-MV beams are mainly attenuated through Compton
scattering, which is proportional to the electron density of the
used material, the effective depth correction for the solid water
phantom with respect to the water phantom should be small.
B. Dose-calculation algorithm
The photon dose-calculation algorithm used in the BrainSCAN v5.21 RTP system is a pencil beam dose computation algorithm, which was introduced(4) and developed(5) by Mohan et al. and implemented by BrainLAB. For a collimator with an arbitrary shape, the ideal dose distribution, IDD(x,y,d), at depth d in an infinite water phantom is calculated by a 2D convolution of the polyenergetic pencil beam kernel, Dp(x - x', y - y', d), and the photon fluence, xxphi(x', y', d):
(1) |
Since Dp(x - x', y -
y', d) is based on a generic spectrum, an optional source
function correction is applied to take into account the
machine-specific adaptations (such as the influence of the finite
size of the source, collimator, and flattening filter scatter,
curvature of the leaf ends, and other smearing effects). A radial
factor function is applied on the photon fluence to account for the
off-axis dependence. The 2D integration is then calculated using
Fourier transform algorithms.
Dose
to a point inside a patient by a shaped beam is calculated from
(2) |
where rrad is the
radiological path length of the beam from the tissue surface to the
calculation point, corrected for tissue density inhomogeneities.
BrainSCAN precalculates the IDD(x,y,d) of each individual
field at the source isocenter distance (SID) in water. The IDD is
dependent on the rrad and lateral
distances at ; is defined similarly. The depth dependence
of the calculated dose is then recalibrated with respect to the
measured beam data by St x TMRPB, , where the total scatter factor
St describes the relative output
factor, and the tissue maximum ratio, TMRPB, is calculated from the measured percentage
depth doses at 100 cm SSD. Since mainly narrow fields are considered
in the BrainSCAN RTP software for IMRS treatment, TMRPB is independent of SSD and can be obtained
by the following transformation equation:
(3) |
where c is the collimator opening, d is the depth, and dcal is the calibration depth of the nominal output factor (MNlin), which is used for obtaining the absolute delivered dose. An adaptive grid algorithm is used to accelerate the dose calculations with a coarse grid in the regions of smooth dose distributions and a fine grid in the regions where the dose distributions are inhomogeneous, such as in the penumbra region.
III. RESULTS
In comparing off-axis percentage profiles, the
calculated field widths (at 50% of the central axis value) for the
6-cm and 14-cm fields at various depths agreed within 1 mm with the
measured widths. The percentage differences between the measured and
calculated values are <2% and <4% of the dmax value for the in-field and the
out-of-field regions, respectively. The differences between the
calculated and measured penumbra width (defined at 20% to 80%) are
<2 mm. The calculated penumbra is steeper and more symmetric at
the 50% values than that measured. The BrainSCAN RTP system used a
steeper penumbra to compensate for the limited spatial resolution of
the scanning ionization chamber on the scanned profiles. This may
not alter the dose uniformity inside the planning target volume
(PTV) and the dose gradient at the edge of the PTV in an IMRT
treatment with multiple X-ray beams where the dose uniformity and
gradient are largely modified by the overlapping of beam. However,
the subject of this quantitative difference is not in the scope of
this paper.
The method used to
evaluate the differences between the calculated and measured
central-axis PDDs with a water phantom at 100 cm SSD is based on (1)
the distance to the nearest point where the calculated and measured
doses are identical and (2) the percentage difference between the
calculated and measured doses at the same measurement point. The
standard deviation (1xxsigma)_ percentage and position uncertainties
of each measured PDD are ±0.5% and ±0.5 mm, respectively. The
position uncertainty of the calculations is ±1.25 mm, while a 2.5-mm
pixel size is used in the pencil beam kernel. The measured and
calculated central-axis PDDs agreed within 1 mm in the buildup
region (0.5 cm to 1.5 cm depth) for all fields. A comparison between
the measured and calculated values for the 10-cm field in the
buildup region is shown in Fig. 1. The periodic fluctuation observed
for the calculated depth dose may be the result of the position
uncertainty in the pencil beam kernel. For depth doses between 1.5
cm and 25 cm, the difference between the calculated and measured
values is less than 2% of the dmax
value for all field sizes.
After validating the depth doses and the off-axis profiles with the phantoms scanned at 100 cm SSD, the accuracy of the MU calculations for a point at 100 cm SAD was examined for various depths and field sizes. The comparisons of the calculations with the measurements performed with water and solid water phantoms are shown in Table 1 for various depths and field sizes. Although the effective depth correction for the solid water phantom with respect to the water phantom should be small (about 1%) as observed for depths less than 10 cm, differences of about 2% were observed at a depth of 20 cm for various field sizes. Therefore, the effective depth correction should be taken into account when using a solid water phantom in measurements at large depths.
FIG. 1. The calculated and measured PDDs for a 10-cm field size are plotted as diamonds and circles, respectively. Measured PDDs with depths shifted ±1 mm are plotted as crosses. |
||
TABLE 1. Extracted monitor units (MUs) from measured output factors with water and solid water phantoms, and MUs calculated by the BrainSCAN RTP for delivering 100 cGy to prescription depths of various field sizes |
Since the water phantom was used to calculate MUs in BrainSCAN, MUs extracted from measured output factors with the water phantom were used to evaluate the MU calculation of the BrainSCAN RTP system. The difference between the calculated and measured MUs increases with depth for a given field size and with field size for a given depth. Large errors of ~5% and ~7% were observed for 20-cm and 30-cm fields, respectively, at a 20-cm depth.
IV. DISCUSSION
Although the measured and calculated PPDs and the
percentage profiles with a 100-cm SSD geometry agree to less than 2%
or within 1 mm, we observed large differences between the calculated
and extracted MUs from measurements with the water phantom.
Examination of the dose calculation algorithm used in BrainSCAN
reveals an error in converting the measured PPDs to TMRs in the
BrainSCAN RTP system. We found that the change of field size was not
taken into account correctly in Eq. (3). Percent depth doses from
larger field sizes were used to calculate the TMRs. As a result, the
calculated values of TMR in BrainSCAN are
overestimated.
The correct
transformation function(6) for Eq. (3)
should be
(4) |
where rt0
(calculated as (r x (SSD + t0)/SSD)) is the field size at reference depth
t0. The Sp is the phantom scatter factor. The field
size rd (calculated as
c x (SSD + d)/100) increases linearly with the depth
d and increases with field size c of the collimator
(MLC/jaws) opening. The inverse square correction is taken into
account by the term . To estimate the errors in calculating the TMR value
using Eq. (3), measured depth doses were normalized with respect to
a 10-cm field size at each depth for the various depths shown in
Fig. 2. The change in field size from the surface to a 20-cm depth
for a small 4-cm field is less than 1 cm, but for a large 20-cm
field, defined at the surface, the field size is 24 cm at a 20-cm
depth. The values of PDDs at a 20-cm depth change about 5% between
the field sizes of 20 cm and 24 cm, as shown in Fig. 2. Therefore,
using a PDD of field size of 24 cm instead of 20 cm in Eq. (3) to
calculate the TMR, as is calculated in BrainSCAN, overestimates the
TMR by about 5%. Since the PDDs increase for larger field sizes, and
the slope of increments is steeper for larger depths as shown in
Fig. 2, the errors resulting from using Eq. (3) instead of Eq. (4)
increase for deeper depths and larger field sizes. In addition, the
phantom scatter factor Sp also
introduces errors (typically about 2% to 3%).
FIG. 2. The measured PDDs as a function of field size for various depths at 100 cm SSD. PDDs are normalized with respect to the measured PDD of a 10 x 10 cm2 field for each depth. |
||
With the correct TMRPB, the dose(6) is calculated by
(5) |
where the collimator scatter factor Sc depends only on the collimator opening. The correct formulas for both TMR and dose calculations have been reported to BrainLAB.
REFERENCES
- Monk JE, Perks JR, Doughty D, Plowman PN. Comparison of a microleaf collimator with a 5-mm-leaf-width collimator for intracranial stereotactic radiotherapy. Int J Radiat Oncol Biol Phys. 2003;57:1443-1449.
- Van Dyk J, Barnett RB, Cygler JE, Shragge PC. Commissioning and quality assurance of treatment planning computers. Int J Radiat Oncol Biol Phys. 1993;26;261-273.
- Sibata CH, Mota HC, Beddar AS, Higgins PD, Shin KH. Influence of detector size in photon beam profile measurements. Phys Med Biol. 1991;36; 621-631.
- Mohan R, Chui C, Lidofsky L. Differential pencil beam dose computation model for photon. Med Phys. 1985;13;64-73.
- Mohan R, Chui C-S. Use of fast Fourier transforms in calculating dose distributions for irregularly shaped field for the three-dimensional treatment planning. Med Phys. 1986;14;70-77.
- Khan FM. The physics of radiation therapy. 3rd ed. Philadelphia (PA): Lippincott Williams & Wilkins; 2003. p178.
© 2005 Am. Coll. Med. Phys.